Noninvasive monitoring of PaCO2 during one-lung ventilation and minimal access surgery in adults: End-tidal versus transcutaneous techniques

Abstract

Background:

Previous studies have suggested that end-tidal CO2 (ET-CO2) may be inaccurate during one-lung ventilation (OLV). This study was performed to compare the accuracy of the noninvasive monitoring of PCO2 using transcutaneous CO2 (TC-CO2) with ET-CO2 in patients undergoing video-assisted thoracoscopic surgery (VATS) during OLV.

Materials and Methods:

In adult patients undergoing thoracoscopic surgical procedures, PCO2 was simultaneously measured with TC-CO2 and ET-CO2 devices and compared with PaCO2.

Results:

The cohort for the study included 15 patients ranging in age from 19 to 71 years and in weight from 76 to 126 kg. During TLV, the difference between the TC-CO2 and the PaCO2 was 3.0 ± 1.8 mmHg and the difference between the ET-CO2 and PaCO2 was 6.2 ± 4.7 mmHg (P=0.02). Linear regression analysis of TC-CO2 vs. PaCO2 resulted in an r2 = 0.6280 and a slope = 0.7650 ± 0.1428, while linear regression analysis of ET-CO2 vs. PaCO2 resulted in an r2 = 0.05528 and a slope = 0.1986 ± 0.1883. During OLV, the difference between the TC-CO2 and PaCO2 was 3.5 ± 1.7 mmHg and the ET-CO2 to PaCO2 difference was 9.6 ± 3.6 mmHg (P=0.03 vs. ET-CO2 to PaCO2 difference during TLV; and P<0.0001 vs. TC-CO2 to PaCO2 difference during OLV). In 13 of the 15 patients, the TC-CO2 value was closer to the actual PaCO2 than the ET-CO2 value (P =0.0001). Linear regression analysis of TC-CO2 vs. PaCO2 resulted in an r2 = 0.7827 and a slope = 0.8142 ± 0.0.07965, while linear regression analysis of ET-CO2 vs. PaCO2 resulted in an r2 = 0.2989 and a slope = 0.3026 ± 0.08605.

Conclusions:

During OLV, TC-CO2 monitoring provides a better estimate of PaCO2 than ET-CO2 in patients undergoing VATS.

Keywords: End-tidal CO2, thoracoscopy, transcutaneous CO2

The measurement of the partial pressure of carbon dioxide in the arterial blood (PaCO2) is used to determine the adequacy of ventilation and to guide changes in mechanical ventilation. Although arterial sampling provides this information, frequent arterial blood gas (ABG) analysis is expensive, time restrictive and provides only intermittent data for what may be a rapidly changing value. To overcome such problems, noninvasive monitors are frequently used to provide a continuous estimate of the PaCO2. In the operating room, end-tidal (ET) CO2 monitoring remains the standard of care for continuous and noninvasive PaCO2 monitoring. However, various factors, including patient positioning, sampling errors and ventilation-perfusion mismatch due to dead space or shunt, may significantly affect the arterial to ET-CO2 gradient.[13]

One-lung ventilation (OLV) is commonly used during thoracic surgery to allow surgical access to the operative lung and to eliminate lung movement during ventilation. The technique is helpful during open thoracic procedures and a necessity for minimal access thoracoscopic procedures. During OLV, the bronchus of the affected lung is occluded by a bronchial blocker or isolated by a double-lumen endotracheal tube, with oxygenation and ventilation supported by the nonoperative lung. Even with effective hypoxic pulmonary vasoconstriction, the technique results in an increase in the shunt fraction and perfusion of the nonventilated lung. Previous studies have demonstrated a significant arterial to ET-CO2 gradient during one-lung OLV.[46] The efficacy of ET-CO2 monitoring is limited during such procedures, suggesting that alternative noninvasive monitors of PaCO2 may be needed. Although used predominantly in the neonatal and pediatric population, there is increasing interest in the use of and recent reports of transcutaneous (TC) CO2 monitoring in the adult population.[59] The current study prospectively compares ET-CO2 and TC-CO2 monitoring during OLV in adults undergoing video-assisted thoracoscopic surgery.

MATERIALS AND METHODS

The study was approval by the Institutional Review Board and the Committee for the Protection of Human Subjects of the University of Missouri. Verbal informed consent was obtained from each patient. The patient population included patients scheduled for minimal access thoracic surgery and OLV, who were 18 years of age or older and in whom intraarterial access was deemed necessary. OLV was provided by either a bronchial blocker or a double lumen endotracheal tube. Effective lung separation was confirmed by clinical auscultation and fiberoptic bronchoscopy prior to the start of the procedure. Intrathoracic CO2 insufflation was not used during the procedure.

Transcutaneous and end-tidal carbon dioxide monitoring

ET-CO2 was measured using an infrared analyzer with a side stream sampler attached at the elbow between the endotracheal tube and the anesthesia circuit. Prior to use, the ET-CO2 device was calibrated according to the manufacturer's recommendations. TC-CO2 was measured with a commercially available TC-CO2 device (Sentec AG, Therwil, Switzerland). This device is based on a Stow-Severinghaus-type CO2 sensor combined with a pulse oximeter and is attached to the patient's earlobe by a low pressure attachment clip.[10,11] The in vitro 90% response time is <50s for the CO2 electrode. Prior to placement, the sensor was prepared and calibrated according to the manufacturer's recommendations. The sensor is calibrated in vitro by using a one-point dry gas calibration with 7% carbon dioxide. The working temperature of the sensor is 42°C. The sensor was cleaned with alcohol and dried before application. One drop of contact gel was applied to the center of the sensor prior to placement on the patient's earlobe. Oxygen saturation values are available immediately, while TC-CO2 values are available after a 2-3 min calibration time.

Data collection and statistical analysis

ABG analyses were obtained following endotracheal intubation during two-lung ventilation (TLV) and as clinically indicated during OLV. When an ABG was obtained, the ET-CO2 and TC-CO2 were simultaneously recorded on a data sheet. Calculation of the absolute difference between the noninvasive monitor (ET-CO2 or TC-CO2) and the PaCO2 was performed. Negative numbers were not used because this could artificially lower the mathematical mean of the differences between the noninvasive monitors of CO2 and the PaCO2. If multiple ABGs were obtained during OLV, the absolute differences between the PaCO2 and the noninvasive monitors were averaged and counted as a single data point. This was done to avoid biasing the data by overrepresentation of any one patient as the number of ABGs varied for each patient. The absolute difference between the ET-CO2 and PaCO2 was compared to the absolute difference between the TC-CO2 and the PaCO2 during TLV and OLV using a non-paired t-test. A contingency table with a Fisher's exact test was used to compare the times that each of the noninvasive monitors was closest to the actual PaCO2. Using the raw numbers from all of the individual sample sets (PaCO2/ET-CO2/TC-CO2) obtained during TLV and OLV, linear regression analysis and Bland-Altman analyses were performed.

RESULTS

The cohort for the study included 15 patients ranging in age from 19 to 71 years (46.9 ± 17.3 years) and in weight from 76 to 126 kg (93.4 ± 12.4 kg). There were 11 men and 4 women. Nineteen sample sets (PaCO2/ET-CO2/TC-CO2) were obtained during TLV and 27 were obtained during OLV. During TLV, the difference between the TC-CO2 and the PaCO2 was 3.0 ± 1.8 mmHg with a range of 0 to 8 mmHg and the difference between the ET-CO2 and the PaCO2 was 6.2 ± 4.7 mmHg with a range of 2 to 18 mmHg (P=0.02 vs. TC-CO2 to PaCO2 difference) [Table 1]. During TLV, the TC-CO2 value was closer to the actual PaCO2 in 10 patients, the ET-CO2 value was closer in 2 patients and there was no difference in 3 patients (P=NS). During TLV, linear regression analysis of TC-CO2 vs. PaCO2 resulted in an r2 = 0.6280 and a slope = 0.7650 ± 0.1428 (95% confidence intervals: 0.4367 to 1.066). During TLV, linear regression analysis of ET-CO2 vs. PaCO2 resulted in an r2 = 0.05528 and a slope = 0.1986 ± 0.1883 (95% confidence intervals: -0.1956 to 0.5928) [Figure 1]. During TLV, Bland-Altman analysis of TC vs. PaCO2 revealed a bias of +2.2 mmHg and a precision of ± 3.0 mmHg and analysis of ET vs. PaCO2 revealed a bias of -7.3 mmHg and a precision of ± 5.5 mmHg.

Table 1
End-tidal and transcutaneous differences vs. PaCO2 during one-lung ventilation and two-lung ventilation
Figure 1
Linear regression analysis during two-lung ventilation of TCCO2 versus PaCO2 (above) and ET-CO2 vs. PaCO2 (below). Analysis of TC-CO2 vs. PaCO2 revealed r2 = 0.6280 and slope = 0.7650 ± 0.1428 (95% confidence intervals: 0.4367 to 1.066). Analysis ...

During OLV, the difference between the TC-CO2 and the PaCO2 was 3.5 ± 1.7 mmHg with a range of 2 to 8 mmHg (P = NS vs. TC-CO2 to PaCO2 difference during TLV). During OLV, the ET-CO2 to PaCO2 difference was 9.6 ± 3.6 mmHg with a range of 2 to 15 mmHg (P = 0.03 vs. ET-CO2 to PaCO2 difference during TLV; and P< 0.0001 vs. TC-CO2 to PaCO2 difference during OLV) [Table 1]. In 13 of the 15 patients, the TC-CO2 value was closer to the actual PaCO2 than the ET-CO2 value (P= 0.0001). During OLV, linear regression analysis of TC-CO2 vs. PaCO2 resulted in an r2 = 0.7827 and a slope = 0.8142 ± 0.07965 (95% confidence intervals: 0.6513 to 0.9771). During OLV, linear regression analysis of ET-CO2 vs. PaCO2 resulted in an r2 = 0.2989 and a slope = 0.3026 ± 0.08605 (95% confidence intervals: 0.1266 to 0.4786) [Figure 2]. During OLV, Bland-Altman analysis of TC vs. PaCO2 revealed a bias of + 2.9 mmHg and a precision of ± 2.5 mmHg and analysis of ET vs. PaCO2 revealed a bias of −10.5 mmHg and a precision of ± 6.2 mmHg.

Figure 2
Linear regression analysis during one-lung ventilation of TC-CO2 versus PaCO2 (above) and ET-CO2 vs. PaCO2 (below). Analysis of TC-CO2 vs. PaCO2 revealed r2 = 0.7827 and slope = 0.8142 ± 0.0.07965 (95% confidence intervals: 0.6513 to 0.9771). ...

No problems occurred with ET-CO2 monitoring during the study period. In two patients, the TC-CO2 electrode had to be repositioned after the initial placement to obtain an effective value. No blistering, erythema or skin changes were noted on the earlobe following use of the TC-CO2 device.

DISCUSSION

The current study demonstrates that noninvasive monitoring of PaCO2 using ET-CO2 devices may be inaccurate during minimal access thoracoscopic surgery and OLV in adults. In our cohort of 15 patients, TC-CO2 monitoring was more accurate than ET-CO2 monitoring (difference of 3.0 ± 1.8 mmHg vs. 6.2 ± 4.7 mmHg) during the baseline state using TLV. Although no significant change was noted in the TC-CO2 to PaCO2 gradient during OLV (3.5 ± 1.7 mmHg), the ET-CO2 to PaCO2 difference increased to 9.6 ± 3.6 mmHg. In 13 of 15 patients, the TC-CO2 was closer to the actual PaCO2 than the ET-CO2 during OLV.

Several factors may be responsible for discrepancies between ET-CO2 and PaCO2, including technical issues with the monitor; and patient-related factors, including ventilation-perfusion mismatch, dead space and true shunt.[12,13] Whitesell et al demonstrated that patients with underlying lung disease had a significantly greater ET-CO2 to PaCO2 gradient when compared with patients with normal baseline pulmonary function (3.3 ± 0.6 mmHg versus 0.8 ± 0.3 mmHg).[13] Patient positioning has also been shown to have an impact on the accuracy of ET-CO2 monitoring.[2,3] With patients undergoing renal or upper ureteral surgery in the supine position, Pansard et al reported that the ET-CO2 to PaCO2 difference was 4.8 ± 3.9 mmHg 10 min after induction and increased to 7.9 ± 3.5 mmHg (P<0.01) 5 min after placement of the patients into the lateral decubitus ‘kidney rest’ position. Similar results were reported by Grenier et al in a cohort of patients undergoing neurosurgical procedures in the lateral decubitus position.[3]

In addition to these factors, the significant physiologic alterations induced by OLV can be expected to alter the relationship between ET-CO2 and PaCO2 values. Ip Yam et al evaluated the accuracy of ET-CO2 during OLV in a cohort of 22 adults undergoing thoracotomy.[4] During TLV, the ET-CO2 to PaCO2 difference was 1.3 ± 0.6 kPa (1 kPa = 7.5 mmHg) and it was 1.2 ± 0.7 kPa during OLV. Even if the difference for subsequent ABG analysis was corrected by subtracting the gradient from the first ABG analysis, the ET-CO2 to PaCO2 difference varied from -1.3 to 1.7 kPa. The authors concluded that the efficacy of ET-CO2 monitoring during OLV even when using corrected values remains questionable.

Two previous studies have evaluated noninvasive PCO2 monitoring during OLV using both TC-CO2 and ET-CO2 devices in patients undergoing open thoracotomy,[5,6] while there are no previous reports of using such monitoring in patients undergoing minimal access surgery. Oshibuchi et al compared the accuracy of TC-CO2 and ET-CO2 monitoring in a cohort of 26 adult patients undergoing OLV for open thoracotomy and pneumonectomy. The transcutaneous device (TCM3 transcutaneous CO2/oxygen device, Radiometer, Copenhagen, Denmark) was applied to the upper part of the patient's dependent arm. Evaluation of the TC-CO2 to PaCO2 difference revealed a bias of -0.4 mmHg and a precision of ± 2.5 mmHg during TLV and a bias of 1.4 mmHg and a precision of ± 4.3 mmHg during OLV. Evaluation of the ET-CO2 to PaCO2 difference revealed a bias of -5.8 mmHg and a precision of ± 4.1 mmHg during TLV and a bias of -7.1 mmHg and a precision of ± 4.6 mmHg during OLV. The authors concluded that TC-CO2 monitoring was an accurate means of evaluating PaCO2 during OLV. Tobias et al used the same transcutaneous device in their study of 15 young adult and pediatric patients (14.1 ± 6.1 years, range - 5 to 28 years) undergoing open thoracotomy.[5] During TLV, the TC-CO2 to PaCO2 difference was 2.5 ± 0.8 mmHg, while the ET-CO2 to PaCO2 difference was 3.9 ± 1.6 mmHg (P = 0.0049). There was a significant increase in the ET-CO2 to PaCO2 gradient during OLV (5.8 ± 2.3 mmHg), while no change was noted in the TC-CO2 to PaCO2 difference (2.7 ± 1.4 mmHg).

The previously reviewed studies of Oshibuchi et al and Tobias, along with the data from the current cohort of adult patients, demonstrate the inaccuracy of ET-CO2 monitoring during OLV and suggest that TC-CO2 monitoring is an effective alternative or adjunct. Until recently, transcutaneous CO2 monitoring was used most commonly in the neonatal and occasionally in the pediatric ICU population; however, there is growing experience with its use in adult patients in both the operating room and the ICU setting.[79,1416] These studies have demonstrated that TC-CO2 is more accurate than ET-CO2 in situations where the continuous monitoring of PaCO2 is vital.

TC-CO2 monitoring may be of particular benefit when the ventilation-perfusion properties of the respiratory system are altered. ET-CO2 measures a sample of gas that contains a mixture of gas exhaled from several areas of the airway and alveoli. Regions with a high ventilation-perfusion ratio (dead space) do not participate in gas exchange and therefore the partial pressure of CO2 is low or absent. During exhalation, the gas from regions of dead space mixes with the gas from areas of normal ventilation-perfusion ratios, resulting in dilution of the ET-CO2 sample and a widening of the ET-CO2 to PaCO2 difference. Alternatively, areas of low ventilation-perfusion ratios (shunt) result in ineffective gas exchange and the addition of blood with a high partial pressure of CO2 to the arterial circulation contributing to the increased ET-CO2 to PaCO2 gradient.[1718]

TC-CO2 monitoring avoids the effect of sampling gas that may be subject to ventilation-perfusion mismatch. Transcutaneous monitoring relies on cutaneous respiration of the diffusion of gases across the skin. The transcutaneous monitor measures the CO2 that is produced by local tissue metabolism and the CO2 released from the blood as it flows through the capillaries near the skin surface. The latter is in direct equilibrium with the capillary CO2 which is in equilibrium with the arterial CO2. Warming of the skin to 42°C by the sensor increases blood flow and CO2 solubility, resulting in an even greater diffusion of CO2 into the skin and equilibration with capillary and arterial PCO2 values.[19,20] The TC-CO2 monitor measures the PCO2 at the epidermis by using an infrared sensor, pH electrode or a Clark-type electrode.[21] Unlike ET-CO2, which typically underestimates actual CO2, the transcutaneous method typically overestimates actual CO2 by 5.2-6.4 mmHg due to the increased CO2 production from local metabolism induced by heating to 42°C. The currently available TC-CO2 devices have an internal correction / calibration factor to correct for the heat-induced changes in CO2 production. TC-CO2 monitoring requires specific training in calibration, preparation, placement and maintenance of the device. Errors in any one of these steps may give false readings. When compared with ET-CO2 monitoring, currently available TC-CO2 monitors require a longer preparation time, including a 5-min calibration period and then an additional 5-10 min equilibration time after placement on the patient. Although not an issue with the TC-CO2 monitor used in the current study, other TC-CO2 monitors may require heating the skin to 44-45°C to ensure accuracy. When this is done, there are occasional reports of superficial burns and skin blistering. Technical and patient-related factors may affect the accuracy of TC-CO2 monitoring. Improper membrane placement on the sensor or damage to the membrane may affect its accuracy. Patient factors, including skin thickness, skin edema and hypoperfusion (decreased cardiac output, hypovolemia or vasoconstriction), may also alter the diffusion of CO2 to the sensor and result in inaccurate readings.[2224] As no continuous noninvasive monitor can be expected to be 100% accurate, periodic calibration with an arterial sample may be indicated.

CONCLUSION

The current study adds to the growing body of knowledge demonstrating the efficacy of TC-CO2 monitoring in the adult population. This study is the first to evaluate the use of TC-CO2 vs. ET-CO2 during OLV in minimal access surgery. The continuous monitoring of PaCO2 may be particularly important during minimal access surgery. In addition to OLV, CO2 insufflation, to facilitate surgical visualization, places patients at a higher risk of hypercarbia than those undergoing open procedures. The combination of increased CO2 from systemic absorption of the insufflated CO2 with alterations in tidal volume imposed by decreasing minute ventilation during OLV makes accurately and continuously monitoring PaCO2 vital during minimal access surgery. TC-CO2 monitoring is not meant to replace ET-CO2 monitoring. Rather, the devices should be used to complement one another, especially in the OR setting. Although our data further demonstrate that TC-CO2 monitoring is more accurate than ET-CO2 monitoring, ET-CO2 remains the standard of care in the OR to document the intratracheal position of the endotracheal tube, to serve as an additional ventilator disconnect monitor and to provide a capnograph for waveform analysis.

Footnotes

Source of Support: Nil

Conflict of Interest: None declared.

REFERENCES

1. Bhavani-Shankar K, Moseley H, Kumar AY, Delph Y. Capnometry and anesthesia. Can J Anesth. 1992;39:617–32. [PubMed]
2. Pansard JL, Chjolley B, Devilliers C, Clergue F, Viars P. Variation in arterial to end-tidal CO2 tension differences during anesthesia in the “kidney rest” lateral decubitus position. Anesth Analg. 1992;75:506–10. [PubMed]
3. Grenier B, Verchere E, Meslie A, Dubreuil M, Siao D, Vandendriessche M, et al. Capnography monitoring during neurosurgery: Reliability in relation to various intraoperative positions. Anesth Analg. 1999;88:43–8. [PubMed]
4. Ip Yam PC, Innes PA, Jackson M, Snowdon SL, Russell GN. Variation in the arterial to end-tidal PCO2 difference during one-lung thoracic anesthesia. Br J Anaesth. 1994;72:21–4. [PubMed]
5. Tobias JD. Non-invasive carbon dioxide monitoring during one-lung ventilation: End-tidal versus transcutaneous techniques. J Cardiothor Vasc Anesth. 2003;17:306–8. [PubMed]
6. Oshibuchi O, Cho S, Hara T, Tomiyasu S, Makita T, Sumikawa K. A comparative evaluation of transcutaneous and dnd-tidal measurements of CO2 in thoracic anesthesia. Anesth Analg. 2003;97:776–9. [PubMed]
7. Bhavani-Shankar K, Steinbrook RA, Mushlin PS, Freiberger D. Transcutaneous PCO2 monitoring during laparoscopic cholecystectomy in pregnancy. Can J Anaesth. 1998;45:164–9. [PubMed]
8. Griffin J, Terry BE, Burton RK, Ray TL, Keller BP, Landrum AL, et al. Comparison of end-tidal and transcutaneous measures of carbon dioxide during general anesthesia in severely obese adults. Br J Anaesth. 2003;91:498–501. [PubMed]
9. McBride DS, Jr, Johnson JO, Tobias JD. Noninvasive carbon dioxide monitoring during neurosurgical procedures in adults: End-tidal versus transcutaneous techniques. South Med J. 2002;95:870–4. [PubMed]
10. Eberhard P, Gisiger PA, Gardaz JP, Spahn DR. Combining transcutaneous blood gas measurement and pulse oximetry. Anesth Analg. 2002;94:S76–80. [PubMed]
11. Rohling R, Biro P. Clinical investigation of a new combined pulse oximetry and carbon dioxide tension sensor in adult anesthesia. J Clin Monit. 1999;15:23–7. [PubMed]
12. Nunn JF, Hill DW. Respiratory dead space and arterial to end tidal CO2 tension difference in anesthetized man. J Appl Physiol. 1960;15:383–9. [PubMed]
13. Whitesell R, Asiddao C, Gollman D, Jablonski J. Relationship between arterial and peak expired carbon dioxide pressure during anesthesia and factors influencing the difference. Anesth Analg. 1981;60:508–12. [PubMed]
14. Reid CW, Martineau RJ, Miller DR, Hull KA, Baines J, Sullivan PJ. A comparison of transcutaneous, end-tidal and arterial measurements of carbon dioxide during general anesthesia. Can J Anaesth. 1992;39:31–6. [PubMed]
15. Tschupp A, Fanconi S. A combined ear sensor for pulse oximetry and carbon dioxide tension monitoring: Accuracy in critically ill children. Anesth Analg. 2003;96:82–4. [PubMed]
16. Bendjelid K, Schutz N, Stotz M, Gerard I, Suter PM, Romand JA. Transcutaneous PCO2 monitoring in critically ill adults: Clinical evaluation of a new sensor. Crit Care Med. 2005;33:2203–6. [PubMed]
17. Isert P. Control of carbon dioxide levels during neuroanesthesia: Current practice and an appraisal of our reliance upon capnography. Anaes Intens Care. 1994;22:435–41. [PubMed]
18. Short JA, Paris ST, Booker PD, Fletcher R. Arterial to end-tidal carbon dioxide tension difference in children with congenital heart disease. Br J Anaesth. 2001;86:349–53. [PubMed]
19. Hansen TN, Sonoda Y, McIlroy MB. Transfer of oxygen, nitrogen and carbon dioxide through normal adult human skin. J Appl Physiol. 1980;49:438–43. [PubMed]
20. Hazinski TA, Severinghaus JW. Transcutaneous analysis of arterial PCO2. Med Instrum. 1982;16:150–3. [PubMed]
21. Clark JS, Votteri B, Ariagno RL, Cheung P, Eichhorn JH, Fallat RJ, et al. Noninvasive assessment of blood gases. Am Rev Respir Dis. 1992;145:220–32. [PubMed]
22. Eletr S, Jimison H, Ream AK, Dolan WM, Rosenthal MH. Cutaneous monitoring of systemic PCO2 on patients in the respiratory intensive care unit being weaned from the ventilator. Acta Anaesth Scand. 1978;68:123–7. [PubMed]
23. Rithalia SV, Ng YN, Tinker J. Measurement of transcutaneous PCO2 in critically ill patients. Resuscitation. 1982;10:13–8. [PubMed]
24. Rithalia SV, Clutton-Brock TH, Tinker J. Characteristics of transcutaeous carbon dioxide tension monitors in normal adults and critically ill patients. Intens Care Med. 1984;10:149–53. [PubMed]

Articles from Journal of Minimal Access Surgery are provided here courtesy of Wolters Kluwer -- Medknow Publications