1.  Direct intra-abdominal pressure monitoring via piezoresistive pressure measurement: a technical note 
BMC Surgery  2009;9:5.
Background
Piezoresistive pressure measurement technique (PRM) has previously been applied for direct IAP measurement in a porcine model using two different devices. Aim of this clinical study was to assess both devices regarding complications, reliability and agreement with IVP in patients undergoing elective abdominal surgery.
Methods
A prospective cohort study was performed in 20 patients randomly scheduled to receive PRM either by a Coach®-probe or an Accurate++®-probe (both MIPM, Mammendorf, Germany). Probes were placed on the greater omentum and passed through the abdominal wall paralleling routine drainages. PRM was compared with IVP measurement by t-testing and by calculating mean difference as well as limits of agreement (LA).
Results
There were no probe related complications. Due to technical limitations, data could be collected in 3/10 patients with Coach® and in 7/10 patients with Accurate++®. Analysis was carried out only for Accurate++®. Mean values did not differ to mean IVP values. Mean difference to IVP was 0.1 ± 2.8 mmHg (LA: -5.5 to 5.6 mmHg).
Conclusion
Direct IAP measurement was clinically uneventful. Although results of Accurate++® were comparable to IVP, the device might be too fragile for IAP measurements in the clinical setting. Local ethical committee trial registration: EK2024
doi:10.1186/1471-2482-9-5
PMCID: PMC2678082  PMID: 19383161
2.  Clinical evaluation of an air-capsule technique for the direct measurement of intra-abdominal pressure after elective abdominal surgery 
BMC Surgery  2008;8:18.
Background
The gold standard for assessment of intraabdominal pressure (IAP) is via intravesicular pressure measurement (IVP). This accepted technique has some inherent problems, e.g. indirectness. Aim of this clinical study was to assess direct IAP measurement using an air-capsule method (ACM) regarding complications risks and agreement with IVP in patients undergoing abdominal surgery.
Methods
A prospective cohort study was performed in 30 patients undergoing elective colonic, hepatic, pancreatic and esophageal resection. For ACM a Probe 3 (Spiegelberg®, Germany) was placed on the greater omentum. It was passed through the abdominal wall paralleling routine drainages. To compare ACM with IVP t-testing was performed and mean difference as well as limits of agreement were calculated.
Results
ACM did not lead to complications particularly with regard to organ lesion or surgical site infection. Mean insertion time of ACM was 4.4 days (min-max: 1–5 days). 168 pairwise measurements were made. Mean ACM value was 7.9 ± 2.7 mmHg while mean IVP was 8.4 ± 3.0 mmHg (n.s). Mean difference was 0.4 mmHg ± 2.2 mmHg. Limits of agreement were -4.1 mmHg to 5.1 mmHg.
Conclusion
Using ACM, direct IAP measurement is feasible and uncomplicated. Associated with relatively low pressure ranges (<17 mmHg), results are comparable to bladder pressure measurement.
doi:10.1186/1471-2482-8-18
PMCID: PMC2575193  PMID: 18925973
3.  Different tissue reaction of oesophagus and diaphragm after mesh hiatoplasty. Results of an animal study 
BMC Surgery  2008;8:7.
Background
Laparoscopic mesh-reinforcement of the hiatal region in the treatment of gastroesophageal reflux disease (GERD) and paraesophageal hernia (PEH) reduces the risk of recurrence. However, there are still controversies about the technique of mesh placement, shape, structure and material. We therefore compared tissue integration and scar formation after implantation of two different polypropylene-meshes in a rabbit model.
Methods
A total of 20 female chinchilla rabbits were included in this study. Two different meshes (Polypropylene PP, Polyglecaprone 25 Composite PP-PG) were implanted on the abdominal diaphragm around the oesophagus. After 3 months the implanted meshes were excised en-bloc. Histological and morphological analyses were carried out accordingly proliferation rate, apoptosis and collagen type I/III ratio.
Results
Regarding proliferation rate of oesophagus PP (9.31 ± 3.4%) and PP-PG (13.26 ± 2.54%) differ in a significant (p = 0.0097) way. In the diaphragm we found a significant (p = 0.00066) difference between PP (9.43 ± 1.45%) and PP-PG (18.73 ± 5.92%) respectively. Comparing oesophagus and diaphragm we could prove a significant difference within PP-PG-group (p = 0.0195). Within PP-group the difference reached no statistical significance (p = 0.88). We found analogous results regarding apoptosis.
Furthermore, there is a significant (p = 0.00013) difference of collagen type I/III ratio in PP-PG (12.28 ± 0.8) compared to PP (8.44 ± 1,63) in case of oesophageal tissue. Concerning diaphragm we found a significant difference (p = 0.000099) between PP-PG (8.85 ± 0.81) and PP (6.32 ± 1.07) as well.
Conclusion
The histologic and morphologic characteristics after prosthetic enforcement of the hiatus in this animal model show a more distinct tissue integration using PP-PG compared to PP. Additionally, different wound healing and remodelling capability influence tissue integration of the mesh in diaphragm and oesophagus.