PMCC PMCC

Search tips
Search criteria

Advanced
Results 1-2 (2)
 

Clipboard (0)
None
Journals
Authors
Year of Publication
Document Types
jtitle_s:("Acta mol")
1.  Right atrial preference pacing algorithm in the prevention of paroxysmal atrial fibrillation in myotonic dystrophy type 1 patients: a long term follow-up study 
Acta Myologica  2012;31(2):139-143.
Atrial Preference Pacing (APP) is a pacemaker (PM) algorithm that works by increasing the atrial pacing rate to achieve continuous suppression of a spontaneous atrial rhythm and prevent supraventricular tachyarrhythmias. We have previously shown that atrial preference pacing may significantly reduce the number and the duration of AF episodes in myotonic dystrophy type 1 (DM1) patients who are paced for standard indications.
However, the role that APP therapies play in the prevention of AF in a long-term period remains still unclear. Aim of the present prospective study was to evaluate whether this beneficial effect is maintained for 24-months follow-up period.
To this aim, 50 patients with Myotonic Dystrophy type 1 who underwent dual-chamber PM implantation for first- and second- degree atrioventricular block, were consecutively enrolled and followed for 2 years. One month later the stabilization period, after the implantation, they were randomized to APP algorithm programmed OFF or ON for 6 months each, using a cross-over design, and remained in the same program for the second year. The results showed that while the number of AF episodes during active treatment (APP ON phases) was lower than that registered during no treatment (APP OFF phases), no statistically significant difference was found in AF episodes duration between the two phases. Furthermore, during the APP OFF and APP ON phases, the percentage of atrial pacing was 0 and 99%, respectively, while the percentage of ventricular pacing did not show differences statistically significant (11 vs. 9%, P = 0.2). Atrial premature beats were significantly higher during APP OFF phases than during APP ON phases. Lead parameters remained stable over time and there were no lead-related complications. Based on these 24-months follow-up data, we can conclude that, in DM1 patients who underwent dual-chamber PM implantation, APP is an efficacy algorithm for preventing paroxysmal AF even in long term periods.
PMCID: PMC3476853  PMID: 23097606
myotonic dystrophy; atrial preference pacing; atrial fibrillation
2.  Cardiac resynchronization improves heart failure in one patient with Myotonic Dystrophy type 1. A case report 
Acta Myologica  2012;31(2):154-155.
We report an improvement in symptoms of heart failure, a reduced left ventricular dysfunction and induced reverse remodelling in one patient with Myotonic Dystrophy type 1, showing an early onset ventricular dysfunction secondary to a complete left bundle branch block (LBBB) who underwent cardioverter defibrillator CRT (ICD- CRT) implantation.
PMCID: PMC3476859  PMID: 23097608
myotonic dystrophy; cardiac resynchronization therapy; sudden death

Results 1-2 (2)