PMCCPMCCPMCC

Search tips
Search criteria 

Advanced

 
Logo of nihpaAbout Author manuscriptsSubmit a manuscriptNIH Public Access; Author Manuscript; Accepted for publication in peer reviewed journal;
 
Magn Reson Med. Author manuscript; available in PMC Jan 1, 2014.
Published in final edited form as:
PMCID: PMC3714345
NIHMSID: NIHMS456713
Improved Slice-Selective Adiabatic Excitation
Priti Balchandani,1* Gary Glover,1 John Pauly,2 and Daniel Spielman1
1Department of Radiology, Stanford University, Stanford, California, USA
2Department of Electrical Engineering, Stanford University, Stanford, California, USA
*Correspondence to: Priti Balchandani, Ph.D., Radiology Department, Mount Sinai School of Medicine, 1470 Madison Avenue, Floor 1, New York, NY 10029. priti.balchandani/at/mssm.edu
Purpose
The purpose of this work is to design an improved Slice-selective Tunable-flip AdiaBatic Low peak-power Excitation (STABLE) pulse with shorter duration and increased off-resonance immunity to make it suitable for use in a greater range of applications and at higher field strengths. An additional aim is to design a variant of this pulse to achieve B1-insensitive, fat-suppressed excitation.
Methods
The adiabatic SLR algorithm was used to generate a more uniform spectral pulse envelope for this improved radiofrequency pulse for adiabatic slice-selective excitation, called STABLE-2. Pulse parameters were adjusted to design a version of STABLE-2 with a spectral null centered on lipids.
Results
In vivo images obtained of the human brain at 3 and 7 T demonstrate that STABLE-2 provides robust, uniform, slice-selective excitation over a range of B1 values. Phantom and in vivo knee images obtained at 3 T demonstrate the effectiveness of STABLE-2 for fat suppression.
Conclusions
STABLE-2 achieves B1-insensitive slice-selective excitation while providing greater off-resonance immunity and a shorter pulse duration, when compared to the original STABLE pulse. In particular, the 9.8-ms STABLE-2 pulse provides slice selectivity over 120 Hz whereas the 21-ms STABLE pulse is limited to 80 Hz off-resonance. B1-Insensitive fat-suppressed excitation may also be achieved by using a variant of this pulse.
Keywords: adiabatic, Shinnar Le-Roux, RF excitation, STABLE, slice-selective, off-resonance, B1-insensitive, fat suppression
Adiabatic radiofrequency (RF) pulses are a powerful means with which to achieve uniform excitation in the presence of a nonuniform B1 field. Although many alternatives exist for slice-selective adiabatic 180° pulses to refocus or invert magnetization, options for slice-selective adiabatic excitation are more limited (13), and require high RF and/or gradient amplitude. A Slice-selective Tunable-flip AdiaBatic Low peak-power Excitation (STABLE) pulse was introduced in 2008 (4), which consists of an oscillating gradient in conjunction with a BIR-4-like RF envelope that is sampled by many short spatial subpulses to achieve spatial selectivity. The pulse functions within gradient and RF amplifier limits of current commercial clinical scanners. However, the long pulse duration of 21 ms and limited off-resonance immunity of 80 Hz make the pulse unsuitable for use in many applications, especially in the presence of increased B0-inhomogeneity at higher field strengths. Recently Moore et al. (5) designed composite pulse envelopes that were numerically optimized to achieve B1- and B0-insensitivity and sampled these envelopes with spatial subpulses to achieve slice-selectivity. The strategy used to integrate slice-selectivity was very similar to that used for STABLE. Using this numerically optimized approach, Moore et al. achieved shorter pulse duration and consequently greater off-resonance immunity. In this work, we present an analytical method to improve upon the original STABLE pulse. We have modified the BIR-4 envelope with a highly truncated adiabatic SLR (6) spectral RF pulse envelope to reduce the STABLE pulse duration and increase off-resonance immunity. Our redesigned STABLE-2 pulse has a duration of 9.8 ms and off-resonance immunity of 120 Hz. In vivo experiments demonstrate B1-insensitivity of the pulse.
Multitransmit systems with custom RF pulses that utilize the “spokes” excitation k-space trajectory have been shown to achieve very uniform transmit B1 profiles (79). This type of setup has been used to obtain in vivo images of the brain at 7 T (10). However, the heterogeneous SAR profiles resulting from multiple RF transmission are not fully understood, currently limiting flip angles to very small values to remain within safety limits. Furthermore, such an approach requires acquisition of subject-specific field maps to generate the custom RF pulses and requires the scanner to be equipped with parallel transmit hardware. Thus, it remains valuable to have single-channel solutions for uniform B1-insensitive, slice-selective RF excitation. STABLE-2 offers such a solution.
We also recognized the potential to exploit the intrinsic spectral nulls that occur in the STABLE spectral profile to suppress signal from particular chemical species, such as fat. We redesigned the spectral envelope of the pulse to have a stop-band encompassing the largest lipid resonances at 3 T and tested the performance of this new version of the STABLE-2 pulse in a fat–water phantom and a human knee. Effective B1-insensitive fat suppression was achieved by the STABLE-2 pulse.
Design Algorithm
Our first step was to design a more uniform spectral adiabatic BIR-4 pulse envelope to reduce the peak RF amplitude of the final STABLE-2 pulse. We used the adiabatic SLR technique described in Ref. 6 to generate an adiabatic full passage pulse. The adiabatic SLR method allows the pulse designer to apply additional quadratic phase across the pulse profile to achieve a more uniform distribution of RF energy.
The adiabatic SLR 180° pulse was designed to have a 516-Hz bandwidth and was truncated to a 4.89-ms duration with a peak RF amplitude of 13 μT. The pulse waveform was truncated well beyond the 5% of peak amplitude as suggested in Ref. 6, and the spectral profile had low selectivity. However, as the final STABLE-2 pulse is meant to be only slice-selective and not spectrally selective, the spectral selectivity of the adiabatic envelope may be low without deleteriously affecting the off-resonance behavior.
The full passage adiabatic SLR pulse was divided into two half passage segments. As in the conventional BIR-4 design, the spectral envelope was made up of four adiabatic SLR half-passage segments, with the first and third segments being time-reversed. A phase discontinuity was introduced between the first and second segments and between the third and fourth segments to produce the desired flip angle, which was 90°. As in the original STABLE design, the adiabatic spectral envelope was sampled by small tip-angle subpulses to provide spatial selectivity. The subpulses used were windowed sinc pulses with a time–bandwidth product of 3. The number of subpulses was chosen to faithfully sample the spectral envelope while maintaining long enough subpulse duration to allow accrual of sufficient gradient area to achieve a slice thickness of ~ 5 mm. Nearly triangular gradient lobes and VERSEd (11) subpulses were used to limit the gradient slew rate to remain below gradient hardware and peripheral nerve stimulation (PNS) limits. The resulting spatial bandwidth was 4600 Hz. The final pulse had a duration of 9.8 ms, peak RF amplitude at adiabatic threshold of 13 μT, 15 spatial subpulses, and slice selectivity over a 120-Hz off-resonance range. As a comparison, the original STABLE pulse had the following pulse parameters: 21-ms duration, 80-Hz off-resonance immunity, peak RF amplitude at adiabatic threshold of 10 μT, 33 subpulses, a spatial bandwidth of 4300 Hz, and 5-mm minimum slice thickness. The amplitude waveform (with the spectral envelope superimposed for illustrative purposes) and the phase and gradient waveforms for the STABLE-2 pulse are shown in Figure 1.
FIG. 1
FIG. 1
RF (a) amplitude, (b) phase, and (c) gradient waveforms for the STABLE-2 RF pulse. In (a), the spectral envelope composed of truncated adiabatic SLR pulse segments is overlaid on the RF pulse amplitude as a dotted line.
The RF energy deposition as defined by the specific absorption rate (SAR) value for the 9.8-ms STABLE-2 pulse was calculated and compared to the original 21-ms STABLE pulse presented in Ref. 4. At adiabatic threshold, the SAR for the STABLE-2 pulse was 92% of the SAR of the original STABLE pulse. Profile shape and maximum pulse amplitude at adiabatic threshold are factors determining the calculated SAR value.
Simulations were performed in MATLAB by calculating the Cayley–Klein parameters, α and β, for the piecewise-constant RF pulse as described in Ref. 12. The expressions for the Cayley–Klein parameters represent a series of rotations and may be used to generate the spectral profile for the pulse, which is given by Eq. 1,
equation M1
[1]
where the initial longitudinal magnetization is assumed to be 1. Note that this is equivalent to the calculations performed by a discrete-time Bloch simulator (12)
The 2D simulated spatial-spectral profiles over a 120-Hz and 1000-Hz range in off-resonance are shown for the 90° STABLE-2 pulse in Figure 2a,b. Vertical cross sections through these 2D profiles yield the selected slice at a particularly frequency. As shown in Figure 2a, the slice profile remains constant for a ±60-Hz shift in resonant frequency.
FIG. 2
FIG. 2
2D plots of slice profile versus off-resonance frequency over (a) a 120-Hz range and (b) a 1000-Hz range.
The simulated slice profile behavior plotted for RF overdrive factors ranging from 1 to 2 is shown in Figure 3. The RF overdrive factor is equal to the amplitude at which we operate the pulse divided by the amplitude of the pulse at adiabatic threshold. Therefore, a RF overdrive factor of 1.2 would mean a 20% increase of pulse amplitude above the adiabatic threshold. The fidelity of the slice is maintained over a 50% increase in B1 above adiabatic threshold, at which point slight signal loss appears at the center of the profile due to overdriving the subpulses.
FIG. 3
FIG. 3
Simulated slice profile produced by a 9.8-ms STABLE-2 pulse for a range of B1 amplitudes. The B1 overdrive factor is equal to the amplitude at which we operate the pulse divided by the amplitude of the pulse at adiabatic threshold. Some loss at the center (more ...)
Leveraging Nulls for Fat Suppression
The STABLE-2 pulse is not spectrally selective in the way that conventional spatial-spectral (SPSP) excitation pulses (13) or adiabatic 180° SPSP pulses (14) are, as it does not produce a spectral profile with well-defined passbands and stop bands. This is because the adiabatic BIR-4 pulse envelope is nonspectrally selective. However, after the pulse envelope is sampled by spatial subpulses, there is aliasing of the nonselective BIR-4 spectral profiles. Adjacent nonspectrally selective aliased profiles interplay in such a way that there are regions of null signal, which can be leveraged for our purposes. Figure 2b demonstrates distortions in the 2D magnetization profile for STABLE-2, which appear beyond a 120-Hz range of off-resonance. We recognized an opportunity to exploit these frequency ranges where the profile “pinches” off to create nulls, to suppress signal from chemical species such as fat. We designed a variant of the original pulse design to achieve this purpose.
Targeting implementation at 3 T, we first redesigned the adiabatic SLR spectral envelope to have a 242-Hz bandwidth. This envelope was divided into segments and sampled with subpulses using the same method that is described for the nonselective STABLE-2 pulse. The time-bandwidth product of the subpulses used was reduced to 1 in order to achieve thinner slices at the cost of less spatial selectivity. The pulse duration was decreased to 9.2 ms to further stretch and shift the profile null to be closer to the fat resonance frequency. We also introduced a phase shift in the initial 180° adiabatic SLR waveform to generate a STABLE-2 pulse with a slightly asymmetric 2D spectral profile with a wider stop band on fat. Amplitude and phase waveforms for the selective STABLE-2 pulse for fat-suppressed excitation are shown in Figure 4a,b. The additional phase is evident in the phase waveform plotted in Figure 4b. Figure 5a shows the behavior of the simulated 2D magnetization profile over a 1000-Hz frequency range. There is a stop band in the 2D profile centered at 410 Hz. When using the pulse in a sequence, we shift the transmit frequency by −20 Hz to better encompass the lipid peaks that are centered at ~440 Hz at 3 T. This reduces the off-resonance immunity of the excitation passband. However, there is still sufficient undistorted slice-selective behavior over a 100-Hz range in frequencies to excite water adiabatically. Alternatively, we could shorten the pulse duration to stretch out the spectral profile of the pulse, but this would require shorter subpulses and increase the minimum achievable slice-thickness as dictated by gradient limitations. Figure 5b shows the 2D profile for the 200–500 Hz frequency range for several different B1 amplitudes ranging from 0 to 50% above adiabatic threshold. The lipid stop band becomes appreciably narrower, as B1 varies beyond 30% above the adiabatic threshold due to bleeding from adjacent side bands.
FIG. 4
FIG. 4
RF (a) amplitude and (b) phase waveforms for STABLE-2 RF pulse designed for fat-suppressed excitation. An adiabatic SLR envelope with a lower spectral BW is used to enlarge the nulls in the profile.
FIG. 5
FIG. 5
2D plots of simulated slice profile versus off-resonance frequency for STABLE-2 RF pulse designed for adiabatic fat-suppressed excitation. (a) Slice profile over a 1000-Hz frequency range showing a stop band/null in the profile used to suppress lipid (more ...)
Experiments to Validate Pulse Performance
In vivo data were initially obtained from the brain of a normal volunteer scanned at 3 T (Echospeed whole-body magnet; GE Healthcare, Waukesha, WI) with a standard birdcage head coil. Axial images of the brain were obtained using the STABLE-2 pulse shown in Figure 1 instead of a conventional windowed sinc pulse for excitation in a gradient-recalled echo (GRE) sequence. Acquisition parameters were: echo time/pulse repetition time (TE/TR) 10/300 ms, 256×128 grid, 2 NEX, 5.5-mm slice, 21×21 cm2 FOV, and 1:17 min scan time. Several such images were obtained with the STABLE-2 pulse scaled, in increments of 10%, from 0 to 50% above the adiabatic threshold. Images of the same axial slice through the brain were also obtained using the GRE sequence with B1 scaled from −20% to +30% of the nominal pulse amplitude (set at prescan). Variations in signal and contrast with B1 scaling were compared for images obtained using the two sequences.
In vivo scans were also performed at 7 T (Echospeed whole-body magnet; GE Healthcare) using a Nova Medical (Wilmington, MA) two-channel quadrature transmit and 32-channel parallel-receive head coil. An axial slice through the brain was acquired using the same STABLE-2 sequence as the 3 T scans and a conventional GRE sequence. Acquisition parameters for both scans were: TE/TR: 9/500 ms, 256 × 256 grid, 1 NEX, 5-mm slice, 22 × 22 cm2 FOV, 2:08 min scan time. The STABLE-2 pulse was scaled, in increments of 10%, to 60% above the RF adiabatic threshold amplitude. A similar experiment was conducted using a standard GRE sequence with a SLR pulse scaled from −40 to +20% of the nominal pulse amplitude.
A transmit B1 map of the same slice was obtained using a Bloch–Siegert (B–S) B1 mapping sequence (15,16) utilizing a 1-ms adiabatic B–S pulse as described in Ref. 17 with the following acquisition parameters: TE/TR: 5/184 ms, FA = 30°, slice thickness = 5 mm and matrix size = 64 × 64, FOV = 22 × 22 cm2 and scan time = 50 s. A receive sensitivity map was also calculated using a method described in Ref. 18, utilizing the measured B–S B1 map and a gaussian-filtered image (to remove spin density). The conventional GRE and STABLE-2 images were divided by the normalized receive sensitivity to compensate for signal attenuation on the receive side. These receive-compensated images were compared and evaluated for signal and contrast uniformity.
Phantom and in vivo experiments were conducted to verify the performance of the STABLE-2 pulse designed for fat-suppressed excitation. The scans were performed on a 3 T scanner (Echospeed whole-body magnet; GE Health-care) using a GE eight-channel receive/quadrature transmit knee coil. First, a cylindrical phantom containing water and several vials of peanut oil was imaged with a GRE sequence utilizing the STABLE-2 pulse for fat-suppressed excitation and then compared to an image obtained using a conventional GRE sequence. Acquisition parameters were: TE/TR: 9.5/1000 ms, 256×128 grid, 4-mm slice, 22×22 cm2 FOV, and 2:08 min scan time. We were able to increase gradient amplitude to achieve 4 mm slice thickness without resulting in PNS due to the reduced time-bandwidth of the subpulses. A frequency offset of −20 Hz was introduced to shift the spectral profile of the pulse, so that the spectral null better encompassed the lipid resonances centered at 440 Hz. Next, we performed a similar experiment in vivo at 3 T. We scanned a knee of a healthy volunteer, so we could assess the effectiveness of the pulse in suppressing fat signal from the bone marrow. As in the phantom experiments, we compared images obtained using STABLE-2 to those obtained using a conventional GRE sequence. We also scaled the STABLE-2 pulse to 25% above adiabatic threshold to test insensitivity to B1 variation. Acquisition parameters were: TE/TR: 9.5/500 ms, 256×256 grid, 1 NEX, 4-mm slice, 22 × 22 cm2 FOV, and 2:08 min scan time.
Figure 6 shows the 3 T axial images of the brain obtained using the STABLE-2 pulse, scaled to several different RF amplitudes. Images obtained using a conventional GRE sequence for a similar range of B1 amplitudes are also shown. The STABLE-2 images exhibit minimal change in SNR and contrast at different B1 values, whereas the GRE images suffer from significant signal and contrast variations.
FIG. 6
FIG. 6
In vivo axial images of the brain of a normal volunteer obtained at 3 T using the STABLE-2 pulse in a GRE sequence to excite and image a 5.5-mm slice (TE/TR = 10/300 ms). Several such images were obtained with the STABLE-2 pulse scaled from 0 to 50% above (more ...)
Figure 7a,b shows axial images of the brain obtained using the conventional GRE and STABLE-2 GRE sequences. The B–S B1 map and the calculated receive sensitivity map of the same slice are shown in Figure 7c,d. All images in Figure 7a have been compensated by the receive sensitivity profile provided in Figure 7d to assess the transmit uniformity of the pulse. The contrast between the gray and white matter in the brain is less for the 7 T images when compared to the 3 T images in Figure 6 due to changes in T1 as field strength increases. This should be decoupled from the assessment of the transmit uniformity achieved by the RF pulses used.
FIG. 7
FIG. 7
In vivo axial images of the brain of a normal volunteer obtained at 7 T using the STABLE-2 pulse in a GRE sequence to excite and image a 5-mm slice (TE/TR = 9/500 ms). a: Images obtained using the GRE sequence with B1 scaled from −40 to +20% of (more ...)
As seen in Figure 7, the conventional GRE images obtained at 7 T suffer from significant signal and contrast changes in regions where the B1 field, shown in Figure 7c, reaches its highest and lowest values. Conversely, the STABLE-2 images demonstrate minimal variation, even at the peaks and troughs of the B1 field map, as RF amplitude is scaled. The first image in the STABLE-2 series suffers some loss in areas of low B1, likely because adiabatic threshold is not yet reached in those areas.
Figure 8a shows an axial image of the cylindrical fat–water phantom obtained using a conventional GRE sequence at 3 T. Signal from the fat in the vials is bright and spatially shifted (due to chemical shift) from the water. Figure 8b shows the same slice excited using the variant of the STABLE-2 pulse with a null on lipid resonances. Fat signal is effectively suppressed.
FIG. 8
FIG. 8
Image of a slice through a cylindrical water phantom containing several vials of oil obtained at 3 T using (a) a standard GRE sequence and (b) the 90° STABLE-2 pulse used for fat-suppressed excitation in a GRE sequence. Scan parameters for both (more ...)
Figure 9 shows the in vivo knee images obtained at 3 T using a conventional GRE and STABLE-2 sequence. The conventional GRE image in Figure 9a shows fat signal in the marrow of the femur, tibia, and patella. Figure 9b,c shows the same slice of the knee imaged using the STABLE-2 pulse with a null on lipids at the nominal B1 and 25% above the nominal B1. Fat signal in the marrow of the knee bones is largely suppressed, enabling better visualization of the articular cartilage in the knee joint.
FIG. 9
FIG. 9
In vivo knee images obtained at 3 T using (a) a standard GRE sequence and a GRE sequence using a 90° STABLE-2 RF pulse designed with a null on the lipid resonances, set to (b) the nominal B1 at prescan, and (c) 25% greater than the Nominal B1 (more ...)
We have designed the STABLE-2 RF pulse, a shorter, more B0-insensitive option for slice-selective adiabatic excitation than the previously proposed STABLE pulse. The shorter pulse duration increases the pulse’s applicability by enabling its use in a wider range of pulse sequences and at higher field strengths. Two variants of the STABLE-2 pulse were investigated: (1) a nonselective version that is designed to achieve robust excitation profiles in the presence of B1 and B0 inhomogeneity and (2) a selective version that is designed to provide effective fat suppression, while retaining B1-insensitive slice-selective excitation.
In vivo data at 3 and 7 T demonstrate that STABLE-2 performs adiabatically over approximately a 50% increase in the B1-amplitude above the adiabatic threshold. The pulse is better suited than the previously proposed STABLE pulse (4) for imaging at high magnetic fields, such as 7 T, due to increased robustness to B0-shifts and a shorter pulse duration enabling shorter echo times.
Intrinsic nulls in the off-resonance profile were also leveraged to suppress signal from fat in the selective version of the pulse. Phantom and in vivo data at 3 T demonstrate that simultaneous adiabatic slice-selective excitation and fat suppression are achieved by this version of STABLE-2. To use STABLE-2 for fat-suppressed excitation at 7 T, the spectral profile would have to be redesigned to achieve a null at the scaled frequency difference between water and fat, which is ~1026 Hz.
STABLE-2 may provide a valuable B1-insensitive alternative to existing fat suppression techniques. A common method for fat suppression is the use of frequency selective presaturation (FATSAT) (19) pulses that are not adiabatic and susceptible to B1-inhomogeneity (20). Water excitation using SPSP pulses results in more robust fat-suppression but loss in water excitation efficiency, as the B1 field varies (21). Short tau inversion recovery (22,23) utilizes an adiabatic 180° pulse for inversion of all spins and delays excitation until an inversion time when fat is passing through a null. Due to the use of an adiabatic pulse, short tau inversion recovery is inherently more resistant to changes in B1. However, as fat suppression is achieved by exploiting the differential T1 value of tissues, SNR efficiency reduces as T1 values converge with increasing field strength. Spectrally selective adiabatic inversion recovery has been proposed, which does not depend on differential T1’s but still requires waiting a inversion time prior to excitation (24). B1-Insensitive RF pulse trains may be used to achieve fat suppression over a limited range of B1 values, but still require a total duration of 77 ms (25). STABLE-2 offers a B1- and T1-insensitive option for water only excitation and requires no additional preparation time. STABLE-2 does have the disadvantage of having greater SAR than a FATSAT pulse or conventional SPSP excitation pulse. However, the pulse functioned well within SAR limits at both 3 and 7 T when integrated into a GRE sequence with a 500-ms TR, as shown in the “Results” section. Also, narrowing of the lipid spectral stop band beyond a 30% increase in B1-amplitude above the adiabatic threshold results in incomplete fat suppression for STABLE-2. Further exploration of new spectral envelopes with a wider stop band on lipids would make this a more robust fat suppression technique.
We investigated the 2D magnetization profile of STABLE pulses for flip angles below 90° and found that off-resonance immunity for all STABLE designs is reduced for excitation flip angles below 90°. Figure 10 shows the simulated slice profile versus off-resonance frequency for a 45° STABLE-2 pulse. The pulse was created by reducing the phase offset between adiabatic half-passage pulse segments to equal the desired flip angle. As the off-resonance frequency approaches ±60 Hz, distortion lobes appear in the 2D profile and the excited magnetization gradually increases to 100%, which deviates from the expected flip angle. Therefore, the pulse only provides the expected flip angle over a narrower frequency range of ~50 Hz.
FIG. 10
FIG. 10
Simulated slice profile versus off-resonance frequency for STABLE-2 RF Pulse designed for a 45° excitation flip angle. The desired flip angle is achieved for approximately a 50-Hz range in off-resonance.
Limited off-resonance immunity is an intrinsic disadvantage of STABLE pulses. The use of insert gradients may allow shorter pulse duration, thereby increasing the frequency range over which undistorted slice-selection is achieved. A higher gradient maximum amplitude and slew rate will also enable thinner minimum slice thickness. However, for some subjects, PNS may occur before reaching these limits. Due to the use of short subpulses to achieve slice selection in all STABLE designs, exciting thin slices remains a challenge. Slice selectivity may also be traded off with slice thickness, that is, subpulses with lower spatial bandwidth may be used to generate thinner slices with less sharp edges. This technique was employed in the version of STABLE-2 for fat suppressed excitation.
STABLE, STABLE-2, and the numerically optimized slice-selective composite pulses proposed by Moore et al. (5) are all part of a family of SPSP-like nonlinear phase pulses, which exhibit adiabatic and slice-selective behavior. The pulses by Moore et al. achieve many of the same design goals as STABLE-2 and offer a powerful alternative for B1-insensitive excitation. We expect that there probably exist more RF pulses that are members of this family that remain to be explored. Similar limits exist on the slice thickness that may be achieved by these pulses, as slice thickness is determined by the subpulse spatial bandwidth and the strength of the gradients used. Moore et al. explored gaussian and sinc subpulses and operated at gradient limits to achieve slices as thin as 2 mm. In the design of STABLE-2, different subpulse waveforms were not fully explored to decrease slice thickness. However, subpulses can easily be adjusted within this design framework to explore these trade offs. Moore et al. were successful in shortening pulse duration to 5 ms and achieved slice-selective behavior over a 200-Hz range with similar B1-immunity but less selective slice profiles than STABLE-2. As STABLE-2 enforces an adiabatic envelope, it is challenging to reduce pulse duration to 5 ms without increasing peak RF amplitude to prohibitively high values. This may be an inherent disadvantage of STABLE-2 when compared to the composite pulses by Moore et al. However, the fact that STABLE-2 uses an adiabatic BIR-4 envelope, which is systematically generated by the adiabatic SLR algorithm (6) makes it possible for the pulse designer to have greater control over the final shape of the spatial-spectral profile. The fat-nulling version of STABLE-2 presented in this work is an example of how pulse parameters may be adjusted to predictably change profile shape.
Contrast differences due to the spin-locked state of the magnetization during the STABLE-2 pulse were not explored in this study. These differences are less pronounced than those observed for the original STABLE pulse due to the shorter duration of the pulse. However, they remain a subject of future study.
Acknowledgments
Grant sponsor: Lucas Foundation; Grant numbers: NIH R01 MH080913, NIH-NINDS K99NS070821, and P41 EB015891; Grant sponsor: GE Healthcare.
The authors thank Mehdi Khalighi for his help with B1 transmit and receive sensitivity maps.
1. Johnson J, Garwood M, Ugurbil K. Slice selection with gradient modulated adiabatic excitation despite the presence of large B1 inhomogeneities. J Magn Reson. 1989;81:653–660.
2. de Graaf RA, Nicolay K, Garwood M. Single-shot, B1-insensitive slice selection with a gradient-modulated adiabatic pulse, BISS-8. Magn Reson Med. 1996;35:652–657. [PubMed]
3. Hsu EW, Reeder SB, MacFall JR. Single-shot, variable flip-angle slice-selective excitation with four gradient-modulated adiabatic half-passage segments. Magn Reson Med. 1998;40:334–340. [PubMed]
4. Balchandani P, Pauly J, Spielman DM. Slice-selective tunable-flip adiaBatic low peak-power excitation (STABLE) pulse. Magn Reson Med. 2008;59:1072–1078. [PMC free article] [PubMed]
5. Moore J, Jankiewicz M, Anderson A, Gore J. Slice-selective excitation with equation M2-insensitive composite pulses. J Magn Reson. 2012;214:200–211. [PMC free article] [PubMed]
6. Balchandani P, Pauly J, Spielman DM. Designing adiabatic RF pulses using the Shinnar-Le Roux algorithm. Magn Reson Med. 2010;64:843–851. [PMC free article] [PubMed]
7. Zhu Y. Parallel excitation with an array of transmit coils. Magn Reson Med. 2004;51:775–784. [PubMed]
8. Katscher U, Bornert P, Leussler C, van den Brink J. Transmit sense. Magn Reson Med. 2003;49:144–150. [PubMed]
9. Grissom W, Yip C, Zhang Z, Stenger V, Fessler J, Noll D. Spatial domain method for the design of RF pulses in multicoil parallel excitation. Magn Reson Med. 2006;56:620–629. [PubMed]
10. Setsompop K, Alagappan V, Gagoski B, Witzel T, Polimeni J, Potthast A, Hebrank F, Fontius U, Schmitt F, Wald L, Adalsteinsson E. Slice-selective RF pulses for in vivo equation M3 inhomogeneity mitigation at 7 tesla using parallel RF excitation with a 16-element coil. Magn Reson Med. 2008;60:1422–1432. [PMC free article] [PubMed]
11. Conolly S, Nishimura D, Macovski A, Glover G. Variable-rate selective excitation. J Magn Reson (1969) 1988;78:440–458.
12. Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm. IEEE Trans Med Imaging. 1991;10:53–65. [PubMed]
13. Meyer CH, Pauly JM, Macovski A, Nishimura DG. Simultaneous spatial and spectral selective excitation. Magn Reson Med. 1990;15:287–304. [PubMed]
14. Conolly S, Pauly J, Nishimura D, Macovski A. Two-dimensional selective adiabatic pulses. Magn Reson Med. 1992;24:302–313. [PubMed]
15. Sacolick L, Wiesinger F, Hancu I, Vogel M. B1 mapping by Bloch-Siegert shift. Magn Reson Med. 2010;63:1315–1322. [PMC free article] [PubMed]
16. Khalighi MM, Rutt BK, Kerr AB. RF pulse optimization for Bloch-Siegert equation M4 mapping. Magn Reson Med. 2012;68:857–862. [PMC free article] [PubMed]
17. Khalighi MM, Rutt BK, Kerr AB. Adiabatic pulse design for Bloch-Siegert equation M5 mapping. Magn Reson Med. 2012 doi: 10.1002/mrm.24507. [PMC free article] [PubMed] [Cross Ref]
18. Khalighi MM, Zeineh M, Rutt BK. Intensity correction at 7 T using Bloch-Siegert equation M6 mapping. Proceedings of the 20th Annual Meeting of ISMRM; Melbourne, Australia. 2012. p. 3361.
19. Haase A, Frahm J, Hanicke W, Matthaei D. 1H NMR chemical shift selective (CHESS) imaging. Phys Med Biol. 1985;30:341–344. [PubMed]
20. Delfaut E, Beltran J, Johnson G, Rousseau J, Marchandise X, Cotten A. Fat suppression in MR imaging: techniques and pitfalls. Radiographics. 1999;19:373–382. [PubMed]
21. Block W, Pauly J, Kerr A, Nishimura D. Consistent fat suppression with compensated spectral-spatial pulses. Magn Reson Med. 2005;38:198–206. [PubMed]
22. Bydder GM, Hajnal JV, Young IR. MRI: use of the inversion recovery pulse sequence. Clin Radiol. 1998;53:159–176. [PubMed]
23. Spielman DM, Pauly JM, Macovski A, Glover GH, Enzmann DRA. Lipid-suppressed single- and multisection proton spectroscopic imaging of the human brain. J Magn Reson Imaging. 1992;2:253–262. [PubMed]
24. Balchandani P, Spielman DM. Fat suppression for 1H MRSI at 7 T using spectrally selective adiabatic inversion recovery. Magn Reson Med. 2008;59:980–988. [PMC free article] [PubMed]
25. Abe T. Fast fat suppression RF pulse train with insensitivity to B1 inhomogeneity for body imaging. Magn Reson Med. 2011;67:464–469. [PubMed]