J Atr Fibrillation. 2017 Apr-May; 9(6): 1600.
Published online 2017 April 30. doi:  10.4022/jafib.1600
PMCID: PMC5673347

Pulmonary Vein Isolation with the Multipolar nMARQ™ Ablation Catheter: Efficacy And Safety In Acute And Long-Term Follow Up

Abstract

Pulmonary vein isolation (PVI) is an established therapy for atrial fibrillation (AF). One challenge in the catheter-based treatment of this arrhythmia is to develop an effective and safe ablation approach to achieve durable and consistent lesions around the PVs. The multipolar irrigated radiofrequency (RF) ablation catheter nMARQTM was designed as a single-shot device with the aim to achieve these goals. This article reviews the current literature with respect to acute- and long- term success rates after PVI with this circular mapping and ablation device. Furthermore, since this device recently became discredited to potential lethal complications, we will also focus on the data available on safety issues with this ablation system.

Keywords: nMARQ, circular ablation catheter, atrial fibrillation, pulmonary vein isolation

Introduction

Pulmonary vein isolation (PVI) is an established method for the treatment of atrial fibrillation (AF). In 1998, Michel Haïssaguerre demonstrated that the pulmonary veins (PV) were an important anatomical structure harboring triggers for the initiation of AF [1]. Thus, the primary endpoint for interventional treatment of AF by ablation is circumferential electrical isolation of the PVs [2]. However, as this procedure is challenging and still time-consuming even for experienced operators, there is a need for workflow optimization, e.g. by novel ablation devices. In this context, so-called “single-shot’ devices have been introduced in order to enable a quick and durable PV isolation, thereby increasing efficacy and safety of PVI procedures. Single-shot devices were developed as a tool aiming to provide circular transmural lesions by simultaneously mapping and ablating at multiple sites around the antra of the PVs via a single-transseptal access point [3].

In 2011, a steerable multi-electrode catheter (8.4 F) with a deflectable tip (nMARQTM, Biosense Webster, Inc., Diamond Bar, Ca, USA) was introduced. The nMARQTM catheter consists of ten irrigated electrodes, and is capable of full integration into the CARTO® electroanatomic mapping system [4] [Figure 1]. Energy delivery duration is set between 30 to 60 seconds, and radiofrequency (RF) ablation can be individually performed over each combination of the 10 electrodes in unipolar mode (maximum 25 W) or bipolar mode between two electrodes (maximum 15 W) [5].

Figure 1.
illustrates the position of the circular nMARQ TM catheter in the left atrium, recording the electroanatomical map in Carto® 3. Yellow-grey: temperature probe; green: RV catheter for phrenic nerve stimulation during ablation of right-sided PVs. ...

Early studies suggested the nMARQTM to be an effective and safe tool for PVI [4], [6]-[8] with one multicenter study confirming a high success rate after nMARQTM procedures [3]. However, the device also presented with some safety concerns arising from some severe complications, questioning the safety of this novel device [9]-[11], and ultimately leading to the interim recall of the 2nd generation nMARQTM catheter [12]. Therefore, we conducted this review of the current literature with respect to mid- and long-term efficacy as well as safety of the nMARQTM ablation device.

Case Report

The electronic databases PubMed and Google Scholar were used to identify potential articles including prospective and retrospective studies, case reports, registries, editorials, and review articles. Search terms included “atrial fibrillation (AF),” “pulmonary vein isolation (PVI),” “circular ablation catheter,” “multipolar ablation catheter,” “single shot device,” and “nMARQTM.” Data from 81 identified articles were reviewed carefully for information regarding ablation with the nMARQTM device. We summarized the data according to the available information on clinical outcome (n=11 studies), procedural parameters (n=14 studies) and safety outcome (n=16 studies).

Clinical outcome

Since the release of the first generation of this catheter in 2011, outcomes following nMARQTM ablation from more than 1400 patients have been reported [3], [4], [6], [7], [9], [13]-[18]. Specifically, our search found 11 published studies, with follow-up (FU) data exceeding 3 months post PVI. The outcome of interest in these studies was generally defined as recurrence of AF or the combination of AF with atrial flutter and atrial tachycardia following a 90-day blanking period after PVI. Results from one multicenter study, and 10 single-center studies reported overall mid-term success rates ranging from 52% to 80.9%. Success rates varied depending on AF type, FU duration, and the concomitant use of antiarrhythmic drugs (AAD) [Table 1].

Table 1
Clinical outcomes of patients investigated in available literature. FU denotes follow-up; AAD antiarrhythmic drugs

For those patients that underwent ablation of paroxysmal AF, recurrence rates were between 22.8% and 35% [3], [16], and are comparable to those obtained by conventional RF, Cryoballoon, or different circumferential RF ablation catheters (PVAC) after one year (between 30.1 - 35.9%) [19]-[21]. The very low recurrence rate of 22.8% reported by Rodriguez et al. may be in part attributable to the fact that all patients were administered AAD during the blanking period [16]. Burri et al. reported recurrence rates of 54% over 15 ± 4 months which were considerably higher than other published studies [13]. The authors suggested that in addition to the slightly longer FU duration compared to other studies, reduced power output (max. 15 watt unipolar), the restricted RF delivery, and the waiving on a circular mapping catheter to confirm PVI, could be causative for worse outcomes in their study [7], [18] (see chapter “acute efficacy”).

Data on success rates after nMARQTM ablation in persistent AF are scarce. The five published studies following patients with persistent AF reported recurrence rates ranging from 30% to 48% [3], [9], [14], [15], [17]. The clinical use of the nMARQTM device has been limited so far in patients with persistent AF. Prior expert consensus documents from the HRS/EHRA/ECAS suggested that for patients with persistent AF “operators should consider more extensive ablation based on linear lesions or complex fractionated electrograms” [22], for which the nMARQTM catheter is not intended. Notably, since Verma et al. reported that ablation strategies beyond conventional PVI did not translate into additional clinical benefit in persistent AF in the STAR-AF-II trial, the use of the single-shot devices, incl. the nMARQTM catheter, re-gain attention for a PVI only treatment in patients with persistent AF [23].

Acute efficacy of nMARQTM guided ablation

Acute durable PVI (acute efficacy) with the nMARQTM device ranged from 83% to 100% of treated patients, with acute efficacy in 95.7% to 100% of targeted veins [4], [18], [24] (see [Table 2]). Wakili et al. reported that 5 of 116 PVs (4.3%) could not successfully be isolated with the nMARQTM catheter [18]. Zellerhoff et al. failed to acutely isolate three PVs (2x RSPV, 1x RIPV), Rodriguez-Entem, 2 PVs (1 RIPV and 1 LIPV), and Scaglione, 1 LSPV [6], [7], [16]. Indicated reasons for isolation failure comprised of difficulties in achieving a transmural lesion at the ridge, significant temperature rise in esophagus, catheter geometry, and limited device maneuverability [7], [18]. According to their single-center experience, Deneke et al. reported that through the routine use of a steerable sheath for catheter access into LA, when appropriate contact force in the LAA ridge region is achieved, all different anatomies of PVs should be treatable by the nMARQTM device [8], [14]. Inconsistent with results from PVI with single-tip catheters and circular mapping catheters (CMC), most of the reported studies did not routinely perform exit block testing to confirm PVI. This was due to challenging intubation of small PVs with the nMARQTM catheter [18].

Table 2
summarizes acute success rates and procedural results with the nMARQTM ablation device. * highlights studies with PVI confirmation with additional circular mapping catheter

With respect to acute success rates of PVI, these results are comparable to those obtained by conventional RF energy [25], PVAC [21], [26] and Cryoballoon ablations [25], [27]. However, most of these studies used the nMARQTM as the intended “single-shot” device, without confirming the PV isolation with a standard CMC. Scaglione and Rosso et al. reported on an overall inconsistency between CMC and nMARQTM signals in 22 of 102 PVs (22%) to 12 of 39 PVs (30%). Additionally, Rosso observed good consistency prior to PVI, but poor concordance after PVI. In all cases these variations led to further RF delivery [7], [24]. Wakili et al. reported on a discrepancy rate of 35% in their study [18] [Figure 2]. Scaglione et al. speculated that persistent PV potentials on the CMC after extinction on nMARQTM suggest persistence of electrical conduction from the PV to the atrium. They suggested that the difference in inter-electrode spacing between CMC and nMARQTM, or the more proximal position of the nMARQTM in the PVs, are causative for significant signal divergence [7], [28], [29]. In order to avoid false-positive PVI results which may impair the outcome of the procedure, Wakili et al. strongly recommended a dual transseptal approach with simultaneous PV potential recordings [18].

Figure 2.
Insufficient signal accuracy of the nMARQTM catheter [18]. The illustration shows intracardiac recordings of consecutive PV mappings by the nMARQ TM catheter and by a CMC of the same vein after ablation; (A) RSPV mapping with nMARQ TM suggests absence ...

Deneke et al. suggested that there may be procedure-related factors influencing the success rates following ablation with the nMARQTM device. In particular, Deneke et al. reported that overall success rates were positively associated with higher maximum energy delivery rates at the posterior wall (25 watt vs. 20 watt) [14]. However, these higher energy delivery rates were likely associated with a higher risk for esophageal thermal damage [5].

Procedural results

The development of the nMARQTM as a single-shot device was driven by the intention to shorten and simplify PVI procedures, increase safety, and reduce radiation dose, all while producing equal (or better) success rates of other ablation devices. Pooled results for periprocedural data are depicted in [Table 2].

Total procedure times ranged from 72 ± 6.5 minutes [9] to 223 ± 53 minutes [5]. Total procedure time in the latter study is likely highest due to four cavotricuspid isthmus ablations, and one ablation of roof dependent LA tachycardia which was performed during the course of PVI. Summarized, total procedure times reported from the nMARQTM device compared well with procedure times obtained from other PVI ablation modalities (Cryoballoon: 136 to 371 min [20], [30], [31]; PVAC: 121 to 137.1 min [32], [33]; RF 140.9 to 165 min [19], [25]). Multiple groups suggested after a learning period a mean reduction in overall procedure time of 19.1% to 62.1% [4], [14], [24]. However, Wakili et al. failed to show a significant nMARQTM ablation learning curve with respect to overall procedure time [18].

Mean fluoroscopy times varied over a broad range, from 1.8 minutes [7] to 35.5 minutes [5]. In the latter study, the prolonged fluoroscopy times may be explained by additional CMC use in order to confirm complete PVI. Ablation with the nMARQTM reveals comparable fluoroscopy times as indicated in literature for other ablation devices (Cryoballoon: 21 to 40 min [19], [30], [31], [34]; PVAC 21 to 33 min [21], [32]; single tip 16.6 to 24 min [19], [25]. A suggest learning curve shows a reduction of 51.5% to 64.5% of total fluoroscopy time [4], [24].

With respect to total RF time, as the number of active electrodes during ablation can individually be varied, the comparison to different PVI modalities is challenging [9]. When reporting on RF duration, the majority of studies reported the total RF duration, without indication of the number of active electrodes. This hampers the direct comparison of RF times to other one shot devices or single-tip catheter approaches. However, total nMARQTM RF times (7.7 to 18.5 min [9], [35]) are slightly longer compared to reported RF durations with conventional single tip catheters (33 min [25]; 21 min [18]). Only three studies used an additional CMC to confirm complete PVI [7], [18], [24]. Wakili et al. reported that the use of an additional CMC to confirm PVI was associated with longer RF durations, and with the identification of 19 of 29 PVs (65.5%) with persisting atrio-PV conduction after nMARQTM ablation (21.0 ± 9.0 vs. 17.6 ± 6.5 min) [18]. Data on analyses of RF times per individual vein is scarce. The available literature provides evidence that RF times needed for PVI are significantly longer in the superior PV compared to RF times needed in the inferior PVs [3], [7], [16]. All but one study indicates that mean RF times with the nMARQ device are longest in the LSPV (191.6 ± 41.9 sec).

Safety

As the nMARQTM catheter has shown to be associated with comparable outcomes to currently available ablation technologies, in respect to recurrence post ablation, a specific focus is placed on safety issues. In general, AF ablation is associated with a incidence of acute complications ranging from <1% to 6% [36].

Esophageal thermal damage

Due to the specific design of circular ablation devices and therefore high energy delivery at the posterior wall, esophageal lesions are of major concern [Table 3]. Esophageal thermal damage (ETD) is considered a precursor of fistulas, even though the causal relation between fistulas and thermal esophageal lesions is largely unclear [14]. Following PVI with a single-tip, a high incidence of thermal lesions have been reported (thermal esophageal damage (11%) and gastroparesis (17%)) [37]. Deneke et al. assessed 136 out of 145 patients with endoscopy after nMARQTM ablation, and report on 7 ulcerous and 22 erythematous lesions after PVI with the nMARQTM [14].

Table 3
An overview of published literature on procedure-related complications with the nMARQTM system. PE denotes pericardial effusion/tamponade; PNP phrenic nerve palsy; AEF atrio-esophageal fistula; SCE silent cerebral lesion; TP temperature probe; EGD esophago-gastro ...

ETD resulting in fistulas can lead to fatal complications. The indicated mortality in literature after development of atrio-esophageal fistula (AEF) is 71% [38]. An overall incidence of 3 of 1417 patients (0.21%) that developed AEF has been derived from the published nMARQTM studies. Of those reported cases of AEF, Vurma et al. reported fatal outcomes following development of AEF in 2 of 327 consecutive patients (0.6%) following ablation with the nMARQTM device. This report led to an immediate recall of the nMARQTM catheter in its 2nd generation [9], [12]. Deneke and Mahida et al. each reported cases of delayed occurrence of AEF, the latter reporting on a delay of 4.5 weeks between PVI procedure and occurrence of first symptoms [3], [11].

Various safety precautions have therefore been suggested in order to avoid thermal esophageal damage. According to initial experience, the use of a thermal probe has been suggested in order to reduce the incidence of thermal damage during AF ablation at the posterior wall [39], [40]. Disagreement remains on the esophageal cut-off temperature during RF delivery, ranging from 39 degrees [41] to 41 degrees Celsius [5], [42].

Considering the recent literature on nMARQTM ablations, only one study [43] suggested a benefit of using a temperature probe during multipolar RF ablation [14], [35], [41], [44]. Consistent with other reports concerning RF ablation, Deneke et al. suggested an increased risk for ETD in patients with thermal probes during RF ablation (21% vs. 0%, p<0.001) [5], [8], [14], [18]. They speculate a possible ‘antenna’ effect of the thermal probe intensifying local energy with heating at the esophageal region, or a stiffening of the esophagus itself avoiding the esophagus to sidestep during catheter pressure [5], [14], [39], [45]. However, in cases of large esophageal diameter, the probe is not able to cover the entire esophageal region (as shown by barium sulfate ingestion), and therefore may lead to an underestimation of the local temperature. This underestimation of temperature may result in a higher risk for esophageal thermal lesion [41]. According to those presented data, the use of thermal probes should therefore be avoided.

Other precautions suggested for ETD prevention comprise a reduction of the maximum power (20 watt [14]), and even lower temperatures when bipolar ablation is performed [41]. Limitation of RF time at the posterior wall is also recommend for ETD prevention [5]. The two cases of AEF reported by Vurma et al. occurred following ablation with a max. temp of 16 to 18 watts (30 sec max. duration for vast of energy deliveries) [9]. It must be mentioned that the report of maximum delivered RF energy is often misleading. In order to avoid ETD, most operators only decrease RF power at the posterior wall.

Finally, the use of general anesthesia has been reported to serve as a risk factor for ETD [46]. Most of the patients undergoing ablation under general anesthesia also had esophageal temperature probes during the procedure. Therefore, the influence of general anesthesia as a risk factor for thermal lesions remains unclear, and needs to be critically questioned.

Thromboembolic complications

Thromboembolic complications are generally considered a major concern with the nMARQTM device, which is based on former negative experience with the circular single-shot ablation PVAC device [47], [48]. Reviewing the current literature on nMARQTM ablations, no stroke or transient ischemic attack (TIA) were reported. However, silent cerebral lesions (SCL), which likely represent small thromboembolic infarctions, have been reported in literature. Varying based on the ablation technology used, SCL were reported in up to 40% of patients after RF ablations [14], [47], [49], [50]. Since embolic-lowering maneuvers have been introduced into clinical practice, the use of the nMARQTM device remains associated with the highest reported incidence of asymptomatic thromboembolic complications [6], [8], [51]. The clinical significance of these SCL is unclear. However, an association between SCL and neuropsychological changes, especially of verbal memory, has been suggested [52], yet other studies have failed to show an association [53], [54].

Out of 16 reported studies on complication rates, six studies performed cerebral imaging (CT\MRI) after PVI to rule out SCL [4], [6], [8], [14], [16], [53]. Two groups found SCL following PVI with the nMARQTM device, ranging from 1 in 19 patients (5%) [53] to 14 in 43 (33%) [55]. However, none presented with any obvious neurological symptoms. The high percentage of 38% post-ablation SCL, as indicated by Deneke et al.[55], might overestimate the real percentage as Sugihara et al. found a high incidence of preexisting SCL before PVI [53]. This high prevalence of pre-existing lesions (12.3-92%) might represent a condition of inappropriate anticoagulation before PVI [50], [51], [54]-[57]. In this context, studies have indicated that the maintenance of preexisting anticoagulation, compared to discontinuation and bridging with heparin, contributed to a reduction of periprocedural cerebral events [58], [59]. In general, different anticoagulation regimens make the comparison of studies dealing with microbubbles during ablation difficult [35]. Kiss et al. demonstrated that nMARQTM ablation was associated with a high incidence of microbubbles. This bubble formation seems to be higher than when compared to ablation with new-generation PVAC devices, or cryoballoon ablation [35]. The assessment of the intensity of micro emboli generation during ablation procedures is measured in the middle cerebral artery by transcranial Doppler [35]. However, this technique of measuring microbubbles by ultrasonic techniques has not been consistently validated with respect to the clinical significance. It remains completely unclear as to whether these microbubbles represent solid particles or gas and how they translate into a manifest clinical finding.

With respect to conditions predisposing to thrombi formation, the specific design of the circular nMARQTM catheter with 10 irrigated electrodes is suspected to be causative for this phenomenon. Csanadi et al. speculated that the high volume flow of irrigation saline solution (6-7ml/ electrode, resulting in 60-70ml/min) can result in bubble formation and subsequent microembolism [60]. Further, charring on the electrodes is thought to be another major source of SCL, arising from former PVAC experience [5]. Shin reported the identification of charring on 3 of 15 cases (20%) with the nMARQTM catheter [4]. Charring was found primarily between electrode 1 and 10. This location is where electrodes are delivering RF energy in close proximity, and is likely the source of a bipolar short circuit resulting in tissue and blood heating [51]. Therefore, Shin et al. recommend RF delivery only with sufficient distance between electrode 1 and 10 on fluoroscopy, 3D visualization, without indication of proximity by artifacts on the corresponding EGMs [4].

Despite existing data from animal studies investigating the PVAC device, there is still discrepancy as to whether the use of unipolar over bipolar RF energy per se could reduce the incidence of microembolism [61]. Nevertheless, in order to reduce the incidence of SCL, abandonment of a bipolar RF energy use is recommended in general now. As catheter manipulations are thought to be a source of microbubble formation, the following precautions should be considered [54], [59]: at least half the calculated bolus dosage of heparin should be given before transseptal passage, continuous flushing of the long LA sheath, and whenever possible, retraction of the sheaths in RA. Additionally, a catheter change over the long LA sheath should be avoided [35]. This however questions the intention of these “single-shot” devices, because in addition to the PVI, an additional CMC may be required [7], [18], [24]. Further, the administration of a proton pump inhibitor should be considered for at least 6 weeks following ablation in order to prevent progression of esophageal thermal damage to ulceration [5], [8].

Other severe complications

Other severe complications including pericardial effusion/tamponade and phrenic nerve palsy (PNP) were reported in 7 out of 16 cited studies [5], [6], [10], [13]-[16], [18]. Pericardial effusion/tamponade was reported in 6 out of 1417 (0.4%) patients, and PNP in 4 of 1417 patients (0.3%). However, the prevalence of PNP following nMARQTM ablation is lower than in the literature for overall PVI procedures, with PNP rates ranging from 0.48% to 11% [62]-[64]. Although injury of the phrenic nerve is reported following various ablation techniques, it has been suggested to be more likely with the Cryoballoon [20], [65]. The exact mechanism of the high rate of PNP after circular PV ablation with the Cryoballoon remains unclear, especially with regard to the lower percentage of PNP after nMARQTM of overall 0.3% [10], [13], [66]. This may be explained in part by a more antral ablation with nMARQTM catheter compared to Cryoballoon (diameter 20 to 35mm vs. 23 or 28mm) [6]. With respect to the underlying mechanism, experimental data suggested a Wallerian degeneration (axonal damage by coagulation), or an injury of the right pericardiophrenic artery, both with the potential for recovery [62], [67], [68].

In order to avoid PNP during nMARQTM ablation, Arroja et al. suggested a further power limitation of 12 to 15 watts, phrenic nerve stimulation on each electrode of the nMARQTM catheter, and continuous phrenic nerve stimulation during RF application [10]. Additionally, Roka et al. reported on a novel technique to prevent PNP, by identifying the overlapping region between right and left atrium. RF applications proximal to this line are suggested to be safe with respect to PNP [62]. In order to rule out pulmonary vein stenosis following PVI, imaging modalities were reported on in five studies [Table 3], and no significant stenosis was mentioned.

Conclusion

The nMARQTM catheter was developed in order to enable fast, durable, and safe PVI by using a single-transseptal approach. As presented, the literature reveals comparable acute and long-term clinical outcomes after AF ablation to single-tip and different other circular ablation catheters. With respect to procedural parameters, current studies failed to provide an evidence for reduction of total RF with the nMARQTM device (see [Table 2]). However, these studies comprised initial experience, scientific evaluation with a learning curve and small patient cohorts in a majority of studies.

Although intended to function as a single-shot device, some issues were presented concerning the catheter’s procedural performance. For example, when using the device via the intended single-transseptal approach without CMC confirmation, this results in insufficient PVI, and may be impairing ablation success rates. In order to perform this additional CMC assessment of complete PVI, it must be advanced in to LA by catheter change, or dual-transseptal approach. These approaches increase complexity and prolong procedure and fluoroscopy times. Further, catheter change in the LA is suggested to be associated with microembolism. Therefore, the intended single-shot character of this device requires investigation in larger prospective trials with strict intention-to-treat designs.

Still despite establishing safety precautions, a high rate of esophageal thermal damage and atrio-esophageal fistulas were reported with the nMARQTM device, especially with the 2nd generation device. This is of major concern as these injuries often result in fatal outcomes. Following the re-launch of a new nMARQTM generation, further investigation into the safety of this device with respect to esophageal thermal damage is absolutely essential prior evaluating the clinical efficacy. In general, based on the early and limited experience with few severe complications associated with the nMARQTM device, a close FU of patients after PVI with all circular mapping devices should be aimed for.

In sum, following the idea of an easy-to-use and efficient ablation tool enabling fast and complete PVI, the nMARQTM catheter has proven feasibility, but still needs further evaluation in order to establish a reliable safety profile before aiming for superiority with respect to procedural and clinical variables in larger trials.

Acknowledgements

Dres. med. Siebermair and Wakili were supported by the German Centre for Cardiovascular Research (DZHK). Dr. Siebermair received by the FöFoLe-Programme of the University of Munich, Munich, Germany. Dr. Wakili received funding from the European Union’s Horizon 2020 research and innovation programme under grant agreement No 633193 (CATCH ME).

Conflict Of Interests

None.

Disclosures

None.

References

1. Haïssaguerre M, Jaïs P, Shah D C, Takahashi A, Hocini M, Quiniou G, Garrigue S, Le Mouroux A, Le Métayer P, Clémenty J. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N. Engl. J. Med. 1998 Sep 03;339 (10):659–66. [PubMed]
2. Calkins Hugh, Kuck Karl Heinz, Cappato Riccardo, Brugada Josep, Camm A John, Chen Shih-Ann, Crijns Harry J G, Damiano Ralph J, Davies D Wyn, DiMarco John, Edgerton James, Ellenbogen Kenneth, Ezekowitz Michael D, Haines David E, Haissaguerre Michel, Hindricks Gerhard, Iesaka Yoshito, Jackman Warren, Jalife José, Jais Pierre, Kalman Jonathan, Keane David, Kim Young-Hoon, Kirchhof Paulus, Klein George, Kottkamp Hans, Kumagai Koichiro, Lindsay Bruce D, Mansour Moussa, Marchlinski Francis E, McCarthy Patrick M, Mont J Lluis, Morady Fred, Nademanee Koonlawee, Nakagawa Hiroshi, Natale Andrea, Nattel Stanley, Packer Douglas L, Pappone Carlo, Prystowsky Eric, Raviele Antonio, Reddy Vivek, Ruskin Jeremy N, Shemin Richard J, Tsao Hsuan-Ming, Wilber David. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012 Apr;9 (4):632–696.e21. [PubMed]
3. Mahida Saagar, Hooks Darren A, Nentwich Karin, Ng G Andre, Grimaldi Massimo, Shin Dong-In, Derval Nicolas, Sacher Frederic, Berte Benjamin, Yamashita Seigo, Denis Arnaud, Hocini Mélèze, Deneke Thomas, Haissaguerre Michel, Jais Pierre. nMARQ Ablation for Atrial Fibrillation: Results from a Multicenter Study. J. Cardiovasc. Electrophysiol. 2015 Jul;26 (7):724–9. [PubMed]
4. Shin Dong-In, Kirmanoglou Kiriakos, Eickholt Christian, Schmidt Jan, Clasen Lukas, Butzbach Britta, Rassaf Tienush, Merx Marc, Kelm Malte, Meyer Christian. Initial results of using a novel irrigated multielectrode mapping and ablation catheter for pulmonary vein isolation. Heart Rhythm. 2014 Mar;11 (3):375–83. [PubMed]
5. Rillig Andreas, Lin Tina, Burchard Andre, Kamioka Masashi, Heeger Christian, Makimoto Hisaki, Metzner Andreas, Wissner Erik, Wohlmuth Peter, Ouyang Feifan, Kuck Karl-Heinz, Tilz Roland Richard. Modified energy settings are mandatory to minimize oesophageal injury using the novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. Europace. 2015 Mar;17 (3):396–402. [PubMed]
6. Zellerhoff Stephan, Daly Matthew, Lim Han S, Denis Arnaud, Komatsu Yuki, Jesel Laurence, Derval Nicolas, Sacher Frédéric, Cochet Hubert, Knecht Sébastien, Yiem Sunthareth, Hocini Mélèze, Haïssaguerre Michel, Jaïs Pierre. Pulmonary vein isolation using a circular, open irrigated mapping and ablation catheter (nMARQ): a report on feasibility and efficacy. Europace. 2014 Sep;16 (9):1296–303. [PubMed]
7. Scaglione Marco, Caponi Domenico, Anselmino Matteo, Di Clemente Francesca, Blandino Alessandro, Ferraris Federico, Di Donna Paolo, Ebrille Elisa, Halimi Franck, Leclercq Jean F, Iunco Costanza, Vaudagna Carloeugenio, Cesarani Federico, Gaita Fiorenzo. Pulmonary vein isolation with a new multipolar irrigated radiofrequency ablation catheter (nMARQ™): feasibility, acute and short-term efficacy, safety, and impact on postablation silent cerebral ischemia. J. Cardiovasc. Electrophysiol. 2014 Dec;25 (12):1299–305. [PubMed]
8. Deneke Thomas, Schade Anja, Müller Patrick, Schmitt Rainer, Christopoulos Georgios, Krug Joachim, Szöllösi Geza, Mügge Andreas, Kerber Sebastian, Nentwich Karin. Acute safety and efficacy of a novel multipolar irrigated radiofrequency ablation catheter for pulmonary vein isolation. J. Cardiovasc. Electrophysiol. 2014 Apr;25 (4):339–45. [PubMed]
9. Vurma Martin, Dang Lam, Brunner-La Rocca Hans-Peter, Sütsch Gabor, Attenhofer-Jost Christine H, Duru Firat, Scharf Christoph. Safety and efficacy of the nMARQ catheter for paroxysmal and persistent atrial fibrillation. Europace. 2016 Aug;18 (8):1164–9. [PubMed]
10. Arroja José David, Zimmermann Marc. Phrenic nerve lesion: a potential complication of the nMARQ ablation technique. Int. J. Cardiol. 2015 Feb 01;180 ():91–2. [PubMed]
11. Deneke Thomas, Schade Anja, Diegeler Anno, Nentwich Karin. Esophago-pericardial fistula complicating atrial fibrillation ablation using a novel irrigated radiofrequency multipolar ablation catheter. J. Cardiovasc. Electrophysiol. 2014 Apr;25 (4):442–3. [PubMed]
12. B Webster, B Webster. Voluntary field removal. Urgent field safety notice. Circulation. 2015;0:0–0.
13. Burri Haran, Park Chan-Il, Poku Nana, Giraudet Philippe, Stettler Carine, Zimmermann Marc. Pulmonary Vein Isolation for Paroxysmal Atrial Fibrillation Using a Circular Multipolar Ablation Catheter: Safety and Efficacy Using Low-Power Settings. J. Cardiovasc. Electrophysiol. 2016 Feb;27 (2):170–4. [PubMed]
14. Deneke Thomas, Müller Patrick, Halbfaß Philipp, Szöllösi Atilla, Roos Markus, Krug Joachim, Fochler Franziska, Schade Anja, Schmitt Rainer, Christopoulos Georgios, Mügge Andreas, Nentwich Karin. Effect of Different Ablation Settings on Acute Complications Using the Novel Irrigated Multipolar Radiofrequency Ablation Catheter (nMARQ). J. Cardiovasc. Electrophysiol. 2015 Oct;26 (10):1063–8. [PubMed]
15. Laish-Farkash Avishag, Khalameizer Vladimir, Fishman Evgeny, Cohen Ornit, Yosefy Chaim, Cohen Iris, Katz Amos. Safety, efficacy, and clinical applicability of pulmonary vein isolation with circular multi-electrode ablation systems: PVAC® vs. nMARQ™ for atrial fibrillation ablation. Europace. 2016 Jun;18 (6):807–14. [PubMed]
16. Rodríguez-Entem Felipe, Expósito Víctor, Rodríguez-Mañero Moisés, González-Enríquez Susana, Fernández-López Xesús Alberte, García-Seara Javier, Martínez-Sande José Luis, Olalla Juan José. Initial experience and treatment of atrial fibrillation using a novel irrigated multielectrode catheter: Results from a prospective two-center study. J Arrhythm. 2016 Apr;32 (2):95–101. [PMC free article] [PubMed]
17. Stabile Giuseppe, De Ruvo Ermenegildo, Grimaldi Massimo, Rovaris Giovanni, Soldati Ezio, Anselmino Matteo, Solimene Francesco, Iuliano Assunta, Sciarra Luigi, Schillaci Vincenzo, Bongiorni Maria Grazia, Gaita Fiorenzo, Calò Leonardo. Safety and efficacy of pulmonary vein isolation using a circular, open-irrigated mapping and ablation catheter: A multicenter registry. Heart Rhythm. 2015 Aug;12 (8):1782–8. [PubMed]
18. Wakili Reza, Siebermair Johannes, Fichtner Stephanie, Sinner Moritz F, Klocker Eva, Olesch Lucia, Hilberath Jan N, Sarai Samira, Clauss Sebastian, Sattler Stefan, Kääb Stefan, Estner Heidi L. One-year clinical outcome after ablation with a novel multipolar irrigated ablation catheter for treatment of atrial fibrillation: potential implications for clinical use. Europace. 2016 Aug;18 (8):1170–8. [PubMed]
19. K.H Kuck, J Brugada. Cryoballoon or Radiofrequency Ablation for Paroxysmal Atrial Fibrillation. (New England Journal of Medicine). 2016;0:2235–2245. [PubMed]
20. Packer Douglas L, Kowal Robert C, Wheelan Kevin R, Irwin James M, Champagne Jean, Guerra Peter G, Dubuc Marc, Reddy Vivek, Nelson Linda, Holcomb Richard G, Lehmann John W, Ruskin Jeremy N. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: first results of the North American Arctic Front (STOP AF) pivotal trial. J. Am. Coll. Cardiol. 2013 Apr 23;61 (16):1713–23. [PubMed]
21. Wieczorek Marcus, Hoeltgen Reinhard, Akin Elvan, Salili Ali Reza, Oral Hakan, Morady Fred. Results of short-term and long-term pulmonary vein isolation for paroxysmal atrial fibrillation using duty-cycled bipolar and unipolar radiofrequency energy. J. Cardiovasc. Electrophysiol. 2010 Apr;21 (4):399–405. [PubMed]
22. Calkins Hugh, Kuck Karl Heinz, Cappato Riccardo, Brugada Josep, Camm A John, Chen Shih-Ann, Crijns Harry J G, Damiano Ralph J, Davies D Wyn, DiMarco John, Edgerton James, Ellenbogen Kenneth, Ezekowitz Michael D, Haines David E, Haissaguerre Michel, Hindricks Gerhard, Iesaka Yoshito, Jackman Warren, Jalife José, Jais Pierre, Kalman Jonathan, Keane David, Kim Young-Hoon, Kirchhof Paulus, Klein George, Kottkamp Hans, Kumagai Koichiro, Lindsay Bruce D, Mansour Moussa, Marchlinski Francis E, McCarthy Patrick M, Mont J Lluis, Morady Fred, Nademanee Koonlawee, Nakagawa Hiroshi, Natale Andrea, Nattel Stanley, Packer Douglas L, Pappone Carlo, Prystowsky Eric, Raviele Antonio, Reddy Vivek, Ruskin Jeremy N, Shemin Richard J, Tsao Hsuan-Ming, Wilber David. 2012 HRS/EHRA/ECAS expert consensus statement on catheter and surgical ablation of atrial fibrillation: recommendations for patient selection, procedural techniques, patient management and follow-up, definitions, endpoints, and research trial design: a report of the Heart Rhythm Society (HRS) Task Force on Catheter and Surgical Ablation of Atrial Fibrillation. Developed in partnership with the European Heart Rhythm Association (EHRA), a registered branch of the European Society of Cardiology (ESC) and the European Cardiac Arrhythmia Society (ECAS); and in collaboration with the American College of Cardiology (ACC), American Heart Association (AHA), the Asia Pacific Heart Rhythm Society (APHRS), and the Society of Thoracic Surgeons (STS). Endorsed by the governing bodies of the American College of Cardiology Foundation, the American Heart Association, the European Cardiac Arrhythmia Society, the European Heart Rhythm Association, the Society of Thoracic Surgeons, the Asia Pacific Heart Rhythm Society, and the Heart Rhythm Society. Heart Rhythm. 2012 Apr;9 (4):632–696.e21. [PubMed]
23. Verma Atul, Jiang Chen-yang, Betts Timothy R, Chen Jian, Deisenhofer Isabel, Mantovan Roberto, Macle Laurent, Morillo Carlos A, Haverkamp Wilhelm, Weerasooriya Rukshen, Albenque Jean-Paul, Nardi Stefano, Menardi Endrj, Novak Paul, Sanders Prashanthan. Approaches to catheter ablation for persistent atrial fibrillation. N. Engl. J. Med. 2015 May 07;372 (19):1812–22. [PubMed]
24. Rosso Raphael, Halkin Amir, Michowitz Yoav, Belhassen Bernard, Glick Aharon, Viskin Sami. Radiofrequency ablation of paroxysmal atrial fibrillation with the new irrigated multipolar nMARQ ablation catheter: verification of intracardiac signals with a second circular mapping catheter. Heart Rhythm. 2014 Apr;11 (4):559–65. [PubMed]
25. Schmidt Martin, Dorwarth Uwe, Andresen Dietrich, Brachmann Johannes, Kuck Karl-Heinz, Kuniss Malte, Lewalter Thorsten, Spitzer Stefan, Willems Stephan, Senges Jochen, Jünger Claus, Hoffmann Ellen. Cryoballoon versus RF ablation in paroxysmal atrial fibrillation: results from the German Ablation Registry. J. Cardiovasc. Electrophysiol. 2014 Jan;25 (1):1–7. [PubMed]
26. De Greef Yves, Tavernier Rene, Schwagten Bruno, De Keulenaer Gilles, Stockman Dirk, Duytschaever Mattias. Impact of radiofrequency characteristics on acute pulmonary vein reconnection and clinical outcome after PVAC ablation. J. Cardiovasc. Electrophysiol. 2013 Mar;24 (3):290–6. [PubMed]
27. Andrade Jason G, Khairy Paul, Guerra Peter G, Deyell Marc W, Rivard Lena, Macle Laurent, Thibault Bernard, Talajic Mario, Roy Denis, Dubuc Marc. Efficacy and safety of cryoballoon ablation for atrial fibrillation: a systematic review of published studies. Heart Rhythm. 2011 Sep;8 (9):1444–51. [PubMed]
28. Wakili Reza, Voigt Niels, Kääb Stefan, Dobrev Dobromir, Nattel Stanley. Recent advances in the molecular pathophysiology of atrial fibrillation. J. Clin. Invest. 2011 Aug;121 (8):2955–68. [PMC free article] [PubMed]
29. Weiss Christian, Gocht Andreas, Willems Stephan, Hoffmann Matthias, Risius Tim, Meinertz Thomas. Impact of the distribution and structure of myocardium in the pulmonary veins for radiofrequency ablation of atrial fibrillation. Pacing Clin Electrophysiol. 2002 Sep;25 (9):1352–6. [PubMed]
30. Bordignon Stefano, Chun K R Julian, Gunawardene Melanie, Fuernkranz Alexander, Urban Verena, Schulte-Hahn Britta, Nowak Bernd, Schmidt Boris. Comparison of balloon catheter ablation technologies for pulmonary vein isolation: the laser versus cryo study. J. Cardiovasc. Electrophysiol. 2013 Sep;24 (9):987–94. [PubMed]
31. Neumann Thomas, Vogt Jürgen, Schumacher Burghard, Dorszewski Anja, Kuniss Malte, Neuser Hans, Kurzidim Klaus, Berkowitsch Alexander, Koller Marcus, Heintze Johannes, Scholz Ursula, Wetzel Ulrike, Schneider Michael A E, Horstkotte Dieter, Hamm Christian W, Pitschner Heinz-Friedrich. Circumferential pulmonary vein isolation with the cryoballoon technique results from a prospective 3-center study. J. Am. Coll. Cardiol. 2008 Jul 22;52 (4):273–8. [PubMed]
32. De Greef Y, Buysschaert I, Schwagten B, Stockman D, Tavernier R, Duytschaever M. Duty-cycled multi-electrode radiofrequency vs. conventional irrigated point-by-point radiofrequency ablation for recurrent atrial fibrillation: comparative 3-year data. Europace. 2014 Jun;16 (6):820–5. [PubMed]
33. Andrade Jason G, Dubuc Marc, Rivard Lena, Guerra Peter G, Mondesert Blandine, Macle Laurent, Thibault Bernard, Talajic Mario, Roy Denis, Khairy Paul. Efficacy and safety of atrial fibrillation ablation with phased radiofrequency energy and multielectrode catheters. Heart Rhythm. 2012 Feb;9 (2):289–96. [PubMed]
34. Klein Gunnar, Oswald Hanno, Gardiwal Ajmal, Lüsebrink Ulrich, Lissel Christoph, Yu Hong, Drexler Helmut. Efficacy of pulmonary vein isolation by cryoballoon ablation in patients with paroxysmal atrial fibrillation. Heart Rhythm. 2008 Jun;5 (6):802–6. [PubMed]
35. Kiss Alexandra, Nagy-Baló Edina, Sándorfi Gábor, Edes István, Csanádi Zoltán. Cerebral microembolization during atrial fibrillation ablation: comparison of different single-shot ablation techniques. Int. J. Cardiol. 2014 Jun 15;174 (2):276–81. [PubMed]
36. Gupta Aakriti, Perera Tharani, Ganesan Anand, Sullivan Thomas, Lau Dennis H, Roberts-Thomson Kurt C, Brooks Anthony G, Sanders Prashanthan. Complications of catheter ablation of atrial fibrillation: a systematic review. Circ Arrhythm Electrophysiol. 2013 Dec;6 (6):1082–8. [PubMed]
37. Knopp H, Halm U, Lamberts R, Knigge I, Zachäus M, Sommer P, Richter S, Bollmann A, Hindricks G, Husser D. Incidental and ablation-induced findings during upper gastrointestinal endoscopy in patients after ablation of atrial fibrillation: a retrospective study of 425 patients. Heart Rhythm. 2014 Apr;11 (4):574–8. [PubMed]
38. Cappato Riccardo, Calkins Hugh, Chen Shih-Ann, Davies Wyn, Iesaka Yoshito, Kalman Jonathan, Kim You-Ho, Klein George, Natale Andrea, Packer Douglas, Skanes Allan, Ambrogi Federico, Biganzoli Elia. Updated worldwide survey on the methods, efficacy, and safety of catheter ablation for human atrial fibrillation. Circ Arrhythm Electrophysiol. 2010 Feb;3 (1):32–8. [PubMed]
39. Deneke Thomas, Bünz Kathrin, Bastian Annely, Päsler Marcus, Anders Helge, Lehmann Rainer, Meuser Wolfgang, de Groot Joris R, Horlitz Marc, Haberkorn Ron, Mügge Andreas, Shin Dong-In. Utility of esophageal temperature monitoring during pulmonary vein isolation for atrial fibrillation using duty-cycled phased radiofrequency ablation. J. Cardiovasc. Electrophysiol. 2011 Mar;22 (3):255–61. [PubMed]
40. Tilz Roland R, Chun K R Julian, Metzner Andreas, Burchard Andre, Wissner Erik, Koektuerk Buelent, Konstantinidou Melanie, Nuyens Dieter, De Potter Tom, Neven Kars, Fürnkranz Alexander, Ouyang Feifan, Schmidt Boris. Unexpected high incidence of esophageal injury following pulmonary vein isolation using robotic navigation. J. Cardiovasc. Electrophysiol. 2010 Aug 01;21 (8):853–8. [PubMed]
41. Di Monaco Antonio, Quadrini Federico, Katsouras Grigorios, Caccavo Vincenzo, Troisi Federica, Quatraro Francesco, Cecere Giacomo, Langialonga Tommaso, Grimaldi Massimo. Ablation of atrial fibrillation and esophageal injury: Role of bipolar and unipolar energy using a novel multipolar irrigated ablation catheter. Heart Rhythm. 2015 Jun;12 (6):1120–7. [PubMed]
42. Halm Ulrich, Gaspar Thomas, Zachäus Markus, Sack Stephan, Arya Arash, Piorkowski Christopher, Knigge Ingrid, Hindricks Gerhard, Husser Daniela. Thermal esophageal lesions after radiofrequency catheter ablation of left atrial arrhythmias. Am. J. Gastroenterol. 2010 Mar;105 (3):551–6. [PubMed]
43. Singh Sheldon M, d'Avila Andre, Doshi Shephal K, Brugge William R, Bedford Rudolph A, Mela Theofanie, Ruskin Jeremy N, Reddy Vivek Y. Esophageal injury and temperature monitoring during atrial fibrillation ablation. Circ Arrhythm Electrophysiol. 2008 Aug;1 (3):162–8. [PubMed]
44. Knecht Sven, Sticherling Christian, Reichlin Tobias, Mühl Aline, Pavlović Nikola, Schaer Beat, Osswald Stefan, Kühne Michael. Reliability of luminal oesophageal temperature monitoring during radiofrequency ablation of atrial fibrillation: insights from probe visualization and oesophageal reconstruction using magnetic resonance imaging. Europace. 2016 Jun 29; () [PubMed]
45. Kuwahara Taishi, Takahashi Atsushi, Takahashi Yoshihide, Okubo Kenji, Takagi Katsumasa, Fujino Tadashi, Kusa Shigeki, Takigawa Masateru, Watari Yuji, Yamao Kazuya, Nakashima Emiko, Kawaguchi Naohiko, Hikita Hiroyuki, Sato Akira, Aonuma Kazutaka. Incidences of esophageal injury during esophageal temperature monitoring: a comparative study of a multi-thermocouple temperature probe and a deflectable temperature probe in atrial fibrillation ablation. J Interv Card Electrophysiol. 2014 Apr;39 (3):251–7. [PubMed]
46. Di Biase Luigi, Saenz Luis Carlos, Burkhardt David J, Vacca Miguel, Elayi Claude S, Barrett Conor D, Horton Rodney, Bai Rong, Siu Alan, Fahmy Tamer S, Patel Dimpi, Armaganijan Luciana, Wu Chia Tung, Kai Sonne, Ching Ching Keong, Phillips Karen, Schweikert Robert A, Cummings Jennifer E, Arruda Mauricio, Saliba Walid I, Dodig Milan, Natale Andrea. Esophageal capsule endoscopy after radiofrequency catheter ablation for atrial fibrillation: documented higher risk of luminal esophageal damage with general anesthesia as compared with conscious sedation. Circ Arrhythm Electrophysiol. 2009 Apr;2 (2):108–12. [PubMed]
47. Herrera Siklódy Claudia, Deneke Thomas, Hocini Mélèze, Lehrmann Heiko, Shin Dong-In, Miyazaki Shinsuke, Henschke Susanne, Fluegel Peter, Schiebeling-Römer Jochen, Bansmann Paul M, Bourdias Thomas, Dousset Vincent, Haïssaguerre Michel, Arentz Thomas. Incidence of asymptomatic intracranial embolic events after pulmonary vein isolation: comparison of different atrial fibrillation ablation technologies in a multicenter study. J. Am. Coll. Cardiol. 2011 Aug 09;58 (7):681–8. [PubMed]
48. F Gaita, J.F Leclercq. Incidence of silent cerebral thromboembolic lesions after atrial fibrillation ablation may change according to technology used: comparison of irrigated radiofrequency, multipolar nonirrigated catheter and cryoballoon. (J Cardiovasc Electrophysiol). 2011;0:961–8. [PubMed]
49. Gaita Fiorenzo, Caponi Domenico, Pianelli Martina, Scaglione Marco, Toso Elisabetta, Cesarani Federico, Boffano Carlo, Gandini Giovanni, Valentini Maria Consuelo, De Ponti Roberto, Halimi Franck, Leclercq Jean François. Radiofrequency catheter ablation of atrial fibrillation: a cause of silent thromboembolism? Magnetic resonance imaging assessment of cerebral thromboembolism in patients undergoing ablation of atrial fibrillation. Circulation. 2010 Oct 26;122 (17):1667–73. [PubMed]
50. Verma Atul, Debruyne Philippe, Nardi Stefano, Deneke Thomas, DeGreef Yves, Spitzer Stefan, Balzer Jörn O, Boersma Lucas. Evaluation and reduction of asymptomatic cerebral embolism in ablation of atrial fibrillation, but high prevalence of chronic silent infarction: results of the evaluation of reduction of asymptomatic cerebral embolism trial. Circ Arrhythm Electrophysiol. 2013 Oct;6 (5):835–42. [PubMed]
51. De Greef Yves, Dekker Lukas, Boersma Lucas, Murray Stephen, Wieczorek Marcus, Spitzer Stefan G, Davidson Neil, Furniss Steve, Hocini Mélèze, Geller J Christoph, Csanádi Zoltan. Low rate of asymptomatic cerebral embolism and improved procedural efficiency with the novel pulmonary vein ablation catheter GOLD: results of the PRECISION GOLD trial. Europace. 2016 May;18 (5):687–95. [PMC free article] [PubMed]
52. Schwarz Niko, Kuniss Malte, Nedelmann Max, Kaps Manfred, Bachmann Georg, Neumann Thomas, Pitschner Heinz-Friedrich, Gerriets Tibo. Neuropsychological decline after catheter ablation of atrial fibrillation. Heart Rhythm. 2010 Dec;7 (12):1761–7. [PubMed]
53. Sugihara Conn, Barlow Neil, Owens Emma, Sallomi David, Sulke Neil. Hyperacute And Chronic Changes In Cerebral Magnetic Resonance Images After Pvac, nmarq And Epicardial Thoracoscopic Surgical Ablation For Paroxysmal Atrial Fibrillation. J Atr Fibrillation. 2016 Dec 3;8 (6) [PMC free article] [PubMed]
54. Deneke Thomas, Jais Pierre, Scaglione Marco, Schmitt Rainer, DI Biase Luigi, Christopoulos Georgios, Schade Anja, Mügge Andreas, Bansmann Martin, Nentwich Karin, Müller Patrick, Krug Joachim, Roos Markus, Halbfass Phillip, Natale Andrea, Gaita Fiorenzo, Haines David. Silent cerebral events/lesions related to atrial fibrillation ablation: a clinical review. J. Cardiovasc. Electrophysiol. 2015 Apr;26 (4):455–63. [PubMed]
55. Deneke Thomas, Shin Dong-In, Balta Osman, Bünz Kathrin, Fassbender Frank, Mügge Andreas, Anders Helge, Horlitz Marc, Päsler Markus, Karthikapallil Sinthu, Arentz Thomas, Beyer Dieter, Bansmann Martin. Postablation asymptomatic cerebral lesions: long-term follow-up using magnetic resonance imaging. Heart Rhythm. 2011 Nov;8 (11):1705–11. [PubMed]
56. Gaita Fiorenzo, Corsinovi Laura, Anselmino Matteo, Raimondo Cristina, Pianelli Martina, Toso Elisabetta, Bergamasco Laura, Boffano Carlo, Valentini Maria Consuelo, Cesarani Federico, Scaglione Marco. Prevalence of silent cerebral ischemia in paroxysmal and persistent atrial fibrillation and correlation with cognitive function. J. Am. Coll. Cardiol. 2013 Nov 19;62 (21):1990–7. [PubMed]
57. Schmidt Boris, Gunawardene Melanie, Krieg Detlef, Bordignon Stefano, Fürnkranz Alexander, Kulikoglu Mehmet, Herrmann Wilfried, Chun K R Julian. A prospective randomized single-center study on the risk of asymptomatic cerebral lesions comparing irrigated radiofrequency current ablation with the cryoballoon and the laser balloon. J. Cardiovasc. Electrophysiol. 2013 Aug;24 (8):869–74. [PubMed]
58. Di Biase Luigi, Burkhardt J David, Santangeli Pasquale, Mohanty Prasant, Sanchez Javier E, Horton Rodney, Gallinghouse G Joseph, Themistoclakis Sakis, Rossillo Antonio, Lakkireddy Dhanunjaya, Reddy Madhu, Hao Steven, Hongo Richard, Beheiry Salwa, Zagrodzky Jason, Rong Bai, Mohanty Sanghamitra, Elayi Claude S, Forleo Giovanni, Pelargonio Gemma, Narducci Maria Lucia, Dello Russo Antonio, Casella Michela, Fassini Gaetano, Tondo Claudio, Schweikert Robert A, Natale Andrea. Periprocedural stroke and bleeding complications in patients undergoing catheter ablation of atrial fibrillation with different anticoagulation management: results from the Role of Coumadin in Preventing Thromboembolism in Atrial Fibrillation (AF) Patients Undergoing Catheter Ablation (COMPARE) randomized trial. Circulation. 2014 Jun 24;129 (25):2638–44. [PubMed]
59. Di Biase Luigi, Gaita Fiorenzo, Toso Elisabetta, Santangeli Pasquale, Mohanty Prasant, Rutledge Neal, Yan Xue, Mohanty Sanghamitra, Trivedi Chintan, Bai Rong, Price Justin, Horton Rodney, Gallinghouse G Joseph, Beheiry Salwa, Zagrodzky Jason, Canby Robert, Leclercq Jean François, Halimi Franck, Scaglione Marco, Cesarani Federico, Faletti Riccardo, Sanchez Javier, Burkhardt J David, Natale Andrea. Does periprocedural anticoagulation management of atrial fibrillation affect the prevalence of silent thromboembolic lesion detected by diffusion cerebral magnetic resonance imaging in patients undergoing radiofrequency atrial fibrillation ablation with open irrigated catheters? Results from a prospective multicenter study. Heart Rhythm. 2014 May;11 (5):791–8. [PubMed]
60. Csanadi Zoltan, Nagy-Baló Edina, Danik Stephan, Barrett Conor, Burkhardt J David, Sanchez Javier, Santangeli Pasquale, Santoro Francesco, Di Biase Luigi, Natale Andrea. Cerebrovascular Complications Related to Atrial Fibrillation Ablation and Strategies for Periprocedural Stroke Prevention. Card Electrophysiol Clin. 2014 Mar;6 (1):111–23. [PubMed]
61. Haines David E, Stewart Mark T, Dahlberg Sarah, Barka Noah D, Condie Cathy, Fiedler Gary R, Kirchhof Nicole A, Halimi Franck, Deneke Thomas. Microembolism and catheter ablation I: a comparison of irrigated radiofrequency and multielectrode-phased radiofrequency catheter ablation of pulmonary vein ostia. Circ Arrhythm Electrophysiol. 2013 Feb;6 (1):16–22. [PubMed]
62. Roka Attila, Heist E Kevin, Refaat Marwan, Ruskin Jeremy, Mansour Moussa. Novel Technique to Prevent Phrenic Nerve Injury During Pulmonary Vein Isolation Using Preprocedural Imaging. J. Cardiovasc. Electrophysiol. 2015 Oct;26 (10):1057–62. [PubMed]
63. Sacher Frédéric, Monahan Kristi H, Thomas Stuart P, Davidson Neil, Adragao Pedro, Sanders Prashanthan, Hocini Mélèze, Takahashi Yoshihide, Rotter Martin, Rostock Thomas, Hsu Li-Fern, Clémenty Jacques, Haïssaguerre Michel, Ross David L, Packer Douglas L, Jaïs Pierre. Phrenic nerve injury after atrial fibrillation catheter ablation: characterization and outcome in a multicenter study. J. Am. Coll. Cardiol. 2006 Jun 20;47 (12):2498–503. [PubMed]
64. Yong Ji Sang, Dewire Jane, Barcelon Bernadette, Philips Binu, Catanzaro John, Nazarian Saman, Cheng Alan, Spragg David, Tandri Harikrishna, Bansal Sandeep, Ashikaga Hiroshi, Rickard Jack, Kolandaivelu Aravindan, Sinha Sunil, Marine Joseph E, Calkins Hugh, Berger Ronald. Phrenic nerve injury: An underrecognized and potentially preventable complication of pulmonary vein isolation using a wide-area circumferential ablation approach. J. Cardiovasc. Electrophysiol. 2013 Oct;24 (10):1086–91. [PubMed]
65. Metzner Andreas, Rausch Peter, Lemes Christine, Reissmann Bruno, Bardyszewski Alexander, Tilz Roland, Rillig Andreas, Mathew Shibu, Deiss Sebastian, Kamioka Masashi, Toennis Tobias, Lin Tina, Ouyang Feifan, Kuck Karl-Heinz, Wissner Erik. The incidence of phrenic nerve injury during pulmonary vein isolation using the second-generation 28 mm cryoballoon. J. Cardiovasc. Electrophysiol. 2014 May;25 (5):466–70. [PubMed]
66. Jais P. Final Outcomes of REVOLUTION: a prospective, multcenter clinical study of pulmonary vein isolationusing the nMARQ circular and crescent multi-electrode irrigated ablation catheter, Presentation. Final Outcomes of REVOLUTION. 2014;0:0–0.
67. Bunch T Jared, Bruce G Keith, Mahapatra Srijoy, Johnson Susan B, Miller Dylan V, Sarabanda Alvaro V, Milton Mark A, Packer Douglas L. Mechanisms of phrenic nerve injury during radiofrequency ablation at the pulmonary vein orifice. J. Cardiovasc. Electrophysiol. 2005 Dec;16 (12):1318–25. [PubMed]
68. Swallow E B, Dayer M J, Oldfield W L, Moxham J, Polkey M I. Right hemi-diaphragm paralysis following cardiac radiofrequency ablation. Respir Med. 2006 Sep;100 (9):1657–9. [PubMed]

Articles from Journal of Atrial Fibrillation are provided here courtesy of CardioFront, LLC