1.  Repeatability of regional myocardial blood flow calculation in 82Rb PET imaging 
Background
We evaluated the repeatability of the calculation of myocardial blood flow (MBF) at rest and pharmacological stress, and calculated the coronary flow reserve (CFR) utilizing 82Rb PET imaging. The aim of the research was to prove high repeatability for global MBF and CFR values and good repeatability for regional MBF and CFR values. The results will have significant impact on cardiac PET imaging in terms of making it more affordable and increasing its use.
Methods
12 normal volunteers were imaged at rest and during pharmacological stress, with 2220 MBq of 82Rb each. A GE Advance PET system was used to acquire dynamic 50-frame studies. MBF was calculated with a 2-compartmental model using a modified PMOD program (PMOD; University Hospital Zurich, Zurich, Switzerland). Two differential equations, describing a 2-compartmental model, were solved by numerical integration and using Levenberg-Marquardt's method for fitting data. The PMOD program defines 16 standard segments and calculates myocardial flow for each segment, as well as average septal, anterior, lateral, inferior and global flow. Repeatability was evaluated according to the method of Bland and Altman.
Results
Global rest and stress MBF, as well as global CFR, showed very good repeatability. No significant differences were found between the paired resting global MBF (0.63 ± 0.13 vs. 0.64 ± 0.13 mL/min/g; mean difference, -1.0% ± 2.6%) and the stress global MBF (1.37 ± 0.23 vs. 1.37 ± 0.24; mean difference, 0.1% ± 2.3%). Global CFR was highly reproducible (2.25 ± 0.56 vs. 2.22 ± 0.54, P = not statistically significant; mean difference, 1.3% ± 14.3%). Repeatability coefficients for global rest MBF were 0.033 (5.2%) and stress MBF 0.062 (4.5%) mL/min/g. Regional rest and stress MBF and CFR have shown good reproducibility. The average per sector repeatability coefficients for rest MBF were 0.056 (8.5%) and stress MBF 0.089 (6.3%) mL/min/g, and average repeatability coefficient for CFR was 0.25 (10.6%).
Conclusion
The results of the study show that software calculation of MBF and CFR with 82Rb myocardial PET imaging is highly repeatable for global values and has good repeatability for regional values.
doi:10.1186/1756-6649-9-2
PMCID: PMC2646684  PMID: 19178700
2.  Non-invasive imaging of atherosclerotic plaque macrophage in a rabbit model with F-18 FDG PET: a histopathological correlation 
Background
Coronary atherosclerosis and its thrombotic complications are the major cause of mortality and morbidity throughout the industrialized world. Thrombosis on disrupted atherosclerotic plaques plays a key role in the onset of acute coronary syndromes. Macrophages density is one of the most critical compositions of plaque in both plaque vulnerability and thrombogenicity upon rupture. It has been shown that macrophages have a high uptake of 18F-FDG (FDG). We studied the correlation of FDG uptake with histopathological macrophage accumulation in atherosclerotic plaques in a rabbit model.
Methods
Atherosclerosis was induced in rabbits (n = 6) by a combination of atherogenic diet and balloon denudation of the aorta. PET imaging was performed at baseline and 2 months after atherogenic diet and coregistered with magnetic resonance (MR) imaging. Normal (n = 3) rabbits served as controls. FDG uptake by the thoracic aorta was expressed as concentration (μCi/ml) and the ratio of aortic uptake-to-blood radioactivity. FDG uptake and RAM-11 antibody positive areas were analyzed in descending aorta.
Results
Atherosclerotic aortas showed significantly higher uptake of FDG than normal aortas. The correlation of aortic FDG uptake with macrophage areas assessed by histopathology was statistically significant although it was not high (r = 0.48, p < 0.0001). When uptake was expressed as the ratio of aortic uptake-to-blood activity, it correlated better (r = 0.80, p < 0.0001) with the macrophage areas, due to the correction for residual blood FDG activity.
Conclusion
PET FDG activity correlated with macrophage content within aortic atherosclerosis. This imaging approach might serve as a useful non-invasive imaging technique and potentially permit monitoring of relative changes in inflammation within the atherosclerotic lesion.