J Diabetes Sci Technol. 2010 September; 4(5): 1146–1155.
Published online 2010 September 1.
PMCID: PMC2956822

Hypoglycemia Prevention via Pump Attenuation and Red-Yellow-Green “Traffic” Lights Using Continuous Glucose Monitoring and Insulin Pump Data

Colleen S Hughes, M.S.,1 Stephen D Patek, Ph.D.,1 Marc D Breton, Ph.D.,2 and Boris P Kovatchev, Ph.D.2

Abstract

Background

Hypoglycemia has been identified as a primary barrier to optimal management of diabetes. This observation, in conjunction with the introduction of continuous glucose monitoring (CGM) devices, has set the stage for achieving tight glycemic control with systems that adjust the insulin pump settings based on measured glucose concentrations. Because system safety and system reliability are key considerations, there is a need for algorithms that reduce the risk of hypoglycemia in closed-loop, open-loop, and advisory-mode systems. More specifically, the algorithm presented here is formulated as a component of the independent safety system module proposed in the modular control-to-range architecture.

Methods

We developed two algorithms for attenuating insulin pump injections, which we refer to as Brakes and Power Brakes: Brakes is a pump attenuation function computed using CGM information only, while Power Brakes is an attenuation function in which a metabolic state observer with insulin input is used in addition to CGM information to inform the level of pump attenuation. These algorithms modulate the insulin pump delivery so that the insulin injection rate is dramatically reduced when the risk of hypoglycemia is high. Additionally, we combined these algorithms with an alert system that indicates a level of hypoglycemic risk to the user.

Results

We demonstrated the effectiveness of Brakes and Power Brakes in reducing the incidence of hypoglycemia in two simulated scenarios: an elevated basal rate scenario and a scenario in which a bolus is delivered for a meal that is skipped. For these scenarios, the incidence of hypoglycemia using Power Brakes was reduced by 88 and 94%, respectively, where we defined hypoglycemia based on the American Diabetes Association guidelines for defining and reporting as 70 mg/dl. In the elevated basal rate scenario, no rebounds above 180 mg/dl (the desired upper limit of the control-to-range protocol) following hypoglycemia were shown to occur. We demonstrated the way the hypoglycemia alert system can trigger the intake of carbohydrates to reduce the incidence of hypoglycemia by 98%.

Conclusions

This article offers, for the first time, a method for smoothly reducing insulin delivery rate to prevent hypoglycemia in individuals with type 1 diabetes mellitus based on a mathematically formal assessment of hypoglycemic risk. In silico, we demonstrate the way this method can prevent hypoglycemia while avoiding hyperglycemia rebounds that exceed 180 mg/dl. In conjunction with the pump attenuation algorithms, this article also proposes a system for alerting an individual of their hypoglycemic risk that contains three “levels” of alerts in the form of a traffic light. This alert system can be used in an advisory mode setting to alert the user to take action when hypoglycemia is imminent (“red” light) or in a closed-loop setting where initiation of rescue action begins when the red light alert is triggered.

Keywords: hypoglycemia, Kalman filter, modeling, risk of hypoglycemia, simulation

Introduction

Hypoglycemia has been identified as a primary barrier to optimal management of diabetes.1,2 In this article, we present two algorithms for hypoglycemia prevention and detection, one that uses continuous glucose monitoring (CGM) information only and one that uses both CGM and insulin injection information. The algorithms continually assess the risk of hypoglycemia and attenuate the insulin pump delivery rate in accordance with this formal risk assessment. In addition, these algorithms are designed to provide a signal to the user indicating one of three possible “levels” of hypoglycemic risk consistent with the formal risk assessment. This signal is given in the form of a traffic light: green light indicates no risk of hypoglycemia, yellow light indicates that a risk of hypoglycemia is present, and red light indicates imminent hypoglycemia, either initiating automated rescue action from the controller in a closed-loop setting or suggesting the need for rescue action to the user in an advisory mode. We consider the issues of user compliance and time of alarm to be outside the scope of this work. Employing our methods, a “safety supervision” function can be added to different types of blood glucose (BG) management methods, including conventional therapy, open-loop and advisory-mode systems, and closed-loop systems, as in the control-to-range architecture of Kovatchev and colleagues.3 Under the assumption that other components of the safety module will supervise insulin boluses, the algorithms presented here serve to modify basal insulin rates.

To outline the remainder of this article, we first provide a background of the current methods employed for hypoglycemia detection and prevention. In the Methods section, we present two algorithms for hypoglycemia prevention: Brakes and Power Brakes. In the Results section, we present in silico results of pump attenuation methods in two instances of hypoglycemia, elevated basal rate and overbolus scenarios, using the University of Virginia/University of Padova Metabolic Simulator described by Kovatchev and colleagues,4 which is accepted by the U.S. Food and Drug Administration.

Background

Current commercial CGM devices contain an alarm system that alerts the user of hypoglycemia (see References 5 and 6 for a thorough review). The methods designed to generate these alarms come in two general types: low-threshold detection and predictive. Hypoglycemia alarms based on low-threshold detection have a BG concentration such that, when the CGM measurements are at or below this low threshold, an alarm is triggered alerting the user of their risk of hypoglycemia. Some commercial CGM devices also contain predictive alarms. The nature of this type of alarm gives the user the opportunity to administer rescue action so that the hypoglycemic event can be avoided. Rescue action to prevent hypoglycemia may take the form of glucagon infusion7,8 or fast-acting oral consumption (e.g., glucose tablets).

Other types of alarms for hypoglycemia are based on a combination of glucose threshold and glucose rate of change. In this case, if the CGM device measurements are reading low and the rate of decrease in CGM device measurements is high, an alarm is triggered.9 There has been extensive study of the effectiveness of alarms for the commercial CGM devices.1014 Various algorithmic methods for predictive alarms have also been proposed based on linear regression,15 time series,16 and optimal estimation theory.17

In order to be effective in reducing the incidence of hypoglycemia, alarms must be accompanied by some type of action. Buckingham and associates18 proposed a hypoglycemia prevention method based on insulin pump shutoff: when the predicted glucose concentration reaches a certain value, the pump halts the delivery of insulin. This method has been tested in a clinical setting18 and has been shown to reduce the incidence of hypoglycemia while avoiding a glucose rebound effect that can occur when the insulin pump is shut off for a period of 90 minutes. The pump shutoff method proposed regulates insulin delivery to reduce the incidence of hypoglycemia, while another method introduced by Choleau et al.19 demonstrates the effectiveness of a glucose infusion to mimic the appearance of glucose from an intragastric load.

In what follows, we describe a detection method that alerts the user of a “level” of hypoglycemic risk as well as two algorithms for hypoglycemia prevention that work by gradually attenuating insulin delivery.

Methods

Here, we present two algorithms for the attenuation of insulin pump delivery: Brakes and Power Brakes. The similarity between these two algorithms is that they both provide a method for smoothly attenuating the insulin pump delivery rate based on a formal assessment of hypoglycemic risk. The difference comes in the information (CGM only or CGM and insulin) that is used to make this formal risk assessment.

Brakes: Based on Continuous Glucose Monitoring Data

In this subsection, we discuss the implementation of an attenuation of the insulin delivery rate computed using CGM measurements alone. We refer to this method as Brakes.

Attenuation Function

The role of Brakes is to adjust the insulin rate commands to avoid hypoglycemia (refer to Figure 1 throughout this section). The brake algorithm is designed to smoothly attenuate the insulin delivery rate of the patient at the current time by monitoring the CGM data, assessing a measure of the patient's current risk of hypoglycemia, R(y(t)), and then computing an attenuation factor [var phi]brackes(R(y(t))):

Figure 1.
Schematic of Brakes.
equation m1
(1)

where y(t) is a measure of the patient's current BG state (described in more detail later), and Γ is a patient-specific “aggressiveness” parameter. The parameter Γ is patient-specific parameter tuned using the University of Virginia/University of Padova Metabolic Simulator. The optimal value of Γ is chosen to avoid hypoglycemia (70 mg/dl) with minimum aggressiveness (based on the American Diabetes Association guidelines for defining and reporting20). Following simulation experiments, we conduct a correlation analysis for Γ with a set of known patient-specific biometric parameters that include total daily insulin (TDI; U), correction factor (CF; mg/dl/U), body weight (kg), and carbohydrate ratio (gCHO/U). Based on this correlation analysis, the per-patient optimal Γ is set according to the following regression formula:

equation m2
(2)

The attenuation factor [var phi]brakes(R(y(t)))) is used to compute the reduced insulin pump rate Jactual(t) by

equation m3
(3)

where Jcommand(t) is the rate of insulin injection (U/h) that the pump would administer without Brakes, and Jactual(t) is the attenuated insulin delivery rate (U/h). In this case, where the risk is computed using CGM information alone, the formula for the risk function is as follows:

equation m4
(4)

where y(t) is a 15 min nonweighted moving average of the CGM device measurements, that is, equation m18, where CGM(t) is defined as the CGM device measurement at time t, equation m19 is the rate of change in CGM assessed over a 15 min moving average window, and this condition on R(y(t)) is set so as to avoid the potential for hyperglycemia rebound. The form of R(y(t)) for y(t) [set membership] (20,120) is adopted from the BG symmetrization function and the low blood glucose index introduced by Kovatchev and colleagues,21 where a transformation of the BG to a “risk scale” is introduced. Figure 2 provides a representative subject example of the attenuation factor [var phi]brakes(R(y(t))) applied to the insulin delivery rate to prevent hypoglycemia.

Figure 2.
Attenuation factor derived from the Brakes algorithm shown with the corresponding BG trace for a representative subject with a noisy CGM input signal.

Power Brakes: Based on Continuous Glucose Monitoring and Insulin Pump Data

In this subsection, we introduce an algorithm that is informed with the history of past insulin delivery rates that allows us to attenuate insulin pump delivery rate in anticipation of hypoglycemia by accounting for the insulin that has been injected. We refer to this as Power Brakes, as the use of both insulin information and prediction increase our ability to prevent hypoglycemia through pump attenuation. (Please refer to Figure 3 throughout this section.) Here, we present a Kalman filtering technique to account for CGM signal noise.

Figure 3.
Schematic of Power Brakes.

Metabolic State Observer Design

The model from which our metabolic state observer is derived is the compartmental “minimal model” of Bergman and colleagues,22 with extensions that account for subcutaneous oral glucose sensing and actuation. The coefficients of the model reflect population average glucose–insulin dynamics with a 1-minute sample time. The discretized, linearized implementation of this model in state space form is

equation m5
(5)

where x(k) is the state of the system k at the the sample time, u(k) = Jactual(k)Jbasal (in mU/min) is the insulin input signal at stage k (held constant between samples)—where Jactual(k) (in mU/min) is the current rate of insulin infusion and Jbasal (in mU/min) is the minimum value over the patient's basal rate profile—and ω(k) = meal(k)mealref (in mg/min) is the ingested glucose disturbance signal at stage k, where meal(k) (in mg/min) is the rate of ingested glucose at stage k and mealref (in mg/min) is the reference meal input value. Matrices A, B and G are derived from a linearized, discretized form of the model, with model parameters chosen to be representative of an average adult patient with type 1 diabetes mellitus (T1DM). Matrix parameters are provided in the Appendix.

The state observer is based on a steady state Kalman filter. Since only CGM device measurements are available at each 1-minute sample period, it is necessary to compute estimates equation m20 of x(k) based on the knowledge of infused insulin rate u(k) and measurements y(k), where

equation m6
(6)

where CGM(k) is the readout of the continuous glucose monitor at stage k and Gref is a reference glucose value, set here at 112.5 mg/dl. We model the measurement signal as

equation m7
(7)

where v(k) (mg/dl) represents the CGM signal noise and is added to the state variable representing plasma glucose. The matrix C is given by

equation m8
(8)

The metabolic state observer is derived from the state space model, which generates x(k) and the measurements y(k) as a Kalman filter, treating the meal disturbance process ω(k) and the noise process v(k) as zero-mean, white Gaussian processes with covariances Rs = 1 (mg/min) and Qs = 5e – 4 (mg/dl), respectively. These covariance parameters are time and state invariant.

The observer itself can be expressed recursively (as a dynamic process) as

equation m9
(9)

where equation m21 refers to the best estimate of using data x(k) collected up to stage k − 1 and equation m22 refers to the best estimate of x(k) given information up to stage k. The filter gain matrix is defined by

equation m10
(10)

where the matrix Pf is the unique stabilizing solution to the algebraic Riccati equation:

equation m11
(11)

(In our implementation, this step is performed in Matlab using the command “kalman.”) We point out that, even though meals ω(k) and sensor noise v(k) are not zero-mean, white Gaussian processes in reality, the resulting Kalman filter is still a stable observer.

We compute an estimate of the (projected) BG concentration at stage k by

equation m12
(12)

where C is defined in Equation (8), τ represents an amount of time (in minutes), and

equation m13
(13)

where Aτ is the A matrix of the state space model raised to the τth power and

equation m14
(14)

We note that the insulin input signal and the meal signal are held at their operating point values of Jbasal and mealref, respectively, over the prediction horizon τ. In computing equation m23, τ is a parameter that must be specified; τ = 0 corresponds to assessing risk based on the best estimate of BG given all the data received up to stage k, while τ > 0 corresponds to an assessment of the future risk of hypoglycemia, giving Power Brakes the opportunity to intervene well before the onset of hypoglycemia. We set τ = 15 minutes.

Attenuation Function

An estimate of the BG concentration allows us to assess the risk of hypoglycemia, informed now by our metabolic state observer, Rm:

equation m15
(15)

where equation m24 is from Equation (13) and equation m25 is the rate of change in the projected glucose over a 15 min moving average window. Given equation m26 we compute the amount of insulin pump attenuation by

equation m16
(16)

where the aggressiveness parameter Γ is as defined in Equation (2).

The attenuation factor [var phi]powerbrakesequation m27 is used to compute the reduced pump rate Jactual(k) by

equation m17
(17)

where Jcommand(k) is the rate of insulin injection (U/h) that the pump would administer without Power Brakes and Jactual(k) is the attenuated insulin delivery rate (U/h).

Hypoglycemia Traffic Lights for Brakes and Power Brakes

This subsection introduces a corresponding traffic light signal designed to alert the user as to their level of hypoglycemic risk. The traffic light works in conjunction with the risk computation as follows, where Kred and Kred,IOB are BG thresholds used to trigger the transition from the yellow to the red light signal for use in conjunction with Brakes and Power Brakes, respectively. For Brakes, the lights are triggered as follows:

  • If R(y(t)) = 0, the green light is triggered.
  • If R(y(t)) > 0 and y(t) ≥ Kred, the yellow light is triggered.
  • If y(t) < Kred, the red light is triggered.

The value of Kred is chosen as 80 mg/dl. For Power Brakes, the lights are triggered similarly, but our threshold for the trigger is based on a Kalman filter state estimate:

  • If equation m28, the green light is triggered
  • If equation m29 and equation m30Kred,IOB, the yellow light is triggered
  • If equation m31 < Kred,IOB, the red light is triggered.

We formulate equation m32 in the same way as equation m33 in Equation (13), except that we hold the insulin input signal Jactual(k) at 0 over the prediction horizon τ so as to trigger the red light when it is predicted that no amount of pump attenuation will allow hypoglycemia to be avoided. The value of Kred,IOB is chosen as 77.5 mg/dl.

For implementation, this signal would appear on the CGM device display in conjunction with an alert sound or vibration. This threshold is designed to allow the user the time to administer rescue carbohydrates prior to the onset of hypoglycemia to mitigate or avoid the event. During recovery from a hypoglycemic event, the red light remains “on” until at least 80% of the past 30 CGM measurements are above 90 mg/dl.

Results

In this section, we present a series of results that demonstrate the ability of both Brakes and Power Brakes to lower the incidence of hypoglycemia. We also present results demonstrating the performance of the red light alarm in detecting hypoglycemia. Experiments are run on the (FDA-accepted) University of Virginia/University of Padova Metabolic Simulator at the University of Virginia with the noise model developed by Breton and Kovatchev.23 We present two scenarios designed to cause hypoglycemia if no action is taken:

  1. Some T1DM patients experience highly variable insulin sensitivity (e.g., after physical activity). For such patients, it happens that his/her basal rate of insulin delivery, which is tuned to achieve fasting euglycemia under normal circumstances, is suddenly too high, putting the patient at risk of hypoglycemia. This change in the effectiveness of insulin has been observed following exercise.24 We simulate this increased sensitivity to insulin by delivering two times the normal basal rate of the subject, where this increase in basal level can be used to simulate the increase in insulin sensitivity following exercise.25 All 100 adult in silico subjects are initialized at glucose concentrations of 150 mg/dl and are subjected to a basal insulin delivery rate that is two times what would be required to achieve their respective fasting glucose concentration of 112.5 mg/dl. Figure 4 shows the results of this scenario in the case when (a) no action is taken, (b) Brakes are applied, (c) Power Brakes are applied.
    Figure 4.
    Elevated basal rate scenario. The plots show the mean BG trace (blue) for 100 subjects and ±1 standard deviation (orange).
  2. Patients often administer premeal insulin boluses in anticipation of the meal they are about to consume. In unusual circumstances, the patient may forget (or otherwise be unable) to eat the anticipated meal. This scenario puts the patient at severe risk of hypoglycemia. We simulate this by administering a premeal bolus for a meal that is missed. The amount of this premeal bolus is designed such that the subject will drop to a BG concentration of 50 mg/dl (from an initial value of 150 mg/dl) if the meal is not taken (we note that the size of this “missed meal” will vary from patient to patient, depending on their sensitivity to insulin). We can also think of this scenario as a case in which a patient administers a bolus that is too large to correct for a positive deviation from a target glucose concentration. The patient may intend to take a bolus in order to return to a glucose concentration of 90 mg/dl, but instead the bolus taken is incorrectly computed to reach a target of 50 mg/dl. Figure 5 shows the results of this scenario in the case when (a) no action is taken, (b) Brakes are applied, and (c) Power Brakes are applied. Figure 6 shows the way in which the red light alarm can be triggered to deliver rescue carbohydrates. In this case, 16 g of rescue carbohydrates are delivered at the time that the red light is triggered.
    Figure 5.
    Missed meal (or over-bolus) scenario. Plots show the mean BG trace (blue) for 100 subjects and ±1 standard deviation (orange).
    Figure 6.
    Missed meal (or overbolus) scenario. Power Brakes is employed to attenuate insulin delivery rate, and the red light alarm triggers the automated administration of 15 g rescue carbohydrates. The left plot shows the mean BG trace (blue) for 100 subjects ...

In the case of both the elevated basal rate and the missed meal scenario, our results indicate the ability of both Brakes and Power Brakes to reduce the incidence of hypoglycemia (when compared to the case where no action is taken). In scenario 1, Brakes works well to gradually attenuate the insulin delivery rate in reaction to a gradual decrease in BG when insulin information is not available. In scenario 2, Power Brakes, using the insulin injection information, is aware that the large bolus delivered in anticipation of a meal will lead to decreasing BG concentrations. This is reflected in the projected estimate of the BG state obtained from our model of glucose–insulin kinetics and metabolic state observer. Power Brakes acts aggressively in scenario 2 by considering the projected effect of the insulin on board. Another key benefit of Power Brakes is its performance during a recovery from a low BG excursion: as soon as predicted BG rate of change is positive, the attenuation effect is released.

In the case of scenario 1, where the basal rate is elevated to two times basal, we experience no rebound glucose concentrations above 180 mg/dl following the pump attenuation for both Brakes and Power Brakes (where, following the recovery from low BG, the basal rate is returned to a normal basal rate) over 12 hours of total simulation time. In scenario 2, where our pump attenuation algorithm responds to an overbolus of insulin, rebounds above 180 mg/dl occur in less than 3% of subjects.

In Tables 1 and 22, we present the sensitivity, false red light alarm, and time to hypoglycemia event statistics for our alarm system in the case of (1) elevated basal rate and (2) overbolus scenarios. For each scenario, we consider the red light alarm designed for both Brakes and Power Brakes, as described in Methods.

Table 1.
Red Light Alarm Statistics for Scenario 1
Table 2.
Red Light Alarm Statistics for Scenario 2

Results from Tables 1 and 22 indicate the ability of the red light alarm to provide an alert that can be used to trigger rescue carbohydrates. We note that the percentage of hypoglycemic episodes detected represents only the percentage of episodes detected prior to the onset of hypoglycemia. Events that trigger the red light alarm at any time following the onset of hypoglycemia are not considered “detected” events, as we assume that the purpose of the red light alarm is to alert the control system (or the user) so that action can be taken to prevent or mitigate the event. Time to event statistics indicate that Power Brakes, because of its ability to account for insulin on board, provides a longer time, on average, between the red light alarm and the hypoglycemic event.

Discussion

In Figure 6, we demonstrate the way in which hypoglycemia can be prevented when the red light alarm triggers action through the delivery of rescue carbohydrates. This ability of Power Brakes to anticipate imminent hypoglycemia allows adequate time for taking preventative action, which avoids the hypoglycemic episode in 98% of cases. In this case of a closed-loop control system, the delivery of rescue action in the form of glucagon can be envisioned as an automated process. In an advisory mode, the red light trigger will prompt an alarm to the user, advising that rescue action be taken. While the issue of user compliance and alarm time of day is deemed outside the scope of the work presented here, we envision the simulation of user delay and/or the probability of user compliance in future work.

The simulation experiments presented here offer insight into the use of Brakes and Power Brakes for in vivo patients in open-loop, advisory-mode, and closed-loop settings. In an advisory-mode setting, communication with the user through the red light alarm gives the user an opportunity to react to increasing hypoglycemic risk. In a future closed-loop setting, the gradual attenuation of the insulin pump delivery rate and the administration of rescue in the form of an automated glucose infusion offer the potential to prevent hypoglycemic events.

Conclusions

We describe two methods to attenuate insulin levels gradually to avoid the risk of hypoglycemia. While Brakes can compute pump attenuation without requiring access to insulin history, Power Brakes uses insulin pump feedback to increase accuracy of the hypoglycemia risk assessment. The pump attenuation method for hypoglycemia prevention presented here is contrasted with pump shutoff methods, as it does not suffer the complexity of deciding (1) exactly when to shut off and (2) exactly when to resume operation, with both decisions being significantly hampered by CGM device noise/errors. Smooth attenuation of the insulin pump delivery rate based on risk allows us to conveniently handle the case when glucose is dropping but is not yet at a dangerously low level. Spurious errors in the CGM device signal can only become spurious errors in the degree of attenuation since there is never a point in time where a “crisp” attenuation decision has to be made.

Additionally, we present an alert system that communicates with the user to indicate a level of hypoglycemic risk. Results indicate the power of the red light alert to trigger action in the form of rescue carbohydrates to prevent a hypoglycemic event.

Abbreviations

BG
blood glucose
CGM
continuous glucose monitoring
T1DM
type 1 diabetes mellitus

Appendix

Matrix model parameters/model equations for the state space model presented in Equation (5):

equation m34

References

1. Cryer PE. Hypoglycaemia: the limiting factor in the glycaemic management of type I and type II diabetes. Diabetologia. 2002;45(7):937–948. [PubMed]
2. Cryer PE, Davis SN, Shamoon H. Hypoglycemia in diabetes. Diabetes Care. 2003;26(6):1902–1912. [PubMed]
3. Kovatchev B, Patek S, Dassau E, Doyle FJ, III, Magni L, De Nicolao G, Cobelli C. Juvenile Diabetes Research Foundation Artificial Pancreas Consortium. Control to range for diabetes: functionality and modular architecture. J Diabetes Sci Technol. 2009;3(5):1058–1065. [PMC free article] [PubMed]
4. Kovatchev BP, Breton M, Dalla Man C, Cobelli C. In silico preclinical trials: a proof of concept in closed-loop control of type 1 diabetes. J Diabetes Sci Technol. 2009;3(1):44–55. [PMC free article] [PubMed]
5. Buckingham B, Caswell K, Wilson DM. Real-time continuous glucose monitoring. Curr Opin Endocrinol Diabetes Obes. 2007;14(4):288–295. [PubMed]
6. Bequette BW. Continuous glucose monitoring: real-time algorithms for calibration, filtering, and alarms. J Diabetes Sci Technol. 2010;4(2):404–418. [PMC free article] [PubMed]
7. Haymond MW, Schreiner B. Mini-dose glucagon rescue for hypoglycemia in children with type 1 diabetes. Diabetes Care. 2001;24(4):643–645. [PubMed]
8. Pearson T. Glucagon as a treatment of severe hypoglycemia: safe and efficacious but underutilized. Diabetes Educ. 2008;34(1):128–134. [PubMed]
9. Chase HP, Beck R, Tamborlane W, Buckingham B, Mauras N, Tsalikian E, Wysocki T, Weinzimer S, Kollman C, Ruedy K, Xing D. A randomized multicenter trial comparing the GlucoWatch Biographer with standard glucose monitoring in children with type 1 diabetes. Diabetes Care. 2005;28(5):1101–1106. [PubMed]
10. Bode B, Gross K, Rikalo N, Schwartz S, Wahl T, Page C, Gross T, Mastrototaro J. Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the Guardian Continuous Monitoring System. Diabetes Technol Ther. 2004;6(2):105–113. [PubMed]
11. McGarraugh G, Bergenstal R. Detection of hypoglycemia with continuous interstitial and traditional blood glucose monitoring using the FreeStyle Navigator Continuous Glucose Monitoring System. Diabetes Technol Ther. 2009;11(3):145–150. [PubMed]
12. Weinstein RL, Schwartz SL, Brazg RL, Bugler JR, Peyser TA, McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator Glucose Monitoring System: comparison with frequent laboratory reference measurements. Diabetes Care. 2007;30(5):1125–1130. [PubMed]
13. Mastrototaro J, Welsh JB, Lee S. Practical considerations in the use of real-time continuous glucose monitoring alerts. J Diabetes Sci Technol. 2010;4(3):733–739. [PMC free article] [PubMed]
14. McGarraugh G. Alarm characterization for continuous glucose monitors used as adjuncts to self-monitoring of blood glucose. J Diabetes Sci Technol. 2010;4(1):41–48. [PMC free article] [PubMed]
15. Cameron F, Niemeyer G, Gundy-Burlet K, Buckingham B. Statistical hypoglycemia prediction. J Diabetes Sci Technol. 2008;2(4):612–621. [PMC free article] [PubMed]
16. Sparacino G, Zanderigo F, Corazza S, Maran A, Facchinetti A, Cobelli C. Glucose concentration can be predicted ahead in time from continuous glucose monitoring sensor time-series. IEEE Trans Biomed Eng. 2007;54(5):931–937. [PubMed]
17. Palerm CC, Willis JP, Desemone J, Bequette BW. Hypoglycemia prediction and detection using optimal estimation. Diabetes Technol Ther. 2005;7(1):3–14. [PubMed]
18. Buckingham B, Cobry E, Clinton P, Gage V, Caswell K, Kunselman E, Cameron F, Chase HP. Preventing hypoglycemia using predictive alarm algorithms and insulin pump suspension. Diabetes Technol Ther. 2009;11(2):93–97. [PMC free article] [PubMed]
19. Choleau C, Dokladal P, Klein JC, Ward WK, Wilson GS, Reach G. Prevention of hypoglycemia using risk assessment with a continuous glucose monitoring system. Diabetes. 2002;51(11):3263–3273. [PubMed]
20. Workgroup on Hypoglycemia, American Diabetes Association. Defining and reporting hypoglycemia in diabetes: a report from the American Diabetes Association Workgroup on Hypoglycemia. Diabetes Care. 2005;28(5):1245–1249. [PubMed]
21. Kovatchev BP, Cox DJ, Gonder-Frederick LA, Clarke W. Symmetrization of the blood glucose measurement scale and its applications. Diabetes Care. 1997;20(11):1655–1658. [PubMed]
22. Bergman RN, Ider YZ, Bowden CR, Cobelli C. Quantitative estimation of insulin sensitivity. Am J Physiol. 1979;236(6):E667–677. [PubMed]
23. Breton M, Kovatchev B. Analysis, modeling, and simulation of the accuracy of continuous glucose sensors. J Diabetes Sci Technol. 2008;2(5):853–862. [PMC free article] [PubMed]
24. Brun JF, Guintrand-Hugret R, Boegner C, Bouix O, Orsetti A. Influence of short-term submaximal exercise on parameters of glucose assimilation analyzed with the minimal model. Metabolism. 1995;44(7):833–840. [PubMed]
25. Dalla Man C, Breton MD, Cobelli C. Physical activity into the meal glucose-insulin model of type 1 diabetes: in silico studies. J Diabetes Sci Technol. 2009;3(1):56–67. [PMC free article] [PubMed]
26. Magni L, Raimondo DM, Dalla Man C, Breton M, Patek S, Nicolao GD, Cobelli C, Kovatchev BP. Evaluating the efficacy of closed-loop glucose regulation via control-variability grid analysis. J Diabetes Sci Technol. 2008;2(4):630–635. [PMC free article] [PubMed]

Articles from Journal of Diabetes Science and Technology are provided here courtesy of Diabetes Technology Society