A treatment-planning comparison of three beam arrangement strategies for stereotactic body radiation therapy for centrally located lung tumors using volumetric-modulated arc therapy

Abstract

The purpose of this study was to determine appropriate beam arrangement for volumetric-modulated arc therapy (VMAT)-based stereotactic body radiation therapy (SBRT) in the treatment of patients with centrally located lung tumors. Fifteen consecutive patients with centrally located lung tumors treated at our institution were enrolled. For each patient, three VMAT plans were generated using two coplanar partial arcs (CP VMAT), two non-coplanar partial arcs (NCP VMAT), and one coplanar full arc (Full VMAT). All plans were designed to deliver 70 Gy in 10 fractions. Target coverage and sparing of organs at risk (OARs) were compared across techniques. PTV coverage was almost identical for all approaches. The whole lung V10Gy was significantly lower with CP VMAT plans than with NCP VMAT plans, whereas no significant differences in the mean lung dose, V5Gy, V20Gy or V40Gy were observed. Full VMAT increased mean contralateral lung V5Gy by 12.57% and 9.15% when compared with NCP VMAT and CP VMAT, respectively. Although NCP VMAT plans best achieved the dose–volume constraints for mediastinal OARs, the absolute differences in dose were small when compared with CP VMAT. These results suggest that partial-arc VMAT may be preferable to minimize unnecessary exposure to the contralateral lung, and use of NCP VMAT should be considered when the dose–volume constraints are not achieved by CP VMAT.

Keywords: lung tumor, stereotactic body radiation therapy, volumetric-modulated arc therapy, planning study

INTRODUCTION

Stereotactic body radiation therapy (SBRT) has proven to be an effective treatment modality for medically inoperable early-stage non–small cell lung cancer (NSCLC) and lung metastases [1, 2]. However, the favorable data on lung SBRT are based mainly on the treatment of peripheral lesions. The use of SBRT for centrally located lung tumors still remains controversial due to the potential of severe toxicity, such as bronchial stenosis, hemoptysis, and fistulas [3]. Therefore, reducing the radiation dose to organs at risk (OARs) in the mediastinum and pulmonary hilum is essential when treating centrally located lung lesions with SBRT.

Volumetric-modulated arc therapy (VMAT) is now widely used clinically because it can deliver intensity-modulated radiation therapy (IMRT) quality plans with superior treatment delivery efficiency. As for lung SBRT, several treatment planning studies have shown that VMAT achieved superior dose conformity and OAR sparing with a shorter treatment time than 3D conformal radiotherapy (3D-CRT) [48] or conventional IMRT [9, 10]. With these promising results, there has been an increase in the use of VMAT in lung SBRT. A patterns-of-care survey for lung SBRT performed in 2012 reported that 47% of radiation oncologists have used rotational IMRT in their practice in the USA [11].

The treatment beam arrangements used for lung SBRT are either coplanar, non-coplanar or a mixture of both. Many institutions use multiple non-coplanar beams for lung SBRT using 3D-CRT [12, 13]. A study comparing beam arrangements for 3D-CRT-based lung SBRT concluded that non-coplanar beams were best when compared with coplanar beams or arcs (volumes at high doses were equivalent between all, but volumes at low doses were better with non-coplanar beams) [14]. VMAT has the ability for non-coplanar arc delivery, but the current implementations of VMAT have focused on coplanar delivery [46, 9, 10, 1519]. In addition, the range of arc used for VMAT-based lung SBRT varies among institutions: some institutions use full arc [4, 6, 8, 10, 16, 17] and others use partial arc to avoid the contralateral lung [5, 9, 15, 16, 18]. However, data for optimal beam arrangements are limited to date.

This study was undertaken to determine appropriate beam arrangement for VMAT-based SBRT in the treatment of patients with centrally located lung tumors that require better OAR sparing when compared with patients with peripheral lesions.

MATERIALS AND METHODS

Patient selection and contouring

This study was approved by the institutional ethics committee, and written informed consent was obtained from all patients. Computed tomography (CT) datasets of 15 consecutive patients with centrally located lung tumors who were treated with VMAT-based SBRT in our institution between November 2011 and February 2015 were included in this study. Central location was defined as the area within 2 cm of critical structures, such as the bronchial tree, great vessels, esophagus, and heart, but without direct involvement of the mediastinal structures. Those OARs located within 2 cm of the tumor were defined as immediately adjacent OARs.

Four-dimensional CT images were acquired in the supine position, with 1.25-mm slice thickness after immobilization with the BodyFix double vacuum immobilization system (Elekta, Stockholm, Sweden). Scans were performed using the Varian Real-time Position Management (RPM) system (Varian Medical Systems, Palo Alto, CA). The gross tumor volume (GTV) was contoured at the lung window level on all 10 respiratory phase CT datasets. These 10 GTVs were fused to form the internal target volume (ITV). The planning target volume (PTV) was generated by adding a 5-mm margin to the ITV in all dimensions to account for set-up errors. Contouring of OARs, including the whole lung, esophagus, heart, spinal cord, bronchial tree, and great vessels (aorta, vena cavae, pulmonary artery, and pulmonary vein), was defined according to the Radiation Therapy Oncology Group (RTOG) protocol 0813 [20]. The bronchial tree consisted of the trachea and the proximal bronchial tree. The contralateral lung was also contoured as a separate structure. Dose constraints used for OARs are listed in Table 1 and were based primarily on the MD Anderson Cancer Center experiences using 70 Gy in 10 fractions [21].

Table 1.
Dose–volume constraints to OARs used for plan evaluation

Treatment planning

All plans prescribed 70 Gy delivered in 10 daily fractions to the PTV. They were normalized so that at least 95% of the PTV received the prescribed dose. Three VMAT plans were generated with two coplanar partial arcs (CP VMAT), two non-coplanar partial arcs (NCP VMAT) and one coplanar full arc (Full VMAT). CP VMAT and NCP VMAT plans consisted of two ipsilateral arcs of 179° with collimator angles of 30° and 330°, and Full VMAT plans consisted of one arc of 358° with a collimator angle of 30°. The couch was set at ± 15° to the straight position for the NCP VMAT plans. The arrangement of the fields for a representative patient is shown in Fig. 1. Mean field sizes by jaws in CP VMAT, NCP VMAT, and Full VMAT were 5.7 × 5.4 cm2, 5.6 × 5.5 cm2, and 5.9 × 5.6 cm2, respectively, where results were reported as the average of the two arcs for CP VMAT and NCP VMAT. All VMAT plans were developed in the Eclipse v.10.0 treatment planning system using 6-MV photons in a Novalis Tx treatment unit (Varian Medical Systems, and BrainLAB AG, Feldkirchen, Germany). The same optimization template was utilized for the three VMAT techniques (Table 2); however, these objectives were modified during the optimization process based on the real-time updated dose–volume histograms (DVHs) of the PTV and OARs. To achieve the highly steep gradient dose to the PTV, we used the normal tissue objective (a parameter that limits dose as a function of distance from the outer border of the PTV) with a priority of 150 and a 0.15 fall-off between the start dose of 105% and the end dose of 60%. In addition, avoidance structures were used for optimization to minimize the dose to OARs. The optimization constraints and priorities of these avoidance structures were not initially applied and were real-time adjusted during the progression of optimization. During the optimization process, the primary goal was to achieve similar PTV coverage for the three VMAT plans, and the secondary goal was to decrease the OAR doses as much as possible. Multiple plans for each VMAT technique were generated as we performed optimization several times, further tightening dose–volume constraints of the OARs each time, and the most preferable plan was selected for a fair comparison. The final dose calculation was performed using the anisotropic analytical algorithm with a 1.25-mm grid resolution.

Table 2.
Initial optimization parameters
Fig. 1.
Representative beam arrangement for (a) two coplanar partial arcs volumetric-modulated arc therapy (VMAT), (b) two noncoplanar partial arcs VMAT and (c) one coplanar full arc VMAT.

Plan evaluation

DVHs for the PTV, whole lung, contralateral lung, esophagus, heart, spinal cord, bronchial tree, and great vessels were calculated for each subject. Parameters chosen for comparison were mean dose, D2%, and D98% for the PTV, where Dn% is the minimal dose delivered to n% of the PTV. The homogeneity index (HI) and conformation number (CN) for the PTV were calculated as below [15, 22]:

HI = (D2% D98%)/Dp × 100%

CN= (PTV encompassed by 95% isodose/PTV)×(PTV encompassed by 95% isodose/95% isodose volume),

where Dp is the prescription dose. For the CN, the first portion is an assessment of target volume coverage by 95% of the prescription dose, and the second portion is an assessment of normal tissue sparing. The CN values range between 0 and 1, with 1 representing the ideal conformity. Mean lung dose, V5Gy, V10Gy, V20Gy and V40Gy for the whole lung and V5Gy for the contralateral lung were compared, where VnGy is the percentage structure volume receiving at least n Gy. The maximal dose received by 0.1 cm3 (D0.1cm3) of the esophagus, heart, spinal cord, bronchial tree, and great vessels was also compared. The number of Monitor Units (MU) needed for each plan was collected to determine the dose-delivery efficiency of each treatment. Patients were divided into two groups based on whether all VMAT techniques met the dose–volume constraints (Group 1) or not (Group 2). The distance from the PTV to its immediately adjacent OAR was recorded for each patient.

Statistical analysis

Dose–volume parameters, HI, CN and MU were compared by Wilcoxon matched-pair signed-rank tests for non-parametric data. The Mann–Whitney U test was used to analyze the difference in the closest distance to the PTV between Groups 1 and 2. Significance was defined at P < 0.05. All reported P-values are two-tailed.

RESULTS

The characteristics of the 15 patients are listed in Table 3. The comparison of dose distributions is illustrated in Fig. 2 for the example case shown in Fig. 1 for each of the three VMAT techniques. A comparison of mean values of the dosimetric parameters of the PTV and the OARs is provided in Table 4.

Table 3.
Patient characteristics (n = 15)
Table 4.
Comparison of mean values of the dosimetric parameters of the PTV and the OARs
Fig. 2.
Dose distributions for two coplanar arcs volumetric-modulated arc therapy (CP VMAT), two noncoplanar partial arcs VMAT (NCP VMAT), and one coplanar full arc VMAT (Full VMAT) for the example case shown in Fig. 1. Planning target volume (PTV) is ...

The three VMAT techniques provided highly conformal dose distributions in terms of target coverage. All plans were almost identical with respect to the resulting dosimetric parameters of the PTV, HI and CN. The NCP VMAT and CP VMAT plans showed better sparing of the whole lung than the Full VMAT plans. The whole lung V10Gy was significantly lower in the CP VMAT plans than in the NCP VMAT plans, whereas no significant differences were observed in the mean lung dose, V5Gy, V20Gy or V40Gy. NCP VMAT plans achieved the best contralateral lung sparing. On the other hand, Full VMAT increased mean contralateral lung V5Gy by 12.57% and 9.15% when compared with NCP VMAT and CP VMAT, respectively. NCP VMAT plans showed better sparing of mediastinal OARs than CP VMAT plans. Although NCP VMAT plans resulted in significantly lower D0.1cm3 to the esophagus, spinal cord, aorta, and vena cavae compared with CP VMAT plans, the absolute differences in mean dose were small: 1.9 Gy, 1.24 Gy, 2 Gy and 1.68 Gy, respectively. CP VMAT consistently provided equivocal or favorable mediastinal OAR sparing when compared with Full VMAT, although the difference was very small in absolute terms. The mean MUs for CP VMAT, NCP VMAT and Full VMAT were 1889 ± 197, 1850 ± 216 and 1832 ± 71, respectively, showing that there were no statistically significant differences between the three techniques.

All dose–volume constraints were met satisfactorily by the NCP VMAT plans for all patients. However, four immediately adjacent OARs in three patients and five immediately adjacent OARs in four patients failed to meet the dose–volume constraints by CP VMAT plans and Full VMAT plans, respectively. Table 5 shows a list of the dosimetric parameters of the immediately adjacent OARs that received more than the threshold value and the distance from the PTV to each OAR in Group 2 patients. The mean PTV to its closest OAR distance was significantly shorter in Group 2 than in Group 1 (0.13 cm vs 0.46 cm, P = 0.004). In 10 cases where the distance between the PTV and closest OAR was 0.25 cm or greater, optimal OAR sparing was achieved through all VMAT techniques. In cases where the distance was shorter than 0.25 cm, it was not possible to achieve optimal OAR sparing through CP VMAT, Full VMAT or both in four of five cases.

Table 5.
Dosimetric parameters of the immediately adjacent OARs that received more than the threshold value and the distance from the PTV to each OAR in Group 2 patients

DISCUSSION

VMAT has recently played an important role in SBRT of lung tumors. In the present study, appropriate beam arrangement for lung SBRT using VMAT was investigated. Since most OARs are removed from the high-dose area in the treatment of peripheral lung tumors, the dose constraints are easily met using 3D-CRT. Thus, there is controversy as to whether VMAT should be routinely used to treat peripheral lung tumors [23]. On the other hand, it is more challenging to generate an acceptable SBRT plan using 3D-CRT for centrally located lung tumors because of the close proximity of critical structures, which can be better handled with IMRT or VMAT. Therefore, 15 consecutive patients with centrally located tumors were selected for this dosimetric study. As for SBRT for centrally located lung cancer, most institutions have adopted more conservative fractionation schemes [21, 24, 25]. Recently, a report from the MD Anderson Cancer Center showed that SBRT with 70 Gy in 10 fractions achieved excellent local control and acceptable toxicity for clinically challenging cases, including centrally located lesions, compared with 50 Gy in 4 fractions [21]. In the current study, this dose fractionation schedule, as well as recommended dose–volume constraints, was followed. However, differing from the recommended dose–volume constraints, we used D0.1cm3 instead of maximum point dose, because maximum point doses are very sensitive to many factors such as dose calculation grid size and algorithm.

Several recent studies have compared VMAT with 3D-CRT for lung SBRT. In general, VMAT has shown better conformity to the PTV than 3D-CRT [4, 68]. The increase in the volume of normal lung that receives a low dose of radiation is a concern associated with VMAT. However, lung V20Gy/12.5Gy/10Gy/5Gy values were significantly lower with VMAT than with 3D-CRT [5]. A recent study indicated that VMAT improved mean lung dose, lung V5Gy and V20Gy [7]. For the contralateral lung, Rauschenbach et al. showed a lower mean dose with VMAT than with 3D-CRT [8], whereas Ong et al. reported that VMAT led to a small increase in V5Gy [6]. Therefore, the results of these studies showed that VMAT outperformed 3D-CRT in most dosimetric aspects. When compared with IMRT, coplanar VMAT achieved treatment plan qualities comparable with the non-coplanar IMRT technique and better than those of coplanar IMRT [9].

Many institutions have reported promising results with SBRT using VMAT in patients with lung tumors. However, there are very few data in the literature on dosimetric comparisons between various beam arrangements. To the best of our knowledge, there is only one planning study comparing coplanar VMAT with non-coplanar VMAT in the setting of lung SBRT. Zhang et al. showed that the dosimetric differences between coplanar and non-coplanar VMAT were not significant [7]. However, the angle separation of the non-coplanar arc was uniformly set at 10°. As they discussed, it was presumed that the dosimetric differences are expected to increase, favoring the non-coplanar VMAT plans, with a larger angle separation. In the present study, the couch was offset ± 15°; thus, the angle separation was set at 30°. This large angle separation may lead to better mediastinal OAR sparing compared with coplanar VMAT.

All three VMAT techniques produced highly conformal dose distributions around the PTV. Li et al. suggested that compromised PTV coverage contributed to local failure after lung SBRT [21]. With respect to PTV coverage, VMAT seems to be an appropriate technique for lung SBRT.

The use of Full VMAT, compared with CP VMAT or NCP VMAT, resulted in increased doses to most of the OARs. In particular, contralateral lung V5Gy was markedly increased with Full VMAT. Ong et al. reported that contralateral lung V5Gy strongly correlates with radiation pneumonitis after lung SBRT using VMAT [16]. In addition, Full VMAT offered only a small advantage over CP VMAT in terms of treatment time, because coplanar beams were used for both techniques. Therefore, it is conceivable that partial-arc VMAT is more suitable for lung SBRT to minimize unnecessary dose deposition to the contralateral lung.

Lung dosimetric parameters in CP VMAT were comparable with those of NCP VMAT: whole lung V5Gy, V20Gy, V40Gy and mean lung dose were identical for the two techniques, whereas whole lung V10Gy was better with CP VMAT, and contralateral lung V5Gy was lower with NCP VMAT. Optimal OAR sparing was achieved in all cases by NCP VMAT, but in only 12 cases by CP VMAT. However, the changes in the quantitative dosimetric differences for mediastinal OARs were small. Furthermore, NCP VMAT generally requires more time to deliver treatments than CP VMAT due to the time required for positioning of the couch. The European Organisation for Research and Treatment of Cancer (EORTC) recommends coplanar beam arrangements for lung SBRT, as long as the dose distribution criteria are met, to keep the treatment time as short as possible [26]. With respect to this study, all CP VMAT plans fulfilled the dose requirements for OARs in all cases where the distance between the PTV and closest OAR was 0.25 cm or greater. Based on the present results, CP VMAT is sufficient for SBRT for centrally located lung tumors when all dose–volume constraints can be achieved. NCP VMAT may be more appropriate only in patients who require better mediastinal OAR sparing of OARs that are very close to the PTV.

This study had several limitations. In practice, the applicable area of the non-coplanar beam direction is limited in lung SBRT using VMAT due to a risk of collision between the couch and the gantry. In the present study, the angle separation of the non-coplanar arc was uniformly set at 30°, the maximum angle at our institution. However, it is likely that the angle separation is smaller depending on the linear accelerator and immobilization system used. In such cases, the benefit of using the non-coplanar beam becomes less. Additionally, evaluating the optimal number of beams is beyond the scope of this study. However, Chi et al. compared OAR dose parameters between 2 and 8-arc VMAT plans in delivering SBRT for centrally located lung tumors; they concluded that increasing the number of arcs in VMAT cannot significantly improve OAR sparing [15].

In conclusion, VMAT techniques provided highly conformal target coverage for SBRT for lung tumors. Considering the increase in dose to the contralateral lung without improvement in mediastinal OAR sparing with Full VMAT, beams that avoid the contralateral lung may be preferable. Although NCP VMAT plans best achieved the dose–volume constraints for mediastinal OARs for centrally located lung tumors, the absolute differences in doses were small when compared with CP VMAT. Therefore, we suggest that adoption of NCP VMAT be considered, in practice, when the dose–volume constraints are not achieved by CP VMAT due to shorter distances between the PTV and OARs.

FUNDING

Funding to pay the Open Access publication charges for this article was provided by K.I. (Tane General Hospital).

REFERENCES

1. Baumann P, Nyman J, Hoyer M et al. Outcome in a prospective phase II trial of medically inoperable stage I non-small-cell lung cancer patients treated with stereotactic body radiotherapy. J Clin Oncol 2009;27:3290–6. [PubMed]
2. Rusthoven KE, Kavanagh BD, Burri SH et al. Multi-institutional phase I/II trial of stereotactic body radiation therapy for lung metastases. J Clin Oncol 2009;27:1579–84. [PubMed]
3. Timmerman R, McGarry R, Yiannoutsos C et al. Excessive toxicity when treating central tumors in a phase II study of stereotactic body radiation therapy for medically inoperable early-stage lung cancer. J Clin Oncol 2006;24:4833–9. [PubMed]
4. Verbakel WF, Senan S, Cuijpers JP et al. Rapid delivery of stereotactic radiotherapy for peripheral lung tumors using volumetric intensity-modulated arcs. Radiother Oncol 2009;93:122–4. [PubMed]
5. McGrath SD, Matuszak MM, Yan D et al. Volumetric modulated arc therapy for delivery of hypofractionated stereotactic lung radiotherapy: a dosimetric and treatment efficiency analysis. Radiother Oncol 2010;95:153–7. [PubMed]
6. Ong CL, Verbakel WF, Cuijpers JP et al. Stereotactic radiotherapy for peripheral lung tumors: a comparison of volumetric modulated arc therapy with 3 other delivery techniques. Radiother Oncol 2010;97:437–42. [PubMed]
7. Zhang GG, Ku L, Dilling TJ et al. Volumetric modulated arc planning for lung stereotactic body radiotherapy using conventional and unflattened photon beams: a dosimetric comparison with 3D technique. Radiat Oncol 2011;6:152. [PMC free article] [PubMed]
8. Rauschenbach BM, Mackowiak L, Malhotra HK A dosimetric comparison of three-dimensional conformal radiotherapy, volumetric-modulated arc therapy, and dynamic conformal arc therapy in the treatment of non-small cell lung cancer using stereotactic body radiotherapy. J Appl Clin Med Phys 2014;15:147–61. [PubMed]
9. Holt A, van Vliet-Vroegindeweij C, Mans A et al. Volumetric-modulated arc therapy for stereotactic body radiotherapy of lung tumors: a comparison with intensity-modulated radiotherapy techniques. Int J Radiat Oncol Biol Phys 2011;81:1560–7. [PubMed]
10. Weyh A, Konski A, Nalichowski A et al. Lung SBRT: dosimetric and delivery comparison of RapidArc, TomoTherapy, and IMRT. J Appl Clin Med Phys 2013;14:3–13. [PubMed]
11. Daly ME, Perks JR, Chen AM Patterns-of-care for thoracic stereotactic body radiotherapy among practicing radiation oncologists in the United States. J Thorac Oncol 2013;8:202–7. [PubMed]
12. Takayama K, Nagata Y, Negoro Y et al. Treatment planning of stereotactic radiotherapy for solitary lung tumor. Int J Radiat Oncol Biol Phys 2005;61:1565–71. [PubMed]
13. Timmerman R, Galvin J, Michalski J et al. Accreditation and quality assurance for Radiation Therapy Oncology Group: multicenter clinical trials using stereotactic body radiation therapy in lung cancer. Acta Oncol 2006;45:779–86. [PubMed]
14. Lim DH, Yi BY, Mirmiran A et al. Optimal beam arrangement for stereotactic body radiation therapy delivery in lung tumors. Acta Oncol 2010;49:219–24. [PubMed]
15. Chi A, Ma P, Fu G et al. Critical structure sparing in stereotactic ablative radiotherapy for central lung lesions: helical tomotherapy vs. volumetric-modulated arc therapy. PLoS One 2013;8:e59729. [PMC free article] [PubMed]
16. Ong CL, Palma D, Verbakel WF et al. Treatment of large Stage I–II lung tumors using stereotactic body radiotherapy (SBRT): planning considerations and early toxicity. Radiother Oncol 2010;97:431–36. [PubMed]
17. Palma DA, Senan S, Haasbeek CJ et al. Radiological and clinical pneumonitis after stereotactic lung radiotherapy: a matched analysis of three-dimensional conformal and volumetric-modulated arc therapy techniques. Int J Radiat Oncol Biol Phys 2011;80:506–13. [PubMed]
18. Navarria P, Ascolese AM, Mancosu P et al. Volumetric modulated arc therapy with flattening filter free (FFF) beams for stereotactic body radiation therapy (SBRT) in patients with medically inoperable early stage non small cell lung cancer (NSCLC). Radiother Oncol 2013;107:414–8. [PubMed]
19. Yamashita H, Haga A, Takahashi W et al. Volumetric modulated arc therapy for lung stereotactic radiation therapy can achieve high local control rates. Radiat Oncol 2014;9:243. [PMC free article] [PubMed]
20. Radiation Therapy Oncology Group RTOG 0813. Seamless Phase I/II Study of Stereotactic Lung Radiotherapy (SBRT) for Early Stage, Centrally Located, Non-small Cell Lung Cancer (NSCLC) in Medically Inoperable Patients. http://www.rtog.org/ClinicalTrials/ProtocolTable.aspx (1 Aug 2015, date last accessed).
21. Li Q, Swanick CW, Allen PK et al. Stereotactic ablative radiotherapy (SABR) using 70 Gy in 10 fractions for non-small cell lung cancer: exploration of clinical indications. Radiother Oncol 2014;112:256–61. [PubMed]
22. Wu Q, Mohan R, Morris M et al. Simultaneous integrated boost intensity-modulated radiotherapy for locally advanced head-and-neck squamous cell carcinomas. I: dosimetric results. Int J Radiat Oncol Biol Phys 2003;56:573–85. [PubMed]
23. Cai J, Malhotra HK, Orton CG A 3D-conformal technique is better than IMRT or VMAT for lung SBRT. Med Phys 2014;41:040601. [PubMed]
24. Haasbeek CJ, Lagerwaard FJ, Slotman BJ et al. Outcomes of stereotactic ablative radiotherapy for centrally located early-stage lung cancer. J Thorac Oncol 2011;6:2036–43. [PubMed]
25. Senthi S, Haasbeek CJ, Slotman BJ et al. Outcome of stereotactic ablative radiotherapy for central lung tumors: a systematic review. Radiother Oncol 2013;106:276–82. [PubMed]
26. Ruysscher DD, Faivre-Finn C, Nestle U et al. European Organisation for Research and Treatment of Cancer recommendations for planning and delivery of high-dose, high-precision radiotherapy for lung cancer. J Clin Oncol 2010;28:5301–10. [PubMed]

Articles from Journal of Radiation Research are provided here courtesy of Oxford University Press