Design of dual multiple aperture devices for dynamical fluence field modulated CT

Abstract

A Multiple Aperture Device (MAD) is a novel x-ray beam modulator that uses binary filtration on a fine scale to spatially modulate an x-ray beam. Using two MADs in series enables a large variety of fluence profiles by shifting the MADS relative to each other. This work details the design and control of dual MADs for a specific class of desired fluence patterns. Specifically, models of MAD operation are integrated into a best fit objective followed by CMA-ES optimization. To illustrate this framework we demonstrate the design process for an abdominal phantom with the goal of uniform detected signal. Achievable fluence profiles show good agreement with target fluence profiles, and the ability to flatten projections when a phantom is scanned is demonstrated. Simulated data reconstruction using traditional tube current modulation (TCM) and MAD filtering with TCM are investigated with the dual MAD system demonstrating more uniformity in noise and illustrating the potential for dose reduction under a maximum noise level constraint.

Keywords: Fluence field modulation, Radiation dose reduction, Dynamic bow-tie filter, Region-of-interest CT, X-ray beam modulation, Patient-specific CT

I. Introduction

X-ray computed tomography has found widespread clinical utility; however, increasing concerns about the risks associated with ionizing radiation have driven the search for exposure reduction strategies. While many algorithmic strategies for producing better images at lower exposures have been developed, there has been relatively little research on innovative hardware-based dose reduction methods. Dose to an individual patient is naturally tied to the particular exposure settings of a CT scanner; however, finding minimum dose strategies is both complex due to the dependence on patient size, anatomical site, etc. and, currently, somewhat limited due to the relative inflexibility of modern CT scanners to control the distribution of x-rays used to image a patient.

Typical clinical scanners permit coarse control of the x-ray beam through exposure settings (tube current and voltage), and many systems have tube current modulation hardware that permits variation of exposure as a function of rotation angle and table position. Control of the spatial distribution of the x-ray beam is typically very limited and is achieved through the introduction of a bow-tie filter. Some systems allow selection from a small number (typically three or fewer) bow-tie filters based on patient size. Typical filters attenuate x-rays at large fan angles to achieve higher fluence levels in the center of the patient (where the attenuation is highest) and lower fluence at the edges (where attenuation is low). Unfortunately, such static beam shaping is limited and cannot account for variability in the width/size of the patient as a function of angle and table position. Similarly, static bow-tie filters can be sensitive to positioning since a well-centered patient is presumed.

Fluence-field modulated (FFM) CT is an area of active research that seeks strategies for dynamic modulation of the spatial distribution of the x-ray beam. Successful implementation of FFM-CT increases acquisition flexibility permitting dose reduction objectives [1] as well as novel data collection strategies (e.g., region-of-interest scans). A number of different FFM strategies have been proposed including the use of heavy metal compounds on paper [2], digital beam attenuators [3], piece-wise linear dynamic bowties [4, 5], and fluid filled attenuators [6, 7].

Due to the severe operational requirements within a CT scanner (e.g. limited space, high rotation speeds, accelerations, etc.), the design of dynamic FFM-CT is a challenge. The authors of this paper have previously proposed a novel beam modulation strategy using Multiple Aperture Devices (MADs) which address some of the challenges associated with practical FFM-CT [8]. In this work, we present a strategy for MAD design to achieve specific desired dynamic fluence objectives.

II. Methods

A. The Multiple Aperture Device Concept

The conceptual operation of a MAD filter is illustrated in Figure 1. The device comprises thin bars of a highly attenuating material (e.g. tungsten) of varying widths and spacing. On a fine scale the MAD acts as a binary filter, either completely blocking or passing the X-ray beam using alternating bars and slots. In this fashion, one can concentrate the amount of x-rays spatially by varying the thickness of the blockers locally.

Fig. 1
Illustration of fluence modulation using Dual MAD filters.

As discussed in [8], the pitch (spacing between blockers) of the MAD device may be designed to minimize high-frequency patterns at the detector. For example, if the focal spot of the x-ray source is assumed to be a rectangle, the MAD pitch may be placed at the first null frequency associated with the focal spot blur MTF. In this fashion, the fine bar pattern of the MAD device is blurred out and is not visible at the detector. Desirable (lower frequency) spatial modulation associated with the variable bar width is still achievable.

A single fluence pattern can be obtained with a single MAD device. With multiple MADs in series, capable of moving with respect to each other, a range of fluence patterns can be obtained since it is the composition of two binary filters. Moreover, small relative displacement of the MADs with respect to each other can induce large changes in the fluence pattern. Because small actuations have a large effect on the x-ray distribution, speed and acceleration requirements can be reduced for device construction. Similarly, because these filters do not rely on variable attenuation using a low atomic number material, the filters can be made very thin and compact. In the following sections, we discuss how to design dual MAD filters for specific fluence objectives.

B. Initial Phantom Study and System Geometry

For initial investigations, we have concentrated on fluence optimization for a single target object; however, the approach may be extended to classes of objects. Specifically, the known object in the simulation study was chosen to be an anthropomorphic phantom body of uniform material (acrylic), as illustrated in Fig. 2. This digital phantom emulates commercially available physical phantoms (QRM GmbH, Morehendorf, Germany) that will be used in subsequent studies.

Fig. 2
Illustration of the simulated CT geometry and phantom.

The system geometry was chosen to emulate a CT scanner's source-to-detector distance and also geometry achievable in a flat-panel-based experimental test bench that is available for subsequent experiments. For our investigations, we considered 360 degree rotation, in steps of 0.5 degree. To create projection data for MAD design and analysis, we used a polyenergetic forward model and Spektr [9], a computational tool for x-ray spectral analysis, corresponding to a tube voltage of 100 kVp with additional filtration (2 mm of Al, 0.2 mm of Cu). The model also includes fluence adjustments to accommodate divergent beam effects.

C. Optimization Procedure

1) MAD parameterization

In order to design a set of MAD filters, the location and dimensions of many MAD attributes must be specified. The elements of a dual MAD design are identified in the illustration in Figure 3. Specifically, the free design parameters include: 1) b0(x), the thickness of each bar as a function of position in MAD0 that locally blocks X-rays; 2) b1(x), the analogous bar function for MAD1; 3) δ(x), a local offset function that specifies the position of individual bars in MAD1 relative to MAD0; and 4) the MAD pitch (e.g., the spacing interval between bars).

Fig. 3
Parameterization of the MAD design.

As mentioned in [8], the MAD pitch may be designed independently of other parameters based on the first null frequency of the focal spot, magnified to the MAD plane. For a rectangular focal spot size, fs, the optimal MAD pitch is

m=fs×(1SMDSDD)

We note that for nonrectangular focal spots, one can similarly find a null or minimal pass frequency to enforce smooth fluence profiles. Additionally, even though Fig. 3 shows MAD0 and MAD1 to be parallel with identical pitch, each of the flat MADs have a slightly different pitch and the bars/slots must be focused to the source due to the diverging x-ray beam.

The last parameter that is important for design is the control parameter Δ, which denotes the relative offset between MAD0 and MAD1. This is the one-dimensional actuation that controls the fluence profile enforced by the MAD filters. In general, this parameter must be part of the design process as well, and is a function of the CT rotation angle and/or table position, which we will denote as Δ(θ).

With MAD pitch specified, the remaining parameters: b0(x), b1(x), δ(x), and Δ(θ) are sought. In [8], these values determined analytically using an “endpoint” design to match two desired profiles by considering the minimum and maximum blocking conditions of a dual MAD system. While this approach is attractive due to its closed-form solution, it fails to provide best fit solutions for a wide range of desired fluence patterns. In this work, we seek that more optimal solution, which may stated as a nonlinear, nonconvex optimization (discussed in next section).

To facilitate optimization, we have chosen to further parameterize the dual MAD design using a low-dimensional set of basis functions. For example, rather than have a parameter for every bar width in MAD0, we presume neighboring bar widths vary smoothly as a function of position. Specifically, we chose to represent our parameters with a small set of Fourier coefficients, cp(ω) such that

p(x)=m(1+ep^(x))p^(x)=F1[cp(ω)]

where p(x) is one of {b0(x), b1(x), δ(x), or Δ(θ)}. Thus, the optimization will focus on finding the optimal coefficients: cb1(ω), cb2(ω), cδ(ω), and cΔ(ω) which are functions of the spatial (or, for Δ, angular) frequencies selected for the basis set.

2) Objective function

To define our optimization objective, we construct a model of the fluence output which is a function of the design and actuation values and can be written in terms of the original parameters or vectors of low-dimensional Fourier coefficients:

M(b0(x),b1(x),δ(x);Δ(θ))M(x,θ;cb1,cb2,cδ,cΔ)

Note that M is a function of spatial location (e.g., a fluence profile) as well as rotation angle.

Using this model, we pose the following optimization:

{c^b1,c^b2,c^δ,c^Δ}=argminθxϵPt(θ,x)t0(θ)M(x,θ;cb1,cb2,cδ,cΔ)M0(θ)2

where t(θ, x) denotes desired fluence patterns as a function of rotation angle. The objective is computed as the mean squared error between the desired and modeled fluence patterns over all projections that intersect the phantom (or patient). As such, X-rays passing outside the phantom (e.g. not contributing to dose) will be ignored in the optimization process.

Also note that both the modeled and desired fluence patterns are normalized by M0(θ) = ΣxM(x, θ;.) and t0(θ) = Σx t(θ, x), respectively. This normalization concentrates the design process on achieving the proper fluence shape. The magnitude of the profile can be adjusted post-design through exposure settings and tube current modulation. While there are many potential desired fluence patterns that one might seek including those that enforce minimum peak variance [10], combined noise and dose objectives[1], or maximize task-based detectability[11], we will focus on fluence patterns that flatten the signal and homogenize noise in projection data.

3) Optimization framework

Covariance Matrix Adaptation Evolution Strategy (CMA-ES) was chosen for the optimization. CMA-ES requires no derivative computations and is well-suited to nonlinear nonconvex optimization since a population of solutions is employed to avoid local optima. We implemented the objective function and the profile modeling function in efficient C++ code including parallelized computation of objective function values (over the population) using OpenMP. The CMA-ES algorithm was initialized to the output of the end-point design process from [8]. A population size of 16 was selected and the stopping criteria for optimization was to stop if successive function evaluations differ less than 10−12 or 10,000 iterations was computed. All the MAD design parameters were constrained to lie between 0 and a single pitch, m, to avoid nonphysical and periodic solutions (e.g. beyond a single cycle of actuation).

III. Results/Discussion

A. Target and achieved fluence profiles

To generate desired fluence patterns, the phantom was rotated 360 degrees in steps of 0.5 degrees. The fluence is simulated at the detector plane, and the fluence profile at the MAD plane required to flatten this fluence is computed using the methodology of Section IIB. The MAD parameterization used 8 Fourier coefficients for each MAD feature (4 total), and only symmetric basis functions were employed to enforce symmetric MAD designs. A subset of these target fluence profiles are shown in Fig. 4. The fluence obtainable with the dual MADs using the CMA-ES optimization is also plotted.

Fig. 4
Normalized target fluence patterns (red) at the MAD plane to flatten the fluence through the phantom at the detector plane, and achievable fluence patterns (blue) using the dual MAD setup. The fluence required is normalized such that the sum is unity. ...

The designed fluence profiles very closely match the desired fluence pattern suggesting that a dual MAD system can match a range of fluence profiles and, in this case, substantially flatten the fluence profiles at the detector for this phantom. For the fluence profiles with a flat top, the achievable CMA-ES profiles show fluctuation on the flat edge. The narrow fluence profiles show slight misalignment error, which is potentially correctable by shifting both MADs together. Such analysis is the topic of ongoing investigations and will likely be important for asymmetric beam profiles (e.g., for miscentered patients, off axis targets, etc.).

B. Estimated MAD design parameters

The optimized MAD design parameters are shown in Fig. 5. Note since the bar widths are directly proportional to the amount of local blockage and consequently inversely proportional to the local fluence, the effect of the MAD0 filter alone is not unlike a traditional bowtie (e.g. more fluence in the center of the field and less at the edges). The MAD1 design is almost the opposite (when acting alone). The bar widths in both MADs span the range of approximately 50 μm to 800 μm. Such designs are largely within the constraints of modern tungsten sintering technology, though features <100 μm can present some challenges (such constraints can potentially be integrated into the design process). The local offset function, δ(x), is predominantly negative, meaning that the MAD1 bars are located to the left of the center position in each MAD period.

Fig. 5
Solutions to the dual MAD design optimization. A) Bar width function, b0(x) for MAD0; B) barwidth function b1(x) for MAD1; C) local offset function δ(x); and D) the actuation control, Δ(θ), as a function of rotation angle.

The actuation control shown in Figure 5D illustrates that MAD1 is displaced between 0.15 mm to 0.4 mm as the projection angle changes from 0 to 360 degrees. This minimal movement of the MADs causes the large change in the fluence patterns seen in Figure 4 and can be attributed to the relatively small MAD pitch. From an implementation standpoint, the potential mechanical advantage is the fast switching speed of the MAD fluence profiles as the CT gantry spins around the patient. The smooth profile of the displacement also reduces the acceleration requirements on the actuator. Though not done here, one could integrate specific acceleration limits as part of the optimization.

C. Achievable Fluence Patterns

It is interesting to note that the design of the previous sections only utilizes part of the actuation control range. Fig. 6 shows the full range of fluence patterns achievable as the second MAD is moved with respect to the first MAD within a single MAD pitch (e.g. one cycle). Recall, that for the selected phantom, only fluence profiles between MAD1 displacements of 0.1 to 0.4 were used.

Fig. 6
Post-filtering fluence profiles at the MAD plane with a full range of control actuation (displacements of MAD1 with respect to MAD0).

However, from the fluence map, it is clear that much sharper fluence patterns can be obtained by changing the displacement to 0.7 mm. This potentially enables other applications such as region-of-interest fluence modulation and suggests additional design flexibility for larger classes of profiles (e.g. more complex objects, multiple classes, etc.).

D. Tube Current Modulation (TCM)

Although a variety of fluence patterns have been demonstrated, practical application and fitting to the desired fluence profiles requires proper scaling. This scaling can be achieved through tube current modulation (TCM). Typical Automatic Exposure Control (AEC) seeks to provide a constant fluence at the center of the detector. We have applied this strategy for the no filter scenario. For the MAD scenario, we applied the same strategy of providing constant fluence at the central detector pixel, through the Dual MAD and phantom. For comparison between the no filter and MAD filtered scenarios we have ensured that the total fluence (i.e., the number of simulated photons) incident on the phantom is constant for the two approaches. Specifically, TCM is scaled to enforce a total of 100,000 photons incident on the phantom.

Fig. 7 shows the TCM required to convert the fluence generated by the MADs to the required target fluence. Without the MAD filter, the TCM is largest when the path length of X-rays through the phantom is largest. The dual MAD filter has maximum attenuation when the fluence profile is narrow. Therefore, more photons are required at 0 and 180 degrees to flatten the fluence with MAD than at 90 or 270 degrees. The MAD requires higher scaling and modulation to generate the same number of photons incident on the phantom.

Fig. 7
Illustration of tube current modulation with and without MAD filters.

E. Simulated Projection Data

Figure 8 and and99 show the fluence profiles with and without the phantom in the field of view for the no filter and MAD filtered scenarios (TCM is used in both cases). In Fig. 8, we see that the no filter scenario can only modulate the per view number of photons through TCM while the dual MAD filter can customize both the shape and intensity of the beam. In Fig. 9, the post-object fluence is more uniform across object projections (the design goal) than the no filter, TCM-only scenario.

Fig. 8
Illustration of the fluence profiles received at the detector with no phantom in the scanner using A) no filters and AEC, and B) using optimally actuated and designed dual MAD filters and AEC. The space occupied by phantom is shown with a dotted red line. ...
Fig. 9
Illustration of the projection data received at the detector with the phantom in the scanner using A) no filters and AEC, and B) using optimally actuated and designed dual MAD filters and AEC. The space occupied by phantom is shown with a dotted red line. ...

F. Simulated Reconstructions

With Poisson noise added to the projection data in Fig. 9, filtered backprojection reconstructions were performed for both filtering scenarios. Results are shown in Figure 10. Both methods show approximately the same average noise level (as expected due to an equal number of incident photons). However, we see much greater noise uniformity in the MAD filtered image. This is significant if a minimum noise level is prescribed to obtain sufficient image quality. The TCM-only case will require more incident photons (hence larger dose) to obtain the same minimum noise level over the entire image.

Fig. 10
A) Filtered backprojection of phantom with no MAD filters and Dual MAD filters. B) Noise image with no MAD filter and Dual MAD filter

G. Ongoing and Future Work

While these initial results suggest that dual MAD filters can successfully achieve a broad class of fluence patterns, we are seeking to extend this work to even larger classes of fluence patterns (e.g. different size patients). Similarly, parallel efforts are working to fabricate physical MAD devices and evaluate performance in an experimental CT system.

Acknowledgments

This work supported, in part, by NIH grant 5U01EB018758.

References

[1] Bartolac S, et al. Fluence field optimization for noise and dose objectives in CT. Med. Phys. 2011;38(2011):S2. [PubMed]
[2] Hasegawa BH, et al. Selective Exposure Radiography using digitally-formed x-ray beam attenuators. Proc. SPIE. 1983;0419:282–288.
[3] Szczykutowicz TP, Mistretta CA. Design of a digital beam attenuation system for computed tomography. Part II. Performance study and initial results. Med. Phys. 2013;40(2013):021906. [PubMed]
[4] Hsieh SS, Pelc NJ. The feasibility of a piecewise-linear dynamic bowtie filter. Med. Phys. 2013;40(3):031910. [PubMed]
[5] Liu F, et al. Dynamic bowtie filter for cone-beam/multi-slice CT. PLoS One. 2014;9(7):1–10. [PMC free article] [PubMed]
[6] Shunhavanich P, Hsieh SS, Pelc NJ. Fluid-filled dynamic bowtie filter: a feasibility study. Proceedings of SPIE. 2015;9412:94121L.
[7] Szczykutowicz TP, Hermus J. Fluid Dynamic Bowtie Attenuators. Proc. SPIE. 2015;9412:1–7.
[8] Webster Stayman J, et al. Fluence-Field Modulated X-ray CT using Multiple Aperture Devices. Proc. SPIE. 2016 [PMC free article] [PubMed]
[9] Siewerdsen JH, et al. Spektr: A computational tool for x-ray spectral analysis and imaging system optimization. Med. Phys. 2004;31(11):3057. [PubMed]
[10] Hsieh SS, Pelc NJ. Algorithms for optimizing CT fluence control. Proc. SPIE. 2014;9033:90330M.
[11] Gang GJ, et al. Task-driven tube current modulation and regularization design in computed tomography with penalized-likelihood reconstruction. SPIE Medical Imaging. 2016 [PMC free article] [PubMed]