1.  Swab culture monitoring of automated endoscope reprocessors after high-level disinfection 
AIM: To conduct a bacterial culture study for monitoring decontamination of automated endoscope reprocessors (AERs) after high-level disinfection (HLD).
METHODS: From February 2006 to January 2011, authors conducted randomized consecutive sampling each month for 7 AERs. Authors collected a total of 420 swab cultures, including 300 cultures from 5 gastroscope AERs, and 120 cultures from 2 colonoscope AERs. Swab cultures were obtained from the residual water from the AERs after a full reprocessing cycle. Samples were cultured to test for aerobic bacteria, anaerobic bacteria, and mycobacterium tuberculosis.
RESULTS: The positive culture rate of the AERs was 2.0% (6/300) for gastroscope AERs and 0.8% (1/120) for colonoscope AERs. All the positive cultures, including 6 from gastroscope and 1 from colonoscope AERs, showed monofloral colonization. Of the gastroscope AER samples, 50% (3/6) were colonized by aerobic bacterial and 50% (3/6) by fungal contaminations.
CONCLUSION: A full reprocessing cycle of an AER with HLD is adequate for disinfection of the machine. Swab culture is a useful method for monitoring AER decontamination after each reprocessing cycle. Fungal contamination of AERs after reprocessing should also be kept in mind.
doi:10.3748/wjg.v18.i14.1660
PMCID: PMC3325533  PMID: 22529696
Automated endoscope reprocessor; Gastrointestinal scope; High-level disinfection; Swab culture; Monitoring; Decontamination
2.  Delayed presentation of intrathoracic esophageal perforation after pneumatic dilation for achalasia 
Pneumatic dilation (PD) is considered to be a safe and effective first line therapy for achalasia. The major adverse event caused by PD is esophageal perforation but an immediate gastrografin test may not always detect a perforation. It has been reported that delayed management of perforation for more than 24 h is associated with high mortality. Surgery is the treatment of choice within 24 h, but the management of delayed perforation remains controversial. Hereby, we report a delayed presentation of intrathoracic esophageal perforation following PD in a 48-year-old woman who suffered from achalasia. She completely recovered after intensive medical care. A review of the literature is also discussed.
doi:10.3748/wjg.15.4461
PMCID: PMC2747072  PMID: 19764103
Intrathoracic esophageal perforation; Delayed presentation; Pneumatic dilation; Esophageal achalasia
3.  Surveillance cultures of samples obtained from biopsy channels and automated endoscope reprocessors after high-level disinfection of gastrointestinal endoscopes 
BMC Gastroenterology  2012;12:120.
Background
The instrument channels of gastrointestinal (GI) endoscopes may be heavily contaminated with bacteria even after high-level disinfection (HLD). The British Society of Gastroenterology guidelines emphasize the benefits of manually brushing endoscope channels and using automated endoscope reprocessors (AERs) for disinfecting endoscopes. In this study, we aimed to assess the effectiveness of decontamination using reprocessors after HLD by comparing the cultured samples obtained from biopsy channels (BCs) of GI endoscopes and the internal surfaces of AERs.
Methods
We conducted a 5-year prospective study. Every month random consecutive sampling was carried out after a complete reprocessing cycle; 420 rinse and swabs samples were collected from BCs and internal surface of AERs, respectively. Of the 420 rinse samples collected from the BC of the GI endoscopes, 300 were obtained from the BCs of gastroscopes and 120 from BCs of colonoscopes. Samples were collected by flushing the BCs with sterile distilled water, and swabbing the residual water from the AERs after reprocessing. These samples were cultured to detect the presence of aerobic and anaerobic bacteria and mycobacteria.
Results
The number of culture-positive samples obtained from BCs (13.6%, 57/420) was significantly higher than that obtained from AERs (1.7%, 7/420). In addition, the number of culture-positive samples obtained from the BCs of gastroscopes (10.7%, 32/300) and colonoscopes (20.8%, 25/120) were significantly higher than that obtained from AER reprocess to gastroscopes (2.0%, 6/300) and AER reprocess to colonoscopes (0.8%, 1/120).
Conclusions
Culturing rinse samples obtained from BCs provides a better indication of the effectiveness of the decontamination of GI endoscopes after HLD than culturing the swab samples obtained from the inner surfaces of AERs as the swab samples only indicate whether the AERs are free from microbial contamination or not.
doi:10.1186/1471-230X-12-120
PMCID: PMC3482587  PMID: 22943739
Surveillance culture monitoring; Gastrointestinal scope; Automated endoscope reprocessor; High-level disinfection reprocessing