PMCC PMCC

Search tips
Search criteria

Advanced
Results 1-1 (1)
 

Clipboard (0)
None

Select a Filter Below

Journals
Authors
Year of Publication
Document Types
author:("oldroyd, J R")
1.  National reference doses for dental cephalometric radiography 
The British Journal of Radiology  2011;84(1008):1121-1124.
Objectives
Diagnostic reference levels (DRLs) are an important tool in the optimisation of clinical radiography. Although national DRLs are provided for many diagnostic procedures including dental intra-oral radiography, there are currently no national DRLs set for cephalometric radiography. In the absence of formal national DRLs, the Health Protection Agency (HPA) has previously published National Reference Doses (NRDs) covering a wide range of diagnostic X-ray examinations. The aim of this study was to determine provisional NRDs for cephalometric radiography.
Methods
Measurements made by the Dental X-ray Protection Service (DXPS) of the HPA, as part of the cephalometric X-ray equipment testing service provided to dentists and dental trade companies throughout the UK, were used to derive provisional NRDs.
Results
Dose–area product measurements were made on 42 X-ray sets. Third quartile dose–area product values for adult and child lateral cephalometric radiography were found to be 41 mGy cm2 and 25 mGy cm2, respectively, with individual measurements ranging from 3 mGy cm2 to 108 mGy cm2.
Conclusion
This report proposes provisional NRDs of 40 mGy cm2 and 25 mGy cm2 for adult and child lateral cephalometric radiographs, respectively; these doses could be considered by employers when establishing their local DRLs.
doi:10.1259/bjr/26420990
PMCID: PMC3473818  PMID: 22101581

Results 1-1 (1)