The experimental arrangement employs a standard pump-probe configuration, shown in
, where the pump is modulated at 1 MHz with an acousto-optic modulator (AOM), and a motorized delay stage controls pump-probe delay,
τ. Probe and pump pulses are supplied by a Ti:sapph oscillator (Spectra-Physics Tsunami), and an intracavity-doubled optical parametric oscillator (Coherent Mira OPO), capable of providing pump and probe wavelength combinations from 520 nm down to 880 nm. The pump and probe pulses are focused onto a sample with a microscope objective (here we use either a 10x / 0.25 NA dry or a 40x / 0.8 NA water immersion) and re-collimated afterwards by 1.1 NA condenser (the high-NA condenser helps to minimize artifacts from thermal lensing and Kerr lensing [
6]).
Cross-phase modulation (XPM) shifts the probe pulse spectrum in a manner that depends on the pump-probe delay
τ [
11]. This shift may be detected by splitting the probe pulse at its center wavelength, and looking for an imbalance between the two spectral halves. After the condenser, the pump is rejected and the probe beam is split at its center wavelength by a dielectric filter (Chroma D800/20m), angle-tuned to set the filter cutoff at the center of the probe spectrum. The transmitted half of the probe spectrum is directed onto a photodiode. Any shift Δ
ω in the probe spectrum caused by XPM interaction with the modulated pump results in 1 MHz modulation of the probe intensity transmitted through the filter; this signal is detected with a lock-in amplifier. Optionally, a high-speed balanced photodiode (NewFocus model 2307) may be used to suppress other nonlinear pump-probe interactions that do not shift the probe spectrum (such as nonlinear absorption).
As the probe pulse propagates through the sample, it accumulates phase from both the linear and nonlinear indexes of refraction:
Here,
n0 is the linear index of refraction,
n2,SPM
Re{
χ(3)(
-ωpr; ωpr,
ω
pr, -
ωpr)} is the nonlinear index mediating self-phase modulation,
n2,XPM
Re{
χ(3)(
-ωpr; ω
pr,
ωpu, -
ωpu)} is the nonlinear index mediating cross-phase modulation,
I{pu,pr}(
t) are the pump and probe pulse temporal intensity profiles,
ω{pu,pr} are the pump and probe angular frequencies,
L is the effective interaction length, and
c is the velocity of light in vacuum. The pump-induced XPM term,
n2,XPMIpu(
t), may be isolated from the other terms by modulating the pump with an AOM, then filtering the photodiode’s output with a lock-in amplifier.
To roughly estimate the expected signal amplitude, consider a pump-probe delay
τ such that the peak of the probe pulse coincides with the steepest slope of the pump pulse (on either the rising or the falling half of the pump pulse). The resulting XPM may be approximately regarded as linear across the probe pulse in time; this linear temporal phase produces a spectral shift in the probe:
The change in detected intensity (on a single photodiode) in a spectrally shifted probe is
For a balanced detector, the intensity incident on one photodiode increases by Δ
S, while the intensity incident on the other photodiode decreases by Δ
S; the total signal in this case is 2Δ
S.
Consider the case where both pump and probe pulses have Gaussian intensity profiles with full-width at half maximum duration
tfwhm. Centering the probe on the rising slope of the pump shifts the probe to higher frequency; centering the probe on the falling slope of the pump shifts the probe to lower frequency. The maximum spectral shift will be imparted at

, where
Of course, this shift applied to only a small portion of the pulse. A more precise numerical calculation is straightforward with the exact pulse shape, as discussed later, but other factors affect the measurement (such as beam spatial mode quality, residual uncompensated dispersion, and spatial overlap) [
12] so this is only an estimate.