Miniaturized Lensless Imaging Systems for Cell and Microorganism Visualization in Point-of-Care Testing

Abstract

Low-cost, robust, and user-friendly diagnostic capabilities at the point-of-care (POC) are critical for treating infectious diseases and preventing their spread in developing countries. Recent advances in micro- and nano-scale technologies have enabled the merger of optical and fluidic technologies (optofluidics) paving the way for cost-effective lensless imaging and diagnosis for POC testing in resource limited settings. Applications of the emerging lensless imaging technologies include detecting and counting cells of interest, which allows rapid and affordable diagnostic decisions. This review presents the advances in lensless imaging and diagnostic systems, and their potential clinical applications in developing countries. The emerging technologies are reviewed from a POC perspective considering cost-effectiveness, portability, sensitivity, throughput and ease of use for resource-limited settings.

Keywords: Microfluidics, lensless imaging, point-of-care, resource-limited settings

1. Introduction

The challenges associated with delivering healthcare in point-of-care (POC) are significant in developing countries [14]. Prevention and treatment of diseases require accurate diagnosis, which is generally achieved with trained personnel and equipment readily available in developed countries. However, the limited availability of qualified personnel, adequate infrastructure and costly medical instruments present significant challenges in treating and preventing the spread of especially the communicable diseases in resource-limited countries [57]. There is a significant need for affordable and simple diagnostic technologies for infectious diseases and for measurement of patients’ health conditions in these countries [6, 8].

World Health Organization (WHO) has recognized the critical requirements and outlined the standards for diagnostic instruments for resource limited settings [7, 911]. According to WHO standards, these devices need to be inexpensive, disposable and easy to use [12]. They should also be functional under heat and humidity, given the lack of refrigeration, reliable electricity, and clean water resources in many developing countries. The diagnostic devices that can be utilized in POC generally require an analysis equipment, which need to be automated and operable within clinically acceptable sensitivity without extensive technical training. A comparison between the needs for resource-limited POC diagnosis and conventional diagnosis has been presented in the light of WHO standards in Table 1, which clearly displays the differences and the specific requirements in these two settings. These specific needs impose challenges and require innovative approaches for efficient delivery of healthcare in POC in resource limited settings [3].

Table 1
The Differences in Needs for Resource-limited Point-of-Care (POC) Diagnosis and Conventional Diagnosis based on World Health Organization (WHO) Standards

Lab-on-a-chip technologies have been emerging in detection and monitoring of infectious diseases at resource-limited settings [3, 5, 1324]. Healthcare personnel including technicians, nurses and physicians can use these devices with minimum training to diagnose patients for infectious diseases or monitor the progress of a disease [2, 5, 25]. Thus, these technologies will allow the healthcare workforce to deliver medical services more efficiently without the need for expensive equipment or extensive training [1, 13, 2426]. Miniaturization and integration of diagnostic devices would allow rapid and reliable high-throughput chemical and biomedical imaging and analysis from a tiny amount of sample such as a fingerprick volume of blood [7, 2731]. Optofluidics takes advantage of integrating microfluidics and microelectronic optical components onto the same platform [32, 33]. Such a platform, with fluidics for sample delivery/capture and lensless optics for sensing and detection, can be applied to areas such as ultra wide-field cell monitoring array [14, 34, 35], digital in-line holography [3642], optofluidic microscopy [4346], and lensless on-chip microscopy [15, 42, 43, 47]. In this review, we provide an overview of the recent literature on lensless imaging technologies with a POC perspective in a resource limited setting. We present the current state of lensless imaging, its advantages, application challenges and the potential for use in POC.

2. Optical and Fluidic Concepts and Considerations for POC Testing

Microscale investigations in life sciences have so far been largely carried out by conventional light microscopes using lenses and visible light (i.e. geometrical optics) [4749]. These technologies and instruments are difficult to miniaturize and require highly trained personnel. Hence they are mostly confined to laboratory settings, and currently they have limited practical use for POC testing at resource limited settings [46]. However, these technologies can be integrated and supplemented with the emerging nano- and microscale methods in fluidics (i.e., nano- and microfluidics), which would allow new and feasible approaches in POC testing and diagnosis. In this section, we will review the current status of optics and fluidics with a resource-limited POC perspective, and highlight the associated challenges, needs and future perspectives.

2.1. Optical Aspects and Components for Imaging

Conventional microscopic imaging and detection technologies primarily consist of four main components: (i) a light source, (ii) optical modulators, (iii) lenses and (iv) a detector, which are sequentially positioned on the spatial beam path [50]. Here, we will discuss each of these components, their roles and relevance for POC applications.

2.1.1. Light sources

Microscopes use various light sources ranging from incandescent light bulbs to solid state light emitting devices (i.e., light emitting diodes, LEDs) [51]. LEDs are miniaturized light sources that are commonly used in compact optical devices without high power demands. However, LEDs emit non-collimated light, whereas high resolution imaging applications require collimated light sources that minimize diffraction. On the other hand, laser is an intense coherent light source with a narrow bandwidth, high spatial coherence and insignificant chromatic aberration. Laser can be easily focused and facilitates high resolution imaging [52]. Therefore, laser diodes that combine the portability, low cost and low power consumption of LEDs and the coherent emission characteristics of lasers would be ideal for on-chip POC diagnostic platforms [53, 54].

2.1.2. Optical modulators

The properties of light (e.g., intensity, path, and wavelength) change as it travels through a medium (e.g., sample of interest), which acts as an optical modulator. The medium reflects, diffracts, and deflects the incident light. Sample characteristics and incident light wavelength affect the sample-light interaction and the observed outcome. For instance, in high resolution fluorescent imaging, fluorescent molecules transfer energy from the incident light to the emitted light at a different (longer) wavelength. Due to this phenomenon, fluorescent-based detection requires expensive optical filters to eliminate the background noise and the wavelengths other than the emission which carries the relevant information. Therefore, it would be challenging to adapt fluorescent imaging in on-chip microscopy for POC oriented applications. However, advances in LED and CCD technologies may overcome these challenges and facilitate feasible utilization of fluorescent luminescence in POC [55, 56].

High resolution imaging can also be achieved without lenses or filters by employing surfaces as optical modulators (i.e., surface plasmon resonance) [5761]. Fabrication of nanoscale structures and features on surfaces has been well established and therefore, enabled the surface plasmon methods for imaging [6269], This imaging method is based on optical energy changes, converting incident light into a surface plasmon resonance through interactions between nanoscale structures and incident waves [57, 5961, 70]. Metals are commonly used for surface-plasmon optical systems as they support surface electric charge waves. Non-metallic materials can also have similar energy transformation properties when used as nanoscale hole arrays on metallic films [71]. High resolution imaging can also be achieved through light phase images which differentiate features according to their geometry using a collimated light source [35, 36, 38]. Since these methods do not require expensive filters and can operate in a lensless setting, they may be considered for applications in POC detection.

2.1.3. Lenses

Lens-based imaging is an optical method where a lens is situated away from the object. Lenses condense and modulate light resulting from the sample of interest (e.g., cells). A lens can be realized by an aperture or using convex or concave optically clear materials. The term aperture in optics refers to an opening that can determine the diffraction angle of a bundle of rays, focusing them on an image plane. The aperture can also determine how much light reaches the image plane. Although the resolution increases with decreased aperture size, images become darker as apertures become narrow due to the constriction of light. Challenges exist regarding both the diffraction limited resolution and practical use of lens-based optical microscopy because of the narrow field of view at high magnifications [72, 73]. Lens-based methods are not capable of imaging nanofeatures (i.e. less than 100 nm) that are smaller than the imaging wavelength (i.e. diffraction limit) [74]. One solution is to send light through a tiny aperture on an opaque screen (e.g., metallic or nonmetallic films) without using conventional lenses (i.e. near optical detection approaches). In this method, the target must remain at a sub-wavelength distance (less than the wavelength of the incident light) to the detector to eliminate diffraction effects caused by incoherent light [72, 73, 75]. However, these approaches present challenges in terms of their applications in POC due to requirements such as: expensive high energy light sources, sensitive detection systems and complex peripheral equipment. Lensless approaches hold great promise in POC testing due to simpler operation and low-cost [13, 25, 42, 43].

2.1.4. Detectors

Detector is an electronic sensor that converts incident light into an electrical current as a function of the light intensity. Photomultiplier tubes (PMTs), avalanche photodiodes (APDs), and PIN (p-type-intrinsic type-n-type) photodiodes are also used as optical detectors. Traditionally, PMTs and APDs have been used in commercial cytometers that measure fluorescence intensity because of their high sensitivity, including the ability to count photons. PMTs are typically chosen for very low intensity light applications because they have large internal gains, and low noise [76]. However, they are difficult to use for point-of-care systems, as they are expensive and require peripheral circuitry and heavy equipment (e.g. high voltage power supply). APDs are more intrinsically sensitive than PMTs (typically 3–4 orders of magnitude higher). However, they are sensitive to temperature changes. The disadvantage of both types of detectors is their cost. When more light is available, PIN photodiodes can be well-suited for detection and integration in a miniaturized device [77]. Different from a single sensor, imaging array using CCD (Charge Coupled Device) and CMOS (Complementary Metal Oxide Semiconductor) are more preferable for wide FOV (Field of View) detection since they do not require a peripheral apparatus to scan the entire sample area. In lensless microscopy, CMOS (high speed and low light sensitivity) and CCDs (low speed and high light sensitivity) can be used as light detectors based on the speed and sensitivity needs. For the wide FOV applications, CCD and CMOS imaging arrays have been employed in lensless systems and have proven to be effective and feasible. Recently, a wide FOV system using arrayed CCD image sensor was adopted for high speed imaging by sacrificing the resolution using sequential frame detection [15, 78].

2.2 General Microfluidics for Biological Applications

Applications of fluidics in the form of nano- and microfluidic systems have been emerging in biomedicine [7991]. In biological applications, the size and operational regimes of fluidic devices can be adjusted to match that of the target, e.g., ~10 μm for cells, 10 ~ 100 nm for macro molecules, and ~10 nm for small single molecules [9296]. In these systems, the flow is driven by pressure which is generated by peripheral syringe pumps, integrated peristaltic pumps, electrokinetic drive, or gravity [43, 9799]. Optical forces can also be used in microfluidic channels such as optical tweezers which can manipulate cells [100102]. Integration of fluidic systems and lensless optics can offer platforms for a wide range of cell manipulation processes, including capture [13], separation, isolation, high-resolution imaging [13, 43, 103], and probing cellular functions at the single-cell level [104].

3. Integration of Optics and Microfluidics

Optofluidic approaches eliminate the need for lenses and filters by benefiting from low-resolution ultra-wide FOV approaches with CCD based detection. In these methods, fluidics is used to immobilize cells and maintain a single layer of cells within confined flow geometry. One example application is cell identification using whole blood. LUCAS system is used with microfluidic channels to detect cells from whole blood [13, 15]. So, both the LUCAS and holographic systems use microfluidic channels and flow [13, 15]. Lensless diagnosis systems can be built onto chips with the existing fabrication techniques [32, 43, 44, 46, 105]. The portability achieved by the lensless operation of lensless optofluidic systems would significantly benefit POC diagnosis devices. On the other hand, the imaging and diagnostic methods targeted for POC need to be robust, user-friendly and operable without extensive training. A comparison of the lensless on-chip imaging technologies has been presented in relation to POC applications in Table 2. The advantages and challenges associated with each method in terms of functionality and application in resource-limited POC have been highlighted discussed in the next sections.

Table 2
Comparison of lensless on-chip imaging platforms for POC applications: Optofluidic Microscope (OFM), Digital Holography and Lensless Ultra Wide-Field Cell Monitoring Array Platform (LUCAS)

3.1 Optofluidic Microscopy (OFM)

Optofluidic microscopes (OFM) integrate microscale fluidics and optics in a single system. An earlier application of this approach aimed to investigate the effects of spaceflight on living organisms (i.e., Caenorhabditis elegans or round worm) [106]. The samples of interest were transferred through microchannels into a microchamber illuminated by an LED light source placed above, and imaged with a CMOS sensor array below. When the chamber was illuminated, the samples casted a shadow on the CMOS array, which recorded the images. A challenge in utilization of this system was caused by the convex corners of the microchamber, which reduced the imaging area and necessitated manual recording of the signal spectrum with a vector signal analyzer. In addition, the use of non-collimated LED illumination led to a low resolution, which could be improved by decreasing the distance between the imaging plane and the sensor array or reducing the illumination distance. In an effort to improve the resolution and performance of this approach, masks of nanometer sized apertures were placed above a CMOS array [43, 44]. This was achieved by coating an opaque metal film layer onto the sensor array with two lines of nano-apertures (Figure 1a) and encasing them into a microfluidic chip (Figure 1b). The system can be operated in an upright position to utilize gravity driven flow and hence eliminate the need for external pumps, which is suitable for elongated samples such as round worms in this case (Figure 1c). However, when spherical or ellipsoidal samples (e.g., cells, bacteria) are imaged, an external pump should be used such as electrokinetic pumps [43]. In the case of imaging cells, a uniform fluid flow bestowed by an external pump would minimize the imaging artifacts due to rotation of the cells. The device is uniformly illuminated from the top with white light (~20 mW/cm2, approximately the intensity of sunlight) using a halogen lamp. The target object is flowed across the aperture array, creating a series of image slices that produce a high resolution image when assembled (Figure 1d). The flow speed of the specimen can be calculated by tracking the time difference between the detection obtained through each aperture, which is then used to reconstruct an image. Since the object passes over each aperture at different time intervals, the time varying light transmission through the aperture constitutes a line trace. When these line traces from all the apertures are assembled an image is constructed (Figure 1d).

Figure 1
Optofluidic Microscope (OFM) system

The challenges associated with OFM approach include the critical dependence of the imaging outcome on the orientation of the sample and the constant flow rate within the channel. In other words, an acceptable image would not be obtained if the sample orientation during imaging changes (i.e., rotation of the sample during flow). A constant and uniform flow rate needs to be achieved inside the channel for this approach to work effectively, which introduces operational difficulties that is not suitable for resource limited POC applications. On the other hand, the aperture array needs to be in perfect alignment with the underlying sensor grid and the resolution is limited by the size of the apertures and the spacing in between. If these challenges are addressed or minimized with improved design and fabrication methods, OFM microscopy would be suitable for POC since it allows portable operation and does not require expensive peripheral equipment.

3.2. Digital Holography

Digital holography is an emerging phase contrast imaging technique that offers both qualitative and quantitative information [35, 39, 107]. In holographic imaging (Figure 2a), samples are illuminated with a spherical wavefront produced by focusing a laser beam from a 1 μm diameter pinhole (Figure 2b). The focused light serves as a reference beam, while the scattered light serves as an object beam. The CCD sensor array records the interference pattern produced by the superposition of the two wavefronts (the hologram). A subsequent digital elaboration allows retrieving the volume information on the sample from a single 2D hologram (Figure 2c–h). In this method, multiple focal planes can be achieved and captured mimicking the focus control of conventional optical microscopes. Digital holography records the entire object in three-dimensional (3D) space whereas a conventional microscope has a 2D focal plane and requires mechanical scanning to create a full image. Due to the complex analysis that can be performed with holography, the orientation of asymmetric yeast cells can be determined by the broken symmetry as shown in the figure (Figure 2c-h). The resolving power in this method is a function of the hologram recording geometry, the CCD dimensions and incident wavelength.

Figure 2
Digital Holography with lensless ultra wide-field operation

Digital holography has been integrated with fluorescent imaging to achieve lensless fluorescent detection without any thin-film filters or mechanical scanning with an ultra-wide field-of-view (2.5 cm × 3.5 cm) [104]. In this approach, the sample was illuminated with an incoherent excitation beam from an LED or a Xenon lamp source located directly above (Figure 3a). Fluorescent excitation was provided from the side by utilizing a rhomboid prism. The incoherent light beam interacted with the sample and undergoing total internal reflection (TIR) at the bottom (Figure 3a). Detection of the fluorescent emission from the excited cells/particles over the entire FOV of the sensor-array (CCD or CMOS) was achieved without using any lenses. Unlike the traditional fluorescent microscopy, a need for expensive interference filters was not a limiting factor for this method. An inexpensive plastic-based absorption filter was used between the sample and detector planes to achieve a better dark-field background. The issue of potential overlapping of diverged fluorescent emission of cells or particles at the sensor plane was addressed by deconvolving the acquired images, which provided a resolution of ~40–50 μm over the entire sensor FOV without the use of a lens. The system was validated by imaging white blood cells as seen in Figure 3b–h.

Figure 3
Digital Holography integrated with fluorescent imaging

Digital holography and lensless fluorescent imaging platforms discussed above are innovative approaches that are transforming the microscopy techniques dramatically by simplifying and reducing the cost of imaging equipment. These methods can provide valuable information about the sample, such as the 3D structure and orientation, at the expense of relatively more complex nature, operation and higher computational needs compared to other on-chip imaging technologies (Table 2). Therefore, the aptness and feasibility of utilizing digital holography based approaches in resource-limited POC is questionable and warrants further investigation.

3.3. Lensless Ultra Wide-Field Cell Monitoring Array Platform (LUCAS)

A wide FOV imaging platform would be an alternative to optical microscopes for detection and counting applications, where high resolution is not critical. Therefore, a “lensless, ultra wide-field cell monitoring array platform” has been developed that can be used to detect and count cells on-chip with a two orders of magnitude wider FOV than conventional light microscopes (Figure 4a–b) [13, 15]. This system has been shown to detect and count thousands of individual cells in real time in seconds. LUCAS system is based on recording the “shadow images” of microscale objects onto an optoelectronic sensor array plane. Microscopic objects (e.g., cells) are placed between two microscope glass slides (in a microfluidic device) and are uniformly illuminated with an incoherent white-light source or a laser beam (Figure 4a). Cell shadow pattern is digitally recorded using a CCD sensor array. A CMOS chip, depending on the application type, can also be used as the optoelectronic sensor array. While the former is preferred for applications where image signal-to-noise (S/N) ratio is important, the latter is the sensor of choice for high speed image acquisition. Each cell’s shadow falling onto the sensor array is recorded as object-specific “signatures”. The cell types, assigned with specific signatures, can easily be differentiated and counted (Figure 4c). The distance between the active region of the sensor array (i.e., the active surface of the CCD chip) and the microscopic object location (e.g., cells) is important to optimize LUCAS operation. The shadow diameter is estimated from diffracted light intensity at the sensor plane. Finally, individual cells are computationally modeled as uniform circular objects with a reduced field-transmission coefficient. This technology has recently been modularized with microfluidic devices, paving the way for lensless cell counting technologies at POC applications. A recent study has also demonstrated the potential of this lensless technology for microfluidic based HIV monitoring applications [13].

Figure 4Figure 4
Integration of Microfluidics with lensless ultra wide-field cell monitoring array platform (LUCAS)

4. Conclusions and Perspectives

The needs and challenges in resource-poor POC diagnosis are distinct from the needs and challenges of conventional diagnostics that can be performed at well-equipped settings by well-equipped personnel (Table 1). The essential characteristics of resource-limited POC detection and diagnostic platforms can be listed as: low-cost, functionality, portability, robustness, simplicity, flexibility, ease of use, and sensitivity. Recent advances in on-chip approaches have enabled and facilitated the integration of optics and microfluidics technologies to create new and more effective platforms (Table 2) that are needed at the POC in developing countries. The lensless imaging technologies are best suited for deployment to areas with limited resources. These platforms are promising since they enable rapid detection and counting of thousands of cells of interest, which has already been utilized for rapid CD4+ tests to monitor HIV [13]. The technology provided mainly the HIV diagnostics as an example since it is one of the greater problems of the world. Such technologies combined with surface chemistry can also be used for multiple other diseases including capture of cancer circulating tumor cells [108], detection of other infectious diseases such as sepsis [109] and bacterial pathogens including E. Coli [35]. The advances and approaches discussed in this review may help improve healthcare in developing countries by allowing healthcare personnel to rapidly diagnose and disseminate information and therefore, prevent the spread of communicable diseases. These enabling technologies will potentially shape the future of clinical medicine and will potentially provide solutions for global health problems.

Acknowledgments

This work was performed at the Demirci Bio-Acoustic MEMS in Medicine (BAMM) Labs at the HST-BWH Center for Bioengineering, Harvard Medical School. This work was partially supported by NIH R21 (EB007707) and New Development Grant of the US Department of the Defense (DOD). We would like to acknowledge the Coulter Foundation Young Investigator Award, NIH (R01AI081534), NIH (R21AI087107), the Center for Integration of Medicine and Innovative Technology (CIMIT) under U.S. Army Medical Research Acquisition Activity Cooperative Agreement and the National Natural Science Foundation Young Scholars Research Fund.

Footnotes

6. Conflict of Interest Statement

The authors have declared no conflict of interest.

References

1. Wang S, Xu F, Demirci U. Advances in developing HIV-1 viral load assays for resource-limited settings. Biotechnol Adv. 2010;28:770–781. [PMC free article] [PubMed]
2. Laksanasopin T, Chin CD, Moore H, Wang J, et al. Microfluidic Point-of-Care Diagnostics for Resource-Poor Environments. Ieee; New York: 2009. [PubMed]
3. Loubiere S, Moatti JP. Economic evaluation of point-of-care diagnostic technologies for infectious diseases. Clinical Microbiology and Infection. 2010;16:1070–1076. [PubMed]
4. Usdin M, Guillerm M, Calmy A. Patient Needs and Point-of-Care Requirements for HIV Load Testing in Resource-Limited Settings. Journal of Infectious Diseases. 2010;201:S73–S77. [PubMed]
5. Chin CD, Linder V, Sia SK. Lab-on-a-chip devices for global health: Past studies and future opportunities. Lab Chip. 2007;7:41–57. [PubMed]
6. Varmus H, Klausner R, Zerhouni E, Acharya T, et al. Grand challenges in global health. Science. 2003;302:398–399. [PMC free article] [PubMed]
7. Martinez AW, Phillips ST, Whitesides GM, Carrilho E. Diagnostics for the Developing World: Microfluidic Paper-Based Analytical Devices. Analytical Chemistry. 2010;82:3–10. [PubMed]
8. Daar AS, Thorsteinsdottir H, Martin DK, Smith AC, et al. Top ten biotechnologies for improving health in developing countries. Nature Genetics. 2002;32:229–232. [PubMed]
9. Yager P, Domingo GJ, Gerdes J. Point-of-care diagnostics for global health. Annual review of biomedical engineering. 2008;10:107–144. [PubMed]
10. WHO. 2007 UNAIDS annual report: Knowing your epidemic. Joint United Nations Programme on HIV/AIDS (UNAIDS); GENEVA. 2008.
11. Gilks CF, Crowley S, Ekpini R, Gove S, et al. The WHO public-health approach to antiretroviral treatment against HIV in resource-limited settings. Lancet. 2006;368:505–510. [PubMed]
12. Peeling RW, Holmes KK, Mabey D, Ronald A. Rapid tests for sexually transmitted infections (STIs): the way forward. Sexually Transmitted Infections. 2006;82:V1–V6. [PMC free article] [PubMed]
13. Moon S, Keles HO, Ozcan A, Khademhosseini A, et al. Integrating microfluidics and lensless imaging for point-of-care testing. Biosensors & bioelectronics. 2009;24:3208–3214. [PMC free article] [PubMed]
14. Su TW, Seo S, Erlinger A, Ozcan A. High-throughput lensfree imaging and characterization of a heterogeneous cell solution on a chip. Biotechnology and bioengineering. 2009;102:856–868. [PMC free article] [PubMed]
15. Ozcan A, Demirci U. Ultra wide-field lens-free monitoring of cells on-chip. Lab on a chip. 2008;8:98–106. [PubMed]
16. Whitesides GM. The origins and the future of microfluidics. Nature. 2006;442:368–373. [PubMed]
17. Sia SK, Kricka LJ. Microfluidics and point-of-care testing. Lab Chip. 2008;8:1982–1983. [PubMed]
18. Paguirigan AL, Beebe DJ. Microfluidics meet cell biology: bridging the gap by validation and application of microscale techniques for cell biological assays. Bioessays. 2008;30:811–821. [PMC free article] [PubMed]
19. Klapperich CM. Microfluidic diagnostics: time for industry standards. Expert review of medical devices. 2009;6:211–213. [PubMed]
20. Yager P, Edwards T, Fu E, Helton K, et al. Microfluidic diagnostic technologies for global public health. Nature. 2006;442:412–418. [PubMed]
21. Cheng X, Irimia D, Dixon M, Sekine K, et al. A microfluidic device for practical label-free CD4+ T cell counting of HIV-infected subjects. Lab-on-a-Chip. 2007;7:170–178. [PMC free article] [PubMed]
22. Cheng X, Irimia D, Dixon M, Ziperstein JC, et al. A Microchip Approach for Practical Label-Free CD4+T-Cell Counting of HIV-Infected Subjects in Resource-Poor Settings. J Acquir Immune Defic Syndr. 2007 [PubMed]
23. Alyassin MA, Moon S, Keles HO, Manzur F, et al. Rapid automated cell quantification on HIV microfluidic devices. Lab Chip. 2009;9:3364–3369. [PMC free article] [PubMed]
24. Lee WG, Kim YG, Chung BG, Demirci U, et al. Nano/Microfluidics for diagnosis of infectious diseases in developing countries. Adv Drug Deliv Rev. 2010;62:449–457. [PMC free article] [PubMed]
25. Moon S, Keles HO, Kim YG, Kuritzkes D, et al. Lensless imaging for point-of-care testing. Conf Proc IEEE Eng Med Biol Soc. 2009;2009:6376–6379. [PMC free article] [PubMed]
26. Lee WG, Demirci U, Khademhosseini A. Microscale electroporation: challenges and perspectives for clinical applications. Integrative Biology. 2009;1:242–251. [PMC free article] [PubMed]
27. Christodoulides N, Floriano PN, Miller CS, Ebersole JL, et al. Lab-on-a-chip methods for point-of-care measurements of salivary biomarkers of periodontitis. In: Malamud D, Niedbala RS, editors. Oral-Based Diagnostics. 2007. pp. 411–428. [PubMed]
28. Christodoulides N, Mohanty S, Miller CS, Langub MC, et al. Application of microchip assay system for the measurement of C-reactive protein in human saliva. Lab Chip. 2005;5:261–269. [PubMed]
29. Jokerst JV, McDevitt JT. Programmable nano-bio-chips: multifunctional clinical tools for use at the point-of-care. Nanomedicine. 2010;5:143–155. [PMC free article] [PubMed]
30. Jokerst JV, Raamanathan A, Christodoulides N, Floriano PN, et al. Nano-bio-chips for high performance multiplexed protein detection: Determinations of cancer biomarkers in serum and saliva using quantum dot bioconjugate labels. Biosens Bioelectron. 2009;24:3622–3629. [PMC free article] [PubMed]
31. Parsa H, Chin CD, Mongkolwisetwara P, Lee BW, et al. Effect of volume-and time-based constraints on capture of analytes in microfluidic heterogeneous immunoassays. Lab Chip. 2008;8:2062–2070. [PubMed]
32. Yang C, Psaltis D. Optofluidics can create small, cheap biophotonic devices. Laser Focus World. 2006;42:85–88.
33. Ozcan A, Demirci U. Rewritable self-assembled long-period gratings in photonic bandgap fibers using microparticles. Optics Communications. 2007;270:225–228.
34. Seo S, Su TW, Erlinger A, Ozca A. Multi-color LUCAS: Lensfree On-chip Cytometry Using Tunable Monochromatic Illumination and Digital Noise Reduction. Cellular and Molecular Bioengineering. 2008;1:146–156.
35. Seo S, Su TW, Tseng DK, Erlinger A, et al. Lensfree holographic imaging for on-chip cytometry and diagnostics. Lab on a chip. 2009;9:777–787. [PMC free article] [PubMed]
36. Gopinathan U, Pedrini G, Osten W. Coherence effects in digital in-line holographic microscopy. Journal of the Optical Society of America. 2008;25:2459–2466. [PubMed]
37. Morlens AS, Gautier J, Rey G, Zeitoun P, et al. Submicrometer digital in-line holographic microscopy at 32 nm with high-order harmonics. Optics letters. 2006;31:3095–3097. [PubMed]
38. Garcia-Sucerquia J, Xu W, Jericho MH, Kreuzer HJ. Immersion digital in-line holographic microscopy. Optics letters. 2006;31:1211–1213. [PubMed]
39. Garcia-Sucerquia J, Xu W, Jericho SK, Klages P, et al. Digital in-line holographic microscopy. Applied optics. 2006;45:836–850. [PubMed]
40. Xu W, Jericho MH, Meinertzhagen IA, Kreuzer HJ. Digital in-line holography of microspheres. Applied optics. 2002;41:5367–5375. [PubMed]
41. Tulandi T, Huang JY, Tan SL. Preservation of female fertility: an essential progress. Obstet Gynecol. 2008;112:1160–1172. [PubMed]
42. Repetto L, Piano E, Pontiggia C. Lensless digital holographic microscope with light-emitting diode illumination. Opt Lett. 2004;29:1132–1134. [PubMed]
43. Cui X, Lee LM, Heng X, Zhong W, et al. Lensless high-resolution on-chip optofluidic microscopes for Caenorhabditis elegans and cell imaging. Proceedings of the National Academy of Sciences of the United States of America. 2008;105:10670–10675. [PubMed]
44. Heng X, Erickson D, Baugh LR, Yaqoob Z, et al. Optofluidic microscopy--a method for implementing a high resolution optical microscope on a chip. Lab Chip. 2006;6:1274–1276. [PubMed]
45. Lee LM, Cui X, Yang C. The application of on-chip optofluidic microscopy for imaging Giardia lamblia trophozoites and cysts. Biomedical microdevices. 2009 [PMC free article] [PubMed]
46. Brennan D, Justice J, Corbett B, McCarthy T, et al. Emerging optofluidic technologies for point-of-care genetic analysis systems: a review. Analytical and bioanalytical chemistry. 2009;395:621–636. [PubMed]
47. Gabor D. A new microscopic principle. Nature. 1948;161:777. [PubMed]
48. Masters BR. Correlation of histology and linear and nonlinear microscopy of the living human cornea. J Biophotonics. 2009;2:127–139. [PubMed]
49. Schawlow AL, Townes CH. Infrared and Optical Masers. Physical Review. 1958;112:1940–1949.
50. Psaltis D. Coherent optical information systems. Science. 2002;298:1359–1363. [PubMed]
51. Schubert EF, Kim JK. Solid-state light sources getting smart. Science. 2005;308:1274–1278. [PubMed]
52. Zhang X, Stuart JN, Sweedler JV. Capillary electrophoresis with wavelength-resolved laser-induced fluorescence detection. Anal Bioanal Chem. 2002;373:332–343. [PubMed]
53. Kuswandi B, Nuriman, Huskens J, Verboom W. Optical sensing systems for microfluidic devices: A review. Analytica Chimica Acta. 2007;601:141–155. [PubMed]
54. Ryvolova M, Preisler J, Brabazon D, Macka M. Portable capillary-based (non-chip) capillary electrophoresis. Trac-Trends Anal Chem. 2010;29:339–353.
55. Huber G, Krankel C, Petermann K. Solid-state lasers: status and future. Journal of the Optical Society of America B-Optical Physics. 2010;27:B93–B105.
56. Kulmala S, Suomi J. Current status of modern analytical luminescence methods. Analytica Chimica Acta. 2003;500:21–69.
57. Barnes WL, Dereux A, Ebbesen TW. Surface plasmon subwavelength optics. Nature. 2003;424:824–830. [PubMed]
58. Link S, El-Sayed MA. Spectral properties and relaxation dynamics of surface plasmon electronic oscillations in gold and silver nanodots and nanorods. Journal of Physical Chemistry B. 1999;103:8410–8426.
59. Mulvaney P. Surface plasmon spectroscopy of nanosized metal particles. Langmuir. 1996;12:788–800.
60. Haes AJ, Van Duyne RP. A nanoscale optical blosensor: Sensitivity and selectivity of an approach based on the localized surface plasmon resonance spectroscopy of triangular silver nanoparticles. Journal of the American Chemical Society. 2002;124:10596–10604. [PubMed]
61. Demirel MC, Kao P, Malvadkar N, Wang H, et al. Bio-organism sensingvia surface enhanced Raman spectroscopy on controlled metal/polymer nanostructured substrates. Biointerphases. 2009;4:35–41. [PubMed]
62. Demirci U, Toner M. Direct etch method for microfludic channel and nanoheight post-fabrication by picoliter droplets. Applied Physics Letters. 2006;88:3.
63. Malvadkar NA, Hancock MJ, Sekeroglu K, Dressick WJ, et al. An engineered anisotropic nanofilm with unidirectional wetting properties. Nat Mater. 2010 advance online publication. [PubMed]
64. Pursel S, Horn MW, Demirel MC, Lakhtakia A. Growth of sculptured polymer submicronwire assemblies by vapor deposition. Polymer. 2005;46:9544–9548.
65. Xiong X, Busnaina A, Selvarasah S, Somu S, et al. Directed assembly of gold nanoparticle nanowires and networks for nanodevices. Applied Physics Letters. 2007;91:3.
66. El-Sayed MA. Some interesting properties of metals confined in time and nanometer space of different shapes. Accounts of Chemical Research. 2001;34:257–264. [PubMed]
67. Kelly KL, Coronado E, Zhao LL, Schatz GC. The optical properties of metal nanoparticles: The influence of size, shape, and dielectric environment. Journal of Physical Chemistry B. 2003;107:668–677.
68. Love JC, Estroff LA, Kriebel JK, Nuzzo RG, et al. Self-assembled monolayers of thiolates on metals as a form of nanotechnology. Chemical Reviews. 2005;105:1103–1169. [PubMed]
69. Rosi NL, Mirkin CA. Nanostructures in biodiagnostics. Chemical Reviews. 2005;105:1547–1562. [PubMed]
70. Ebbesen TW, Lezec HJ, Ghaemi HF, Thio T, et al. Extraordinary optical transmission through sub-wavelength hole arrays. Nature. 1998;391:667–669.
71. Lezec HJ, Thio T. Diffracted evanescent wave model for enhanced and suppressed optical transmission through subwavelength hole arrays. Optics express. 2004;12:3629–3651. [PubMed]
72. Hell SW. Far-field optical nanoscopy. Science. 2007;316:1153–1158. [PubMed]
73. Lewis A, Taha H, Strinkovski A, Manevitch A, et al. Near-field optics: from subwavelength illumination to nanometric shadowing. Nature biotechnology. 2003;21:1378–1386. [PubMed]
74. Thio T. A bright future for subwavelength light sources -Generating tiny points of light for such things as storing data on optical disks is aided by a new theory involving evanescent waves. American Scientist. 2006;94:40–47.
75. Oshikane Y, Kataoka T, Okuda M, Hara S, et al. Observation of nanostructure by scanning near-field optical microscope with small sphere probe. Science and Technology of Advanced Materials. 2007;8:181–185.
76. Roth JM, Murphy TE, Xu C. Ultrasensitive and high-dynamic-range two-photon absorption in a GaAs photomultiplier tube. Opt Lett. 2002;27:2076–2078. [PubMed]
77. Tian Z, Quick NR, Kar A. Laser endotaxy in silicon carbide and PIN diode fabrication. Journal of Laser Applications. 2008;20:106–115.
78. Bub G, Tecza M, Helmes M, Lee P, et al. Temporal pixel multiplexing for simultaneous high-speed, high-resolution imaging. Nat Meth. 2010;7:209–211. [PMC free article] [PubMed]
79. Andersson H, van den Berg A. Microfluidic devices for cellomics: a review. Sensors and Actuators B-Chemical. 2003;92:315–325.
80. Gardeniers H, Van den Berg A. Micro-and nanofluidic devices for environmental and biomedical applications. International Journalof Environmental Analytical Chemistry. 2004;84:809–819.
81. Samel B, Nock V, Russom A, Griss P, et al. A disposable lab-on-a-chip platform with embedded fluid actuators for active nanoliter liquid handling. Biomed Microdevices. 2007;9:61–67. [PubMed]
82. Sethu P, Mastrangelo CH. Cast epoxy-based microfluidic systems and their application in biotechnology. Sensors and Actuators B-Chemical. 2004;98:337–346.
83. Tasoglu S, Demirci U, Muradoglu M. The effect of soluble surfactant on the transient motion of a buoyancy-driven bubble. Physics of Fluids. 2008;20
84. Tasoglu S, Kaynak G, Szeri AJ, Demirci U, et al. Impact of a compound droplet on a flat surface: A model for single cell epitaxy. Physics of Fluids. 2010;22 [PubMed]
85. Huh D, Gu W, Kamotani Y, Grotberg JB, et al. Microfluidics for flow cytometric analysis of cells and particles. Physiol Meas. 2005;26:R73–R98. [PubMed]
86. Douville N, Huh D, Takayama S. DNA linearization through confinement in nanofluidic channels. Analytical and bioanalytical chemistry. 2008;391:2395–2409. [PubMed]
87. Huh D, Mills KL, Zhu XY, Burns MA, et al. Tuneable elastomeric nanochannels for nanofluidic manipulation. Nature Materials. 2007;6:424–428. [PubMed]
88. Mills KL, Huh D, Takayama S, Thouless MD. Instantaneous fabrication of arrays of normally closed, adjustable, and reversible nanochannels by tunnel cracking. Lab on a chip. 10:1627–1630. [PubMed]
89. Takayama S, Tung YC, Chueh BH, Asme Biological micro/nanofluidics. 2008
90. Choi JY, Seo TS. An integrated microdevice for high-performance short tandem repeat genotyping. Biotechnology Journal. 2009;4:1530–1541. [PubMed]
91. Hattori K, Sugiura S, Kanamori T. On-chip cell culture on a microarray of extracellular matrix with surface modification of poly(dimethylsiloxane) Biotechnology Journal. 2010;5:463–469. [PubMed]
92. Koo JM, Kleinstreuer C. Liquid flow in microchannels: experimental observations and computational analyses of microfluidics effects. Journal of Micromechanics and Microengineering. 2003;13:568–579.
93. Neto C, Evans DR, Bonaccurso E, Butt HJ, et al. Boundary slip in Newtonian liquids: a review of experimental studies. Reports on Progress in Physics. 2005;68:2859–2897.
94. Adiga SP, Brenner DW. Flow control through polymer-grafted smart nanofluidic channels: Molecular dynamics simulations. Nano Letters. 2005;5:2509–2514. [PubMed]
95. Squires TM, Quake SR. Microfluidics: Fluid physics at the nanoliter scale. Reviews of Modern Physics. 2005;77:977–1026.
96. Yi CQ, Li CW, Ji SL, Yang MS. Microfluidics technology for manipulation and analysis of biological cells. Analytica Chimica Acta. 2006;560:1–23.
97. Chen L, Lee S, Choo J, Lee EK. Continuous dynamic flow micropumps for microfluid manipulation. Journal of Micromechanics and Microengineering. 2008;18
98. He X, Hauan S. Microfluidic modeling and design for continuous flow in electrokinetic mixing-reaction channels. Aiche Journal. 2006;52:3842–3851.
99. Shui L, Eijkel JCT, van den Berg A. Multiphase flow in microfluidic systems -Control and applications of droplets and interfaces. Advances in Colloid and Interface Science. 2007;133:35–49. [PubMed]
100. Pine J, Chow G. Moving live dissociated neurons with an optical tweezer. IEEE Trans Biomed Eng. 2009;56:1184–1188. [PubMed]
101. Heng X, Hsiao E, Psaltis D, Yang C. An optical tweezer actuated, nanoaperture-grid based Optofluidic Microscope implementation method. Opt Express. 2007;15:16367–16375. [PubMed]
102. Ghosh S, Sharma P, Bhattacharya S. Surface modes of a sessile water drop: an optical tweezer based study. Rev Sci Instrum. 2007;78:115110. [PubMed]
103. Kim YG, Moon S, Kuritzkes DR, Demirci U. Quantum dot-based HIV capture and imaging in a microfluidic channel. Biosensors & bioelectronics. 2009;25:253–258. [PMC free article] [PubMed]
104. Coskun AF, Su TW, Ozcan A. Wide field-of-view lens-free fluorescent imaging on a chip. Lab Chip. 2010;10:824–827. [PMC free article] [PubMed]
105. Yang C, Heng X, Cui X, Psaltis D. Optofluidic Microscope-Fitting a Microscope onto a Sensor Chip. Springer; US: 2007.
106. Lange D, Storment CW, Conley CA, Kovacs GTA. A microfluidic shadow imaging system for the study of the nematode Caenorhabditis elegans in space. Sensors and Actuators B-Chemical. 2005;107:904–914.
107. Di J, Zhao J, Jiang H, Zhang P, et al. High resolution digital holographic microscopy with a wide field of view based on a synthetic aperture technique and use of linear CCD scanning. Applied optics. 2008;47:5654–5659. [PubMed]
108. Stott SL, Lee RJ, Nagrath S, Yu M, et al. Isolation and characterization of circulating tumor cells from patients with localized and metastatic prostate cancer. Sci Transl Med. 2010;2:25ra23. [PMC free article] [PubMed]
109. Kotz KT, Xiao W, Miller-Graziano C, Qian WJ, et al. Clinical microfluidics for neutrophil genomics and proteomics. Nat Med. 2010;16:1042–1047. [PMC free article] [PubMed]