Dissociation of a Bipolar Prosthesis after Right Hip Hemiarthroplasty

Introduction

We report the case of an 84-year-old woman with a dissociated bipolar component of her right hip hemiarthroplasty. Dissociation of components is a rare but increasingly recognized consequence of modular componentry. Radiologists should be aware of the distinction between dislocation of a prosthesis and dissociation of its bipolar components, and be able to describe these complications accurately in their interpretations.

Case Report

We report the case of an 84-year-old woman who fell, resulting in a fracture of her right femoral neck. This fracture was treated with placement of a bipolar hemiarthroplasty. Two weeks later, a routine post-operative radiograph demonstrated an abnormal appearance of her arthroplasty, despite no history of hip pain, leg numbness, or weakness.

Initial right hip radiographs demonstrated dissociation of the components of the bipolar hemiarthroplasty, with inferomedial displacement of the acetabular cup component, and with the prosthetic femoral head abutting the roof of the acetabulum (Figure 1). There was no evidence of fracture or loosening of the femoral component of the prosthesis. The patient underwent successful conversion to total hip replacement and had an uneventful postoperative course.

Figure 1
Anteroposterior radiograph of the pelvis demonstrates a bipolar hip arthroplasty, with a cemented femoral stem. The acetabular cup component of the arthroplasty has dissociated from the femoral component and is dislocated inferomedially. The head of the ...

Discussion

Although interpositional hip arthroplasty was first performed in the 1800's (1), it did not become commonplace, or widely successful until the significant refinements by Charley in the 1960's (2, 3). Since then, hip replacement surgery has become a multi-billion dollar industry, with close to 300,000 total hip arthroplasties performed each year in the United States (2). Osteoarthritis and significantly displaced femoral neck fracture represent two of the more common indications for hip replacement surgery (4, 5, 6, 7).

Painful loosening and acetabular erosion often occurred after the use of the early unipolar femoral endoprostheses (6). In the early 1970's, bipolar femoral prostheses were introduced, with the goal of reducing these complications. These prostheses are designed to allow movement to occur, not only between the patient's native acetabulum and the prosthesis, but also at a joint within the prosthesis itself (8). Their purported advantages over unipolar prostheses include less wear and erosion of the acetabulum, lower risk of dislocation, variable head sizes and neck lengths, and an increased range of motion (5, 7, 9). However, recent studies have shown no significant differences between unipolar and bipolar prostheses regarding complication rates and functional outcome (7, 8, 10).

Complications from hip arthroplasty are rare, but result in significantly increased costs and patient morbidity. Dislocation of the hemiarthroplasty from the native acetabulum is a well-documented, albeit rare, complication and is reported to occur in 1.2% to 3.4% of cases (11, 12).

A related but distinct complication of bipolar arthroplasties is the development of component dissociation. These dissociations are quite uncommon. When they do occur, they frequently co-exist with dislocation of the prosthesis from the native acetabulum. There have been relatively few prior reports of this serious complication. Georgiou et al. recently reported five cases of dissociation involving elderly patients (ages 75 to 86) with a variable postoperative period from two months to ten years (13). Four of these five patients denied significant insult or trauma, similar to the case we describe. Four other case reports have been published describing various dissociations of bipolar components, in a total of eight additional patients (14, 15, 16, 17).

The exact mechanisms behind component dissociation are variable. The most commonly identifiable factor is the behavior of the prosthesis during the force of dislocation itself and during subsequent attempts at closed reduction of a dislocated prosthesis. If the femoral head component locks against the acetabular rim during dislocation, and the component sizing is not accurate, simultaneous dissociation can occur (13). This mechanism is plausible in our case, with the prosthetic head locking on the superior acetabular rim during dislocation leading to varus displacement of the acetabular cup. Another possible etiology in our case is an intra-acetabular dislocation/dissociation as a result of a weak or defective polyethylene locking ring (16). A defective locking ring can allow slippage to occur such that dissociation may take place without traditional extra-acetabular dislocation.

Dissociation during attempted closed reduction typically results from a “bottle-opener” effect wherein the cup locks on the posterior acetabular rim while traction is applied to the limb (9, 15). These forces can lead to a valgus displacement of the acetabular cup. During reduction, care should be taken to ensure proper fluoroscopic visualization, patient sedation, and avoidance of twisting forces. Post-reduction radiographs should always confirm placement and any variance in component seating should raise suspicion of potential disassociation.

When component dissociation occurs, the treatment of choice is undoubtedly open reduction. The orthopedist should entertain possible modification of prosthetic components or, as in our case, convert to total hip arthroplasty (THA). Conversion to THA alleviates future alignment problems, which is often advantageous given the typical age group of these patients.

In conclusion, we report the radiographic appearance of a dissociated bipolar hip arthroplasty, a rare but increasingly recognized consequence of modular componentry. To avoid this complication, orthopaedic surgeons make great efforts to assure proper fit of the arthroplasty components. Additionally, when dislocation of the bipolar arthroplasty does occur, every effort should be made to avoid dissociation of the components during closed reduction. Radiologists should be aware of the distinction between dislocation of a prosthesis and dissociation of its bipolar components, and be able to describe these complications accurately in their interpretations.

Footnotes

Published: November 22, 2007

References

1. Gomez PF, Morcuende JA. Early attempts at hip arthroplasty–1700s to 1950s. Iowa Orthop J. 2005;25:25–29. [PubMed] [PubMed]
2. Gomez PF, Morcuende JA. A historical and economic perspective on Sir John Charnley, Chas F. Thackray Limited, and the early arthoplasty industry. Iowa Orthop J. 2005;25:30–37. [PubMed] [PubMed]
3. Toledo-Pereyra LH. John Charnley–father of modern total hip replacement. J Invest Surg. 2004;17:299–301. [PubMed] [PubMed]
4. St Clair SF, Higuera C, Krebs V, Tadross NA, Dumpe J, Barsoum WK. Hip and knee arthroplasty in the geriatric population. Clin Geriatr Med. 2006;22:515–533. [PubMed] [PubMed]
5. Goldhill VB, Lyden JP, Cornell CN, Bochner RM. Bipolar hemiarthroplasty for fracture of the femoral neck. J Orthop Trauma. 1991;5:318–324. [PubMed] [PubMed]
6. Cornell CN, Levine D, O'Doherty J, Lyden J. Unipolar versus bipolar hemiarthroplasty for the treatment of femoral neck fractures in the elderly. Clin Orthop Relat Res. 1998:67–71. [PubMed] [PubMed]
7. Ong BC, Maurer SG, Aharonoff GB, Zuckerman JD, Koval KJ. Unipolar versus bipolar hemiarthroplasty: functional outcome after femoral neck fracture at a minimum of thirty-six months of follow-up. J Orthop Trauma. 2002;16:317–322. [PubMed] [PubMed]
8. Parker MJ, Gurusamy K. Arthroplasties (with and without bone cement) for proximal femoral fractures in adults. Cochrane Database Syst Rev. 2006;3 [PubMed] CD001706. [PubMed]
9. Star MJ, Colwell CW, Jr., Donaldson WF, 3rd, Walker RH. Dissociation of modular hip arthroplasty components after dislocation. A report of three cases at differing dissociation levels. Clin Orthop Relat Res. 1992:111–115. [PubMed] [PubMed]
10. Raia FJ, Chapman CB, Herrera MF, Schweppe MW, Michelsen CB, Rosenwasser MP. Unipolar or bipolar hemiarthroplasty for femoral neck fractures in the elderly? Clin Orthop Relat Res. 2003:259–265. [PubMed] [PubMed]
11. Sierra RJ, Schleck CD, Cabanela ME. Dislocation of bipolar hemiarthroplasty: rate, contributing factors, and outcome. Clin Orthop Relat Res. 2006;442:230–238. [PubMed] [PubMed]
12. Varley J, Parker MJ. Stability of hip hemiarthroplasties. Int Orthop. 2004;28:274–277. [PubMed] [PubMed]
13. Georgiou G, Siapkara A, Dimitrakopoulou A, Provelengios S, Dounis E. Dissociation of bipolar hemiarthroplasty of the hip after dislocation. A report of five different cases and review of literature. Injury. 2006;37:162–168. [PubMed] [PubMed]
14. Corteel J, Putz P. [Dislocation-dissociation of a bipolar hip prosthesis] Acta Orthop Belg. 1996;62:173–176. [PubMed] [PubMed]
15. Loubignac F, Boissier F. [Cup dissociation after reduction of a dislocated hip hemiarthroplasty] Rev Chir Orthop Reparatrice Appar Mot. 1997;84:469–472. [PubMed] [PubMed]
16. Tabutin J, Damotte A. [Progressive intra-acetabular dislocation of bipolar hip prostheses: four cases] Rev Chir Orthop Reparatrice Appar Mot. 2004;90:79–82. [PubMed] [PubMed]
17. Tanaka K, Nakayama Y, Murashige R. A dislocation of the inner head in bipolar prosthesis with a self-centering system: a case report. J Nippon Med Sch. 2002;69:192–195. [PubMed] [PubMed]

Articles from Radiology Case Reports are provided here courtesy of Elsevier