Radiol Case Rep. 2009; 4(2): 278.
Published online 2016 January 5. doi:  10.2484/rcr.v4i2.278
PMCID: PMC4898000

Catheter-Induced Pulmonary Artery Dissection and Contrast Extravasation During Pulmonary Angiography

Daniel T. Ginat, M.D., M.S.1 and Lawrence Sahler, M.D.1

Abstract

Pulmonary artery injury during pulmonary angiography is an uncommon complication. We describe two cases of iatrogenic pulmonary artery extravasation that resulted from the use of pigtail catheters for pulmonary angiography. No further complications occurred and no treatment was necessary in either case. Thus, mild contrast extravasation appears to be a self-limited complication that likely requires little more than short-term observation.

Abbreviations: CT, computed tomography

Introduction

Although CT angiography has now largely supplanted conventional catheter pulmonary angiography for the assessment of pulmonary embolism, catheter angiography has long been considered the gold standard modality [1]. Pulmonary angiography has generally been regarded as a safe procedure [2, 3, 4, 5, 6]. Major complications of pulmonary angiography most commonly include renal dysfunction requiring hemodialysis, significant hemorrhage requiring transfusion, cardiac perforation, major arrhythmias, cardiac arrest, and severe contrast reactions. Such complications occur in 0.3 to 1% of cases of pulmonary angiography and death directly related to the procedure occurs in 0.2 to 0.5% [2, 3, 5, 6]. Minor complications, such as transient angina, small hematomas at the puncture site, urticaria, and drug-responsive renal dysfunction, hypotension, and respiratory distress result in 1.4 to 4.8% [2, 4]. Catheter-induced intramural deposition of contrast is a rarely reported complication of pulmonary angiography [7, 8]. In this report, we describe two cases of iatrogenic contrast extravasation from the pulmonary arteries during pulmonary angiography.

Case Report 1

A 58 year-old woman with a history of melanoma presented with several days of progressive cough and shortness of breath. She had been discharged from the hospital two days earlier following resection of a right thigh melanoma. On readmission, an arterial blood gas yielded a pO2 of 67mmHg on room air. A ventilation-perfusion scan was performed and interpreted as intermediate probability for pulmonary embolus. Selective bilateral pulmonary angiograms were performed using an 8 French pigtail catheter (Cook Inc, Bloomington, IN), through which 55ml of contrast was injected at a rate of 27 milliliters per second. The left pulmonary arteriogram demonstrated no evidence of thromboembolic disease. The catheter was easily re-directed in to the right pulmonary artery over a 0.038” soft tip guide wire (Cook Inc, Bloomington, IN). The right pulmonary angiogram showed a right lower lobe pulmonary embolus (Figure 1A). In addition, extraluminal accumulation of contrast adjecent to the right main pulmonary artery occurred and persisted at 45 minutes (Figure 1B). The patient was asymptomatic and a follow-up chest X-ray 8 hours after the procedure demonstrated resolution of the contrast stain. The remainder of the patient’s hospital course was uneventful. Anticoagulation was initiated and a Greenfield filter was inserted.

Figure 1A
58-year-old woman with suspected pulmonary embolism. Pulmonary angiogram of the right pulmonary artery obtained shortly after injection of contrast reveals an embolism within the right lower lobe pulmonary artery (arrow).
Figure 1B
58-year-old woman with suspected pulmonary embolism. Pulmonary angiogram at the completion of contrast injection demonstrates an a large area of contrast extravasation (encircled) in the region of the right pulmonary artery with extension into the middle ...

Case Report 2

A 25-year-old, 22-weeks pregnant female presented with vaginal bleeding. Two days after delivery of a nonviable fetus, the patient experienced shortness of breath and chest pain. An arterial blood gas sample revealed a pO2 of 62mmHg. A ventilation-perfusion scan was performed and interpreted as low probability. Selective bilateral pulmonary angiograms were performed as described in the previous case. No pulmonary embolus was identified. However, a small amount of subintimal contrast extravasation was identified in the superior margin of the right pulmonary artery (Figure 2A, Figure 2B). There was no clinically significant sequelae or abnormalities on follow-up chest X-ray.

Figure 2A
25-year-old woman with suspected pulmonary embolism. Pulmonary angiogram suggests the presence of a small collection subintimal contrast within the superior margin of the right pulmonary artery (arrow). Otherwise, the right pulmonary artery branches are ...
Figure 2B
25-year-old woman with suspected pulmonary embolism. A delayed fluoroscopic image after pulmonary angiography reveals a persistent contrast blush, which confirms right pulmonary artery injury resulting in contrast extravasation into the vessel wall (arrow). ...

Discussion

Subintimal contrast staining is considered a rare minor complication of pulmonary angiography, occurring in less than 0.4% of cases [6]. In a series of 122 pulmonary angiograms using pigtail pulmonary-artery-seeking catheters, one case of right pulmonary artery extravasation occurred without sequelae [9]. At our institution, among 626 pulmonary angiograms performed over the course of 20 years, only two such cases were encountered. In contrast, pulmonary artery rupture occurs more frequently during placement of pulmonary artery catheters without fluoroscopic guidance [10].

With regard to our cases, the distal positioning of the catheter and perhaps fluctuations in the catheter position during deep inspiration may have wedged the pigtail creating a localized high-pressure jet at the end hole. This may have led to subintimal contrast injection and subsequent contrast extravasation. This phenomenon has been described in association with the use of a wedge-shaped pulmonary catheter during angiography of an interlobular pulmonary artery [11]. This complication is preventable using fluoroscopic guidance, particularly through cautious positioning of the catheter in such a way that there is adequate range of motion for the tip and preventing the end-hole from abutting the arterial wall. In addition, the catheter should be sufficiently undersized for the selected pulmonary artery branch. The benign course of our patients and the lack of morbidity reported in the literature suggest that observation alone may suffice.

Footnotes

Published: May 31, 2009

References

1. Newman GE. Pulmonary angiography in pulmonary embolic disease. J Thorac Imaging. 1989 Oct;4(4):28–39. [PubMed] [PubMed]
2. Zuckerman DA, Sterling KM, Oser RF. Safety of pulmonary angiography in the 1990s. J Vasc Interv Radiol. 1996 Mar-Apr;7(2):199–205. [PubMed] [PubMed]
3. Hudson ER, Smith TP, McDermott VG. Pulmonary angiography performed with iopamidol: complications in 1,434 patients. Radiology. 1996 Jan;198(1):61–65. [PubMed] [PubMed]
4. Nilsson T, Carlsson A, Mâre K. Pulmonary angiography: a safe procedure with modern contrast media and technique. Eur Radiol. 1998;8(1):86–89. [PubMed] [PubMed]
5. Mills SR, Jackson DC, Older RA. The incidence, etiologies, and avoidance of complications of pulmonary angiography in a large series. Radiology. 1980 Aug;136(2):295–299. [PubMed] [PubMed]
6. Stein PD, Athanasoulis C, Alavi A. Complications and validity of pulmonary angiography in acute pulmonary embolism. Circulation. 1992 Feb;85(2):462–468. [PubMed] [PubMed]
7. Bookstein J, Sigmann J. Intramural deposition of contrast agent during selective angiocardiography. Radiology. 1963 Dec;81:932–938. [PubMed] [PubMed]
8. Sumita S, Ujike Y, Namiki A. Rupture of pulmonary artery induced by balloon occlusion pulmonary angiography. Intensive Care Med. 1995 Jan;21(1):79–81. [PubMed] [PubMed]
9. Stein M, Winter J, Grollman J., Jr The value of the pulmonary-artery-seeking catheter in percutaneous selective pulmonary arteriography. Radiology. 1975 Feb;114(2):299–304. [PubMed] [PubMed]
10. Abreu A, Campos M, Krieger B. Pulmonary artery rupture induced by a pulmonary artery catheter: a case report and review of the literature. J Intensive Care Med. 2004 Sep-Oct;19(5):291–296. [PubMed] [PubMed]
11. Ray C, Jr, Rivitz S, Fan C, Geller S. Re: complication of pulmonary arteriography: contrast extravasation from a wedged pigtail-shaped catheter. Cardiovasc Intervent Radiol. 1996 Jul-Aug;19(4):305–306. [PubMed] [PubMed]

Articles from Radiology Case Reports are provided here courtesy of Elsevier