F1000 Med Rep. 2009; 1: 1.
Published online 2009 January 21. doi:  10.3410/M1-1
PMCID: PMC2920698

Radiation exposure in cardiac multislice spiral computed tomography (MSCT)

Abstract

Computed tomographic (CT) imaging involves exposure to ionizing radiation, with a consequent potential risk to the patient. Here we review the technical factors that affect the radiation dose in cardiac CT diagnostic procedures, the parameters that are used to measure and compare radiation doses, the magnitude of radiation exposure, and how to efficiently reduce it.

Introduction and context

The use of computed tomographic (CT) diagnostic procedures in cardiology is steadily increasing and is expected to grow even more if the technique evolves to become a primary diagnostic modality for coronary artery disease. As CT involves the use of ionizing radiation in the form of X-rays, the issue of radiation exposure in coronary imaging by multislice spiral computed tomography (MSCT) has attracted considerable attention. The risks and benefits of each CT coronary angiogram (CTCA) should be carefully weighed, and the ALARA principle (as low as reasonably achievable) to reduce radiation dose to the patient should always be respected. It is essential that the ordering physicians and the technicians who perform CT examinations have adequate knowledge of the radiation exposure of the procedures and understand how to minimize the dosage while preserving the quality of the diagnostic image. Here we review the technical factors that affect the radiation dose, the parameters that are used to measure and compare radiation doses, the magnitude of radiation exposure and how to efficiently reduce it.

Radiation-dose terminology

Radiation exposure in CT is generally quantified by the tube-current-time product or the mAs-value. It does not, however, directly quantify the absorbed radiation dose. Absorbed dose is the amount of radiation energy per unit mass tissue deposited in a patient's body by X-ray photons. The most useful parameters for comparing radiation doses are the computed tomography dose index (CTDIvol), the dose-length product (DLP) and the effective dose (E) (Table 1).

Table 1.
Parameters of radiation dose

The CTDIvol [measured in units of grays (Gy)] represents the average radiation dose in the x, y, z directions and estimates the average radiation within the irradiated volume of one CT acquisition. The value of CTDIvol can be obtained by standardized phantom measurements with an ionization chamber. It compensates for the overlap between slices, by normalizing to the pitch (pitch is the table advancement per rotation divided by the collimated detector width) CTDIvol enables ready comparison between the radiation doses from different CT scanners. The DLP represents the integrated radiation dose over all slices from an entire CT examination. The DLP is calculated as DLP = CTDIvol x scan length.

The effective dose (E) reflects the risk of potential biological injury of radiation, which depends on the tissue that is exposed. Biological risks include deterministic effects (skin burns, and so on) and stochastic effects (carcinogenesis). A good estimation of effective dose can be obtained from E = k x DLP, where k is a conversion coefficient which depends on the body region that is scanned; for the chest k is 0.017 mSv x mGy-1 x cm-1. The effective dose allows for a crude comparison between different CT scanners and scan protocols, but is only a rough estimate of the true risk. The latter should be obtained from patient-specific organ doses and risk factors. Here, E is calculated as the weighted mean of relevant organ values according to the International Commission of Radiological Protection [1]. In cardiac CT the organs receiving the highest doses are the female breasts, lungs, liver and esophagus.

Cardiac CT examination and dose reduction

Radiation exposure varies with types of CT scanner and scan protocols for cardiac examinations. Variables that play a role in radiation exposure and measures to minimize the radiation exposure are listed in Table 2. Following the ALARA principle, all measures should be taken to reduce radiation exposure to a minimum while preserving diagnostic-quality (rather than esthetic) images. Obviously, the number of investigations is important. The justification for referring a patient for a CT scan should include a careful balancing of the expected benefits of information provided and the (long-term) risk.

Table 2.
Variables associated with radiation exposure

A CT scan should not be performed if the same diagnostic information can be obtained by an alternative modality that does not use ionizing radiation. Of particular importance, always avoid unintentional repeat examinations.

Image noise increases and image quality decreases if the number of photons that is used for image reconstruction decreases. The number of X-ray photons generated by the X-ray tube per reconstructed slice depends on tube current (mA), tube voltage (kV) and pitch, which can be adjusted to achieve diagnostic-image quality with the lowest settings. Improved spatial resolution with newer-generation 64-slice scanners requires an increase in tube current. Lower mA settings can be used in slim patients. Radiation dose varies with the square of the voltage (in kilovolts), and image contrast improves by decreasing tube voltage in patients with low body mass index (BMI). This approach has resulted in dose reductions of up to 88% in cardiac CT examinations [2]. However, low-voltage settings are less preferable in CT coronary angiography as image noise also increases, resulting in degradation of the coronary images. The pitch (which determines the degree of slice overlap) should be selected to obtain optimal image quality and should be adjusted to heart rate, if possible. The scan length should be individualized to the size of the heart. Most cardiac CT scans are performed using retrospectively gated helical (RGH) data acquisition. With RGH, the patient moves through the gantry at a constant speed, and data are acquired and available throughout the cardiac cycle. The tube current can be modulated during scanning by selecting a pulsing window according to the electrocardiogram (for example, during end-diastole) during which the patient is exposed to maximum radiation, while tube current is lowered during the remaining cardiac phases. This can reduce overall effective dose by 30-50% [2,3] and is most effective in the case of low heart rates.

Figure 1.
Comparison of different CT scan protocols
Figure 2.
Example of retrospectively gated dual-source CT coronary angiography

Recent advances

Another technique recently introduced for CT coronary angiography is prospectively gated axial (PGA) data acquisition. This technique uses a step-and-shoot mode: the table is stationary during data acquisition and then moves to the next position for the second scan. X-ray generation is switched off between two consecutive scans, resulting in an overall low dose to the patient (1.1-3.0 mSv) [4]. The downside of this technique is that it is only applicable in patients with low heart rates (pre-scan beta-blockers are required for patients with higher heart rates) and regular, very stable heart rhythms. Estimates of the effective dose for conventional coronary angiography, cardiac nuclear imaging, and various cardiac CT scan protocols are listed in Table 3.

Table 3.
Estimates of effective dose (E)

Radiation exposure and risk estimation

Cardiac CT examination is associated with a (low) lifetime-attributable risk (LAR) of incidence of and mortality from cancer. The LAR estimates of cancer have been derived from studies of atomic-bomb survivors (those who received low doses of radiation ranging from 5 to 150 mSv, mean dose 40 mSv) [5] and a large-scale study of 400,000 radiation workers in the nuclear industry (receiving a dose of approximately 20 mSv) [6]. These studies revealed a significant association between radiation dose and mortality from cancer. Einstein et al. [7] showed that 64-slice cardiac CTCA is associated with a non-negligible LAR of cancer and, most importantly, that the LAR estimates of cancer incidence attached to CT examinations are highly dependent on the age and gender of the patient (Table 4), with younger patients and females having a higher LAR of cancer incidence. The risk can, however, be significantly reduced using dose-reduction techniques.

Table 4.
Lifetime-attributable risk of cancer following 64-slice CT coronary angiography

Implications for clinical practice

Cardiac CT examinations are, inevitably, associated with a certain radiation-induced risk of cancer, which is age and gender dependent. The risk is higher in younger patients and women. The risk of cancer must be decreased by minimizing radiation exposure to as low as possible while maintaining diagnostic image quality. The ordering physician should be aware of this cancer risk and should carefully consider the potential diagnostic benefits of the investigation, which should outweigh the associated risk of cancer. The technician performing the CT scan should have a thorough knowledge of the CT scanner specifications and should select a scan protocol that generates images with a minimal radiation dose and the required image quality. But most importantly, at all times avoid cardiac CT examinations that are not justified by medical need [8].

Acknowledgments

The authors would like to thank Marcel van Straten, Department of Radiology, Erasmus Medical Center, for stimulating discussions and contributions to this review.

Abbreviation

ALARA
as low as reasonably achievable
BMI
body mass index
CT
computed tomographic
CTCA
computed tomographic coronary angiogram
CTDIvol
computed tomography dose index
DLP
dose-length product
E
effective dose
LAR
lifetime-attributable risk
MSCT
multislice spiral computed tomography
PGA
prospectively gated axial
RGH
respectively gated helical

Notes

The electronic version of this article is the complete one and can be found at: http://F1000.com/Reports/Medicine/content/1/1

Notes

Competing interests

The authors declare that they have no competing interests.

References

1. 1990 Recommendations of the International Commission on Radiological Protection (IRCP Publication 60) Ann ICRP. 1991;21:1305–10.
2. Abada HT, Larchez C, Daoud B, Sigal-Cinqualbre A, Paul JF. MDCT of the coronary arteries: feasibility of low-dose CT with ECG-pulsed tube current modulation to reduce radiation dose. Am J Roentgenol. 2006;186(2):S387–90. doi: 10.2214/AJR.05.0216. [PubMed] [Cross Ref]
3. Jakobs TF, Becker CR, Ohnesorge B, Flohr T, Suess C, Schoepf UJ, Reiser MF. Multislice helical CT of the heart with retrospective ECG gating: reduction of radiation exposure by ECG-controlled tube current modulation. Eur Radiol. 2002;12:1081–6. doi: 10.1007/s00330-001-1278-x. [PubMed] [Cross Ref]
4. Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA, Veit-Haibach P, Tatsugami F, von Schulthess GK, Kaufmann PA. Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J. 2008;29:191–7. doi: 10.1093/eurheartj/ehm613. [PubMed] [Cross Ref] F1000 Factor 6.0 Must Read
Evaluated by Pim de Feyter 7 Jul 2008
5. Board on Radiation Effects Research Health Risks from Exposure to Low Levels of Ionizing Radiation: BEIR VII. Washington, DC: National Academies Press; 2005.
6. Cardis E, Vrijheid M, Blettner M, Gilbert E, Hakama M, Hill C, Howe G, Kaldor J, Muirhead CR, Schubauer-Berigan M, Yoshimura T, Bermann F, Cowper G, Fix J, Hacker C, Heinmiller B, Marshall M, Thierry-Chef I, Utterback D, Ahn YO, Amoros E, Ashmore P, Auvinen A, Bae JM, Bernar J, Biau A, Combalot E, Deboodt P, Diez Sacristan A, Eklöf M, et al. The 15-country collaborative study of cancer risk among radiation workers in the nuclear industry: estimates of radiation-related cancer risks. Radiat Res. 2007;167:396–416. doi: 10.1667/RR0553.1. [PubMed] [Cross Ref]
7. Einstein AJ, Henzlova MJ, Rajagopalan S. Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA. 2007;298:317–23. doi: 10.1001/jama.298.3.317. [PubMed] [Cross Ref]Changes Clinical Practice
F1000 Factor 6.4 Must Read
Evaluated by Eugenio Picano 17 Aug 2007, Paul Stein 24 Dec 2007
8. Brenner DJ, Hall EJ. Computed tomography - an increasing source of radiation exposure. N Engl J Med. 2007;357:2277–84. doi: 10.1056/NEJMra072149. [PubMed] [Cross Ref]
9. Einstein AJ, Moser KW, Thompson RC, Cerqueira MD, Henzlova MJ. Radiation dose from cardiac imaging. Circulation. 2007;116:1290–305. doi: 10.1161/CIRCULATIONAHA.107.688101. [PubMed] [Cross Ref]
10. Hunold P, Vogt FM, Schmermund A, Debatin JF, Kerkhoff G, Budde T, Erbel R, Ewen K, Barkhausen J. Radiation exposure during cardiac CT: effective dose at multi-detector row CT and electron-beam CT. Radiology. 2003;226:145–52. doi: 10.1148/radiol.2261011365. [PubMed] [Cross Ref]
11. Flohr TG, Schoepf UJ, Kuettner A, Halliburton S, Bruder H, Suess C, Schmidt B, Hofmann L, Yucel EK, Schaller S, Ohnesorge BM. Advances in cardiac imaging with 16-section CT systems. Acad Radiol. 2003;10:386–401. doi: 10.1016/S1076-6332(03)80027-2. [PubMed] [Cross Ref]
12. Morin RL, Gerber TC, McCollough CH. Radiation dose in computed tomography of the heart. Circulation. 2003;107:917–22. doi: 10.1161/01.CIR.0000048965.56529.C2. [PubMed] [Cross Ref]
13. Trabold T, Buchgeister M, Küttner A, Heuschmid M, Kopp AF, Schröder S, Claussen CD. Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo. 2003;175:1051–55. [PubMed]
14. Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S, Kastrati A, Schömig A. Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation. 2006;113:1305–10. doi: 10.1161/CIRCULATIONAHA.105.602490. [PubMed] [Cross Ref]
15. Mollet NR, Cademartiri F, van Mieghem CA, Runza G, McFadden EP, Baks T, Serruys PW, Krestin GP, de Feyter PJ. High-resolution spiral computed tomography coronary angiography in patients referred for diagnostic conventional coronary angiography. Circulation. 2005;112:2318–23. doi: 10.1161/CIRCULATIONAHA.105.533471. [PubMed] [Cross Ref] F1000 Factor 3.0 Recommended
Evaluated by Stephan Achenbach 26 Jan 2006
16. Weustink AC, Meijboom WB, Mollet NR, Otsuka M, Pugliese F, van Mieghem C, Malago R, van Pelt N, Dijkshoorn ML, Cademartiri F, Krestin GP, de Feyter PJ. Reliable high-speed coronary computed tomography in symptomatic patients. J Am Coll Cardiol. 2007;50:786–94. doi: 10.1016/j.jacc.2007.04.068. [PubMed] [Cross Ref]
17. Ropers U, Karakaya S, Wechsel M, Anders K, Ropers D, Kuettner A, Daniel WG, Kalender W, Bautz W, Achenbach S. Randomized comparison of dual source computed tomography and 64-slice multi-detector computed tomography for the detection of coronary artery stenoses. [Abstract] Circulation. 2006;114:448.
18. Earls JP, Berman EL, Urban BA, Curry CA, Lane JL, Jennings RS, McCulloch CC, Hsieh J, Londt JH. Prospectively gated transverse coronary CT angiography versus retrospectively gated helical technique. Radiology. 2008;246:742–53. doi: 10.1148/radiol.2463070989. [PubMed] [Cross Ref]Changes Clinical Practice
F1000 Factor 6.0 Must Read
Evaluated by Stephan Achenbach 26 Mar 2008
19. Rybicki FJ, Otero HJ, Steigner ML, Vorobiof G, Nallamshetty L, Mitsouras D, Ersoy H, Mather RT, Judy PF, Cai T, Coyner K, Schultz K, Whitmore AG, Di Carli MF. Initial evaluation of coronary images from 320-detector row computed tomography. Int J Cardiovasc Imaging. 2008;24:535–46. doi: 10.1007/s10554-008-9308-2. [PubMed] [Cross Ref]

Articles from F1000 Medicine Reports are provided here courtesy of Faculty of 1000 Ltd