Use of Caval Subtraction 2D Phase-Contrast MR Imaging to Measure Total Liver and Hepatic Arterial Blood Flow: Preclinical Validation and Initial Clinical Translation

Manil D. Chouhan, MRCS, FRCR, PhD, Rajeshwar P. Mookerjee, FRCP, PhD, Alan Bainbridge, PhD, Simon Walker-Samuel, PhD, Nathan Davies, PhD, Steve Halligan, MRCP, FRCR, MD, Mark F. Lythgoe, PhD, and Stuart A. Taylor, MRCP, FRCR, MDcorresponding author

Abstract

Caval subtraction phase-contrast MR imaging is technically feasible and may offer a reproducible and clinically viable method for measuring total liver blood flow and hepatic arterial flow.

Abstract

Purpose

To validate caval subtraction two-dimensional (2D) phase-contrast magnetic resonance (MR) imaging measurements of total liver blood flow (TLBF) and hepatic arterial fraction in an animal model and evaluate consistency and reproducibility in humans.

Materials and Methods

Approval from the institutional ethical committee for animal care and research ethics was obtained. Fifteen Sprague-Dawley rats underwent 2D phase-contrast MR imaging of the portal vein (PV) and infrahepatic and suprahepatic inferior vena cava (IVC). TLBF and hepatic arterial flow were estimated by subtracting infrahepatic from suprahepatic IVC flow and PV flow from estimated TLBF, respectively. Direct PV transit-time ultrasonography (US) and fluorescent microsphere measurements of hepatic arterial fraction were the standards of reference. Thereafter, consistency of caval subtraction phase-contrast MR imaging–derived TLBF and hepatic arterial flow was assessed in 13 volunteers (mean age, 28.3 years ± 1.4) against directly measured phase-contrast MR imaging PV and proper hepatic arterial inflow; reproducibility was measured after 7 days. Bland-Altman analysis of agreement and coefficient of variation comparisons were undertaken.

Results

There was good agreement between PV flow measured with phase-contrast MR imaging and that measured with transit-time US (mean difference, −3.5 mL/min/100 g; 95% limits of agreement [LOA], ±61.3 mL/min/100 g). Hepatic arterial fraction obtained with caval subtraction agreed well with those with fluorescent microspheres (mean difference, 4.2%; 95% LOA, ±20.5%). Good consistency was demonstrated between TLBF in humans measured with caval subtraction and direct inflow phase-contrast MR imaging (mean difference, −1.3 mL/min/100 g; 95% LOA, ±23.1 mL/min/100 g). TLBF reproducibility at 7 days was similar between the two methods (95% LOA, ±31.6 mL/min/100 g vs ±29.6 mL/min/100 g).

Conclusion

Caval subtraction phase-contrast MR imaging is a simple and clinically viable method for measuring TLBF and hepatic arterial flow.

Online supplemental material is available for this article.

Introduction

The dual portal venous (PV) and hepatic arterial blood supply to the liver makes it difficult to assess hemodynamic complications of liver disease. To date, invasive methods still remain the standard of reference. Imaging-based hepatic hemodynamic assessment has the potential to yield useful and meaningful biomarkers of portal hypertension and chronic liver disease (1).

Two-dimensional (2D) phase-contrast magnetic resonance (MR) imaging is an established, widely available, and validated method for the noninvasive measurement of large-vessel bulk flow (25). Several investigators have used phase-contrast MR imaging to study PV portal vein flow, although few have reported data from the proper hepatic artery (69) because evaluation is challenging in routine clinical practice, primarily because of small vessel size, tortuosity, and anatomic variation. Furthermore, low signal-to-noise ratio, vessel orthogonality, partial volume averaging errors, intravoxel phase dispersion, and spatial misregistration can confound estimation of pulsatile flow in small arteries and have impeded both research and clinical use of phase-contrast MR imaging for the measurement of total liver blood flow (TLBF).

Recognizing these challenges, we propose an alternative application of 2D two-dimensional phase-contrast MR imaging to measure TLBF total liver blood flow and proper hepatic arterial flow by using caval subtraction. TLBF total liver blood flow and hepatic arterial flow fraction could be valuable hemodynamic biomarkers of liver disease. The purpose of this study was to validate caval subtraction 2D two-dimensional phase-contrast MR imaging measurements of TLBF total liver blood flow and hepatic arterial fraction in a preclinical animal model and evaluate consistency and reproducibility in humans.

Materials and Methods

Background Theory

From the application of the principle of conservation of mass to flow (Q), for a fixed tissue volume:

equation image

where Qin is blood flow into the organ and Qout is blood flow out of the organ. On the basis of the anatomic configuration of the liver, the intrahepatic inferior vena cava (IVC) receives blood entirely from the hepatic venous system. Outflow TLBF total liver blood flow (Qout, equivalent to QTLBF) can therefore be estimated by measuring bulk flow in the suprahepatic subcardiac portion of the IVC inferior vena cava (Qsuprahepatic) and then subtracting flow from the infrahepatic, suprarenal portion of the IVC inferior vena cava (Qinfrahepatic) (Fig E1a [online]), as follows:

equation image

Hepatic arterial flow (QHA) can then be estimated by subtracting directly measured PV portal vein flow (QPV) from the outflow TLBF total liver blood flow (Fig E1b [online], as follows:

equation image

Preclinical Validation

Subjects and preparation.—All experiments were conducted according to the home office guidelines under the U.K. Animals in Scientific Procedures Act (1986) after approval from the Animal Care Ethical Committee of University College London. Experiments were performed between September 12 and October 17, 2013, on healthy male Sprague-Dawley rats (Charles River UK, Margate, England) with normal liver function (weight, 250–300 g).

The study cohort consisted of 15 healthy animals treated with sham laparotomy (n = 11) or bile duct ligation (BDL) (n = 4). The latter group was included to test the feasibility of performing phase-contrast MR imaging in an animal model of portal hypertension. BDL bile duct ligation and sham surgery were conducted as described previously (10). After recovery, animals were maintained for 4–5 weeks before undergoing the experimental protocol. All procedures were performed by the study coordinator (M.D.C., a radiology research fellow qualified in animal handling with 4 years of experience).

Standards of reference.—For rats that underwent sham operation (n = 11), laparotomy was performed and a 2-mm transit-time US probe (Transonic Systems, Ithaca, NY) was placed around the PV portal vein. PV portal vein flow readings were obtained with transit-time US after 10–15 minutes, once the animal was stable. Extensive adhesions around the porta hepatis and associated high risk of traumatic vessel injury precluded transit-time US validation in the BDL bile duct ligation group.

After transit-time US measurement, 15-μm polystyrene fluorescent microspheres (FluoSpheres; Life Technologies, Warrington, England) suspended in heparinized saline were administered transcutaneously into the left ventricle over approximately 10 seconds under US guidance (Terason; Teratech, Burlington, Mass) (Fig E2 [online]). Animals were then transferred to the imager for phase-contrast MR imaging. The animal was then sacrificed and organs were explanted for microsphere processing using an adapted protocol (11), as summarized in Appendix E1 (online). To ensure adequate central mixing of microspheres, data exceeding 20% difference in microsphere content between right and left kidneys were excluded.

Two-dimensional cine phase-contrast MR imaging.—Temperature was monitored with a rectal probe (SA Instruments, New York, NY), with core body temperature maintained between 36°C and 38°C. Cardiac monitoring was undertaken by using a triple-electrode single-lead system (SA Instruments). Imaging was performed by using a 9.4-T unit (Agilent Technologies, Oxford, England), with sequence parameters listed in the Table. Axial and angled coronal gradient-echo images were used to identify the IVC inferior vena cava and PV portal vein, and phase-contrast image planes were planned to ensure orthogonality with the subcardiac portion of the suprahepatic IVC inferior vena cava, suprarenal portion of the infrahepatic IVC inferior vena cava, and PV portal vein.

Table thumbnail
Sequence Parameters

Cardiac- and respiratory-gated 2D two-dimensional cine phase-contrast MR imaging was performed with 2-mm-thick sections, a 10° flip angle, and a 192 × 192 (frequency encoding × phase encoding) acquisition matrix. On the basis of initial pilot work in four animals (not presented in this study), velocity encoding (Venc) settings of 33 cm/sec for PV portal vein and infrahepatic IVC inferior vena cava flows and 66 cm/sec for suprahepatic IVC inferior vena cava flows were applied (Table). Phase maps were acquired at each Venc velocity encoding setting with opposite flow-encoding directions. Correction for background phase errors was achieved by subtracting phase maps with opposing flow-encoding directions, with the assumption that the phase of stationary spins was identical in each image. Acquisition time for each phase-contrast MR imaging measurement was usually less than 10 minutes. Regions of interest were positioned manually by the study coordinator on each vessel for each frame of the cardiac cycle, and flow quantification was performed by using in-house-developed Matlab code (MathWorks, Natick, Mass) on the basis of established algorithms (12). All PV portal vein flow, estimated TLBF total liver blood flow, and hepatic arterial flow measurements were normalized to explanted liver weight.

Human Translation

Two aspects of the caval subtraction technique were then tested in healthy volunteers: (a) consistency with directly measured phase-contrast MR imaging PV portal vein and hepatic arterial inflow and (b) measurement reproducibility after 7 days.

Subjects and preparation.—University College London Ethics Committee approval was obtained, and participants provided informed written consent. Volunteers were recruited by means of advertisement within the university campus and were eligible if they (a) had no contraindications to MR imaging, (b) were not taking any long-term medication (excluding the oral contraceptive pill), and (c) had no documented history of previous liver or gastrointestinal disease. Fourteen volunteers were screened, one of whom was excluded because of claustrophobia. The final cohort consisted of seven men (mean age ± standard deviation, 26.5 years ± 1.36) and six women (mean age, 31.2 years ± 2.62) who underwent imaging between June 14 and July 25, 2013. Participants fasted for 6 hours before MR imaging and avoided caffeinated fluids.

Two-dimensional cine phase-contrast MR imaging.—Phase-contrast MR imaging was performed with a 3.0-T unit (Achieva; Philips Healthcare, Best, the Netherlands) and a 16-channel body coil (SENSE XL-Torso, Philips Healthcare). Imaging parameters are given in the Table.

Coronal (upper abdomen), sagittal (abdominal great vessels), and oblique (PV portal vein) breath-hold balanced steady-state free precession images were acquired. Two-dimensional phase-contrast MR imaging with expiratory breath hold and retrospective cardiac gating was planned in two planes by the study coordinator to ensure orthogonality to the target vessel. Studies were performed through the PV portal vein (Venc velocity encoding = 40 cm/sec), proper hepatic artery (Venc velocity encoding = 60 cm/sec), infrahepatic IVC inferior vena cava (above the renal veins, below the hepatic IVC inferior vena cava; Venc velocity encoding = 60 cm/sec), and suprahepatic IVC inferior vena cava (above the hepatic venous inflow, below the right atrial junction; Venc velocity encoding = 80 cm/sec). Where hepatic arterial anatomy varied (n = 2), measurements were made as close as possible to the origin of the hepatic artery. Images were reviewed for aliasing and Venc velocity encoding settings increased by 20 cm/sec when appropriate. Data were acquired by using the unit’s clinical flow quantification implementation. Phase maps were acquired at each Venc velocity encoding setting with opposite flow-encoding directions. Correction for background phase errors was achieved by subtracting phase maps with opposing flow-encoding directions, with the assumption that the phase of stationary spins was identical in each image. A local phase-correction filter was also applied to correct for phase errors induced by eddy currents. The acquisition time for each measurement was less than 20 seconds. Each phase-contrast MR imaging study was repeated three times. Flow quantification was performed by using freely available software (Segment; Medviso, Lund, Sweden) and the mean of triplicate measurements used for analysis. Caval subtraction TLBF total liver blood flow, PV portal vein flow, hepatic arterial flow, and hepatic arterial fraction were calculated (Eqq [2, 3]) and compared with direct phase-contrast MR imaging of PV portal vein and hepatic arterial inflow.

Liver volume was estimated by using steady-state free precession coronal images with 5-mm-thick sections. Segmentation was performed manually by the study coordinator using software (Amira Resolve RT; Visage Imaging, Berlin, Germany). A tissue density of 1.0 g/mL was assumed (13).

Seven days after the original study, subjects underwent repeat imaging with identical preparation and MR imaging protocol at a comparable time of the day. All analyses and quantification were performed by the study coordinator (M.D.C., with 5 years of experience in abdominal imaging).

Statistical Analysis

Data normality was confirmed with Kolmogorov-Smirnov testing. All bulk flow measurements obtained at phase-contrast MR imaging were normalized to liver weight or volume. Comparison between measurements derived from caval subtraction phase-contrast MR imaging and standards of reference (transit-time US, microspheres, direct inflow phase-contrast MR imaging) and 7-day reproducibility studies were assessed by using Bland-Altman analysis of agreement, with calculation of 95% limits of agreement (LOA). Coefficients of variation were also calculated and compared by using methods described by Forkman (14). Because of the small number of animals in the sham and BDL bile duct ligation groups that underwent validation with microsphere analysis, validation analysis was pooled across both cohorts and the Mann-Whitney U test used for comparison of PV portal vein and relative hepatic arterial flow. Data are expressed as means ± standard errors, and P < .05 was indicative of a statistically significant difference.

Results

Preclinical Cohort

Across 15 animals, the mean body weight was 451.7 g ± 9.0 and the wet liver mass was 24.7 g ± 2.0. All animals in the BDL bile duct ligation group had evidence of cirrhosis at histopathologic examination.

Technical feasibility of caval subtraction phase-contrast MR imaging.—Electrocardiographically and respiratory-gated cine phase-contrast MR imaging flow studies through the cardiac cycle demonstrated physiologic flow profiles through the PV portal vein and infrahepatic and suprahepatic IVC inferior vena cava (Fig E3 [online]).

Validation of PV and relative hepatic arterial flow.—Comparing phase-contrast MR imaging versus transit-time US (n = 11), the mean difference between PV portal vein flow directly measured with phase-contrast MR imaging (mean, 182.8 mL/min/100 g ± 9.9) and transit-time US (mean, 186.3 mL/min/100 g ± 11.9) was −3.5 mL/min/100 g ± 9.4, with 95% LOA limits of agreement of ±61.3 mL/min/100 g (Fig 1a). The coefficient of variation for PV flow was similar for transit-time US (21.2%) compared with phase-contrast MR imaging (17.9%) (F10,10 = 1.38, P = .31).

Figure 1a:
Validation of phase-contrast MR imaging PV portal vein flow and caval subtraction phase-contrast MR imaging–derived hepatic arterial (HA) fraction in sham-operated (■) and BDL bile duct ligation (▲) rats. (a, c) Bland-Altman plots ...
Figure 1b:
Validation of phase-contrast MR imaging PV portal vein flow and caval subtraction phase-contrast MR imaging–derived hepatic arterial (HA) fraction in sham-operated (■) and BDL bile duct ligation (▲) rats. (a, c) Bland-Altman plots ...
Figure 1d:
Validation of phase-contrast MR imaging PV portal vein flow and caval subtraction phase-contrast MR imaging–derived hepatic arterial (HA) fraction in sham-operated (■) and BDL bile duct ligation (▲) rats. (a, c) Bland-Altman plots ...

In the comparison of phase-contrast MR imaging with microspheres, seven animals had inadequate central mixing of microspheres and were excluded, leaving eight for analysis (sham, n = 4; BDL, n = 4). The mean difference between relative hepatic arterial flow derived from phase-contrast MR imaging with caval subtraction (Eq [3]; mean, 35.3% ± 11.6) and that calculated from the microsphere distribution analysis (mean, 31.1% ± 9.8) was 4.2% ± 3.7, with 95% LOA limits of agreement of ±20.5% (Fig 1c). The coefficient of variation was similar for caval subtraction phase-contrast MR imaging (93.0%) and microsphere analysis (89.2%) (F7,7 = 0.95, P = .52). Relative hepatic arterial flow was greater and PV portal vein flow was lower in animals with cirrhosis and portal hypertension (relative hepatic arterial flow: 50.3% ± 13.5 in BDL bile duct ligation group vs 11.8% ± 4.3 in sham group; PV portal vein flow: 94.3 mL/min/100 g ± 28.8 in BDL bile duct ligation group vs 167.0 mL/min/100 g ± 20.3 in sham group), but these differences were not statistically significant (hepatic arterial fraction, P = .0571; PV portal vein flow, P = .200).

Figure 1c:
Validation of phase-contrast MR imaging PV portal vein flow and caval subtraction phase-contrast MR imaging–derived hepatic arterial (HA) fraction in sham-operated (■) and BDL bile duct ligation (▲) rats. (a, c) Bland-Altman plots ...

Clinical Cohort

The mean liver volume was 1211.0 mL ± 52.9. Two subjects declined subsequent examination, leaving 11 in the reproducibility cohort.

Technical feasibility of caval subtraction phase-contrast MR imaging.—Electrocardiographically and respiratory-gated cine phase-contrast MR imaging flow studies through the cardiac cycle demonstrated physiologic flow profiles through the PV portal vein, infrahepatic IVC inferior vena cava, and suprahepatic IVC inferior vena cava (Fig E4 [online]).

Caval subtraction phase-contrast MR imaging versus direct phase-contrast MR imaging.—The mean difference between TLBF total liver blood flow measured with caval subtraction phase-contrast MR imaging (mean, 71.1 mL/min/100 g ± 3.3) and that measured with direct phase-contrast MR imaging (sum of PV portal vein and common hepatic arterial flow; mean, 72.5 mL/min/100 g ± 3.3) and between calculated hepatic arterial flow (Eq [3]; mean, 14.0 mL/min/100 g ± 3.5) and direct phase-contrast MR imaging measured hepatic arterial flow (mean, 15.3 mL/min/100 g ± 2.4) was −1.3 mL/min/100 g ± 2.4. The 95% LOA limits of agreement for caval subtraction versus direct inflow phase-contrast MR imaging were ±23.1 mL/min/100 g for both TLBF total liver blood flow and hepatic arterial flow (range, 40.5–100.9 mL/min/100 g and −20.7 to 54.9 mL/min/100 g, respectively) (Fig 2a, ,2c).2c). The coefficient of variation for caval subtraction phase-contrast MR imaging TLBF total liver blood flow (23.0%) was similar to that for direct inflow TLBF total liver blood flow (22.2%) (F23,23 = 0.94, P = .56). Hepatic arterial flow with caval subtraction phase-contrast MR imaging (123.2%) was higher than that with direct phase-contrast MR imaging (75.6%), although this did not reach statistical significance (F23,23 = 1.68, P = .11).

Figure 2a:
Consistency of caval subtraction phase-contrast MR imaging. Caval subtraction phase-contrast MR imaging TLBF total liver blood flow and hepatic arterial flow in healthy volunteers were compared with contemporaneous inflow phase-contrast MR imaging measurements. Data ...
Figure 2c:
Consistency of caval subtraction phase-contrast MR imaging. Caval subtraction phase-contrast MR imaging TLBF total liver blood flow and hepatic arterial flow in healthy volunteers were compared with contemporaneous inflow phase-contrast MR imaging measurements. Data ...
Figure 2b:
Consistency of caval subtraction phase-contrast MR imaging. Caval subtraction phase-contrast MR imaging TLBF total liver blood flow and hepatic arterial flow in healthy volunteers were compared with contemporaneous inflow phase-contrast MR imaging measurements. Data ...
Figure 2d:
Consistency of caval subtraction phase-contrast MR imaging. Caval subtraction phase-contrast MR imaging TLBF total liver blood flow and hepatic arterial flow in healthy volunteers were compared with contemporaneous inflow phase-contrast MR imaging measurements. Data ...

Seven-day reproducibility.—The mean differences between TLBF total liver blood flow measured with caval subtraction phase-contrast MR imaging and hepatic arterial flow measurements obtained 7 days apart were −8.5 mL/min/100 g ± 4.9 and 7.3 mL/min/100 g ± 4.4, respectively. The 95% LOA limits of agreement were ±31.6 mL/min/100 g for TLBF total liver blood flow and ±28.8 mL/min/100 g for hepatic arterial flow. The mean difference between directly measured phase-contrast MR imaging TLBF total liver blood flow and hepatic arterial flow measurements obtained 7 days apart were −2.3 mL/min/100 g ± 4.5 and 1.1 mL/min/100 g ± 3.0, respectively. The caval subtraction 95% LOA limits of agreement were ±31.6 mL/min/100 g for TLBF total liver blood flow and ±28.8 mL/min/100 g for hepatic arterial flow. The 95% LOA limits of agreement were ±29.6 mL/min/100 g for TLBF total liver blood flow and ±19.5 mL/min/100 g (range, 4.9–44.6 mL/min/100 g) for hepatic arterial flow.

Discussion

Accurate assessment of proper hepatic arterial flow (and, consequently, TLBF total liver blood flow) with use of phase-contrast MR imaging is challenging in both preclinical and clinical contexts. In rodents, the vessel itself is extremely small—less than 1 mm in diameter (1517)—which introduces measurement errors, even at high field strengths. Technical challenges have confounded attempts to measure proper hepatic arterial flow in humans with clinical systems (69). Adequate signal-to-noise ratio is less of an issue because the vessel is larger, but partial voluming and spatial resolution remain problematic, particularly at 1.5 T (7). Furthermore, frequent anatomic variation (1820) complicates measurement and requires costly radiologic expertise for planning, which substantially impedes clinical implementation.

We have demonstrated that caval subtraction phase-contrast MR imaging provides a relatively simple strategy for overcoming these limitations in both rodents and humans. We found that phase-contrast MR imaging is technically feasible in animals (even in a cirrhotic BDL bile duct ligation model), and phase-contrast MR imaging PV portal vein flow estimates showed encouraging agreement with directly measured transit-time US flow (21). Thereafter, with the caval subtraction technique, reasonable agreement was obtained for hepatic arterial fraction against a microsphere standard of reference. Unfortunately, quantification of absolute hepatic arterial flow was not undertaken because simultaneous peripheral arterial sampling proved unreliable in pilot experiments.

For human translation, we chose not to use transcutaneous Doppler US measurements of PV portal vein and hepatic arterial flow because reported reproducibility is poor (8,2224). Instead, direct phase-contrast MR imaging measurements of PV portal vein and hepatic arterial flow were used to test consistency of the caval subtraction phase-contrast MR imaging technique. Good agreement for TLBF total liver blood flow was demonstrated between the two methods, and although hepatic arterial flow agreement was less impressive, the level of disagreement was not contingent on measurement value (ie, there was no systematic bias). Hepatic arterial flow measurements suffer from error propagation as inaccuracies in IVC inferior vena cava and PV portal vein flow are summated during subtraction. This can result in nonphysiologic results such as negative estimates of hepatic arterial flow (particularly when true hepatic arterial flow is low, as would be expected in healthy volunteers).

Measurements obtained with caval subtraction phase-contrast MR imaging were, however, reassuringly similar to the direct measurements obtained with inflow phase-contrast MR imaging, although variable for both techniques, likely due to natural variation in vessel flow rates contingent on differing subject hydration, for example.

To place the levels of agreement for caval subtraction technique with standards of reference into clinical context, it is known that TLBF total liver blood flow can vary by as much as 58% between health and disease (16,25,26), which is much greater than the expected error range found in the current study.

Our study has limitations. Although hepatic venous pressure gradient and portal venous pressure are clinically useful hepatic hemodynamic parameters for determining both management and prognosis (27), the value of absolute flow parameters in clinical practice remains unclear. Caval subtraction phase-contrast MR imaging may also have limitations in the assessment of some patients with chronic liver disease: The presence of large extrahepatic portosystemic shunts (eg, recanalized umbilical vein or gastric varices), retrograde PV portal vein flow, or venous outflow obstruction (Budd-Chiari syndrome) are likely to compromise simple caval subtraction assessment of TLBF total liver blood flow or hepatic arterial flow. In addition, gating to a specific phase of respiration can potentially introduce (systematic) errors, particularly if caval blood flow at different levels is variably influenced by respiration phase, a phenomenon that was not investigated in this study. Finally, studies in anesthetized animals and healthy volunteers represent ideal conditions to test the technique given the compliance with imaging protocols. Studies translating this method into potentially less compliant, unwell patients with chronic liver disease are planned and are necessary to determine the value of caval subtraction phase-contrast MR imaging in clinical practice.

Caval subtraction phase-contrast MR imaging is a simple and rapid technique, amenable to technologist-led phase-contrast MR imaging planning in clinical practice. It could also be used for noninvasive validation of more complex methods such as four-dimensional phase-contrast MR imaging. In summary, we have demonstrated that caval subtraction phase-contrast MR imaging is technically feasible and may offer a reproducible and clinically viable method for measuring TLBF total liver blood flow and hepatic arterial flow.

Advances in Knowledge

  • ■ In a rodent model, portal venous flow measurement obtained with two-dimensional (2D) phase-contrast MR imaging demonstrates good agreement with invasive transit-time US (mean difference, −3.5 mL/min/100 g; Bland-Altman 95% limits of agreement [LOA], ±61.3 mL/min/100 g).
  • ■ In a rodent model, hepatic arterial fraction measurement obtained with caval subtraction 2D two-dimensional phase-contrast MR imaging demonstrates good agreement with an invasive microsphere standard of reference (mean difference, 4.2%; 95% LOA limits of agreement, ±20.5%).
  • ■ In human volunteers, total liver blood flow (TLBF) estimated with caval subtraction 2D two-dimensional phase-contrast MR imaging shows good agreement with that calculated from direct inflow phase-contrast MR imaging (mean difference, −1.3 mL/min/100 g; 95% LOA limits of agreement, ±23.1 mL/min/100 g).
  • ■ In human volunteers, caval subtraction TLBF total liver blood flow and directly measured inflow phase-contrast MR imaging TLBF total liver blood flow at 7 days were similar (95% LOA limits of agreement, ±31.6 mL/min/100 g vs ±29.6 mL/min/100 g).
  • ■ Caval subtraction 2D two-dimensional phase-contrast MR imaging is a noninvasive, simple, and rapid technique for measuring total liver and hepatic arterial blood flow.

APPENDIX

Appendix E1

SUPPLEMENTAL FIGURES

Figure E1:

Figure E2:

Acknowledgments

We are grateful for the assistance of Raj Ramasawmy, PhD, and Tom Roberts, PhD, in the development and postprocessing of preclinical phase-contrast MR imaging sequences and Abe Habtieson, MSc, for preparing rats for sham treatment and bile duct ligation. Finally, we are very grateful for the assistance of the clinical MR radiographers who conducted imaging of healthy volunteers.

Received August 7, 2015; revision requested October 15; revision received November 17; accepted December 23; final version accepted January 13, 2016.

The study was supported by the National Institute for Health Research, University College London Hospitals Biomedical Research Centre. S.H. and S.A.T. are senior investigators for the National Institute for Health Research, and M.D.C. is a clinical lecturer for the National Institute for Health Research. M.D.C. was supported by the Wellcome Trust (grant WT092186).

Disclosures of Conflicts of Interest: M.D.C. disclosed no relevant relationships. R.P.M. disclosed no relevant relationships. A.B. disclosed no relevant relationships. S.W.S. disclosed no relevant relationships. N.D. disclosed no relevant relationships. S.H. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: received money as an expert witness in the field of gastrointestinal radiology; institution received multiple grants from research funders; received payment for lectures including service on speakers bureaus from Bracco. Other relationships: disclosed no relevant relationships. M.F.L. disclosed no relevant relationships. S.A.T. Activities related to the present article: disclosed no relevant relationships. Activities not related to the present article: received personal fees from Robarts PLC. Other relationships: disclosed no relevant relationships.

Abbreviations:

BDL
bile duct ligation
IVC
inferior vena cava
LOA
limits of agreement
PV
portal vein
TLBF
total liver blood flow
2D
two-dimensional
Venc
velocity encoding

References

1. Pandharipande PV, , Krinsky GA, , Rusinek H, , Lee VS.. Perfusion imaging of the liver: current challenges and future goals. Radiology 2005 ;234(3):661–673. [PubMed]
2. Lotz J, , Meier C, , Leppert A, , Galanski M.. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. RadioGraphics 2002 ;22(3):651–671. [PubMed]
3. Gouya H, , Vignaux O, , Sogni P, , et al. . Chronic liver disease: systemic and splanchnic venous flow mapping with optimized cine phase-contrast MR imaging validated in a phantom model and prospectively evaluated in patients. Radiology 2011 ;261(1):144–155. [PubMed]
4. Jajamovich GH, , Dyvorne H, , Donnerhack C, , Taouli B.. Quantitative liver MRI combining phase contrast imaging, elastography, and DWI: assessment of reproducibility and postprandial effect at 3.0 T. PLoS One 2014 ;9(5):e97355. [PMC free article] [PubMed]
5. Stankovic Z, , Frydrychowicz A, , Csatari Z, , et al. . MR-based visualization and quantification of three-dimensional flow characteristics in the portal venous system. J Magn Reson Imaging 2010 ;32(2):466–475. [PubMed]
6. Jin N, , Lewandowski RJ, , Omary RA, , Larson AC.. Respiratory self-gating for free-breathing abdominal phase-contrast blood flow measurements. J Magn Reson Imaging 2009 ;29(4):860–868. [PMC free article] [PubMed]
7. Wilson DJ, , Ridgway JP, , Evans JA, , Robinson P.. Measurement of hepatic arterial flow using phase contrast magnetic resonance imaging. Phys Med Biol 2009 ;54(19):N439–N449. [PubMed]
8. Yzet T, , Bouzerar R, , Allart JD, , et al. . Hepatic vascular flow measurements by phase contrast MRI and Doppler echography: a comparative and reproducibility study. J Magn Reson Imaging 2010 ;31(3):579–588. [PubMed]
9. Yzet T, , Bouzerar R, , Baledent O, , et al. . Dynamic measurements of total hepatic blood flow with phase contrast MRI. Eur J Radiol 2010 ;73(1):119–124. [PubMed]
10. Harry D, , Anand R, , Holt S, , et al. . Increased sensitivity to endotoxemia in the bile duct-ligated cirrhotic rat. Hepatology 1999 ;30(5):1198–1205. [PubMed]
11. Van Oosterhout MF, , Willigers HM, , Reneman RS, , Prinzen FW.. Fluorescent microspheres to measure organ perfusion: validation of a simplified sample processing technique. Am J Physiol 1995 ;269(2 Pt 2):H725–H733. [PubMed]
12. Haacke EM.. Magnetic resonance imaging: physical principles and sequence design. New York, NY: Wiley, 1999 .
13. Kuo PC, , Li K, , Alfrey EJ, , Jeffrey RB, , Garcia G, , Dafoe DC.. Magnetic resonance imaging and hepatic hemodynamics: correlation with metabolic function in liver transplantation candidates. Surgery 1995 ;117(4):373–379. [PubMed]
14. Forkman J.. Estimator and tests for common coefficients of variation in normal distributions. Commun Stat Theory Methods 2009 ;38(2):233–251.
15. Lautt WW, , Legare DJ, , d’Almeida MS.. Adenosine as putative regulator of hepatic arterial flow (the buffer response). Am J Physiol 1985 ;248(3 Pt 2):H331–H338. [PubMed]
16. Richter S, , Mücke I, , Menger MD, , Vollmar B.. Impact of intrinsic blood flow regulation in cirrhosis: maintenance of hepatic arterial buffer response. Am J Physiol Gastrointest Liver Physiol 2000 ;279(2):G454–G462. [PubMed]
17. Richter S, , Olinger A, , Hildebrandt U, , Menger MD, , Vollmar B.. Loss of physiologic hepatic blood flow control (“hepatic arterial buffer response”) during CO2-pneumoperitoneum in the rat. Anesth Analg 2001 ;93(4):872–877. [PubMed]
18. Covey AM, , Brody LA, , Maluccio MA, , Getrajdman GI, , Brown KT.. Variant hepatic arterial anatomy revisited: digital subtraction angiography performed in 600 patients. Radiology 2002 ;224(2):542–547. [PubMed]
19. Hiatt JR, , Gabbay J, , Busuttil RW.. Surgical anatomy of the hepatic arteries in 1000 cases. Ann Surg 1994 ;220(1):50–52. [PubMed]
20. Song SY, , Chung JW, , Yin YH, , et al. . Celiac axis and common hepatic artery variations in 5002 patients: systematic analysis with spiral CT and DSA. Radiology 2010 ;255(1):278–288. [PubMed]
21. D’Almeida MS, , Cailmail S, , Lebrec D.. Validation of transit-time ultrasound flow probes to directly measure portal blood flow in conscious rats. Am J Physiol 1996 ;271(6 Pt 2):H2701–H2709. [PubMed]
22. Gauthier TP, , Muhammad A, , Wasan HS, , Abel PD, , Leen EL.. Reproducibility of quantitative assessment of altered hepatic hemodynamics with dynamic contrast-enhanced ultrasound. J Ultrasound Med 2012 ;31(9):1413–1420. [PubMed]
23. Hübner GH, , Steudel N, , Reissmann A, , et al. . Hepatic arterial Doppler sonography in patients with cirrhosis and controls: observer and equipment variability with use of the ultrasonic contrast agent SHU 508A. Z Gastroenterol 2005 ;43(7):639–645. [PubMed]
24. Vermeulen MA, , Ligthart-Melis GC, , Buijsman R, , et al. . Accurate perioperative flow measurement of the portal vein and hepatic and renal artery: a role for preoperative MRI? Eur J Radiol 2012 ;81(9):2042–2048. [PubMed]
25. Mehta G, , Mookerjee RP, , Sharma V, , Jalan R.. Systemic inflammation is associated with increased intrahepatic resistance and mortality in alcohol-related acute-on-chronic liver failure. Liver Int 2015 ;35(3):724–734. [PubMed]
26. Iwasa M, , Nakamura K, , Nakagawa T, , et al. . Single photon emission computed tomography to determine effective hepatic blood flow and intrahepatic shunting. Hepatology 1995 ;21(2):359–365. [PubMed]
27. Thalheimer U, , Mela M, , Patch D, , Burroughs AK.. Targeting portal pressure measurements: a critical reappraisal. Hepatology 2004 ;39(2):286–290. [PubMed]

Articles from RSNA Selective Deposit are provided here courtesy of Radiological Society of North America