PMCC PMCC

Search tips
Search criteria

Advanced
Results 1-1 (1)
 

Clipboard (0)
None
Journals
Authors
Year of Publication
Document Types
1.  BRIEF REPORT: How Well Do Clinic-Based Blood Pressure Measurements Agree with the Mercury Standard? 
BACKGROUND
Obtaining accurate blood pressure (BP) readings is a challenge faced by health professionals. Clinical trials implement strict protocols, whereas clinical practices and studies that assess quality of care utilize a less rigorous protocol for BP measurement.
OBJECTIVE
To examine agreement between real-time clinic-based assessment of BP and the standard mercury assessment of BP.
DESIGN
Prospective reliability study.
PATIENTS
One hundred patients with an International Classification of Diseases—9th edition code for hypertension were enrolled.
MEASURES
Two BP measurements were obtained with the Hawksley random-zero mercury sphygmomanometer and averaged. The clinic-based BP was extracted from the computerized medical records.
RESULTS
Agreement between the mercury and clinic-based systolic blood pressure (SBP) was good, intraclass correlation coefficient (ICC)=0.91 (95% confidence interval (CI): 0.83 to 0.94); the agreement for the mercury and clinic-based diastolic blood pressure (DBP) was satisfactory, ICC=0.77 (95% CI: 0.62 to 0.86). Overall, clinic-based readings overestimated the mercury readings, with a mean overestimation of 8.3 mmHg for SBP and 7.1 mmHg for DBP. Based on the clinic-based measure, 21% of patients were misdiagnosed with uncontrolled hypertension.
CONCLUSIONS
Health professionals should be aware of this potential difference when utilizing clinic-based BP values for making treatment decisions and/or assessing quality of care.
doi:10.1111/j.1525-1497.2005.0105.x
PMCID: PMC1490157  PMID: 16050862
blood pressure measurement assessment; clinic method; mercury device

Results 1-1 (1)