Evaluation of Effect of Specimen-Handling Parameters for Plasma Preparation Tubes on Viral Load Measurements Obtained by Using the Abbott RealTime HIV-1 Load Assay[down-pointing small open triangle]

Abstract

HIV-1 viral load testing is essential to the management of HIV-1-infected patients, and proper specimen handling ensures accurate viral load (VL) results. This study was performed to (i) evaluate the effect of freezing plasma in situ in BD Vacutainer plasma preparation tubes (PPT) on the accuracy of HIV-1 viral load results using the Abbott RealTime HIV-1 assay and (ii) evaluate the effect of whole-blood storage in the PPT for 6 h at room temperature prior to centrifugation (PPT6H) rather than 2 h as specified in the PPT product insert. Of the 64 HIV-positive subjects evaluated, 29 had average viral load counts of >40 copies/ml in at least one of the tubes tested and 35 subjects had a result of either “undetected target” or “below the limit of quantification” (LOQ) for all or some of the tubes regardless of handling condition. For the 29 subjects with VLs that were >LOQ, the mean biases between plasma from Vacutainer K2EDTA tubes and plasma frozen in situ in PPT and between K2EDTA tube plasma and plasma from PPT6H (log10 copies/ml) were 0.005 and −0.001, respectively, and r2 was >0.92 for all correlations. We conclude that VLs determined from plasma frozen in situ in PPT are equivalent to VLs in K2EDTA tube plasma and can be used for accurate quantification of HIV-1 RNA in the Abbott RealTime HIV-1 assay. Furthermore specimens collected in PPT can be stored for 6 h at room temperature with no effect on viral load results as measured by the Abbott RealTime HIV-1 assay.

Accurate quantification of human immunodeficiency virus type 1 (HIV-1), also referred to as HIV-1 viral load (VL) testing, is essential for effective management of HIV-1-infected patients. The BD Vacutainer plasma preparation tube (PPT) (BD Preanalytical Systems, Franklin Lakes, NJ) was developed to facilitate the handling of plasma specimens used for HIV-1 viral load testing. The PPT contains a K2EDTA additive and a polyester gel which, upon centrifugation, forms a barrier that separates blood cells from plasma, allowing storage, freezing, and shipment of the plasma specimen in situ in PPT. Use of PPT also results in a decrease in the amount of labor required to process specimens, elimination of a potential source of error in specimen labeling, and reduction in the risk of HIV exposure associated with the transfer of separated plasma to secondary tubes for shipping.

Earlier studies using PPT for specimen collection demonstrated compatibility of this tube with HIV-1 VL tests (1, 4, 5). Over the past few years, however, investigators have reported that freezing plasma in situ in a PPT produced higher HIV-1 viral load results compared to plasma from K2EDTA tubes when tested in the Cobas Amplicor HIV-1 monitor system (Roche Molecular Diagnostics, Pleasanton, CA) (2, 3, 11). Elevated levels of HIV-1 VL from PPT frozen in situ after centrifugation were first reported by Squires et al., who showed that PPT yielded higher HIV RNA levels than K2EDTA tubes. The disparity in quantification, seen in both standard and ultrasensitive assays, was more apparent in specimens with VLs between the limit of quantification (LOQ), 50 copies/ml, and 1,000 copies/ml and more pronounced in specimens close to or below 50 copies/ml (3, 10, 11). In fact, studies that compared plasma aspirated from K2EDTA tubes with plasma frozen in situ in PPT show that VLs that are clearly below the LOQ in the former are quantifiable in the latter (3, 8, 11). Such discrepant results for plasma collected in different tubes from the same subject could be interpreted as virological failure, and these results, therefore, have therapeutic implications.

Additional studies reported that specimens which were aspirated and transferred to another tube after centrifugation did not show patterns of artificially increased viral load (8, 10). Similarly, recentrifugation of specimens transported in PPT at 4°C eliminated the inaccurate quantification of HIV seen in plasmas frozen in situ in PPT (6, 9).

Such observations led investigators to associate elevated viral loads in plasma in PPT with the presence of cell-associated nucleic acids released from cells trapped within or adhered to the surface of the gel barrier in PPT (6). Depletion of cellular material by recentrifugation of separated, unfrozen plasma from PPT resulted in lower VLs, indicating that cell-associated nucleic acid contributed to the elevation in VL. This finding was compatible with an earlier study which had shown that the HIV proviral DNA associated with cellular (genomic) DNA was in part responsible for the increase in viral load (13).

The aim of this study was to evaluate the performance of PPT with the new automated RealTime HIV-1 assay (Abbott Molecular, Des Plaines, IL) for the quantification of HIV-1 viral load using the Abbott m2000sp and m2000rt systems for sample processing and amplification/detection, respectively. Specifically, we were interested in determining the effect of freezing plasma in situ in PPT on HIV-1 viral load analyzed using the Abbott RealTime HIV-1 assay. In addition, the study evaluated the effect on viral load of whole-blood storage in PPT for 6 h before centrifugation rather than 2 h, as specified in the PPT product insert.

MATERIALS AND METHODS

Patient population.

All 68 subjects of this study were consenting HIV-positive adults who had been previously shown to have viral load results in the range of 50 to 5,000 copies/ml or below 50 copies/ml within the 60 days prior to enrollment in the study. Study monitors were blind with regard to patient status, CD4 count, drug therapy, sex, and date of diagnosis of infection.

Assay design.

A schematic of the study protocol is depicted in Fig. Fig.1.1. Following hospital procedures for routine venipuncture, blood was collected into four tubes: one 10-ml BD Vacutainer K2EDTA tube (catalog no. 366643) and three 8.5-ml BD Vacutainer plasma preparation tubes (PPT; catalog no. 362799). Tubes were drawn to a randomization schedule, inverted 8 to 10 times immediately after blood collection, and transferred within 2 h postcollection to the laboratory for processing.

FIG. 1.
Study design.

On the day of phlebotomy and within 2 h after blood collection, the K2EDTA tube (Fig. (Fig.1,1, tube A) and two PPTs (Fig. (Fig.1,1, tubes B and C) were centrifuged for 10 min at 1,100 × g to separate the plasma. At 6 h after blood collection, a PPT (Fig. (Fig.1,1, tube D) was centrifuged for 10 min at 1,100 × g to separate the plasma. After centrifugation, three 1.1-ml aliquots of plasma were transferred from each of tubes A, B, and D into secondary tubes A1 to A3, B1 to B3, and D1 to D3, respectively, and frozen at −70°C until the day of analysis. The plasma specimen in PPT C was frozen in situ in PPT at −70°C until the day of analysis.

On the day of analysis, plasma aliquots A1 to A3, B1 to B3, and D1 to D3 were thawed, processed, and analyzed for viral load in triplicate using the Abbott RealTime HIV-1 assay. At the same time, PPT C, containing in situ frozen plasma, was thawed and inverted 8 to 10 times and three 1.1-ml aliquots of plasma were transferred into secondary tubes C1 to C3. All plasma aliquots were vortexed three times for 2 to 3 s, centrifuged at 2,000 × g for 5 min, and quantified in the Abbott RealTime HIV-1 assay in accordance with the manufacturer's instructions.

The average of the three aliquots from each specimen was used as the VL for that tube. If the VL result for an aliquot was “undetected target,” a value of 0 was used to calculate the average. The lower limit of quantification (LOQ) for the assay is 40 copies/ml. In this study, a “detected” result for specimens indicated that the viral load was below the LOQ (<40copies/ml) but above the “undetected target” level. Viral loads for these specimens were calculated using the assay standard curve (as suggested by the manufacturer) for statistical purposes only.

Statistical analysis.

Tubes for which all three aliquots had detectable virus and the average VL was at least 40 copies/ml were used for estimating mean biases among tubes. Analysis of variance (ANOVA)-type models with tube type and subject effects were used to estimate mean biases between PPT B, PPT C, and PPT D and the K2EDTA tube.

The two one-sided test procedure (TOST) was applied to calculate the 95% confidence limits of PPT B, PPT C, and PPT D mean biases versus the K2EDTA tube. Log-log linear regression plots were constructed for comparing PPT B, C, and D to K2EDTA tube A.

Subjects were excluded from the tube bias calculations when the average VL of three aliquots was below the LOQ. For these subjects with very low viral loads, a Cochran Q test was performed and a P value was calculated to determine whether the rates of detection for them were different for the four tubes.

Additionally, cross tabulations of tubes with VLs below and above the LOQ were constructed. The exact unconditional McNemar test was used to compare the proportions of subjects with VLs below the LOQ between tubes to determine whether one tube is more likely than another to classify a subject as above or below the LOQ.

RESULTS

A total of 68 subjects participated in the study, of which 64 subjects had complete data sets. Specimens from all 64 subjects were run in triplicate, and the average VL from these three specimens was used for the analysis. Of the 64 subjects, 29 subjects had average VLs that were >LOQ in at least one of the tube types collected. Thirty-five of these 64 subjects had VLs which were below the LOQ or for which the target was undetected in all or most of the tubes collected. The mean biases and correlation coefficients of viral loads between tubes were calculated for the 29 subjects with viral loads that were >LOQ. Data from the remaining 35 subjects were used to determine whether the rate of target detection for these very low viral load subjects would differ depending upon the tube or handling protocol for collection and storage. From all 64 subjects, the data were used to compare the tubes for the proportion of subjects with results that were <LOQ.

Mean bias between tubes.

Summary statistics for log10 VL for the 29 subjects included in the bias calculations are shown in Table Table1.1. The mean log10 viral loads in all four tubes were comparable, as were the medians and the ranges of the VLs in each category. Two-thirds of all specimens in the study had VLs below 1,000 copies/ml. Freezing the plasma in the PPT in situ did not affect the viral load, compared to that for K2EDTA or PPT plasma. Moreover, there was no difference in VL if the specimen was centrifuged and aspirated at 2 h or 6 h postphlebotomy in any of the 87 specimens from the 29 subjects with quantifiable VLs. The mean bias for the log10 VL of these subjects for each PPT (tubes B, C, and D) compared to K2EDTA plasma (tube A) is given in Table Table2.2. All comparisons showed that the VLs in each PPT, irrespective of handling conditions, were equivalent to the corresponding VLs in both the K2EDTA tube and the other PPT.

TABLE 1.
Summary statistics for viral load count
TABLE 2.
Estimated mean bias of PPT versus the K2EDTA tube

ANOVA-type models with tube type and subject effects were used to estimate mean biases between PPT (tubes B, C, and D) and the K2EDTA tube (tube A). As seen in Fig. Fig.2,2, the bias between most tubes was <0.5 log10 copies/ml and was consistent through the entire study range. As expected, the VL showed more variability for viral load counts of <100 copies/ml.

FIG. 2.
Scatterplot showing the difference in VLs for PPT (tube B), PPT with plasma frozen in situ (PPT in situ; tube C), and whole-blood storage in PPT for 6 h at room temperature prior to centrifugation (PPT6H; tube D) versus the K2EDTA tube (tube A). Specimens ...

Correlation of VLs between tubes.

Viral loads in the 29 quantifiable specimens were compared in the log-log linear regression plots for PPT (tube B, C, or D) versus the K2EDTA tube (Fig. (Fig.33 to to5).5). The VL from plasma frozen in situ in PPT correlated strongly to the VL from PPT processed 2 to 6 h postphlebotomy as well as to the VL from the K2EDTA plasma. The r2 values for all comparisons were calculated at >0.92. The mean log10 differences between the K2EDTA tube and the PPT (tubes B, C, and D) are 0.02, −0.01, and −0.01, respectively, and the agreement limits (±2 standard deviations [SD]) observed in the Bland-Altman plots are ± 0.52, ± 0.46, and ± 0.39, respectively.

FIG. 3.
Log-log linear regression of PPT (tube B) versus the K2EDTA tube (tube A). Correlation coefficients greater than 92% are shown. S, slope; R-Sq, regression coefficient; R-Sq(adj), regression coefficient (adjusted).
FIG. 5.
Log-log linear regression of PPT6H (tube D) versus the K2EDTA tube (tube A). Correlation coefficients greater than 92% are shown. S, slope; R-Sq, regression coefficient; R-Sq(adj), regression coefficient (adjusted).

Analysis of VL from specimens below the LOQ.

Thirty-five of the 64 subjects had average viral loads below the LOQ. The data from these 35 subjects are summarized in Table Table3.3. The K2EDTA tube result for 15 of these 35 subjects was “undetected target,” and the corresponding PPT (tube C) had this result for 12 subjects. Similarly, the K2EDTA tube results for 20 subjects and the PPT (tube C) results for 17 subjects were classified as “some aliquots detected.” The four different specimen types did not show significant differences between the numbers of specimens with a result of “detected” or “undetected.” A Cochran Q test, performed to evaluate the rate of detection for these very low VLs, showed a P value of 0.33, indicating no statistical difference among the four tubes. To further examine any variability in the specimens, the exact unconditional McNemar test was used to compare the proportions of subjects with VLs under 40 copies/ml between the four tubes (Table (Table4).4). Between-tube agreement was demonstrated for all comparisons, and no significant differences were observed between any tubes. The McNemar tests showed that no tube was more likely than another to classify a subject as above or below the LOQ.

TABLE 3.
Classification of the 35 subjects with VLs <LOQ
TABLE 4.
Tube comparison with McNemar P values

Finally, to determine whether tube type or handling condition had an effect on the distribution of viral loads in the subject population, the frequency distribution of viral loads at different levels was plotted as a function of the tube and handling condition. As seen in Fig. Fig.6,6, frequency distributions for all four collection conditions were similar.

FIG. 6.
Distribution of viral load versus tube type for all 68 subjects clearly shows that most of the specimens in the study had viral load counts close to or below the LOQ of the assay.

DISCUSSION

The current study reports on the effects on viral load results of three handling conditions for HIV-1-positive specimens collected in PPT using the Abbott RealTime HIV-1 viral load assay. In particular, the viral load for each of the test conditions in PPT were compared to values obtained from K2EDTA specimens from the same subject. Our results show that there was virtually no difference in VLs between any of the tubes in the study and that PPT can be used for the determination of VL in the Abbott RealTime HIV-1 assay.

Freezing of plasma in situ in PPT did not result in any difference in the VLs of specimens collected in K2EDTA tubes versus PPT when VLs were quantified in the Abbott HIV-1 RealTime assay, and both correlation and agreement results for the K2EDTA tubes and PPT (tube B, C, or D) at lower VLs (40 to 1,000 copies/ml) were excellent. Our observations clearly contrast with previous reports that demonstrate elevated VL in plasma frozen in situ in PPT (3, 10, 11). To the best of our knowledge, this is the first report that documents that plasma frozen in situ in PPT can be used for the determination of HIV-1 VL even in specimens with low VL values.

Discrepancies in VL results obtained after freezing the plasma in situ in PPT are attributed to the presence and amplification of cellular proviral DNA in the extracted RNA that is used for testing (6), although some investigators believe that platelets associated with the virus could play a role in the process (9). In fact our own studies (13) identified the presence of integrated proviral DNA in the plasmas frozen in situ in PPT, suggesting that this proviral DNA, upon amplification in the Roche Cobas monitor assay, yielded quantifiable VL results. Furthermore, since the Amplicor HIV-1 monitor assay employs total nucleic acid extraction in an assay that does not discriminate between amplification of HIV-1 viral RNA and proviral DNA, the presence of cellular, proviral DNA in the plasma is likely to yield elevated VL results.

The Abbott HIV-1 RealTime assay uses automated extraction of viral nucleic acid, which is performed on the m2000sp using purification reagents specific for RNA, followed by amplification with primers and probes that target the integrase region of the HIV-1 genome. The fact that we did not observe elevated VLs in plasma frozen in situ in PPT raises the likelihood that this fully automated method employs a more “RNA-centric” mode of nucleic acid extraction, with minimal contaminating amplifiable proviral DNA. The elevated VLs in PPT measured in the Roche Cobas Ampliprep TaqMan assay appear to be cell associated, indicating that proviral DNA may contribute to the results of HIV quantification. (6, 7). Since VL assays are used as the standard of care for monitoring patients on highly active antiretroviral therapy (HAART), a potentially incorrect result at the time of quantification could have an effect on patient care (7, 11, 12). The accurate quantification of HIV-1 RNA using the Abbott HIV-1 RealTime assay from plasma frozen in situ in PPT, as established in our study, confirms the use of the assay for the reliable measurement of HIV-1 VL.

Timely separation of plasma from whole blood is recommended for optimal assessment of HIV viral load, as the RNases present in whole blood have the potential to destroy RNA. The PPT product insert (BD, Franklin Lakes, NJ) states that the plasma in PPT ideally should be separated by centrifugation within 2 h of phlebotomy. The results of our study have clearly shown that centrifugation of plasma 6 h postphlebotomy does not affect the quantification of HIV RNA in the Abbott HIV-1 RealTime assay, allowing for convenience in the centrifugation and subsequent transportation of PPT for VL testing.

For subjects with VLs above the assay LOQ of 40 copies/ml, ANOVA analysis to estimate the mean bias between tubes and log-log linear regression analysis showed that all handling conditions tested were equivalent with 95% confidence. For subjects with VLs that were <LOQ for the assay, cross tabulation results and the McNemar P test showed no difference in the rates of detection among the four conditions tested. Further, there was no difference among tubes in the proportions of readings over or under the LOQ for the assay. For all subjects, there was no difference in the frequency distributions of VLs regardless of which tube type or handling condition was used for specimen collection and processing.

In conclusion, using the Abbott RealTime HIV-1 viral load assay, specimens can be collected in BD Vacutainer plasma preparation tubes and centrifuged within 6 h of phlebotomy with no effect on viral load results. Furthermore, plasma separated in PPT can be stored frozen in situ until the time of testing with no effect on viral load results.

FIG. 4.
Log-log linear regression of PPT in situ (tube C) versus the K2EDTA tube (tube A). Correlation coefficients greater than 92% are shown. S, slope; R-Sq, regression coefficient; R-Sq(adj), regression coefficient (adjusted).

Acknowledgments

We thank Valentin Parvu (Corporate Statistics Department, BD, Franklin Lakes, NJ) for providing the statistical analysis and interpretation of study data and Paula Belem for technical assistance.

Footnotes

[down-pointing small open triangle]Published ahead of print on 19 May 2010.

The authors have paid a fee to allow immediate free access to this article.

REFERENCES

1. Elbeik, T., P. Nassos, P. Kipnis, B. Haller, and V. L. Ng. 2005. Evaluation of the VACUTAINER PPT plasma preparation tube for use with the Bayer VERSANT assay for quantification of human immunodeficiency virus type 1 RNA. J. Clin. Microbiol. 43:3769-3771. [PMC free article] [PubMed]
2. García-Bujalance, S., C. Ladron de Guevara, J. González-García, J. R. Arribas, F. Zamora, and A. Gutiérrez. 2007. Elevation of viral load by PCR and use of plasma preparation tubes for quantification of human immunodeficiency virus type 1. J. Microbiol. Methods 69:384-386. [PubMed]
3. Griffith, B. P., and R. M. Donald. 2006. Increased levels of HIV RNA detected in samples with viral loads close to the detection limit collected in plasma preparation tubes (PPT). J. Clin. Virol. 35:197-200. [PubMed]
4. Holodniy, M. 2006. HIV-1 viral load quantitation: a 17-year perspective. J. Infect. Dis. 194:S38-S44. [PubMed]
5. Holodniy, M., L. Rainen, S. Herman, and B. Yen-Lieberman. 2000. Stability of plasma human immunodeficiency virus load in VACUTAINER PPT plasma preparation tubes during overnight shipment. J. Clin. Microbiol. 38:323-326. [PMC free article] [PubMed]
6. Kran, A.-M. B., T. O. Jonassen, M. Sannes, K. Jakobsen, A. Lind, A. Maeland, and M. Holberg-Petersen. 2009. Overestimation of human immunodeficiency virus type 1 load caused by the presence of cells in plasma from plasma preparation tubes. J. Clin. Microbiol. 47:2170-2174. [PMC free article] [PubMed]
7. Lima, V., R. Harrigan, and J. S. G. Montaner. 2009. Increased reporting of detectable plasma HIV-1 RNA levels at the critical threshold of 50 copies per milliliter with the Taqman assay in comparison to the Amplicor assay. J. Acquir. Immune Defic. Syndr. 51:3-6. [PubMed]
8. Rebeiro, P. F., A. Kheshti, S. S. Bebawy, S. E. Stinnette, H. Erdem, Y. W. Tang, T. R. Sterling, S. P. Raffanti, and R. T. D'Aquila. 2008. Increased detectability of plasma HIV-1 RNA after introduction of a new assay and altered specimen-processing procedures. Clin. Infect. Dis. 47:1354-1357. [PMC free article] [PubMed]
9. Sabino, E. C., N. J. Gaburo, J. P. Leite, M. Viecili, and L. G. Rosenfeld. 2004. Platelets may affect detection and quantitation of HIV RNA in plasma samples with low viral loads. J. Acquir. Immune Defic. Syndr. 37:1432-1433. [PubMed]
10. Salimnia, H., E. C. Moore, L. R. Crane, R. D. MacArthur, and M. R. Fairfax. 2005. Discordance between viral loads determined by Roche COBAS AMPLICOR human immunodeficiency virus type 1 monitor (version 1.5) standard and ultrasensitive assays caused by freezing patient plasma in centrifuged Becton-Dickinson Vacutainer brand plasma preparation tubes. J. Clin. Microbiol. 43:4635-4639. [PMC free article] [PubMed]
11. Squires, K., A. Lazzarin, J. M. Gatell, W. G. Powderly, V. Pokrovskiy, J.-F. Delfraissy, J. Jemsek, A. Rivero, W. Rozenbaum, S. Schrader, M. Sension, A. Vibhagool, A. Thiry, and M. Giordano. 2004. Comparison of once-daily atazanavir with efavirenz, each in combination with fixed-dose zidovudine and lamivudine, as initial therapy for patients infected with HIV. J. Acquir. Immune Defic. Syndr. 36:1011-1019. [PubMed]
12. Stosor, V., F. J. Palella, Jr., B. Berzins, M. Till, A. Leake, J. S. Chmiel, and R. L. Murphy. 2005. Transient viremia in HIV-infected patients and use of plasma preparation tubes. Clin. Infect. Dis. 41:1671-1674. [PubMed]
13. Wan, H., A. Seth, L. Rainen, and H. Fernandes. 2010. Co-amplification of HIV proviral DNA and viral RNA in assays used for quantification of HIV-1 RNA. J. Clin. Microbiol. 48:2186-2190. [PMC free article] [PubMed]

Articles from Journal of Clinical Microbiology are provided here courtesy of American Society for Microbiology (ASM)