Coherent Raman scanning fiber endoscopy

Abstract

Coherent Raman scattering methods allow for label-free imaging of tissue with chemical contrast and high spatial and temporal resolution. However, their imaging depth in scattering tissue is limited to less than 1 mm, requiring the development of endoscopes to obtain images deep inside the body. Here, we describe a coherent Raman endoscope that provides stimulated Raman scattering images at seven frames per second using a miniaturized fiber scanner, a custom-designed objective lens, and an optimized scheme for collection of scattered light from the tissue. We characterize the system and demonstrate chemical selectivity in mouse tissue images.

Optical imaging methods are vital in medical diagnostics because they offer high spatial resolution and excellent contrast at high acquisition speeds compared to widely used radiological imaging techniques such as magnetic resonance imaging, position emission tomography, or x-ray computed tomography. Optical disease diagnosis can be enhanced in sensitivity by the application of exogenous labels to obtain molecularly specific information from the tissue, but these are sometimes toxic. In addition, most biological tissues exhibit strong absorption and scattering of light, so the penetration of high-resolution optical methods is limited to a few millimeters, restricting optical diagnostics.

Endoscopy circumvents the penetration depth limit by bringing an optical system into proximity with tissues of interest, even inside a living patient. Optical endoscopes with white light contrast have revolutionized surgery, allowing minimally invasive procedures to be performed inside the body [1]. Advanced endoscopic contrast modes have been reported for wide-field fluorescence and narrowband imaging [2,3], confocal reflectance [4] and fluorescence [5], and two-photon fluorescence [3,68].

The combination of novel label-free optical imaging modalities with endoscopic devices presents new opportunities for in situ diagnostics. Coherent Raman scattering (CRS) [911] methods offer label-free chemical contrast with microscopic resolution and have the potential to provide in situ virtual pathology on fresh tissue without the excision, fixation, slicing, and staining procedure required by traditional histopathology [12,13]. However, coherent Raman methods have required laser scanning microscopes. Such systems can be miniaturized [14], but the use of optical fiber delivery is highly desirable for flexibility in clinical applications. Several groups have reported optical modeling [15] and progress toward realizing optical-fiber-delivered CRS imaging systems, but required either sample scanning [16] or scanning mirrors [17] and long scan times. Here, we report a miniaturized coherent Raman imaging system based on a compact scanning fiber endoscope (SFE) [18], a step toward translating coherent Raman imaging methods into clinical practice. In CRS, which refers to both coherent anti-Stokes Raman scattering (CARS) [9,10] and stimulated Raman scattering (SRS) [11,1922], laser beams at two frequencies, called the pump, ωp, and Stokes, ωS, frequencies, are used to illuminate the sample. When the difference frequency between the two beams is tuned to match an intrinsic molecular vibrational frequency in the sample, ωvib, several nonlinear interactions occur. New light is generated at the anti-Stokes frequency, ωas = 2ωpωS, by the CARS process, and some intensity is transferred from the pump to the Stokes beam by the SRS process [Fig. 1(a)]. Both processes offer chemical selectivity, high spatial resolution, and video-rate imaging speeds in vivo in living animals [23] and humans [22]. SRS offers more desirable contrast that depends linearly on the target species concentration and is free from the image artifacts and nonresonant background that plague CARS imaging [11], but has thus far not been implemented in an endoscope or miniature microscope format.

Fig. 1
(a) Energy diagrams of the SRS and CARS processes. (b) Schematic of the device: the Stokes laser beam is amplitude modulated using an EOM driven by a 10 MHz waveform, temporally delayed, and combined with the pump beam on a DM. The beams pass through ...

Because SRS imaging involves measurement of small intensity changes in the pump beam, we make use of a highly sensitive modulation transfer scheme in which we amplitude modulate the Stokes beam at a high reference frequency (e.g., 10 MHz), illuminate the sample with both the pump and Stokes beams, and then detect only the pump beam with a photodiode (PD) behind an optical filter (OF) that blocks the Stokes beam. The PD signal is demodulated by a lock-in amplifier (LIA) tuned to the reference frequency that only detects amplitude modulation when SRS occurs in the sample [11].

Images are formed by scanning the laser over the sample and recording the modulation of the pump beam as a function of position. In this Letter, the laser system that excites the sample is similar to one described recently [22]. It is based on a passively mode-locked Nd:YVO4 laser (picoTrain, high Q laser), which provides 7 ps pulses at a repetition rate of 80 MHz. A portion of the output at 1064 nm is the Stokes beam for the CRS processes, and the remainder of the output is frequency doubled and used to synchronously pump an optical parametric oscillator (OPO; Levante Emerald, APE GmbH), which generates a tunable pump beam near 800 nm.

The Stokes beam is modulated by an electro-optic modulator (EOM; Raicol Crystals, Ltd.) at 10 MHz and combined in space on a dichroic mirror (DM; 1064dcrb, Chroma Technology) with the OPO output. The two beams are linearly polarized and, after passing through an achromatic wave plate (AHWP05M-980, Thorlabs), they are coupled into a single-mode polarization-maintaining (PM) optical fiber (PM780-HP, Thorlabs).

The optical fiber delivers the laser beams to the sample [Fig. 1(b)], and the image is scanned using a miniature piezoelectric (PZ) actuator to drive the fiber tip at its mechanical resonance frequency in an expanding spiral [Fig. 1(c)]. A scan rate of seven frames per second can be achieved. The scan pattern is actuated by passing the optical fiber through a small (<0.5 mm diameter) piezoelectrically active ceramic tube that is radially patterned with four electrodes. To achieve the spiral scan, one pair of opposing electrodes is driven with a sine wave and the other is driven with a cosine wave, both with linearly expanding envelope functions. The scan driver electronics track the fiber tip for image reconstruction. This SFE provides a unique combination of small package size, high frame rate, and adaptability to a variety of laser scanning techniques.

The laser light from the fiber output is focused onto the sample by a gradient index (GRIN) objective lens (OL). Single element GRIN lenses have large axial chromatic aberrations, which significantly degrade the nonlinear signal-generation efficiency. For this reason, we utilized an OL consisting of two GRIN elements sandwiching a diffractive optic, which was designed to have zero axial chromatic aberration at the chosen wavelengths, an NA of 0.3, a free working distance of 150 μm, and an outer diameter of 1.4 mm. Typical power levels of ~130 mW are used, which is similar to those previously used for in vivo imaging [23].

For light collection, we observed that the modulated pump beam that travels forward from the focus is redirected back toward the tissue surface by multiple scattering within turbid tissue. Nonsequential ray-tracing simulations [Fig. 1(d) and [22]] show that using a large area PD surrounding the excitation lens allows for optimal collection efficiency. In this study, we utilized a 10 mm × 10 mm PD, but simulations indicate that a 5 mm diameter circular PD would achieve ~75% of the present collection efficiency. Thus, we surrounded the lens with a back-biased silicon PIN PD (FDS1010, Thorlabs) through which we drilled a 2 mm diameter hole for the OL [Fig. 1(c)]. A custom-designed OF identical to one in a previous report [22] blocked the modulated Stokes beam from reaching the PD. As observed previously [16], because picosecond pulses are employed in CRS, ordinary silica fibers can be used for light delivery over a length of 1 m without self-phase modulation broadening the optical spectrum [Fig. 2(a)] beyond the intrinsic Raman linewidth (~1 nm at 800 nm).

Fig. 2
(a) Optical spectrum of the pump laser beam at the fiber input (black) and propagating through 1 m of PMF with the indicated power level at the fiber output. Even with 240 mW of power output, the FWHM is 0.85 nm, indicating that we can still resolve individual ...

We verified the spatial resolution of the system with transmission mode CARS images of 2 μm diameter beads [Fig. 2(b)]. Bead sizes of 1.3 μm (FWHM) demonstrate the ability to clearly resolve 2 μm objects across the 80 μm diameter field of view. By axially translating the beads in 1 μm steps, we measured the axial FWHM of 6.5 μm.

We then tested the ability of the system to obtain SRS images in mouse skin tissue at seven frames per second (with a 500 pixel diameter field of view). This required the use of a custom LIA [22] to process the SRS signal. With this system, we could tune into the CH2 stretching vibration of lipids [Fig. 3(a)] to record an image of the sebaceous gland [Fig. 3(b)] and subcutaneous fat in mouse skin at a depth of ~50 μm [Fig. 3(c)]. Image contrast disappeared and the signal level dropped to the demodulator noise floor when the OL was not focused in the tissue sample, indicating a negligible background from nonlinear effects within the fiber.

Fig. 3
CRS images with the endoscope. (a) Raman spectra of mouse skin (black), a representative lipid compound (oleic acid, red), and a representative protein compound (soy protein extract, dotted magenta). (b) SRS image of a sebaceous gland approximately 30 ...

We tested the chemical selectivity of the system by recording images of mouse hair near the surface of the tissue. Hair is known to be rich in protein and coated with a layer of lipids that are secreted by the mouse skin. By tuning the laser system to match the CH3 stretching resonance at 2950 cm−1 [Figs. 3(d) and 3(f)] or the CH2 stretching resonance at 2845 cm−1 [Figs. 3(e) and 3(g)], we could see in the SRS images either the protein-rich center of the hair [Fig. 3(d) indicated by the “H” labels) or the lipid-rich coating of the hair [Fig. 3(e)]. This tuning was previously observed in a bulk SRS microscope [22] and demonstrates that the chemical contrast in tissue matches the behavior of the Raman spectrum [Fig. 3(a)].

However, simultaneous forward CARS detection [Figs. 3(f) and 3(g)] did not manifest this behavior because the nonresonant background of CARS limits the chemical selectivity of the technique and does not allow clear discrimination between lipids and proteins based on CH2 and CH3 stretching images. This demonstrates the importance of obtaining SRS contrast because recent work utilizing this contrast has shown that SRS has pathology-like diagnostic capability in fresh tissue [13].

In summary, the system described here demonstrates that the technical challenges for CRS endoscopy, fiber delivery, laser scanning, focusing, light collection, and demodulation can be solved. Improvements to the device, particularly the use of fiber lasers, will increase ease of use, facilitating clinical translation of CRS.

Acknowledgments

We thank C. Lee and D. Melville for design and fabrication of the SFE at the University of Washington; B. Messerschmidt of GrinTech GmbH for design and fabrication of the GRIN lens; and M. Roeffaers, G. Holtom, D. Fu, and X. Zhang for helpful discussions. C. W. Freudiger acknowledges support from a Boehringer Ingelheim Fonds Ph.D. scholarship. This work was supported by National Institutes of Health (NIH) grants to E. J. Seibel (R33 CA 094303 and R21 EB012666) and a T-R01 grant (1R01EB010244-01) to X. S. Xie, the University of Washington Mechanical Engineering Royalty Fund, and the Harvard University Office of Technology Development Accelerator Fund.

References

1. Mack MJ. J Am Med Assoc. 2001;285:568. [PubMed]
2. Gono K. IEEE J Quantum Electron. 2008;14:62.
3. Lee CM, Engelbrecht CJ, Soper TD, Helmchen F, Seibel EJ. J Biophoton. 2010;3:385. [PMC free article] [PubMed]
4. Boudoux C, Yun S, Oh W, White W, Iftimia N, Shishkov M, Bouma B, Tearney G. Opt Express. 2005;13:8214. [PubMed]
5. Kiesslich R, Burg J, Vieth M, Gnaendiger J, Enders M, Delaney P, Polglase A, McLaren W, Janell D, Thomas S. Gastroenterology. 2004;127:706. [PubMed]
6. Llewellyn ME, Barretto RPJ, Delp SL, Schnitzer MJ. Nature. 2008;454:784. [PMC free article] [PubMed]
7. Hoy CL, Durr NJ, Chen P, Piyawattanametha W, Ra H, Solgaard O, Ben-Yakar A. Opt Express. 2008;16:9996. [PMC free article] [PubMed]
8. Kim P, Puoris’haag M, Côté D, Lin CP, Yun SH. J Biomed Opt. 2008;13:010501. [PMC free article] [PubMed]
9. Zumbusch A, Holtom GR, Xie XS. Phys Rev Lett. 1999;82:4142.
10. Evans CL, Xie XS. Annu Rev Anal Chem. 2008;1:883. [PubMed]
11. Freudiger CW, Min W, Saar BG, Lu S, Holtom GR, He C, Tsai JC, Kang JX, Xie XS. Science. 2008;322:1857. [PMC free article] [PubMed]
12. Evans CL, Xu X, Kesari S, Xie XS, Wong STC, Young GS. Opt Express. 2007;15:12076. [PubMed]
13. Freudiger CW, Saar BG, Orringer DA, Pfannl R, Zeng Q, Ottoboni L, Chen T, Ying W, Folkerth RD, French CA, Welch WR, Waeber C, Sims JR, De Jager PL, Sagher O, Philbert MA, Xu X, Kesari S, Xie XS, Young GS. Stain-free histopathology with stimulated Raman scattering microscopy. submitted to Proc Natl Acad Sci USA
14. Murugkar S, Smith B, Srivastava P, Moica A, Naji M, Brideau C, Stys PK, Anis H. Opt Express. 2010;18:23796. [PubMed]
15. Veilleux I, Doucet M, Coté P, Verreault S, Fortin M, Paradis P, Leclair S, Da Costa RS, Wilson BC, Seibel EJ. Proc SPIE. 2010;7558:75580D.
16. Légaré F, Evans CL, Ganikhanov F, Xie XS. Opt Express. 2006;14:4427. [PubMed]
17. Balu M, Liu G, Chen Z, Tromberg BJ, Potma EO. Opt Express. 2010;18:2380. [PMC free article] [PubMed]
18. Seibel EJ, Brown CM, Dominitz JA, Kimmey MB. Gastrointest Endosc Clin N Am. 2008;18:467. [PMC free article] [PubMed]
19. Ploetz E, Laimgruber S, Berner S, Zinth W, Gilch P. Appl Phys B. 2007;87:389.
20. Ozeki Y, Dake F, Kajiyama S, Fukui K, Itoh K. Opt Express. 2009;17:3651. [PubMed]
21. Nandakumar P, Kovalev A, Volkmer A. New J Phys. 2009;11:033026.
22. Saar BG, Freudiger CW, Reichman J, Stanley CM, Holtom GR, Xie XS. Science. 2010;330:1368. [PMC free article] [PubMed]
23. Evans CL, Potma EO, Puoris’haag M, Côté D, Lin CP, Xie XS. Proc Natl Acad Sci USA. 2005;102:16807. [PubMed]