A Tale of Two Compartments: Interstitial Versus Blood Glucose Monitoring

Abstract

Self-monitoring of blood glucose was described as one of the most important advancements in diabetes management since the invention of insulin in 1920. Recent advances in glucose sensor technology for measuring interstitial glucose concentrations have challenged the dominance of glucose meters in diabetes management, while raising questions about the relationships between interstitial and blood glucose levels. This article will review the differences between interstitial and blood glucose and some of the challenges in measuring interstitial glucose levels accurately.

Not All Blood Glucose Is Created Equal

Blood glucose is generally measured as the venous plasma level. There is a 3–5 mg/mL difference between arterial and venous levels, with higher differences in the postprandial state.1 Levels are higher in the arterial blood because some of the glucose diffuses from the plasma to interstitial fluid (IF) as blood circulates through the capillary system. Arterial blood glucose and capillary blood glucose have been shown to be almost identical in concentration,2 even though the distribution of the glucose to the systemic capillaries does not occur instantaneously.

Glucose concentrations measured by glucose meters are whole blood levels, which can differ from plasma glucose levels by up to 11% (plasma higher). Abnormal hematocrit concentrations can result in falsely low (hematocrit >50%) or high (hematocrit <40%) glucose levels.3 Any delay in processing or transportation of samples can decrease glucose levels by 5–7%/h.4,5 Glucose meters use enzyme-based amperometric biosensors to measure glucose concentrations. Glucose oxidase oxidizes glucose to gluconolactone while reducing oxygen to H2O2. Other mediators like ascorbic acid, uric acid, acetaminophen, and salicylic acid can falsify the results by nonspecifically oxidizing H2O2.3,6

For glucose meter measurements, a skin-pricking device is used to access the dermal capillary plexus. Human skin consists of two layers—epidermis and dermis—residing above the adipose and muscle tissue. Epidermis is an avascular epithelial membrane. It has enzymes with glucose metabolizing effect.7 Moreover, glucose is formed from the breakdown of ceramide at the stratum granulosum–corneum interface.8 Dermis comprises many arterioles, venules, and capillaries, including a deep vascular plexus interfacing dermis and the subcutaneous tissue (Fig. 1a). Another vascular plexus located 0.3–0.6 mm from the skin surface is formed by the feeding vessels arising from the deep vascular plexus. It supplies the blood flow to the dermis and epidermis with the help of small capillary loops branching from the superficial plexus.9 The blood sampled from the skin prick comes from the capillaries of dermis with a small amount of blood from cut arterioles and venules providing a mix concentration. Blood flow to the skin is controlled by many factors, including autonomic nervous system, temperature, hormonal changes during menstrual cycle for females, and chemical inputs.10

FIG. 1.
Skin layers with the magnified IF space. (a) Vasculature in different skin layers with the CGM inserted into the subcutaneous tissue. (b) Diffusion of glucose from plasma to IF is in proportion to the concentration in each compartment. IF glucose is cleared ...

Capillary blood glucose levels at the fingertip have been shown to correlate well with systemic arterial blood glucose levels.11 During times of blood glucose stability, identical glucose levels were demonstrated from alternate sites (e.g., forearm) as compared with finger tip samples.12 However, at times of rapid change, mainly due to blood flow variability, levels from alternate sites differ considerably.12,13 Capillary blood glucose measured from the forearm is lower than fingertip values at times of rapid increases (>2 mg/dL/min) in systemic blood and higher during rapid decreases.12 Samples from the dorsal forearm have been shown to correspond better to fingertip values when compared with volar forearm samples.14 The only exception for the alternate site testing is the palm. The skin type of the palm is in the same skin category, hairless or glabrous skin, as the fingertip, and they share the same amount of blood flow, which is considerably more (five to 20 times) than the blood flow to most alternate sites like the forearm. In that respect, blood flow to forearm and abdomen upper dermal region has been reported to be comparable.11

IF constitutes approximately 45% of the volume fraction of human skin, with blood vessels contributing to the 5% of the skin volume.9 IF is a relatively passive medium that has one-third of the total protein concentration as compared to plasma with an average albumin/globulin ratio of 1.85.15

The total body volume of the interstitial space is three times that of plasma; however, IF compartments around the cells are microscopic.9 IF bathes the cells and feeds them with nutrients, including glucose, by providing a corridor between the capillaries and the cell. There is less IF in the subcutaneous tissue than in the dermis. Adipose tissue, just below the dermis, is richly vascularized with capillary walls that are relatively thinner (0.03 μm vs. 0.1 μm) than the capillaries of the dermis.1619 The basal membranes of the capillaries are in direct contact with the adipose cell cytoplasmic membrane.16 The size of adipocytes might affect the amount of IF in the subcutaneous tissue, suggesting that adiposity might have an affect on IF glucose concentrations.

The Relationship Between Plasma Glucose and Interstitial Glucose

Plasma and IF have different characteristics and should be considered as separate glucose compartments. We will concentrate on the IF–plasma glucose relationship mainly in the dermal/subcutaneous tissue rather than other tissues as this is the area of interest for current continuous glucose monitoring techniques.

Glucose is transferred from the capillary endothelium to the IF by simple diffusion across a concentration gradient without the need of an active transporter.20 Blood flow to the area dictates the amount of glucose delivered. Interstitial glucose values are determined by the rate of glucose diffusion from plasma to the IF and the rate of glucose uptake by subcutaneous tissue cells. Thus, the metabolic rate of the adjacent cells and other factors, like insulin, affecting glucose uptake by cells, the glucose supply from the blood vessel, blood flow to the area, and the permeability of the capillary that can be altered by many factors, including nerve stimulation, influence the interstitial glucose levels.21 The time required for glucose to diffuse from the capillary to the tissue plays an important role in the lag time between changes in plasma and interstitial glucose levels, but the lag during rapid changes of blood glucose is likely due to the magnitude of concentration differences in various tissues at a time of rapid change.22

A two-compartment model, described by Steil, Rebrin, and co-workers,23,24 provides an insight to the glucose dynamics between the plasma and IF compartments. As shown in Figure 1b, the equation characterizing IF glucose was described as follows: dV2G2/dt = K21V1G1  (K12 + K02)V2G2, where G1 =plasma glucose concentration, G2 = IF glucose concentration, K12 = forward flux rate for glucose transport across the capillary, K21 = reverse flux rate for glucose transport across the capillary, K02 = glucose uptake into the subcutaneous tissue, V1 = volume of the plasma, and V2 = volume of the IF.

A major confounding factor in evaluating the dynamics of changes in IF glucose concentrations has been the complexity of direct sampling methods, including insertion of wicks, blister formation, lymph sampling, and ultrafiltration. Microdialysis is an indirect method of estimating IF glucose values. Lönnroth et al.25 was the first to use this method to show that IF glucose was almost identical to venous plasma glucose in healthy individuals during steady state. Jannson et al.26 demonstrated an increase in lag time between IF and plasma glucose when there is a rapid rise in the plasma glucose level. The data of Jensen et al.27 revealed lower IF glucose levels than plasma glucose during clamp experiments extracting IF by suction blister technique. There are relatively limited data on dermal IF glucose levels. Bantle and Thomas28 demonstrated no significant difference between the dermal and plasma pre- and postprandial glucose levels in subjects with type 1 diabetes. A plasma and dermal interstitial glucose concentration lag time of 10–20 min was reported by Stout et al.29 A recent modeling study using skin biopsy specimens of the lower leg revealed a lag time of 1–3 min in the dermis.30 In addition to the physiological lag time, delayed availability of the glucose level depending upon the time required to collect and/or transfer the sample has been an a problem with previous methods of IF glucose sampling. Certain noninvasive methods involving optical technologies (kromoscopy, spectroscopy, etc.) to assess glucose in the IF have been in the works with their own challenges, including motion artifacts and inaccuracies stemming from inter-individual differences.31

Various minimally invasive or invasive methods have been used to detect glucose in IF with many of these techniques tested under different conditions (glucose infusion, glucose and insulin infusion, oral glucose tolerance test) in animal and human (with and without diabetes) studies. A wide span of accuracy has been reported from these studies, which could be due to differences in experimental conditions, methods/techniques, species, and characteristics of subjects.32 In general, IF and plasma glucose variations were evaluated in two different conditions: steady state and non–steady state. Under steady-state conditions, IF glucose generally correlated with the blood glucose with a lag time reported to be between 0 and 45 min and an average lag of 8–10 min.9 Increasing the blood flow to the interstitial glucose sampling site by applying controlled pressure has been shown to decrease the lag time between the blood and interstitial glucose at times of increasing plasma glucose levels.33 The reported gradient between interstitial and plasma glucose concentrations has varied between 20%34 to 110%.28 During the time of decreasing glucose, interstitial glucose may fall in advance of plasma glucose34 and reach nadir values that are lower than corresponding venous glucose levels.35 Interstitial glucose levels have been shown to remain below plasma glucose concentrations for prolonged periods of time after correction of insulin-induced hypoglycemia.36 These findings could be explained by the push–pull phenomenon during which the glucose is pushed from the blood to the interstitial space at times of increased blood glucose, and later on glucose being pulled from the IF to the surrounding cells during decreasing blood glucose levels.37 This phenomenon has been a matter of debate for some time, in light of data failing to support the push–pull phenomenon and instead reporting compensation of enhanced uptake of glucose in the IF by increased plasma glucose delivery and lack of glucose removal effect of insulin in the adipose IF.23,24

Current continuous glucose monitoring systems have the advantage of direct insertion of electrochemical sensors into the IF space rather than transporting the sampled fluid outside the body to detect glucose concentrations. Software programs have been designed to accommodate the lag in IF glucose readings. Despite the advances in the making of sensors with new and improved designs and materials, sensor insertion causes trauma to the insertion site. It can disrupt the tissue structure, provoking an inflammatory reaction that can consume glucose followed by a repair process.3840 The interaction of the sensor with the traumatized microenvironment warrants the need for a waiting period for the sensor signal to stabilize, and that period varies depending on the sensor type.21

Because continuous glucose sensor manufacturing has not progressed to the accuracy and precision of blood glucose meter strips, sensor glucose signals must be calibrated against corresponding blood glucose meter levels. Such calibrations transforms the sensor signal into a glucose value and assumes that the plasma-to-IF glucose gradient remains relatively constant.41 This assumption will not be valid if sensors are calibrated during rapid changes in plasma glucose, which is a major source of sensor error. The effect of sensor lag on performance is most obviously seen during periods of rapid glucose rate of change (either up or down). Sensor levels may trail glucose levels by 5–10 min during periods of rapid change, but the most important effect on lag is to introduce error during calibration, which affects long-term sensor performance. Moreover, changes in plasma–interstitial gradient in certain physiological conditions, like insulin-induced hypoglycemia, may be misinterpreted as sensor inaccuracy.41

Changes in sensor function that contribute to drift in sensor signal over time are due to biocompatibility problems like biofouling (obstruction of fluid exchange after nonspecific protein adsorption), passivation of electrodes (weakening of signal by reduction in conductivity), and degeneration.42,43 The surface of the electrode can become covered with cells or other substances,21 and the sensor can be damaged by the effects of proteolytic enzymes and free radicals.38,40 Inflammatory cells can consume glucose around the implanted sensor,39 and later on during the wound healing, more capillaries can supply glucose to the area.44 Bleeding from the skin also can interfere with the sensor function.34 Certain new technical changes like coating have demonstrated some success in preventing such sensor damage from inflammatory cells.39,44 Recalibration at fixed intervals is currently required to deal with problems related to signal drift.21

Another question that needs to be addressed is if subcutaneous IF glucose measurements parallel the central nervous system glucose levels during hypo- or hyperglycemia. Nielsen et al.45 demonstrated no differences between subcutaneous adipose tissue, muscle, and central nervous system blood and interstitial glucose levels in an animal study during hyperglycemia. Despite the lack of glucose measurement delay between these tissues during hypoglycemia, there was reduced magnitude of excursions in the central nervous system that correlated better with interstitial glucose.

Timeline for Continuous Glucose Monitors (CGMs)

Food and Drug Administration-approved and commercially available CGMs have been introduced over the last decade. The first one on the market was the MiniMed (Northridge, CA) Continuous Glucose Monitoring System (CGMS®), which stored glucose readings every 5 min up to 3 days. Sensor glucose values were available only retrospectively after downloading the sensor without the convenience of real-time values. The GlucoWatch® 2 Biographer (Cygnus Inc., Redwood City, CA) was the first CGM with real-time glucose values. However, inaccurate readings, false alarms, local irritation on the insertion site with lack of improvement in glycemic control, and hypoglycemic episodes destined the GlucoWatch to become a part of CGM history rather than a popular device in use.46,47

All three of the current CGM devices in common use provide sensor glucose levels in real time and use glucose oxidase-based electrochemical methods. Sensor signals are transmitted by wireless radiofrequency telemetry to the receiver. As noted above, interference with glucose readings by the sensor can occur with certain substances like glutathione, ascorbic acid, uric acid, paracetamol, isoniazid, and salicylate, which may become co-oxidized at the sensor and lead to overestimation of glucose levels.21 Recent advances in the sensor technology were reported to overcome this problem.4850 The osmium complex in the wired enzyme sensor is designed to react at a relatively low potential, 40 mV, compared to the 500 mV required to reduce H2O2. Substances such as uric acid or acetaminophen react at higher voltages, and the interference is minimized by the low operating potential.50

Overall percentage of error for the CGM runs around 15%. As mentioned before, accuracy depends on multiple factors like current glucose concentration and rate of change of glucose values, with poor correlation during hypoglycemia and times of rapid change. Percentage of error for individual sensors have been reported as 17% for the Guardian® REAL-Time (Minimed, Northridge, CA),51 11–16% for DexCom STS (DexCom, San Diego, CA),52,53 and 12–14% for the Navigator (Abbott Diabetes Care, Alameda, CA).48,49

The CGM has been declaring its use as a valuable tool for the management of diabetes therapy slowly but surely for the past couple of years. This underscores the importance of correctly interpreting and utilizing the glucose sensor data to make insulin adjustments. Patients might have a tendency to bolus insulin excessively when they experience high blood glucose alarms without taking into consideration the residual insulin from the premeal bolus or respond to declining glucose levels by decreasing the basal insulin infusion unnecessarily.54 Even though there is not a true and tried guideline for modifying insulin dose depending on the sensor glucose levels, the feasibility of certain algorithms has been shown to be effective.55 According to this algorithm the insulin bolus dose should be adjusted by 20% (decrease with declining glucose levels and increase with raising levels) for glucose changes >2 mg/dL. An adjustment of 10% insulin bolus is recommended for a change in glucose levels of ±1–2 mg/dL.55

Clinical studies demonstrating the benefit of using CGM as an adjunct to detect glucose trends for various patient populations with and without diabetes are likely to broaden CGM use in medicine. Sensors have been used successfully to detect abnormal glucose levels in the intensive care setting despite the confounding factors like edema, hypothermia, or multiple medication use.56 Outpatient studies including the use of CGM for diagnosing gestational diabetes and later on guiding therapy for pregnant women demonstrated promising results suggesting the possible future use of CGM in these settings.5760

What Does the Future Hold?

The introduction of new continuous glucose monitoring systems has opened up a new compartment that has not been heretofore accessible—the IF space. Most clinicians are very comfortable in interpreting changes in plasma glucose because the intravascular space has been readily accessible. Changes in plasma glucose usually precede changes in interstitial glucose and thus provide an earlier warning signal regarding evolving hypo- and hyperglycemia. Nevertheless, there are a number of circumstances where there are discrepancies between plasma glucose levels and clinical symptoms. A prime example is persistence of impaired cognition for prolonged periods of time after correction of hypoglycemia, which may be related to delays in the correction of IF glucose concentrations. In circumstances such as these, it can be argued that accurate and precise measurements of interstitial glucose levels may be more important clinically. Advances in sensor technology that do away with the need to calibrate sensor signals against plasma glucose levels will eliminate a major source of sensor error and allow us to measure the concentrations of glucose that are directly available to cells much more accurately. It will also require clinicians to develop a new set of metrics to evaluate normal, as well as clinically relevant high and low, interstitial glucose levels. Even the terminology may need to change. For example, in the future, glucopenia rather than hypoglycemia may be a major obstacle to successful treatment of type 1 diabetes.

Acknowledgments

This study is made possible by grants from NIH-CTSA (RR 023423), NIH-K12 (DK063709), and the Stephen I. Morse Pediatric Diabetes Research Fund.

Author Disclosure Statement

E.C. does not have any disclosures. W.V.T. is a member of the advisory board to Medtronic and Abbott Diabetes Care and a member of the speaker's bureau for Medtronic.

References

1. Glassberg BY. The arteriovenous difference in blood sugar content. Arch Intern Med. 1930;46:605–609.
2. Foster B. Studies in carbohydrate metabolism. J Biol Chem. 1922;1:291–301.
3. Burrin JM. Alberti KG. What is blood glucose: can it be measured? Diabet Med. 1990;7:199–206. [PubMed]
4. Weissman M. Klein B. Evaluation of glucose determinations in untreated serum samples. Clin Chem. 1958;4:420–422. [PubMed]
5. Chan AY. Swaminathan R. Cockram CS. Effectiveness of sodium fluoride as a preservative of glucose in blood. Clin Chem. 1989;35:315–317. [PubMed]
6. Roe JN. Biosensor development. Pharm Res. 1992;9:835–844. [PubMed]
7. Schurer NY. Elias PM. The biochemistry and function of stratum corneum lipids. Adv Lipid Res. 1991;24:27–56. [PubMed]
8. Cunningham DD. Young DF. Measurements of glucose on the skin surface, in stratum corneum and in transcutaneous extracts: implications for physiological sampling. Clin Chem Lab Med. 2003;41:1224–1228. [PubMed]
9. Roe JN. Smoller BR. Bloodless glucose measurements. Crit Rev Ther Drug Carrier Syst. 1998;15:199–241. [PubMed]
10. Yum SI. Roe J. Capillary blood sampling for self-monitoring of blood glucose. Diabetes Technol Ther. 1999;1:29–37. [PubMed]
11. Koschinsky T. Jungheim K. Heinemann L. Glucose sensors and the alternate site testing-like phenomenon: relationship between rapid blood glucose changes and glucose sensor signals. Diabetes Technol Ther. 2003;5:829–842. [PubMed]
12. Jungheim K. Koschinsky T. Glucose monitoring at the arm: risky delays of hypoglycemia and hyperglycemia detection. Diabetes Care. 2002;25:956–960. [PubMed]
13. Jungheim K. Koschinsky T. Risky delay of hypoglycemia detection by glucose monitoring at the arm. Diabetes Care. 2001;24:1303–1306. [PubMed]
14. Monfre SL. Hazen KH. Fischer JS. Welch M. Hocker-Smith L. Guttridge RJ. Ruchti TL. Physiologic differences between volar and dorsal capillary forearm glucose concentrations and finger stick glucose concentrations in diabetes [abstract] Diabetes. 2002;51(Suppl 2):A125.
15. Collison ME. Stout PJ. Glushko TS. Pokela KN. Mullins-Hirte DJ. Racchini JR. Walter MA. Mecca SP. Rundquist J. Allen JJ. Hilgers ME. Hoegh TB. Analytical characterization of electrochemical biosensor test strips for measurement of glucose in low-volume interstitial fluid samples. Clin Chem. 1999;45:1665–1673. [PubMed]
16. Schoonen AJ. Wientjes KJ. A model for transport of glucose in adipose tissue to a microdialysis probe. Diabetes Technol Ther. 2003;5:589–598. [PubMed]
17. Ryan TJ. Curri SB. Blood vessels and lymphatics. Clin Dermatol. 1989;7:25–36. [PubMed]
18. Braverman IM. Keh-Yen A. Ultrastructure of the human dermal microcirculation. III. The vessels in the mid- and lower dermis and subcutaneous fat. J Invest Dermatol. 1981;77:297–304. [PubMed]
19. Braverman IM. Yen A. Ultrastructure of the human dermal microcirculation. II. The capillary loops of the dermal papillae. J Invest Dermatol. 1977;68:44–52. [PubMed]
20. Zierler K. Whole body glucose metabolism. Am J Physiol. 1999;276:E409–E426. [PubMed]
21. Koschinsky T. Heinemann L. Sensors for glucose monitoring: technical and clinical aspects. Diabetes Metab Res Rev. 2001;17:113–123. [PubMed]
22. Van Der Valk PR. Van Der Schatte Olivier-Steding I. Wientjes KJ. Schoonen AJ. Hoogenberg K. Alternative-site blood glucose measurement at the abdomen. Diabetes Care. 2002;25:2114–2115. [PubMed]
23. Steil GM. Rebrin K. Hariri F. Jinagonda S. Tadros S. Darwin C. Saad MF. Interstitial fluid glucose dynamics during insulin-induced hypoglycaemia. Diabetologia. 2005;48:1833–1840. [PubMed]
24. Rebrin K. Steil GM. van Antwerp WP. Mastrototaro JJ. Subcutaneous glucose predicts plasma glucose independent of insulin: implications for continuous monitoring. Am J Physiol. 1999;277:E561–E571. [PubMed]
25. Lönnroth P. Jansson PA. Smith U. A microdialysis method allowing characterization of intercellular water space in humans. Am J Physiol. 1987;253:E228–E231. [PubMed]
26. Jansson PA. Fowelin J. Smith U. Lonnroth P. Characterization by microdialysis of intracellular glucose level in subcutaneous tissue in humans. Am J Physiol. 1988;255:E218–E220. [PubMed]
27. Jensen BM. Bjerring P. Christiansen JS. Orskov H. Glucose content in human skin: relationship with blood glucose levels. Scand J Clin Lab Invest. 1995;55:427–432. [PubMed]
28. Bantle JP. Thomas W. Glucose measurement in patients with diabetes mellitus with dermal interstitial fluid. J Lab Clin Med. 1997;130:436–441. [PubMed]
29. Stout PJ. Peled N. Erickson BJ. Hilgers ME. Racchini JR. Hoegh TB. Comparison of glucose levels in dermal interstitial fluid and finger capillary blood. Diabetes Technol Ther. 2001;3:81–90. [PubMed]
30. Groenendaal W. Schmidt KA. von Basum G. van Riel NA. Hilbers PA. Modeling glucose and water dynamics in human skin. Diabetes Technol Ther. 2008;10:283–293. [PubMed]
31. Oliver NS. Toumazou C. Cass AE. Johnston DG. Glucose sensors: a review of current and emerging technology. Diabet Med. 2009;26:195–196. [PubMed]
32. Wilinska ME. Bodenlenz M. Chassin LJ. Schaller HC. Schaupp LA. Pieber TR. Hovorka R. Interstitial glucose kinetics in subjects with type 1 diabetes under physiologic conditions. Metabolism. 2004;53:1484–1491. [PubMed]
33. Stout PJ. Racchini JR. Hilgers ME. A novel approach to mitigating the physiological lag between blood and interstitial fluid glucose measurements. Diabetes Technol Ther. 2004;6:635–644. [PubMed]
34. Sternberg F. Meyerhoff C. Mennel FJ. Mayer H. Bischof F. Pfeiffer EF. Does fall in tissue glucose precede fall in blood glucose? Diabetologia. 1996;39:609–612. [PubMed]
35. Caplin NJ. O'Leary P. Bulsara M. Davis EA. Jones TW. Subcutaneous glucose sensor values closely parallel blood glucose during insulin-induced hypoglycaemia. Diabet Med. 2003;20:238–241. [PubMed]
36. Aussedat B. Thomé-Duret V. Reach G. Lemmonier F. Klein JC. Hu Y. Wilson GS. A user-friendly method for calibrating a subcutaneous glucose sensor-based hypoglycaemic alarm. Biosens Bioelectron. 1997;12:1061–1071. [PubMed]
37. Thomé-Duret V. Reach G. Gangnerau MN. Lemonnier F. Klein JC. Zhang Y. Hu Y. Wilson GS. Use of a subcutaneous glucose sensor to detect decreases in glucose concentration prior to observation in blood. Anal Chem. 1996;68:3822–3826. [PubMed]
38. Clark LC., Jr Spokane RB. Homan MM. Sudan R. Miller M. Long-term stability of electroenzymatic glucose sensors implanted in mice. An update. ASAIO Trans. 1988;34:259–265. [PubMed]
39. Gerritsen M. Jansen JA. Kros A. Vriezema DM. Sommerdijk NAJM. Nolte RJM. Lutterman JA. Van Hövell SWFM. Van der Gaag A. Influence of inflammatory cells an serum on the performace of implantable glucose sensors. J Biomed Materials Res. 2001;54:69–75. [PubMed]
40. Kyrolainen M. Rigsby P. Eddy S. Vadgama P. Bio-/haemocompatibility: implications and outcomes for sensors. Acta Anaesthesiol Scand Suppl. 1995;104:55–60. [PubMed]
41. Rebrin K. Steil GM. Can interstitial glucose assessment replace blood glucose measurements? Diabetes Technol Ther. 2000;2:461–472. [PubMed]
42. Boyne MS. Silver DM. Kaplan J. Saudek CD. Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor. Diabetes. 2003;52:2790–2794. [PubMed]
43. Mang A. Pill J. Gretz N. Kränzlin B. Buck H. Schoemaker M. Petrich W. Biocompatibility of an electrochemical sensor for continuous glucose monitoring in subcutaneous tissue. Diabetes Technol Ther. 2005;7:163–73. [PubMed]
44. Gerritsen M. Jansen JA. Kros A. Vriezema DM. Sommerdijk NA. Nolte RJ. Lutterman JA. Van Hövell SW. Van der Gaag A. Influence of inflammatory cells and serum on the performance of implantable glucose sensors. J Biomed Mater Res. 2001;54:69–75. [PubMed]
45. Wang J. In vivo glucose monitoring: towards ‘Sense and Act’ feedback-loop individualized medical systems. Talanta. 2008;75:636–641. [PubMed]
46. Nielsen JK. Djurhuus CB. Gravholt CH. Carus AC. Granild-Jensen J. Orskov H. Christiansen JS. Continuous glucose monitoring in interstitial subcutaneous adipose tissue and skeletal muscle reflects excursions in cerebral cortex. Diabetes. 2005;54:1635–1639. [PubMed]
47. Diabetes Research in Children Network (DirecNet) Study Group. Youth and parent satisfaction with clinical use of the GlucoWatch G2 Biographer in the management of pediatric type 1 diabetes. Diabetes Care. 2005;28:1929–1935. [PMC free article] [PubMed]
48. Chase HP. Beck R. Tamborlane W. Buckingham B. Mauras N. Tsalikian E. Wysocki T. Weinzimer S. Kollman C. Ruedy K. Xing D. A randomized multicenter trial comparing the GlucoWatch Biographer with standard glucose monitoring in children with type 1 diabetes. Diabetes Care. 2005;28:1101–1106. [PubMed]
49. Weinstein RL. Schwartz SL. Brazg RL. Bugler JR. Peyser TA. McGarraugh GV. Accuracy of the 5-day FreeStyle Navigator continuous glucose monitoring system: comparison with frequent laboratory reference measurements. Diabetes Care. 2007;30:1125–1130. [PubMed]
50. Wilson DM. Beck RW. Tamborlane WV. Dontchev MJ. Kollman C. Chase P. Fox LA. Ruedy KJ. Tsalikian E. Weinzimer SA. DirecNet Study Group: The accuracy of the FreeStyle Navigator continuous glucose monitoring system in children with type 1 diabetes. Diabetes Care. 2007;30:59–64. [PMC free article] [PubMed]
51. Feldman B. Brazg R. Schwartz S. Weinstein R. A continuous glucose sensor based on Wired Enzyme technology—results from a 3-day trial in patients with type 1 diabetes. Diabetes Technol Ther. 2003;5:769–779. [PubMed]
52. Bode B. Gross K. Rikalo N. Schwartz S. Wahl T. Page C. Gross T. Mastrototaro J. Alarms based on real-time sensor glucose values alert patients to hypo- and hyperglycemia: the Guardian continuous monitoring system. Diabetes Technol Ther. 2004;6:105–113. [PubMed]
53. Garg S. Zisser H. Schwartz S. Bailey T. Kaplan R. Ellis S. Jovanovic L. Improvement in glycemic excursions with a transcutaneous, real-time continuous glucose sensor: a randomized controlled trial. Diabetes Care. 2006;29:44–50. [PubMed]
54. Garg S. Jovanovic L. Relationship of fasting and hourly blood glucose levels to HbA1c values: safety, accuracy, and improvements in glucose profiles obtained using a 7-day continuous glucose sensor. Diabetes Care. 2006;29:2644–2649. [PubMed]
55. Wolpert HA. The nuts and bolts of achieving end points with real-time continuous glucose monitoring. Diabetes Care. 2008;31(Suppl 2):S146–S149. [PubMed]
56. Diabetes Research In Children Network (DirecNet) Study Group. Buckingham B. Xing D. Weinzimer S. Fiallo-Scharer R. Kollman C. Mauras N. Tsalikian E. Tamborlane W. Wysocki T. Ruedy K. Beck R. Use of the DirecNet Applied Treatment Algorithm (DATA) for diabetes management with a real-time continuous glucose monitor (the FreeStyle Navigator) Pediatr Diabetes. 2008;9:142–147. [PMC free article] [PubMed]
57. Piper HG. Alexander JL. Shukla A. Pigula F. Costello JM. Laussen PC. Jaksic T. Agus MS. Real-time continuous glucose monitoring in pediatric patients during and after cardiac surgery. Pediatrics. 2006;118:1176–1184. [PubMed]
58. Bühling KJ. Winkel T. Wolf C. Kurzidim B. Mahmoudi M. Wohlfarth K. Wäscher C. Schink T. Dudenhausen JW. Optimal timing for postprandial glucose measurement in pregnant women with diabetes and a non-diabetic pregnant population evaluated by the Continuous Glucose Monitoring System (CGMS) J Perinat Med. 2005;33:125–131. [PubMed]
59. Kerssen A. de Valk HW. Visser GH. The Continuous Glucose Monitoring System during pregnancy of women with type 1 diabetes mellitus: accuracy assessment. Diabetes Technol Ther. 2004;6:645–651. [PubMed]
60. Kestila KK. Ekblad UU. Ronnemaa T. Continuous glucose monitoring versus self-monitoring of blood glucose in the treatment of gestational diabetes mellitus. Diabetes Res Clin Pract. 2007;77:174–179. [PubMed]
61. McLachlan K. Jenkins A. O'Neal D. The role of continuous glucose monitoring in clinical decision-making in diabetes in pregnancy. Aust N Z J Obstet Gynaecol. 2007;47:186–190. [PubMed]

Articles from Diabetes Technology & Therapeutics are provided here courtesy of Mary Ann Liebert, Inc.