Derivations that enable the testing of fetal urine production as a method of fetal surveillance

Abstract

Purpose

To calculate the measurement error of the hourly fetal urine production rate (HFUPR) and evaluate the implication of different methods for measuring the HFUPR, i.e. ellipsoid versus sum-of-cylinders method.

Methods

The calculation was based on sonographic documentation of the increased bladder volumes during the filling phase, the bladder volume measurement error and the number and time points of bladder image capture.

Results

The probability of a false pathological reading was excluded (0%) with the sum-of-cylinders method for gestational ages of ≥30 weeks. With the ellipsoid method, the risk was higher. The maximum changes which could be exclusively explained by measurement error were four to five times greater with the ellipsoid method compared with the sum-of-cylinders method.

Conclusions

The present paper illustrates a careful evaluation of the HFUPR measurement error and the implications of using different ultrasound methods for bladder volume estimations.

Electronic supplementary material

The online version of this article (doi:10.1007/s00404-009-1242-6) contains supplementary material, which is available to authorized users.

Keywords: Fetus, Urinary bladder, Ultrasonography, Organ volume, Reliability

Introduction

The skillful management and timing of delivery in high-risk pregnancies is important in order to avoid a poor perinatal outcome [5, 16]. The challenge is the early identification of abnormal events. Different procedures to detect these abnormal events are available [6, 18]. However, the fetal urine production rate is a parameter that has not been utilised in clinical practice, even though several studies demonstrate a reduction in blood flow in the fetal renal arteries and urine production rate in compromised fetuses [810, 14, 19].

When utilising the hourly fetal urine production rate (HFUPR) to detect abnormal events during pregnancy, reliable reference HFUPR values for normality are needed. Several attempts have been made to assess these values. However, previous ultrasound studies have presented widely varying reference ranges for fetuses of identical gestational ages. For example, at term, the presented values have varied from as much as 28, 34, 51 and 71–125 mL/h [1, 7, 11, 12, 17]. The huge variation was also present, regardless of whether the 2D or 3D technique was used, suggesting considerable measurement errors. Furthermore, detailed data on the total measurement error in these studies are missing. This is unsatisfactory and a systematic investigation of the accuracy of HFUPR measurements is therefore needed.

When carrying out a systematic investigation of the accuracy of HFUPR measurements, it is important to recognize that the accuracy of HFUPR measurement in living fetuses cannot be assessed directly but only indirectly. The measurement error is made when assessing the volume of the bladder and only here. This measurement error will also be incorporated in the HFUPR estimation, which is based on a sequence of bladder images during the filling phase. There are some other factors that influence the HFUPR measurement error, i.e. the magnitude of the HFUPR and the number and time points of bladder image capture (Fig. 1).

Fig. 1
Initially, the bladder was filled up to 19 ml before emptying and a new filling phase began. For this 32-week fetus, the HFUPR of 33 mL/h was estimated by regression function and extrapolation to a time span of 1 h

In the current paper we illustrate the mathematical derivation of a formula to assess the HFUPR measurement error using these factors and the implication of using different methods for bladder volume. The objective of the paper was to enable responses to two clinically important questions: (1) what is the probability that the HFUPR will be falsely classified at the 2.5th percentile or a lower value, even though the true HFUPR is at a higher percentile point? and (2) if we are to use the HFUPR in the evaluation of at-risk pregnancies by daily measurements, we need to know how much of the change can be explained exclusively by measurement error.

Methods and subjects

In spite of a constant volume, varying estimates will be found when assessing the volume of the fetal urinary bladder, due to measurement error [14]. This error was assumed to have a normal distribution and the standard deviation (SD) was therefore used as a measurement of estimation error (by us denoted SDVOLUME). This volume measurement error will be incorporated in the HFUPR estimation and the SD was consequently also utilized as the measurement of HFUPR estimation error (by us denoted SDHFUPR).

To obtain a generally applicable formula for estimating SDHFUPR, a mathematical derivation was applied (see Statistics section). Although it is an original derivation it can be scrutinized by each statistician without difficulty. It is pure mathematics. In clinical practice, there are two fundamental questions that need to be answered: (1) what is the probability that the HFUPR will be falsely classified at the 2.5th percentile or a lower value, even though the true HFUPR is at a higher percentile point? and (2) If we are to use the HFUPR in the evaluation of at-risk pregnancies by daily measurements, we need to know how much of the change can be explained exclusively by measurement error. These questions were answered by two supplementary formulas.

To illustrate the implication of the calculated SDHFUPR, relevant values of HFUPR, SDVOLUME number and time points of bladder image capture were required. The reference range of the HFUPR in an extensive 2D study (comprising 358 uncomplicated fetuses) was used and the 2.5th and 10th percentile point were calculated from a diagram in that study [15]. To date, detailed information on SDVOLUME in living fetuses has been exclusively presented in studies of 2D ultrasound. To compare the implication of using different bladder volume methods (the ellipsoid or the sum-of-cylinders method), the calculated SDVOLUME in two other 2D studies was applied [2, 4].

Statistics

The measurement error when estimating the HFUPR

Assuming that x1, x2,… are the time points for bladder volume estimations during the filling phase (small letters for constants) and Y1, Y2,… are the estimated bladder volumes (capital letters for random variables), the HFUPR can be calculated as the coefficient of regression:

equation M1
1

This derivation of the coefficient of regression can be found in a number of statistical manuals.

The SD of the bladder volume measurements was assumed to be a linear function of volume.

equation M2
2

The volume Yi at the time point xi can be approximated as

equation M3
3

and

equation M4
4

The different measurement errors, which are involved in each volume estimation during a filling phase, are regarded as statistically independent random variables.

When calculating SDHFUPR, the corresponding variances are used. The variance in the HFUPR according to Eq. 1 equals the variance in equation M5 The factors equation M6 and equation M7 can be regarded as constants in this application.

Only the factor (Yi) must be taken into consideration and the variance [SD(Yi)]2 is used.

However, Yi is the estimated bladder volume and

equation M8
5

according to Eqs. 4 and 5, equation M9

Referring to Eq. 1, the variance in the measurement error when estimating the HFUPR is: equation M10and according to Eqs. 4 and 5:

equation M11

and

equation M12
6

Using the valid values for the constants a and b, the mean time for bladder image documentation equation M13 and the time points when images were captured, the SDHFUPR turned out to be an approximately linear function of the HFUPR and can be expressed as:

equation M14
7

The values of the constants c and d depend on the SDVOLUME and time points for bladder image capture.

The statistical derivations relating to the two clinical examples

  1. What is the maximum change in the HFUPR that could be caused exclusively by measurement errors?
    Assuming that the actual HFUPR = x on two occasions, what is the probability that the difference between these two determinations (in spite of a constant HFUPR) is at least as extreme as a quantity Δ? Assuming that Φ is the distribution function of the standardized normal distribution, the probability of assessing an HFUPR, which is a magnitude of Δ less than 0, is:
    equation M15
    8
    However, in this example, equation M16 whereasequation M17 depends on the evaluation of a difference and SDHFUPR = c + d × x according to (7). The probability that the difference between two determinations is at least as extreme as a quantity Δ is: equation M18 For example, the probability attains a value of 5% if
    equation M19
    9
  2. What is the probability that the HFUPR will be falsely classified at the 2.5th percentile or a lower value, whereas the true HFUPR is at the 10th percentile point?
    When the true HFUPR is exactly at the 10th percentile point (x10), the measured HFUPR will have a normal distribution with the mean value (x10). In line with the discussion in the previous example, the difference is x2.5 − x10 and the standard deviation SD = SDHFUPR = c + d × x10. The probability of obtaining an HFUPR value less than the 2.5th percentile can be calculated as
    equation M20
    10

Results

The SD for the estimated HFUPR (SDHFUPR) depends on the method of bladder volume estimation (SDVOLUME), the magnitude of the HFUPR and the number and time points of bladder image capture. The derivation was presented in Formulas 6 and 7 in the Statistics section. The new constants c and d in Eq. 7 are crucial for the further derivations (Fig. 2; Table 1).

Fig. 2
This calculation of the HFUPR measurement error (SDHFUPR) was based on images captured at five, 10, 15, 20, 25 and 30 min after an emptying phase, when applying the ellipsoid and the sum-of-cylinders method (different SDVOLUME). The SDHFUPR ranged ...
Table 1
The influence of different numbers and times of image capture is illustrated for some situations, when calculating the SDHFUPR (SDHFUPR = c + d ×  HFUPR)

The probability of falsely reading the 10th percentile point as a pathological value (i.e. below the 2.5th percentile point) was excluded (0%) with the sum-of-cylinders method for gestational ages of ≥30 weeks (see Formula 10, Statistics section). With the ellipsoid method, the risk was higher (Table 2).

Table 2
Shows the probability of false HFUPR readings at the 2.5th percentile point or a lower level, even though the true HFUPR was at the 10th percentile point

Using Formula 9 (Statistics section), the maximum changes which might be exclusively explained by measurement error were four to five times greater (25–30%) with the ellipsoid method compared with the sum-of-cylinders method (6%) (Table 3).

Table 3
The magnitude of change in the estimated HFUPR from the initial 10th percentile point, which might be caused exclusively by measurement error, was calculated

Discussion

A starting point for the present paper is the linear relationship between bladder volume and the measurement error: SDvolume = a + b × volume. This relationship has been thoroughly documented in three previous studies [2, 3, 4]. When all the cases in these studies are included, the distribution of the SD and residuals supports a linear relationship, which is a prerequisite for using a linear regression function [13] (Figs. 3 and and44).

Fig. 3
The SD was calculated when estimating the bladder volume of 120 fetuses. Different methods were used, which gave rise to 222 relationships between SD and bladder volume. The maximum and minimum bladder volumes were 80.5 and 0.1 mL, respectively. ...
Fig. 4
The distribution of the residuals supports a linear relationship between the SD and bladder volume based on 222 relationships between the SD and bladder volume. The maximum and minimum of the residuals were 8.2 and −1.9 mL, respectively. ...

The main finding in the present paper is a general formula for calculating the SDHFUPR, which is valid when using 2D ultrasound. As we understand it, this measurement error can only be assessed indirectly by this kind of calculation. The measurement error is made when assessing the volume of the bladder—and only then. When the linear relationship between the bladder volume and measurement error is taken into consideration, there is a specific SDVOLUME related to each estimated volume. The magnitude of the constants a and b depends on the volume assessment method that is used, e.g. 2D ultrasound (ellipsoid and sum-of-cylinders method). The a and b for the ellipsoid method are 0.36516 and 0.09978, whereas they are 0.29911 and 0.02788 for the sum-of-cylinders method. For 3D ultrasound, the relationship between SDVOLUME and the estimated volume is as yet unknown. However, the current paper illustrates the usefulness of this relationship and the implication of using different 2D methods for volume estimation.

When utilizing the HFUPR for fetal surveillance, we need to know whether the estimated HUFPR is pathologically low, i.e. below the 2.5th percentile point. We therefore need to answer the question: What is the risk of false readings at the 2.5th percentile point, for example, even though the true HFUPR is at a higher percentile point? (see Formula 10 in Statistics).

Furthermore, we need to answer the question: How much of an observed HFUPR change (for example, during daily controls) can be explained exclusively by measurement error? (see Formula 9 in Statistics).

In some previous ultrasound studies, the quality of bladder volume as well as HFUPR measurements has been addressed [7, 11, 12, 17]. These studies lack detailed measurement error data and are not sufficient as a basis for HFUPR evaluation. With the assistance of our mathematical derivations of the SDHFUPR, however, the answers to the two questions are within reach. Only three data are needed to obtain the SDHFUPR: (1) the relationship between bladder volume and SDVOLUME for the estimation method that is used (determines the constants a and b), (2) the time points of image capture and (3) the estimated HFUPR. According to Statistics, there is also a linear relationship between SDHFUPR and the HFUPR. This relationship SDHFUPR = c + d × HFUPR can be utilised for further calculations of the constants c and d, which provides the answers to the aforementioned questions. Moreover, the implication of using different methods for bladder volume is illustrated.

Conclusions

When SDHFUPR, the magnitude of the HFUPR and the time points of bladder image capture are taken into account, there is a probability of falsely classifying the HFUPR at the 2.5th percentile or a lower value, even though the true HFUPR at the 10th percentile point may be estimated. Moreover, when evaluating at-risk pregnancies by daily HFUPR measurements, we can assess how much of the change can be explained exclusively by measurement error.

Electronic supplementary material

Acknowledgments

We would like to thank Ingemar Kjellmer, Professor at the Department of Paediatrics, Queen Silvia’s Children’s Hospital, Goteborg. Thank you for your support and enthusiasm for the subject. This study was supported by grants from the North Elvsborg Foundation for Medical Development and Research and the Goteborg Medical Society.

Conflict of interest statement None.

Open Access This article is distributed under the terms of the Creative Commons Attribution Noncommercial License which permits any noncommercial use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.

References

1. Campbell S, Wladimiroff JW, Dewhurst CJ. The antenatal measurement of fetal urine production. J Obstet Gynaecol Br Commonw. 1973;80:680–686. [PubMed]
2. Fagerquist M, Fagerquist U, Oden A, Blomberg SG. Fetal urine production and accuracy when estimating fetal urinary bladder volume. Ultrasound Obstet Gynecol. 2001;17:132–139. doi: 10.1046/j.1469-0705.2001.00294.x. [PubMed] [Cross Ref]
3. Fagerquist M, Fagerquist U, Steyskal H, Oden A, Blomberg SG. Accuracy in estimating fetal urinary bladder volume using a modified ultrasound technique. Ultrasound Obstet Gynecol. 2002;19:371–379. doi: 10.1046/j.1469-0705.2002.00629.x. [PubMed] [Cross Ref]
4. Fagerquist M, Fagerquist U, Oden A, Blomberg SG. Estimation of fetal urinary bladder volume using the sum-of-cylinders method vs. the ellipsoid formula. Ultrasound Obstet Gynecol. 2003;22:67–73. doi: 10.1002/uog.139. [PubMed] [Cross Ref]
5. Ferrazzi E, Bozzo M, Rigano S, Bellotti M, Morabito A, Pardi G, Battaglia FC, Galan HL. Temporal sequence of abnormal Doppler changes in the peripheral and central circulatory systems of the severely growth-restricted fetus. Ultrasound Obstet Gynecol. 2002;19:140–146. doi: 10.1046/j.0960-7692.2002.00627.x. [PubMed] [Cross Ref]
6. Hajek Z, Srp B, Haddad el R, Drbohlav P, Parizek A, Liska K, Paskova A, Zvarova J, Simeckova A. Analysis of present diagnostic methods of intrapartum fetal hypoxia. Ceska Gynekol. 2005;70:22–26. [PubMed]
7. Lee SM, Park SK, Shim SS, Jun JK, Park JS, Syn HC. Measurement of fetal urine production by three-dimensional ultrasonography in normal pregnancy. Ultrasound Obstet Gynecol. 2007;30:281–286. doi: 10.1002/uog.4038. [PubMed] [Cross Ref]
8. Mikovic Z, Mandic V, Djukic M, Egic A, Filimonovic D, Cerovic N, Popovac M. Longitudinal analysis of arterial Doppler parameters in growth retarded fetuses. Srp Arh Celok Lek. 2003;131:21–25. doi: 10.2298/SARH0302021M. [PubMed] [Cross Ref]
9. Miura H. Evaluation of fetal renal arterial blood flow waveforms with pulsed Doppler flowmetry and the correlation to estimated fetal body weight, fetal urine production rate and amniotic fluid volume. Nippon Sanka Fujinka Gakkai Zasshi. 1991;43:1647–1652. [PubMed]
10. Nicolaides KH, Peters MT, Vyas S, Rabinowitz R, Rosen DJ, Campbell S. Relation of rate of urine production to oxygen tension in small-for-gestational-age fetuses. Am J Obstet Gynecol. 1990;162:387–391. [PubMed]
11. Rabinowitz R, Peters MT, Vyas S, Campbell S, Nicolaides KH. Measurement of fetal urine production in normal pregnancy by real-time ultrasonography. Am J Obstet Gynecol. 1989;161:1264–1266. [PubMed]
12. Shin T, Koyanagi T, Hara K, Kubota S, Nakano H. Development of urine production and urination in the human fetus assessed by real-time ultrasound. Asia Oceania J Obstet Gynaecol. 1987;13:473–479. doi: 10.1111/j.1447-0756.1987.tb00294.x. [PubMed] [Cross Ref]
13. Skrepnek GH. Regression Methods in the Empiric Analysis of Health Care Data. J Manag Care Pharm. 2005;11(3):240–251. [PubMed]
14. Stigter RH, Mulder EJ, Bruinse HW, Visser GH. Doppler studies on the fetal renal artery in the severely growth-restricted fetus. Ultrasound Obstet Gynecol. 2001;18:141–145. doi: 10.1046/j.1469-0705.2001.00493.x. [PubMed] [Cross Ref]
15. Takeuchi H, Koyanagi T, Yoshizato T, Takashima T, Satoh S, Nakano H. Fetal urine production at different gestational ages: correlation to various compromised fetuses in utero. Early Hum Dev. 1994;40:1–11. doi: 10.1016/0378-3782(94)90094-9. [PubMed] [Cross Ref]
16. Thornton JG, Hornbuckle J, Vail A, Spiegelhalter DJ, Levene M. Infant wellbeing at 2 years of age in the Growth Restriction Intervention Trial (GRIT): multicentred randomised controlled trial. Lancet. 2004;364:513–520. doi: 10.1016/S0140-6736(04)16809-8. [PubMed] [Cross Ref]
17. Touboul C, Boulvain M, Picone O, Levaillant JM, Frydman R, Senat MV. Normal fetal urine production rate estimated with 3-dimensional ultrasonography using the rotational technique (virtual organ computer-aided analysis) Am J Obstet Gynecol. 2008;199(1):57. doi: 10.1016/j.ajog.2007.12.012. [PubMed] [Cross Ref]
18. Turan S, Turan OM, Berg C, Moyano D, Bhide A, Bower S, Thilaganathan B, Gembruch U, Nicolaides K, Harman C, Baschat AA. Computerized fetal heart rate analysis, Doppler ultrasound and biophysical profile score in the prediction of acid–base status of growth-restricted fetuses. Ultrasound Obstet Gynecol. 2007;30:750–756. doi: 10.1002/uog.4101. [PubMed] [Cross Ref]
19. Vyas S, Nicolaides KH, Campbell S. Renal artery flow-velocity waveforms in normal and hypoxemic fetuses. Am J Obstet Gynecol. 1989;161:168–172. [PubMed]

Articles from Springer Open Choice are provided here courtesy of Springer