J Med Phys. 2007 Oct-Dec; 32(4): 175–184.
PMCID: PMC3014103

Computer simulations and models for the performance characteristics of spectrally equivalent X-ray beams in medical diagnostic radiology

Abstract

In order to achieve uniformity in radiological imaging, it is recommended that the concept of equivalence in shape (quality) and size (quantity) of clinical Xray beams should be used for carrying out the comparative evaluation of image and patient dose. When used under the same irradiation geometry, X-ray beams that are strictly or relatively equivalent in terms of shape and size will produce identical or relatively identical image quality and patient dose. Simple mathematical models and software program EQSPECT.FOR were developed for the comparative evaluation of the performance characteristics in terms of contrast (C), contrast to noise ratio (CNR) and figure-of-merit (FOM = CNR2/DOSE) for spectrally equivalent beams transmitted through filter materials referred to as conventional and k-edged. At the same value of operating potential (kVp), results show that spectrally equivalent beam transmitted through conventional filter with higher atomic number (Z-value) in comparison with that transmitted through conventional filter with lower Z-value resulted in the same value of C and FOM. However, in comparison with the spectrally equivalent beam transmitted through filter of lower Z-value, the beam through filter of higher Z-value produced higher value of CNR and DOSE at equal tube loading (mAs) and kVp. Under the condition of equivalence of spectrum, at scaled (or reduced) tube loading and same kVp, filter materials of higher Z-value can produce the same values of C, CNR, DOSE and FOM as filter materials of lower Z-value. Unlike the case of comparison of spectrally equivalent beam transmitted through one conventional filter and that through another conventional filter, it is not possible to derive simple mathematical formulations for the relative performance of spectrally equivalent beam transmitted through a given conventional filter material and that through kedge filter material.

Keywords: Contrast, elemental filters, medical diagnostic radiology

The use of tube voltage (peak kilovoltage, kVp), effective energy, homogeneity coefficient and half-value layer (HVL) as measures of beam quality and quantity in medical diagnostic radiology continues to produce diversities in results obtained for image quality and patient dose. As a result of this, the use of equivalent spectral as a measure of beam quality and as a tool for the standardization of medical X-ray imaging has been reported.[12] Among other factors, the quality and quantity of X-ray beams for medical diagnostic purpose depend on filter material and filter thickness. The standardization of beam quality can be achieved by providing a ranking for filter materials and filter thicknesses when the technology of X-ray tube is the same. By careful choice of thickness, two different elemental filters can be made to transmit beams that are simultaneously of the same/close shape (quantity) and size (quality).[36] When used under the same irradiation condition, beams that are strictly equivalent or relatively equivalent in terms of shape and size will produce identical or relatively identical image quality and patient dose. The results of theoretical and experimental approaches to the comparative ranking of the transmissions of filter materials on the basis of closeness of hardening (size) and attenuation (quantity) properties have been reported earlier in literature.[79] Due to the differences in attenuation and absorption properties, two different elemental filters that can independently produce beams of exactly equal shape and size do not exist.

In this paper, computer simulations were carried out to obtain spectrally equivalent X-ray beams using elemental filters - which include those referred to as conventional, such as aluminum and copper; and k-edge, such as gadolinium and tungsten. Analytical models similar to those earlier reported[6] for quantitative parameters were derived for the evaluation of differences in parameters for image quality, such as contrast-to-noise ratio (CNR) and figure-of-merit (FOM) for filter materials that transmit spectrally equivalent beams. For the purpose of clinical assessment of the performance characteristics of spectrally equivalent beams, iterative computer software EQSPECT.FOR was developed for the evaluation of these models. Also, by least square methods, parameters which can be used for the evaluation of thicknesses of other filter materials that are ‘spectrally’ equivalent to a specified thickness of copper are presented. The definitions of some of the terms used in the computer simulations and analytical formulations that were derived are presented in Table 1.

Table 1
Glossary of terms used in computer simulations and mathematical formulations

Theoretical formulations

In this work, T (tm) of filter material m at ‘spectrally’ equivalent thickness of tm is given by

T(tm)=E=1V[ɸ0(Ei)exp{-μm(Ei)tm}]E=1Vɸ0(Ei)

C is defined as

C=Ea-EbEb
(2)

Ea and Eb were calculated respectively using equations of the forms[10]

Ea=i[ɸ0(Ei)exp(μmtmμwtp-μctcEi(μLi/ρμdi/ρ){1 -exp(μdiρtd)}]
(3a)

and

Eb=i[ɸ0(Ei)exp(μmtmμwtp)Ei(μLi/ρμdi/ρ){1exp(μdiρtd)}]
(3b)

CNR and FOM were respectively calculated as

CNR=EaEbEb
(4)

and

FOM=CNR2DOSE.
(5)

The absorbed dose in patient (water phantom of thickness tw centimeters) was determined using equation of the form[11]

DOSE=ε(tw)tw
(6)

where ε(tw) could be obtained using an equation of the form

ε(tw)=i=1i=i maxEiɸ0(Ei)exp{μm(Ei)tm}   [n=0n=5βn(Ei)twn]
(7)

Details of the derivation of Eq. (7) can be found in the work reported by Okunade.[11]

From the earlier work reported,[6] for a given filter material m in comparison with aluminum filter at spectrally equivalent thickness,

δ=1+=λmtAl1+λmtAl   =SAlSCuSAl=mAAlmAmmAAl   =mAsAlmAsmmAsAl=HUAlHUmHUAl   =TAlTmTAl=DOSEAl DOSEmDOSEAl 
(8)

and

TmTAl=DOSEmDOSEAl=HUAlHUm   =mAsAlmAsm=mAAlmAm=SAlSm    =1+
(9)

where [increment] = λm tAl.

The values of λm resulting in best fit to Eq. (9) are reported in Table 1 of Okunade[6] for filter materials m with atomic numbers ranging from 12 to 39 and kVp ranging between 50 and 140 when reference filter is assumed to be aluminum.

From the simulated data and curve-fitting exercises carried out in this work,

TmTAl=[CNRm]2[CNRAl]2=1+
(10)

Calculations carried out using copper as reference filter in comparison with filter material m resulted in equation of the form

TmTCu=[CNRm]2[CNRCu]2=1+
(11)

where [increment]= λm tcu

The values of λm resulting in best fit to Eq. (11) are reported in Table 2 of this text. Thus from Eqs. (5), (9) and (10), for spectrally equivalent beams emanating from filter material m and aluminum filter at ‘spectrally’ equivalent thicknesses respectively, FOM is of the form

Table 2
Values of λm resulting in best fit to Eq. (11). r2 gives the range of coefficient of regression
FOM=[CNRm]2DOSEm=[CNRAl]2DOSEAl
(12)

and the CNR for beam transmitted through filter material m in comparison with that through a reference filter (aluminum) is of the form

CNRm=CNRAl×1+
(13)

In order to obtain spectral that is exactly equal in shape and size (or that will produce exactly the same values of transmittance, T, contrast, C, dose, DOSE, contrast-to-noise ratio, CNR, and figure-of-merit, FOM), the spectral data generated from filter material, say copper, in comparison with those from aluminum at ‘spectrally’ equivalent thickness have to be scaled across the entire energy spectrum by using the factor α. Mathematically, for this, we can write,

ɸAlEi=αɸCuEi
(14)

where α=11+

Note that the values of number of photons (per energy bin Ei) transmitted by a filter with higher Z-value are greater than those transmitted by filter of lower Z-values when the tube is operated at the same kVp and tube loading. The use of this scaled spectral results in models of the form

ɸref(Ei)=α×ɸm(Ei)Tref=α×TmDOSEref=α×DOSEmmAsm=α×mAsrefmAm=α×mArefsm=α×srefHUm=α×HUref}
(15)

The thickness of a given material, teq, that is ‘spectrally’ equivalent to a given aluminum thickness tAl is of the form[6]

teq=tAln=0n=4ωnVn
(16)

The values of ωn resulting in best fits to Eq. (16) when Z-value ranges between 12 and 39 are obtainable from Table 5 of Okunade.[6] When Z-values are 12, 13 and 14, the value of the thickness (teq) of other filter material that is ‘spectrally’ equivalent to copper thickness tcu can be obtained from equation of the form

Table 5
Comparison of the performance of filters relative to a 0.200-mm Cu filter at ‘spectrally’ equivalent thicknesses
teq=tCun=0n=4κnVn
(17)

where κn is given by

κn=ωn+αntCu
(17a)

When Z-values are 23, 26, 28, 30, 32 and 39, the value of the thickness (teq) of other filter material that is ‘spectrally’ equivalent to copper thickness tcu can be obtained from equation of the form

teq=tCun=0n=4ωnVn
(18)

The values of ωn and αn resulting in best fits to Eqs. (17) and (18) are shown in Table 3 of this text.

Table 3
Values of ωn and αn resulting in best fit to Eqs. (17-18). r2 is the coefficient of regression

Methods

Simulations of beams of the same/close shape and size, contrast, contrast-to-noise ratio, dose and figure-of-merit

In order to simulate X-ray beams of the same/close shape and size (spectrally equivalent beams), calculations were carried out for the matching of hardening and attenuation from aluminum and copper filters (reference filters) and other filter materials that are different from aluminum and copper. The matching exercises involved the use of the numerical algorithms earlier reported by Jennings.[8] The thicknesses of aluminum filter used were between 1 and 6 mm at 1 mm increments while those of copper were between 0.05 and 0.3 mm at 0.05 mm increments. The incident X-ray spectrals on the filter materials were determined by using polynomial functions earlier reported by Boone and Seibert.[12] Unfiltered tungsten anode spectra [phi]0(Ei) in units of photons/mm2 at energy Ei in keV were calculated for values of kVp ranging between 50 and 140 in 10 kVp steps.

The transmission, contrast, contrast-to-noise ratio, figure-of-merit and dose were calculated using equations presented in the theory section above. A total of 10 values of kVp, 8 alternative conventional filter materials (other than aluminum/copper) and 6 reference filter thicknesses of aluminum/copper were considered. Simulations were carried out for the evaluation of T, C, CNR, FOM and DOSE for water phantom of thicknesses ranging between 5 and 30 cm at 5 cm increments. Also considered in these simulations are three contrast media - namely, calcium, barium and iodine; and three intensifying screens, which include CaWO4, CsI and Gd2O2S. By similar simulation exercise, the k-edge filter materials investigated in comparison with aluminum/copper filters include lanthanium, gadolinium, holmium, thulium and tungsten. The interaction data of photons for water (patient phantom) and those for elemental filters that were considered were taken from McMaster et al.,[13] while those for the contrast media and intensifying screens were from Hubbell and Seltzer.[14] All the models reported in this work were fitted using the curve-fit least-square routines implemented in the computer package GRAF4WIN.[15] Figure 1 shows the schematic diagram for the implementation of the computer simulations.

Figure 1
Schematic diagram for the implementation of the computer simulations

Computer code EQSPECT.FOR

A FORTRAN source codes EQSPECT.FOR was developed for the execution of the simulations of same/close shape and size of beams, C, CNR, FOM and DOSE. In these computer codes, the algorithms reported[12] for the simulations of energy distribution of X-ray beams permit the specification of values of kVp other than those of multiples of 10 kV (say 86 kV). The interpolations for values of λm at values of kVp other than multiples of 10 kVp were carried out using cubic spline computer routines (named DEFSAL and DEFSCU) that were embedded in EQSPECT.FOR. A brief description of the main steps in the execution of EQSPECT.FOR is presented in Table 4.

Table 4
Main steps in the execution of the computer program EQSPECT.FORa

Results

Figure 2 shows the results of comparison of T, CNR, [CNR]2, C, DOSE and FOM for beams transmitted through aluminum/copper and selected alternative filter materials at ‘spectrally’ equivalent thicknesses. Specifically, for peak voltage of 86 kV, Figure 3 shows the results of comparison of the spectra transmitted through 3.7 mm Al and 9.3 mm Al and those through the respective spectrally equivalent thicknesses of copper, 0.1 mm Cu and 0.25 mm Cu, at equal tube loading. Table 5 shows the results of comparison of performance characteristics of other filter materials relative to copper filter. For the purpose of comparison with the work of Nagel,[9] these results are for 0.2 mm Cu filter, imaging of patient phantom of thickness 20 cm and contrast simulated with iodine (10 mg/cm2) and Gd2O2S intensifying screen (80 mg/cm2). Figure 4 shows the comparison of spectral shape and size obtained by forcing the same shape and size for aluminum/copper and gadolinium filters using the algorithms reported by Jennings.[8] The results of the comparison of transmission, contrast, CNR, [CNR2] DOSE and FOM obtained by the matching of spectral shape and size of beams transmitted by aluminum/copper and lanthanium are shown in Figure 5. To further provide verification of the validity of the formulations (Eqs. 815) presented in this work, the results of comparison carried out for 0.088 mm Gd filter and 0.097 mm Cu filter using EQSPECT.FOR are presented in Table 6.

Figure 2
Ratios of T, DOSE, [CNR]2, CNR, contrast and FOM for beams transmitted through aluminum/copper filter (denominator) and selected alternative filter materials (numerator) at ‘spectrally’ equivalent thicknesses. This is for intensifying ...
Figure 3
Comparison of the shape and size of transmitted spectral energy distribution at ‘spectrally’ equivalent thicknesses of aluminum and copper filter materials. The spectrals compared are (i) without scaling, ΦAl(Ei) and Φ ...
Figure 4
Comparison of the shape and size of transmitted spectral energy distribution at ‘spectrally’ equivalent thicknesses of aluminum/copper and gadolinium filters. The values of the minimum root-mean-square error in the matching of hardening ...
Figure 5
Ratios of T, DOSE, [CNR]2, CNR, contrast and FOM for beams transmitted through aluminum/copper filter (denominator) and lanthanium filter (numerator) at ‘spectrally’ equivalent thicknesses. This is for intensifying screen of 80 mg/cm−2 ...
Table 6
Performance characteristics of 0.097 mm Cu filter relative to 0.088 mm Gd filter

Discussion

The value of CNR for the beam filtered by 0.1 mm Cu exceeds that of the beam filtered by 3.7 mm Al by 7.0% for the imaging of a 20-cm patient phantom with the contrast simulated with iodine (12 mg/cm2) and CsI intensifying screen (60 mg/cm2). For this same imaging, the value of CNR for beam filtered by 0.25 mm Cu exceeds that for beam filtered by 9.3 mm Al by 19.0%. For the case of 3.7 mm Al and 0.10 mm Cu filtrations, the values of TCu, DOSECu and [CNRCu]2 exceed those of TAl, DOSEAl and [CNRAl]2 by 14.5% respectively while CCu [congruent with] CAl and FOMCu [congruent with] FOMAl. This is in good agreement with the result from earlier work,[8] which reported a surplus of 14.8% for TCu over TAl. At spectrally equivalent thicknesses, the difference in exposure time obtained experimentally for these aluminum and copper filters to produce beam of exactly the same shape and size was 15.6%, with the former filter requiring more exposure time.[8] This is in reasonable agreement with the theoretical formulations [Eqs. (8) and (9) of this text], which yield 14.5% as the value of [increment] × 100% (where [increment] = λm tref). With the beams from these filters (copper and aluminum at ‘spectrally’ equivalent thicknesses) numerically made exactly equal in shape and size {[phi]Al(Ei) = α[phi]Cu(Ei)}, the results obtained by simulations were such that TCu [congruent with] TAl, DOSECu [congruent with] DOSEAl, CNRCu [congruent with] CNRAl, [CNRCu]2 [congruent with] [CNRAl]2, CCu [congruent with] CAl and FOMCu [congruent with] FOMAl.

In the case of beam filtered by 9.3 mm Al in comparison with that filtered by 0.25 mm Cu and for spectral without scaling (same mAs and heat capacity), the values of TCu, DOSECu and [CNRCu]2 exceed those of TAl, DOSEAl and [CNRAl]2 by 41.9% respectively while CCu [congruent with] CAl and FOMCu [congruent with] FOMAl. Theoretical result presented[8] shows this surplus value to be 41.6%. With the transmitted spectrum through the copper filter theoretically scaled {[phi]Al(Ei = α[phi]Cu (Ei)} in comparison with aluminum filter, the results obtained by simulations were such that TCu [congruent with] TAl, DOSECu [congruent with] DOSEAl, [CNRCu]2 [congruent with] [CNRAl]2, CNRCu [congruent with] CNRAl, CCu [congruent with] CAl and FOMCu [congruent with] FOMAl.

Results of spectrometric measurements carried out by Jennings[8] show that 4.08 mmAl, 0.11 mmCu and 0.10 mm Y transmit spectrally equivalent beams. For 4.08 mm Al in comparison with 0.11 mm Cu and without the scaling of spectral (same mAs and heat capacity), using Eqs. (811), the values of TCu, DOSECu and [CNRCu]2 exceed those of TAl, DOSEAl and [CNRAl]2 by 16.2% respectively while CCu [congruent with] CAl and FOMCu [congruent with] FOMAl. The same surplus was found for yttrium in comparison with aluminum at the respective ‘spectrally’ equivalent thicknesses of 0.10 mm Y and 4.08 mm Al while CY [congruent with] CAl and FOMY [congruent with] FOMAl. This is in good agreement with the results from the work of Jennings,[8] which reported a difference of 15.3% (experimental) and 16.7% (theoretical) for TCu and TAl with copper transmitting surplus. Also, reported by the same author[8] was a difference of 15.4% (experimental) and 16.4% (theoretical) for TY and TAl with yttrium filter transmitting surplus. With the transmitted spectrum through the copper filter scaled (mAs and heat capacity scaled by the factor α), the results obtained from the simulations carried out in this work were such that TAl [congruent with] TCu [congruent with] TY, DOSEAl [congruent with] DOSECu [congruent with] DOSEY, [CNRAl]2 [congruent with] [CNRCu]2 [congruent with] [CNRY]2, CAl [congruent with] CCu [congruent with] CY and FOMAl [congruent with] FOMCu [congruent with] FOMY.

Using scaled spectrum {α ΦCu (Ei)}, operating the tube with copper filter at reduced tube loading/heat capacity (α × mAsAl or 1.35α × mAsAl × kVp) results in same value of T, DOSE, C, CNR and FOM when compared with operating the tube with aluminum filter at higher tube loading/heat capacity (mAsAl or 1.35 × mAsAl × kVp). Note that whether the beams are scaled or not scaled, once the spectral shape is the same, the values of contrast and FOM will be the same. However, beams filtered by materials with higher values of atomic numbers are more intense (bigger in shape) and produce higher values of CNR and dose to patient than those filtered by material of lower atomic numbers at ‘spectrally’ equivalent thicknesses when the tube is operated at the same values of kVp, tube loading and heat capacity [Figures [Figures22 and and3].3]. For a given filter material when compared with aluminum/copper at ‘spectrally’ equivalent thicknesses, only the shapes of transmitted spectra are the same. Depending on the Z-value, the sizes are not the same when the tube is operated at same values of kVp, tube loading and heat capacity [Figure 3]. At respective ‘spectrally’ equivalent thicknesses, the benefits derivable from the use of higher Z-value filter material in comparison with those from the use of lower Z-value include higher intensity, lower filter-generated scattered radiation and shorter tube current or exposure time. This can translate into an increase in the lifespan of tube (more filament hours by operating at reduced mAs) and reduction in the cost of dealing with the detrimental effects of heat and scattered radiation. Also, extra images could be obtained using the gain in exposure time and tube loading/heat capacity.

The spectrum obtained by numerically scaling the values of α[phi]Cu (Ei) using the factor α (Eq. 14) is in agreement with that transmitted by 3.7 mm Al [Figure 3]. The scaled spectrum obtained from copper filter {α[phi]Cu(Ei)} and that directly obtained from aluminum {[phi]Al(Ei)} were found to produce the same contrast, CNR, DOSE and FOM. Though not carried out in this work, experimentally at the same kVp, the spectra obtained from filter material with higher Z-value (say copper) in comparison with those with lower Z-value (say aluminum) could be obtained at the scaled intensity and dose (scaled to that obtainable from aluminum filter) by operating the tube at scaled tube loading and heat capacity. This scaling factor provides for the means of obtaining strictly equivalent (same shape and size) spectra from two different elemental filter materials (conventional filters) at the same value of kVp [Figure 3].

The validity of Eq. (15) had been reported in Okunade[6] using the case of peak voltage of 86 kV and comparison of 0.10 mm Cu with the corresponding ‘spectrally’ equivalent aluminum thickness, 3.7 mm Al (reference filter). Also, Figure 3 shows that the formulation presented as TAl = α × TCu is valid. Hence all the formulations in Eq. (15) are valid. Table 5 shows that the values of ‘spectrally’ equivalent thicknesses obtained using Eqs.(1618) are in good agreement with those earlier reported in literature.[89] Apart from providing comparison for aluminum/copper and any given filter material irrespective of atomic number, EQSPECT.FOR can be used to obtain relative performance for any arbitrary pair of filter materials, say silicon and iron. This is accomplishable by specifying the reference filter material as silicon and the alternative filter material as iron.

Generally, it is well known that it is not possible to have a complete spectral matching for conventional and k-edge filter materials. The k-edge discontinuity interferes substantially in the range of useful energies, resulting in beams with significant difference in shape and size. Unlike the case of the comparison of conventional filter materials (aluminum/copper) with other conventional filter materials, large values of the root-mean-square errors[8] were obtained when conventional filter materials were compared with k-edge filter materials. This implies a large difference between the actual transmission of aluminum/copper and the scaled transmission of the k-edge filters. It is not possible to derive simple mathematical formulations of the form with factors α, δ and [increment] for the comparison of performance of beams from conventional filter materials with those from k-edge filter materials since spectrals transmitted by these pair of filter materials differ in quality and quantity [Figure 4]. Theoretically, the transmission below the k-edge in terms of quality and quantity could not be matched in these comparisons. However, using Eq. (9), the results presented in Table 6 for the comparison of the overall quantity of photons in terms of the ratios of T and DOSE from beams transmitted by 0.097 mm Cu and 0.088 mm Gd are in reasonable agreement with experimentally measured ratio of exposure time reported by Nagel.[9]

Conclusions

Models and software program have been presented for the comparative evaluation of the performance characteristics (contrast, contrast-to-noise ratio and FOM) of beams transmitted at spectral equivalent thicknesses by elemental filter materials (conventional and k-edge filters). Under this condition of spectral equivalence, the mathematical formulations presented show that when compared with one of lower Z-value, conventional filter material with higher Z-value offers opportunity for extended exposure time. This extended exposure time can translate into acquiring extra images using the gain in exposure time, tube loading and heat capacity. The computer software EQSPECT.FOR developed in this work can be used as a tool for the comparative investigation of filter materials with a view to determine alternative beams that could yield optimum performance in clinical radiological practice. The results from this comparative investigation can aid in the design of X-ray machines in such a way as to produce a reference shape and size of X-ray beam with which others can be compared. This could assist in the achievement of some degree of uniformity in clinical X-ray diagnostic practice.

Footnotes

Source of Support: Nil

Conflict of Interest: None declared.

References

1. Boone JM. Equivalent spectra as an index of beam quality. Med Phys. 1986;13:861–8. [PubMed]
2. Boone JM. Three parameter equivalent spectra as an index of beam quality. Med Phys. 1988;15:304–10. [PubMed]
3. Thoraeus R. The study of the ionization method for measuring the intensity and absorption of Roentgen rays and efficiency of different filters used in therapy. Acta Radiol. 1932;15:1–86.
4. Thoraeus R. Tin filters in roentgen therapy. Acta Radiol. 1934;15:225–36.
5. Okunade AA. Numerical models for comparing filter materials for diagnostic radiology. Radiat Phys Isot. 2002;65:1–9.
6. Okunade AA. Parameterized algorithm for quantitative differentials in spectrally equivalent medical diagnostic x-ray beams. Med Phys. 2005;32:1785–95. [PubMed]
7. Nagel HD. Aluminum equivalence of materials used in diagnostic radiology and its dependence on beam quality. Phys Med Biol. 1986;31:1381–99. [PubMed]
8. Jennings RJ. A method for comparing beam-hardening filter materials for diagnostic radiology. Med Phys. 1988;15:588–99. [PubMed]
9. Nagel HD. Comparison of the performance of characteristics of conventional and k-edge filters in general diagnostic radiology. Phys Med Biol. 1989;31:1269–87.
10. Birch R, Marshall M, Ardran GM. The Hospital Physicist’s Association Scientific Report Series -30. 1979. Catalogue of spectral data for diagnostic x-rays.
11. Okunade AA. Effective dose as a limiting quantity for the evaluation of primary barriers for diagnostic x-ray facilities. Health Phys. 2005;89:S100–16. [PubMed]
12. Boone JM, Seibert JA. An accurate method for computer-generated tungsten anode x-ray spectra from 30 to 140 kV. Med Phys. 1997;24:1661–70. [PubMed]
13. McMaster WH, Del Grande NK, Mallett JH, Hubbell JH. Compilation of X-ray cross section”. University of California, Lawrence Radiation Laboratory Report: 1969. No UCRL-50174, Sec, II, Rev 1 (NTIS)
14. Hubbell JH, Seltzer SM. National Institute of Standard and Technology Internal Report NISTIR. 1995. Tables of x-ray mass attenuation coefficients and mass energy-absorption coefficients 1 keV to 20 MeV for elements 1-92 and 48 additional substances of dosimetric interest; p. 5632.
15. GRAF4WIN. A computer program for graphics and curve fitting GRAPHER(™) for Windows v1.2 Golden Software Incorporated. Golden, CO: USA; 1993-1994.

Articles from Journal of Medical Physics are provided here courtesy of Medknow Publications