Anterior-segment imaging for assessment of glaucoma

Abstract

This article summarizes the physics, technology and clinical application of ultrasound biomicroscopy (UBM) and optical coherence tomography (OCT) for assessment of the anterior segment in glaucoma. UBM systems use frequencies ranging from approximately 35 to 80 MHz, as compared with typical 10-MHz systems used for general-purpose ophthalmic imaging. OCT systems use low-coherence, near-infrared light to provide detailed images of anterior segment structures at resolutions exceeding that of UBM. Both technologies allow visualization of the iridocorneal angle and, thus, can contribute to the diagnosis and management of glaucoma. OCT systems are advantageous, being noncontact proceedures and providing finer resolution than UBM, but UBM systems are superior for the visualization of retroiridal structures, including the ciliary body, posterior chamber and zonules, which can provide crucial diagnostic information for the assessment of glaucoma.

Keywords: anterior segment, eye, glaucoma, imaging, optical coherence tomography, ultrasound biomicroscopy

Glaucoma is a disease that is essentially an optic neuropathy. Great strides have been made in the last decade in imaging the retinal nerve fiber layer and the optic nerve using optical coherence tomography (OCT). OCT is a noninvasive technique allowing the visualization and measurement of these structures for the assessment of glaucoma progression [15]. While damage to the optic and retinal nerves is responsible for the degradation of vision, in most instances this damage is associated with elevated intraocular pressure (IOP). Elevation of IOP results from an imbalance between the production and drainage of aqueous fluid, specifically, increased outflow resistance. Thus, assessment of the anterior segment plays a major role in the diagnosis and treatment of glaucoma.

The traditional tools for the assessment of the angle and anterior segment are the slit-lamp and gonioscopy, in combination with a Goldmann contact lens and classification using the system introduced by Shaffer [6]. Several newer technologies now exist for imaging of the anterior segment, including scanning Scheimpflug (e.g., Pentacam, Oculus, Inc., Lynnwood, WA, USA), and scanning slit-lamp systems (e.g., Orbscan, Orbtek Inc., Salt Lake City, UT, USA). The Pentacam can provide rapid and reproducible measurements of anterior chamber depth and angle width [7], but is limited to visualization of only the angle approach. Scheimpflug photography also does not display the retroiridal structures or the ciliary body, which are of great interest in glaucoma diagnosis [8]. The scanning peripheral anterior chamber depth analyzer is another optically based system. It has high reproducibility in screening for narrow angles, as with other systems using visible light, but may have difficulty in assessing the bottom of the angle, as can be accomplished by OCT and ultrasound biomicroscopy (UBM) [9]. While these visible light systems are undoubtedly useful, especially for screening for narrow angles, OCT (based on near-infrared light) and UBM systems allow imaging of the full-angle anatomy and, in the case of UBM, visualization of retroiridal structures and the ciliary body. Thus, these technologies provide optimal means for the assessment of the anterior segment in glaucoma.

The development of the ultrasound biomicroscope in the early 1990s introduced high-resolution cross-sectional imaging for assessment of the angle and ciliary body. The more recent development of anterior segment OCT (AS-OCT) further enhanced our ability to visualize the angle and other structures relevant to glaucoma diagnosis and management. In this article, we will review and compare the physics of UBM and OCT systems, and describe their contribution to glaucoma evaluation and management.

Ultrasound principles & technology

In ultrasonic imaging, the transducer emits a short acoustic pulse in response to a voltage transient generated by a pulser. As the pulse propagates through a tissue and encounters anatomic structures differing in acoustic impedance (density × speed of sound) from their surroundings, echoes are generated. When echoes reach the transducer, they are converted back into voltages and amplified. The range of each echo is proportional to the time delay between pulse emission and echo return, specifically, r = ct/2, where r is the range, c is the speed of sound (1532 m/s at 37°C in normal saline) and t is the time. Each pulse/echo event thus provides information along one line of sight. By mechanically scanning the probe, information along an ordered series of lines of sight is obtained. By converting echo amplitude into pixel intensity, a 2D cross-sectional B-scan image is then formed.

Ultrasound resolution improves with increasing center frequency and with bandwidth. Bandwidth describes the range of frequencies present in an ultrasound pulse. As the number of cycles in a pulse increases, bandwidth decreases and resolution degrades. Thus, high resolution demands both high frequency and broad bandwidth.

Ultrasonic imaging of the eye was introduced in the 1950s by Mundt and Hughes (A-scan) [10] and Baum and Greenwood (B-scan) [11], and 10-MHz mechanically scanned ultrasound systems have long since been used for general-purpose imaging of the eye. Such systems are generally designed to be used in contact with the eye or eyelid using a gel or viscous fluid (e.g., methyl-cellulose), as ultrasound will not otherwise propagate through the air–tissue interface. The 10-MHz ophthalmic B-scanners use a probe whose focus falls on the retina or posterior vitreous. In such a contact B-scanner, the anterior segment is poorly visualized, both owing to the limited resolution (~150 μm at 10 MHz) and because the anterior segment is in the defocused near-field of the probe. While an immersion exam will allow the probe to be offset from the eye sufficiently to place the anterior segment in the focal plane, the limited resolution at 10 MHz reduces diagnostic capacity.

Until the early 1990s, diagnostic ultrasound transducer material was largely limited to lead zirconate titanate crystal. To achieve higher frequencies, the crystal could be made thinner, but above approximately 15 MHz, this process resulted in unacceptable fragility due to the zirconate titanate’s brittleness. By the early 1990s, however, novel polymeric transducer materials (e.g., polyvinylidene fluoride) became available which, for the first time, allowed the production of relatively inexpensive high-frequency transducers [12,13]. Polyvinylidene fluoride membranes could be made sufficiently thin to support frequencies as high as 100 MHz (wavelength: 15 μm) and beyond, and also provided excellent bandwidth. Ultrasound systems utilizing probes of approximately 35 MHz or more have come to be known as UBM or ‘very high-frequency ultrasound systems’. Owing to the fact that acoustic attenuation increases exponentially with frequency, such systems are limited to the superficial tissues of the anterior segment.

In the early 1990s, Foster and Pavlin developed the first practical UBM system for imaging of the eye [14,15]. The ultrasound bio-microscope (Zeiss-Humphrey Instruments, San Leandro, CA, USA) was a commercial outgrowth of this work, allowing dissemination of UBM technology and widespread clinical application. Coleman, Silverman and their group at Cornell Medical College (New York, NY, USA) independently developed a UBM system emphasizing the processing of raw radiofrequency echo data acquired in sequential planes suitable for 3D analysis, especially corneal biometric analysis [16]. This system was also eventually commercialized as the Artemis-2 system (Ultralink, LLC, St Petersburg, FL, USA), which uniquely incorporated an optical subsystem for fixation, centration and display of eye position.

The Zeiss-Humphrey (later Paradigm Medical Industries, Salt Lake City, UT, USA) UBM consisted of a 50-MHz probe supported by an articulated arm. The probe provided a scan rate of 8 Hz, giving real-time imaging, with scans consisting of 256 lines of sight (vectors) over a 5 × 5-mm field. This allowed imaging of anatomic areas of interest in the anterior segment, but not the whole anterior segment in a single scan. After topical anesthesia, the probe was coupled to the eye using a plastic scleral shell that held the eyelids open and was filled with normal saline or methylcellulose for acoustic coupling.

Ultrasound biomicroscopy systems are now produced by numerous companies, including Quantel Medical Instruments (Bozeman, MT, USA), Optikon (Rome, Italy), Sonomed (Lake Success, NY, USA) and OPKO (Miami, FL, USA). IScience Interventional (Menlo Park, CA, USA) manufactures an 80-MHz scanner for high-resolution imaging of the angle and Schlemm’s canal. Arcscan, Inc. (Golden, CO, USA) is developing a successor to the Artemis-2. Handheld UBM probes are now often equipped with acoustically transparent, fluid-filled ‘bubble tips’ that can be placed directly onto the globe. These obviate the use of water-baths or scleral shells for acoustic coupling, greatly simplifying the examination and allowing the patient to be examined in a sitting position. In some cases, these instruments provide scan width sufficient to encompass the entire anterior segment, from angle-to-angle and sulcus-to-sulcus in a single image.

OCT principles & technology

Optical coherence tomography of the eye was first described by Huang and coworkers at the Massachusetts Institute of Technology (Boston, MA, USA) in 1991 [17]. OCT is sometimes referred to as an optical analog of ultrasound, but detects optical backscatter instead of acoustic backscatter. 2D OCT images are similarly known as B-scans [18]. As in ultrasound imaging, OCT images are constructed by combining data from multiple closely spaced lines of sight. Optical backscatter results from the presence of refractive index discontinuities, which, in soft tissues, may include connective tissues, cytoplasmic organelles, cell nuclei and melanin granules [19]. However, owing to the fact that the speed of light is so much faster than that of sound, the time delay between the emission of a light pulse and detection of a reflection is too brief to measure directly. Hence, rather than using a pulse–echo technique, a low-coherence light beam (typically a superluminescent diode emitting in the near-infrared region) is split into reference (mirror) and sample (tissue) paths, and then recombined interferometrically. In time-domain OCT, the reference mirror is mechanically scanned in the range axis, and this allows determination of the range to optical reflections along the tissue path, which are represented by interference fringes in the OCT signal.

As with ultrasound, the resolution of OCT is affected by bandwidth and wavelength. OCT imaging, especially of the retina, is also affected by the optical absorption of water, which makes up most of the vitreous. Absorption is low at 820 nm, allowing this wavelength to readily probe the retina. This technique has had a deep impact on the evaluation of the retina, macula and optic nerve by providing real-time images with less than 10-μm resolution.

In the more recently developed frequency (Fourier) domain OCT, the broadband signal is broken into a spectrum using a grating or linear detector array, and depth is determined from the Fourier transform of the spectrum without motion along the reference arm [20]. This allows data along one line of sight to be acquired virtually instantaneously, with the acquisition speed of the photodetector becoming the rate-limiting factor. An alternative implementation involves the use of a frequency-swept light source [21].

The first use of OCT for anterior segment imaging was reported by Izatt and coworkers, who developed a slit-lamp-mounted, 830-nm time-domain device [22]. Spectral domain OCT imaging of the anterior segment at 1310 nm has also been described [23]. Here, the longer wavelength is advantageous in that higher light intensities may be used, improving the signal-to-noise ratio, since most light entering the eye will be absorbed by the vitreous, protecting the retina from damage [24].

The early commercial OCTs introduced by Zeiss utilized the time-domain method for imaging the retina at 820 nm in the near-infrared region. The Visante anterior segment OCT system (Carl Zeiss Meditec, Dublin, CA, USA) is also a time-domain device, but operating at 1310 nm. It supports several modes, including high-resolution cornea, corneal pachymetry and anterior segment. The latter mode consists of 256 vectors, with images acquired in 0.125 s. Resolution is 18 μm axially by 60 μm laterally. The SL-OCT (Heidelberg Engineering, GmbH, Dossenheim, Germany) combines an OCT with a Haag–Streit slit lamp for improved assessment of the iridocorneal angle [25]. The RTVue (Optoview, Inc., Fremont, CA, USA) is a frequency-domain OCT utilizing an 840-nm wavelength. The instrument, originally designed for retinal imaging, utilizes an optional lens for anterior segment imaging. The improved acquisition speed (26,000 vectors/s) and signal-to-noise ratio obtained with frequency-domain processing plus the higher resolution (5 μm axially × 15 μm laterally), obtainable at 840 nm, provides high-quality images of anterior-segment structures. A recent report also describes modification of the Cirrus HD-OCT (Carl Zeiss Meditec), a frequency-domain instrument designed for retinal imaging, with a 60-diopter lens, adapting it for imaging of the anterior segment, providing scan rate and resolution comparable to that of the RTVue [26]. The improved resolution of these 840-nm systems allows resolution of structures such as Descemet’s membrane, the trabecular meshwork and Schwalbe’s line.

Table 1 compares UBM and AS-OCT in terms of resolution and usage. Comparative AS-OCT and UBM images of the anterior segment are presented in Figure 1. In both imaging modes, there is a noticeable bowing forward of the iris plane with slight narrowing of the angles. The UBM image shows greater detail on retroiridal structures, including the ciliary body, zonules and the lens surface. Several recent studies have compared anterior-segment OCT with UBM in detail [2729]. Garcia and Rosen, using an OTI AS-OCT (Ophthalmic Technologies, Inc., Toronto, ON, Canada) concluded that UBM was advantageous for viewing structures posterior to the iris (whose pigment epithelium strongly absorbs light) and the sclera (which strongly scatters light), and for light–dark studies in glaucoma [27], as previously described by Gazzard et al. [30]. Nolan described the advantage of UBM for the detection of cyclodialysis clefts in the ciliary body, and the value of AS-OCT as a noncontact examination technique for rapid follow-up following treatment [28]. Dada et al. compared biometric parameters related to the anterior chamber and angle obtained with UBM and OCT, and found them to be comparable [29].

Figure 1
(A) Comparative Artemis – ultrasound biomicroscopy and (B) Visante – optical coherence tomography images of the right eye (undilated) of a 57-year-old male
Table 1
Comparative properties of anterior segment 840-nm frequency-domain optical coherence tomography and 50 MHz ultrasound biomicroscopy systems.

Clinical application of UBM & AS-OCT for glaucoma diagnosis & management

Pavlin and coworkers carried out the first clinical UBM studies of the anterior segment in glaucoma in the early 1990s [31,32]. Pavlin and others demonstrated the utility of UBM in characterizing several forms of glaucoma, including plateau iris syndrome [33] and pupillary block [34], which together constitute the most common forms of primary angle-closure glaucoma [35].

In pupillary block (Figures 2 & 3), pressure in the posterior chamber is elevated relative to that of the anterior chamber owing to impairment of the flow of aqueous through the pupil. This results in forward bowing of the iris from the root to the pupil margin in the presence of a formed posterior chamber. Laser peripheral iridotomy will result in iris flattening by relieving the anterior–posterior chamber pressure differential. UBM has been shown to be of value in elucidating the etiology of pupillary block glaucoma [34,36,37]. UBM can also be useful for the diagnostic imaging of malignant glaucoma (Figure 4), which is characterized by a forward movement of the iris–lens diaphragm and a very shallow anterior chamber, typically following glaucoma surgery, chronic angle-closure glaucoma or pseudoexfoliation [38]. The primary consideration in the differentiation of pupillary block and malignant glaucoma is the presence of a formed posterior chamber in the former, which can be demonstrated by UBM.

Figure 2
Ultrasound biomicroscopy image of relative pupillary block
Figure 3
A 52-year-old Hispanic male with neovascular glaucoma and uncontrolled diabetes
Figure 4
Ultrasound biomicroscopy images of malignant glaucoma

In plateau iris syndrome (Figures 5 & 6), the ciliary body is anteriorly positioned and possibly enlarged, compressing the iridocorneal angle and placing the peripheral iris in apposition to the trabecular meshwork, impairing outflow [33,39]. UBM and OCT will demonstrate little iris bowing, but rather a steep rise in the iris near its point of insertion. Anterior positioning of the ciliary processes and absence of the sulcus are best visualized with UBM. In some cases, multiple ciliary body cysts, also detectable by UBM, can cause a similar effect on the angle (Figure 7) [4042]. UBM has shown that iridociliary cysts are more common than initially thought, and allows their progression to be followed over time. Although ciliary body cysts are common and generally benign, impaired outflow may result when multiple cysts further narrow an angle that is already compromised by age-related anterior-chamber shallowing [43]. UBM should be performed in all atypical cases of narrow-angle glaucoma. When an iridociliary cyst is detected, management should include UBM examination of both eyes over 360° to rule out any potential angle compromise. Patients with multiple bilateral cysts compromising the angle should be followed as glaucoma suspects. However, if less than 1 clock-hour of angle is narrowed, no follow-up is required.

Figure 5
Ultrasound biomicroscopy image of temporal quadrant in patient with plateau iris
Figure 6
Series of ultrasound biomicroscopy scans covering each quadrant in patient with plateau iris
Figure 7
Focal angle closure due to presence of iridociliary cyst (arrow) pushing the iris anteriorly.

Sihota and coworkers demonstrated the UBM to be instrumental in characterizing the anatomical features of eyes with primary angle-closure glaucoma: thinner iris, shorter trabecular ciliary process distance and narrow angle [44]. Acute primary angle-closure glaucoma seems to have the narrowest angle recess of all conditions. The major advantage of UBM remains that it provides images of the position, angulation and anatomical variants of the ciliary body and peripheral iris.

Pigment-dispersion glaucoma results from the dissemination of pigment granules from the posterior of the iris as a result of friction between this surface and the zonules and/or lens. The deposition of particles in the trabecular meshwork can impair outflow and lead to an increase in IOP. The UBM appearance of pigment dispersion glaucoma was first described by Pavlin [45]. Pigment dispersion syndrome typically demonstrates an open angle and iris concavity (Figure 8) consistent with the hypothesis that iris–zonular chafing is responsible for the dispersion of pigment particles [46,47]. To account for the concave iris geometry, it has been suggested that a reverse pupillary block is present, as the concavity can often be reversed with iridotomy [48]. Detry-Morel et al. described the combined use of OCT and UBM for the diagnostic imaging of pseudophakic secondary pigment dispersion glaucoma, using OCT to demonstrate intermittent iris–intraocular lens contact through a dilated pupil [49]. Yip et al. found the Pentacam to be useful for the diagnosis of pigment dispersion glaucoma, reporting that the posterior cornea exhibited a significantly flatter radius of curvature than in normal eyes, and that this effect increased with pigment loss [50]. He hypothesized that corneal biomechanics may play a role in the etiology of this syndrome.

Figure 8
Ultrasound biomicroscopy images of pigmentary glaucoma

Choroidal thickening, leading to increased vitreous cavity pressure, has been suggested to be a potential precipitating factor in angle closure [51]. Kumar [52] and Sakai [53] utilized UBM to document the presence of choroidal effusion in patients with primary angle-closure glaucoma and acute angle closure, both pre- and postiridotomy. Gazzard et al. were among the first to describe the association of primary acute angle closure with suprachoroidal fluid [54].

Yao et al. used UBM under dark conditions to investigate the occurrence of appositional angle closure in postiridotomy fellow eyes of patients with acute-angle closure [55]. Over one third of fellow eyes showed appositional angle closure, indicative of a narrower angle, a more anterior position of the ciliary body and a thicker peripheral iris, suggesting increased risk for progressive angle closure in the fellow eyes.

Angle-recession glaucoma may occur following ocular trauma. Comparative UBM and AS-OCT images in a case of angle recession are shown in Figure 9. Ozdal compared UBM with gonioscopy following such trauma and reported UBM to be diagnostically useful in imaging angle recession, cyclodialysis, zonular deficiency, lens dislocation and synechiae [56]. Kawana et al. described the use of a prototype 3D swept-source, 1310-nm OCT to image post-traumatic angle recession [57]. OCT is advantageous in not requiring physical contact with a traumatized eye, but UBM enables the evaluation of the ciliary body, lens and zonules. UBM is a valuable tool in the detection of intraocular foreign bodies affecting the angle or ciliary body after perforating trauma. The noninvasive nature (especially using a bubble-tip set-up) and good penetration of UBM are helpful in the accurate characterization and localization of intraocular foreign bodies, and surgical planing.

Figure 9
(A) Ultrasound biomicroscopy and (B) optical coherence tomography images in angle-recession glaucoma in horizontal plane of left eye in 45-year-old male patient with history of trauma

Ultrasound biomicroscopy and OCT are both useful tools for the planning and guidance of glaucoma surgery [58], including the evaluation of filtering blebs [59,60], sclerectomy [61,62] and canaloplasty [63,64], as well as the diagnosis and evaluation of postoperative complications. Figure 10 shows 80-MHz UBM images of the angle pre- and postcanaloplasty.

Figure 10
Two 80-MHz ultrasound biomicroscopy images of the angle pre- and 6 months postcanaloplasty

Ultrasound biomicroscopy and OCT are superior to slit lamp for the evaluation of bleb function and failure, allowing a demonstration of flattened, encapsulated and cystic avascular thin-walled blebs. In patients with filtering blebs, the AS-OCT can be performed immediately after surgery because nothing touches the eye for scanning. By contrast, the UBM procedure has some risk of bleb wall damage and infection through the eyecups used. The height, wall thickness, apposition of the scleral flap to sclera and the patency of the internal ostium can be assessed. Bleb assessment with UBM or AS-OCT can influence clinical decision-making regarding laser suture lysis following trabeculectomy [65]. AS-OCT images of filter blebs are shown in Figures 11 & 12.

Figure 11
Anterior-segment optical coherence tomography images of filter bleb in 48-year-old Hispanic female with chronic open-angle chronic glaucoma
Figure 12
Optical coherence tomography images of filter blebs post-trabeculectomy in the right (A) and left (B) eyes of an 89-year-old male

Bochmann et al. described the use of UBM for identifying narrow-diameter (<100 μm) iridotomy sites, which were then retreated based on the UBM findings [66]. Ishikawa et al. demonstrated that indentation of the cornea by a small eyecup can result in angle widening [67]. Following upon this work, Matsunaga et al. described the use of a special UBM eyecup that allowed simultaneous corneal compression with UBM observation of the angle configuration before and after compression [68]. This allowed a differentiation of appositional angle closure from synechial closure. Carillo demonstrated the usefulness of UBM to diagnose obstruction of an Ahmed valve by the iris [69]. UBM may also potentially offer intraoperative guidance in the anterior segment as a supplement to gonioscopy and intraocular microendoscopy [70].

Postpenetrating keratoplasty glaucoma is a major cause of graft failure and the most common cause of irreversible visual loss after keratoplasty. UBM allows imaging of anterior-segment anatomy in the presence of corneal opacity. The visualization of synechiae and secondary angle closure by UBM in such cases can be a valuable tool for planning filtering surgery or the implantation of drainage devices [71].

In glaucoma shunts, the AS-OCT technique allows visualization of the shunt cross-section within the anterior chamber, even, in some instances, in the presence of corneal haze, as demonstrated in Figure 13 in an eye following postpenetrating keratoplasty. By contrast, the UBM is able to evaluate glaucoma tube shunts placed beneath the sclera or iris plane (Figure 14). Rothman et al. demonstrated that UBM is instrumental in diagnosing the presence and cause of tube obstruction, as it can detect focal obstructions of Baerveldt tubes caused by kinking at the scleral entry site after pars plana insertion [72].

Figure 13
Good positioning with no endothelial touch of a shunt tube (bottom arrow) seen in cross-section with optical coherence tomography in anterior chamber of a 73-year-old patient with hazy cornea postpenetrating keratoplasty
Figure 14
Ultrasound biomicroscopy views of Molteno tube (arrows) before and after entering anterior chamber

In 1992, Pavlin described UBM biometric criteria that could be used for reproducible measurement of various anterior segment structures [31]. These included the angle opening at 250 and 500 μm from the scleral spur, scleral thickness at the spur, trabecular–ciliary process distance and iris thickness at specific positions. Tello et al. reported on the reproducibility of these measures in 1994 [73]. Ishikawa, Liebman and Ritch described further criteria in 2000, especially numeric descriptors of angle geometry [74]. These criteria are of importance in allowing the definition of reproducible criteria for characterizing different glaucoma types and for documenting change with time or with treatment. Marchini, for instance, used UBM to biometrically compare different forms of angle closure glaucoma [75,76], and Sihota et al. applied these criteria for comparing subtypes of primary angle-closure glaucoma [44]. Ramani used UBM to compare anterior-segment biometry between primary angle-closure suspects and age-matched controls, including parameters such as trabecular–ciliary process distance, iris thickness and angle width (in degrees) [77]. Nonaka et al. described the use of UBM for the measurement of iris convexity in primary angle-closure glaucoma [78]. Many criteria developed for the assessment of the anterior segment were immediately applicable to AS-OCT when this technology became available. Currently, visualization of the angle in the Visante’s high-resolution mode provides clinicians with optimal anatomical and numerical information on angle anatomy. Frequency-domain OCT provides especially good resolution, enabling the utilization of Schwalbe’s line, for instance, as a biometric landmark (Figure 15). A recent comparative UBM and AS-OCT study by Radhakrishnan et al. on 31 eyes of normal and narrow-angle subjects found that the anterior-chamber angle parameters measured by both AS-OCT and UBM had similar mean values, reproducibility and sensitivity–specificity profiles [23].

Figure 15
Narrow angle imaged with RTVue Fourier-domain optical coherence tomography with corneal adaptor module

Table 2 describes a variety of biometric parameters that have been developed for OCT and/or UBM for assessment of the angle and anterior segment in glaucoma. Some of these are illustrated in Figure 16.

Figure 16
Diagram illustrating several biometric descriptors of the angle, including angle-opening distance, iris thickness, trabecular-ciliary process distance, scleral spur–iris insertion distance and angle recess area
Table 2
Biometric parameters used in optical coherence tomography and ultrasound biomicroscopy for characterizing the angle and anterior segment (numerous other parameters beyond this list have been described).

Numerous clinical studies utilizing OCT and/or UBM have made use of these biometric parameters. Sakata et al. compared open-angle subjects with patients with occludable angles on gonioscopy following laser iridotomy [79]. Over half of occludable-angle eyes had appositional angle closure, although this was also found in approximately 20% of normal eyes. A long ciliary process with no ciliary sulcus was observed in 61% of occludable-angle eyes, and also in 32% of control eyes. Gazzard et al. described anterior segment morphology changes following laser iridotomy in Asian eyes [80]. Using UBM, Dada et al. compared changes in anterior-chamber anatomy in patients with primary angle closure and primary angle-closure glaucoma following iridotomy, demonstrating widening of the anterior-chamber angle and a deepening of the anterior chamber in eyes with primary angle closure, but no significant change in eyes with primary angle-closure glaucoma [81]. Dada also described the UBM changes occurring during a Valsalva maneuver: narrowing of the anterior-chamber angle recess, thickening of the ciliary body and increased iris thickness. In eyes anatomically predisposed to primary angle closure, the Valsalva maneuver may lead to angle closure. The lack of response to iridotomy in primary angle-closure glaucoma is attributed to the closure of the angle by extensive peripheral anterior synechiae (PAS). Yoo et al. compared anterior segment biometric parameters in angle-closure suspect eyes with and without PAS [82]. They report that the trabecular–ciliary process distance was significantly shorter in the superior quadrant of PAS eyes when compared with the eyes without PAS, suggesting that the shorter distance from trabecular meshwork to the ciliary body or the anterior placement of ciliary process may play a role in the development of PAS in angle-closure glaucoma suspect eyes. Also studying the effect of iridotomy on angle-closure suspect eyes, He et al. found iridotomy to result in a significant increase in angle width, but with some iridotrabecular contact in 59% of eyes with a patent iridotomy [83]. This was associated with smaller angle dimensions and a thicker iris, both of which may play a causative role in maintaining angle closure after iridotomy. Kaushik et al. compared UBM and gonioscopy in evaluating changes in angle anatomy following laser iridotomy [84]. They report that the angle significantly widened in the quadrant with iridotomy and the quadrant furthest away in patients of chronic angle closure with established glaucomatous damage. This change was much better appreciated by UBM than gonioscopy. OCT has also recently been applied for assessment of the angle and iridotomy. In 2005, Chalita et al. described the use of a 1300-nm prototype OCT system for the evaluation of a patient with bilateral occludable angle closure pre- and postiridotomy [85]. They demonstrated the iridotomy site as well as an increase in the trabecular–iris area following treatment. More recently, Lei et al. used the Visante to study the effects of iridotomy in primary angle closure, concluding that iridotomy leads not only to an increase in peripheral anterior chamber depth and anterior chamber volume, but also to an increase in central anterior-chamber depth [86].

Traditional methods to assess the iridocorneal angle and iris include gonioscopy exam and the van Herick method. Gonioscopy requires a diagnostic contact lens; also, the angle grading is subjective, depending on the good visualization of specific angle structures. The Van Herick technique assesses peripheral anterior chamber depth in relation to corneal thickness. By contrast, both AS-OCT and the UBM provide angle visualization. AS-OCT can be employed as a screening tool for primary angle-closure glaucoma, as it is a noncontact procedure and is less dependent on examiner skills than UBM.

Limitations of the UBM include the requirement of a coupling medium and the patient being scanned in a supine position, which might theoretically lead the iris diaphragm to fall back and change the depth of the anterior chamber and the angle opening. Ishikawa et al. demonstrated that inadvertent pressure on the eyecup while scanning can influence the angle configuration [74]. Another limitation is represented by the difficulty in pinpointing the exact location of the UBM scan plane (as there is no fixed reference point for the handheld probe), so the angle region measured is located and referred to as nasal, temporal, superior, inferior and not in exact degrees of an arc (the Artemis system is an exception to this, as all UBM images obtained with this system contain embedded information regarding the orientation of scan planes and position of each vector). Compared with OCT, UBM might be more time consuming and require a skilled operator to obtain high-quality precision images. Nevertheless, these limitations are outweighed by the benefit of UBM for visualizing the ciliary body, zonules and posterior chamber, making it an essential tool in defining the mechanism of closure in angle-closure glaucomas. UBM remains the gold standard in cases of plateau iris configuration and imaging of ciliary processes. Its accuracy and ability to visualize behind a clouded cornea makes it very useful in the preoperative assessment of anterior-segment pathology, contributing to optimal surgical planning. The UBM can be used to perform a darkroom provocative test, which elicits information on whether the angle automatically closes in dark conditions. In plateau iris, UBM typically demonstrates an anteriorly directed ciliary body, no ciliary sulcus, a central flat iris plane and commonly iridocorneal contact. UBM is the superior imaging device for diagnosing cyclodialysis clefts (Figure 17) and anterior suprachoroidal effusions. The clefts are not always visible on gonioscopy, but UBM can confirm the diagnosis and estimate the extent of involvement.

Figure 17
Ciliary body separation (arrow) and cyclodialysis cleft demonstrated by ultrasound biomicroscopy in an 89-year-old pseudophakic veteran

Expert commentary

Ultrasound biomicroscopy and OCT can both provide detailed views of anterior segment structures associated with impaired outflow. OCT is advantageous, especially in the case of spectral-domain systems, owing to its superb resolution, high speed and noninvasive character, which allows it to be performed in the immediate postoperative period. UBM is advantageous in that it provides better penetration through opaque or cloudy media than OCT, allowing improved depiction of the ciliary body, retro-iridal structures and the anterior chamber in the presence of corneal edema, scars or hyphema. While neither technique is a substitute for gonioscopy, both permit an objective, quantitative description of angle anatomy using a variety of standard biometric parameters for the screening and management of glaucoma.

Five-year view

Several new technologies are being developed for both OCT and UBM. The UBM system being developed by Arcscan, Inc. employs compound scanning for improved depiction of the lens capsule, zonules and other anterior segment structures (Figure 18). High-frequency annular arrays have been developed for improved sensitivity and depth of field [87]. Linear arrays have also been developed, enabling scanning without mechanical motion of the probe [88]. VisualSonics, Inc. (Toronto, ON, Canada) recently introduced a UBM for small animal research with linear arrays of up to 70 MHz providing up to 1000 frames/s. This relative speed increase compared with current single-element mechanically scanned UBM systems is comparable to that of Fourier over time-domain OCT systems. 40-MHz pulsed Doppler has been shown to be applicable to the eye within safe exposure limits [89] and this may offer an avenue for assessing ciliary body perfusion in glaucoma. OCT systems continue to advance in speed, resolution and penetration. This in combination with eye tracking enables the 3D reconstruction of anterior-segment structures [57,60,90]. OCT birefringence imaging of ocular tissues has been introduced [91,92]. This utilizes the alteration of light polarization by fibrous tissue structures to noninvasively detect alterations in such tissues. Yasuno et al. recently described the use of 3D polarization-sensitive OCT to assess changes in the anterior segment following glaucoma surgery (trabeculectomy blebs and laser iridotomy sites) [93]. OCT scanning speeds of over 300,000 vectors/s have been achieved [94,95]. These continuing advances in UBM and OCT technology will be translated from the laboratory to commercial instrumentation in the coming years, offering increasing diagnostic capabilities for the assessment of the anterior segment and management of glaucoma.

Figure 18
Anterior segment image of normal anterior segment obtained with Artemis-3 prototype

Learning objectives

Upon completion of this activity, participants should be able to:

  • Describe the characteristics of optical coherence tomography (OCT)
  • Describe the disadvantages of scanning Scheimpflug photography in the investigation of glaucoma
  • List types of glaucoma for which ultrasound biomicroscopy (UBM) is particularly useful
  • Identify cases of glaucoma suitable for OCT or UBM
  • Identify the uses of OCT and UBM in glaucoma management

Key issues

  • Both optical coherence tomography (OCT) and ultrasound biomicroscopy (UBM) provide real-time cross-sectional images of the angle and anterior segment.
  • Standard biometric descriptors of angle and anterior segment anatomy allow a quantitative approach to glaucoma diagnosis and management.
  • OCT advantages versus UBM:
    • – Higher resolution
    • – Faster scan rate
    • – Noncontract method, allowing immediate postoperative use
  • UBM advantages versus OCT:
    • – Ability to image in the presence of cloudy or opaque media
    • – Ability to visualize ciliary body, anterior vitreous, posterior chamber and zonules

Acknowledgments

We wish to acknowledge the generous support of Janet and Barry Lang (Roxana Ursea), and that of the Dyson Foundation (Ronald H Silverman).

Footnotes

For reprint orders, please contact moc.sweiver-trepxe@stnirper

Financial & competing interests disclosure

Editor

Elisa Manzotti, Editorial Director, Future Science Group, London, UK.

Disclosure: Elisa Manzotti has disclosed no relevant financial relationships.

CME Author

Désirée Lie, MD, MSEd

Clinical Professor, Family Medicine, University of California, Irvine, Orange, CA, USA; Director of Research and Patient Development, Family Medicine, University of California, Irvine, Medical Center, Rossmoor, CA, USA

Disclosure: Désirée Lie has disclosed the following relevant financial relationship: she has served as a nonproduct speaker for ‘Topics in Health’ for Merck Speaker Services.

Authors and Credentials

Roxana Ursea, MD, MSEd

Assistant Professor of Ophthalmology; Director, Cornea and Refractive Surgery Division, Department of Ophthalmology, University of Arizona, Tucson, AZ, USA

Disclosure: Roxana Ursea, MD, has disclosed no relevant financial relationships.

Ronald H Silverman, PhD

Professor, Department of Ophthalmology, Weill Cornell Medical College, NY, USA

Disclosure: Ronald H Silverman, PhD, has disclosed a financial interest in ArcScan, Inc., and has partly been funded by NIH grant EB000238.

Contributor Information

Roxana Ursea, Assistant Professor of Ophthalmology, Director, Cornea and Refractive Surgery Division, Department of Ophthalmology, University of Arizona, 655 N. Alvernon Way, Suite 108, Tucson, AZ 85711, USA, Tel.: +1 520 322 3800 ext. 204, Fax: +1 520 321 3665.

Ronald H Silverman, Professor, Department of Ophthalmology, Weill Cornell Medical College, 1300 York Ave, Room LC303, NY 10065, USA, Tel.: +1 212 746 6106, Fax: +1 212 746 8101 and FL Lizzi Center for Biomedical Engineering, Riverside Research Institute, 156 William Street, 9th floor, NY 10038, USA, Tel.: +1 212 502 1748, Fax: +1 212 502 1729.

References

Papers of special note have been highlighted as:

• of interest

1. Jaffe GJ, Caprioli J. Optical coherence tomography to detect and manage retinal disease and glaucoma. Am J Ophthalmol. 2004;137:156–169. [PubMed]
2. Medeiros FA, Zangwill LM, Bowd C, Vessani RM, Susanna R, Jr, Weinreb RN. Evaluation of retinal nerve fiber layer, optic nerve head, and macular thickness measurements for glaucoma detection using optical coherence tomography. Am J Ophthalmol. 2005;139:44–55. [PubMed]
3. Schuman JS, Hee MR, Puliafito CA, et al. Quantification of nerve fiber layer thickness in normal and glaucomatous eyes using optical coherence tomography. Arch Ophthalmol. 1995;113:586–596. [PubMed]
4. Greaney MJ, Hoffman DC, Garway-Heath DF, Nakla M, Coleman AL, Caprioli J. Comparison of optic nerve imaging methods to distinguish normal eyes from those with glaucoma. Invest Ophthalmol Vis Sci. 2002;43:140–145. [PubMed]
5. Sakata LM, DeLeon-Ortega J, Sakata V, Girkin CA. Optical coherence tomography of the retina and optic nerve – a review. Clin Exp Ophthalmol. 2009;37:90–99. [PubMed]
6. Shaffer RN. A new classification of the glaucomas. Trans Am Ophthalmol Soc. 1960;58:219–225. [PMC free article] [PubMed]
7. Rabsilber TM, Khoramnia R, Auffarth GU. Anterior chamber measurements using Pentacam rotating Scheimpflug camera. J Cat Refr Surg. 2006;32:456–459. [PubMed]
8. Friedman DS, He M. Anterior chamber angle assessment techniques. Surv Ophthalmol. 2008;53:250–273. [PubMed]
9. Kashiwagi K, Kashiwagi F, Toda Y, Osada K, Tsumura T, Tsukahara S. A newly developed peripheral anterior chamber depth analysis system: principle, accuracy, and reproducibility. Br J Ophthalmol. 2004;88:1030–1035. [PMC free article] [PubMed]
10. Mundt GH, Hughes WF. Ultrasonics in ocular diagnosis. Am J Ophthalmol. 1956;42:488–498. [PubMed]
11. Baum G, Greenwood I. The application of ultrasonic locating techniques to ophthalmology – part 2. Ultrasonic visualization of soft tissues. Arch Ophthalmol. 1958;60:263–279. [PubMed]
12. Sherar MD, Foster FST. The design and fabrication of high frequency poly(vinylide fluoride) transducers. Ultrason Imag. 1989;11:75–94. [PubMed]
13. Foster FS, Harasiewicz KA, Sherar MD. A history of medical and biological imaging with polyvinylidene fluoride (PVDF) transducers. IEEE Trans Ultrason Ferro Freq Contr. 2000;47:1363–1371. [PubMed]
14. Pavlin CJ, Sherar MD, Foster FS. Subsurface ultrasound microscopic imaging of the intact eye. Ophthalmology. 1990;97:244–250. [PubMed]
15 •. Pavlin CJ, Harasiewicz K, Sherar MD, Foster FS. Clinical use of ultrasound biomicroscopy. Ophthalmology. 1991;98:287–295. Classic paper describing ultrasound biomicroscopy (UBM) imaging of the eye. [PubMed]
16. Reinstein DZ, Silverman RH, Coleman DJ. High-frequency ultrasound measurement of the thickness of the corneal epithelium. Refract Corn Surg. 1993;9:385–387. [PubMed]
17 •. Huang D, Swanson EA, Lin CP, et al. Optical coherence tomography. Science. 1991;254:1178–1181. Describes optical coherence tomography (OCT) imaging of the retina. [PubMed]
18. Hee MR, Izatt JA, Swanson EA, Huang D, Schumn JS, Lin CP. Optical coherence tomography of the human retina. Arch Ophthalmol. 1995;113:325–332. [PubMed]
19. Tuchin V. Light Scattering Methods and Instruments for Medical diagnosis. 2. SPIE Press; WA, USA: 2007. Tissue optics.
20. Fercher AF, Hitzenberger CK, Kamp G, El-Zaiat SY. Measurement of intraocular distances by backscattering spectral interferometry. Optics Comm. 1995;117:43–48.
21. Lee ECW, de Boer JF, Mujat M, Lim H, Yun SH. In vivo optical frequency domain imaging of human retina and choroid. Optics Exp. 2006;14:4403–4411. [PubMed]
22 •. Izatt JA, Hee MR, Swanson MS, et al. Micrometer scale resolution imaging of the anterior eye in vivo with optical coherence tomography. Arch Ophthalmol. 1994;112:1584–1589. Early paper describing anterior-segment OCT. [PubMed]
23. Radhakrishnan S, Rollins AM, Roth JE, et al. Real-time optical coherence tomography of the anterior segment at 1310 nm. Arch Ophthalmol. 2001;119:1179–1185. [PubMed]
24 •. Radhakrishnan S, Goldsmith J, Huang D, et al. Comparison of optical coherence tomography and ultrasound biomicroscopy for detection of narrow anterior chamber angles. Arch Ophthalmol. 2005;123:1053–1059. Useful review comparing UBM versus OCT for assessment of the angle. [PubMed]
25. Wirbelauer C, Karandish A, Haberle H, Pham DT. Noncontact goniometry with optical coherence tomography. Arch Ophthalmol. 2005;123:179–185. [PubMed]
26 •. Wong H-T, Lim CM, Sakata LM, et al. High-definition optical coherence tomography imaging of the iridocorneal angle of the eye. Arch Ophthalmol. 2009;127:256–260. Impressive images of the angle made with high-definition (3–5-μm axial resolution) OCT. [PubMed]
27. Garcia JP, Jr, Rosen RB. Anterior segment imaging: optical coherence tomography versus ultrasound biomicroscopy. Ophthal Surg Laser Imag. 2008;39:476–484. [PubMed]
28. Nolan W. Anterior segment imaging: ultrasound biomicroscopy and anterior segment optical coherence tomography. Curr Opin Ophthalmol. 2008;19:115–121. [PubMed]
29. Dada T, Sihota R, Gadia R, Aggarwal A, Mandal S, Gupta V. Comparison of anterior segment optical coherence tomography and ultrasound biomicroscopy for assessment of the anterior segment. J Cat Ref Surg. 2007;33:837–840. [PubMed]
30. Gazzard G, Foster PJ, Friedman DS, Khaw PT, Seah SKL. Light to dark physiological variation in irido–trabecular angle width. Br J Ophthalmol. 2004;88:1357–1482. [PubMed]
31 •. Pavlin CJ, Harasiewicz K, Foster FS. Ultrasound biomicroscopy of anterior segment structures in normal and glaucomatous eyes. Am J Ophthalmol. 1992;113:381–389. Describes what has become the standard biometric parameter set for characterization of the angle. [PubMed]
32. Pavlin CJ, Foster FS. Ultrasound biomicroscopy in glaucoma. Acta Ophthalmologica. 1992;204(Suppl):7–9. [PubMed]
33. Pavlin CJ, Ritch R, Foster FS. Ultrasound biomicroscopy in plateau iris syndrome. Am J Ophthalmol. 1992;113:390–395. [PubMed]
34. Aslanides IM, Libre PE, Silverman RH, et al. High frequency ultrasound imaging in pupillary block glaucoma. Br J Ophthalmol. 1995;79:972–976. [PMC free article] [PubMed]
35. Mandell MA, Pavlin CJ, Weisbrod DJ, Simpson ER. Anterior chamber depth in plateau iris syndrome and pupillary block as measured by ultrasound biomicroscopy. Am J Ophthalmol. 2003;136:900–903. [PubMed]
36. Satish S, MacKinnon JR, Atta HR. Role of ultrasound biomicroscopy in managing pseudophakic pupillary block glaucoma. J Cat Refr Surg. 2000;26:1836–1838. [PubMed]
37. Kobayashi H, Hirose M, Kobayashi K. Ultrasound biomicroscopic analysis of pseudophakic pupillary block glaucoma induced by Soemmering’s ring. Br J Ophthalmol. 2000;84:1142–1146. [PMC free article] [PubMed]
38. Trope GE, Pavlin CJ, Bau A, et al. Malignant glaucoma. Clinical ultrasound biomicroscopic features. Ophthalmology. 1994;101:1030–1035. [PubMed]
39. Kumar RS, Baskaran M, Chew PTK, et al. Prevalence of plateau iris in primary angle closure suspects. Ophthalmology. 2008;115:430–434. [PubMed]
40. Tanihara H, Akita J, Honjo M, Honda Y. Angle closure caused by multiple, bilateral iridociliary cysts. Acta Ophthalmol Scand. 1997;75:216–217. [PubMed]
41. Kuchenbecker J, Motschmann M, Schmitz K, Behrens-Baumann W. Laser iridocystotomy for bilateral acute-angle closure glaucoma secondary to iris cysts. Am J Ophthalmol. 2000;129:391–393. [PubMed]
42. McWhae JA, Rinke M, Crichton AC, Van Wyngaarden C. Multiple bilateral iridociliary cysts: ultrasound biomicroscopy and clinical characteristics. Can J Ophthalmol. 2007;42:268–271. [PubMed]
43. Katsimpris JM, Petropoulos IK, Sunaric-Mégevand G. Ultrasound biomicroscopy evaluation of angle closure in a patient with multiple and bilateral iridociliary cysts. Klin Monatsbl Augenheilkd. 2007;224:324–327. [PubMed]
44 •. Sihota R, Dada T, Gupta R, Lakshminarayan P, Pandey RM. Ultrasound biomicroscopy in the subtypes of primary angle-closure glaucoma. J Glaucoma. 2005;14:387–391. Very useful paper describing comparative differences in UBM-determined biometric parameters between normal eyes, open-angle glaucoma (subacute and acute) and chronic primary angle-closure glaucoma. [PubMed]
45. Pavlin CJ, Macken P, Trope GE, et al. Accommodation and iridotomy in the pigment dispersion syndrome. Ophthalmic Surg Lasers. 1996;27:113–120. [PubMed]
46. Potash SD, Tello C, Liebmann J, Ritch R. Ultrasound biomicroscopy in pigment dispersion syndrome. Ophthalmology. 1994;101:332–339. [PubMed]
47. Kanadani FN, Dorziraj S, Langlich AM, et al. Ultrasound biomicroscopy in asymmetric pigment dispersion syndrome and pigmentary glaucoma. Arch Ophthalmol. 2006;124:1573–1576. [PubMed]
48. Karickhoff JR. Pigment dispersion syndrome and pigmentary glaucoma: a new mechanism concept, a new treatment, and a new technique. Ophthalmic Surg. 1992;23:269–277. [PubMed]
49. Detry-Morel ML, van Acker E, Pourjavan S, Levi N, De Potter P. Anterior segment imaging using optical coherence tomography and ultrasound biomicroscopy in secondary pigmentary glaucoma associated with in-the-bag intraocular lens. J Cataract Refract Surg. 2007;32:1866–1869. [PubMed]
50. Yip LW, Sothornwit N, Berkowitz J, Mikelberg FS. A comparison of intraocular differences in patients with pigment dispersion syndrome. J Glaucoma. 2009;18:1–5. [PubMed]
51. Quigley HA, Friedman DS, Congdon NG. Possible mechanisms of primary angle closure and malignant glaucoma. J Glaucoma. 2003;12:167–180. [PubMed]
52. Kumar RS, Quck D, Lee KY, et al. Confirmation of the presence of uveal effusion in Asian eyes with primary angle-closure glaucoma. Arch Ophthalmol. 2008;126:1647–1651. [PubMed]
53. Sakai H, Morine-Shinjyo S, Shinzato M, Nakamura Y, Sakai M, Sawaguchi S. Uveal effusion in primary angle-closure glaucoma. Ophthalmology. 2005;112:413–419. [PubMed]
54. Gazzard G, Friedman DS, Devereux J, Seah S. Primary acute-angle closure glaucoma is associated with supra-choroidal fluid in Chinese patients. Eye. 2001;15(Pt 3):358–360. [PubMed]
55. Yao B, Wu L, Zhang C, Wang X. Ultrasound biomicroscopic features associated with angle closure in fellow eyes of acute primary angle closure after laser iridotomy. Ophthalmology. 2009;116:444–448. [PubMed]
56. Özdal MPC, Mansour M, Deschênes J. Ultrasound biomicroscopic evaluation of the traumatized eyes. Eye. 2003;17:467–472. [PubMed]
57. Kawana K, Yasuno Y, Yatagai T, Oshika T. High-speed, swept-source optical coherence tomography: a 3D view of anterior chamber angle recession. Acta Ophthalmol Scand. 2007;85:684–685. [PubMed]
58. Yamamoto T, Sakuma T, Kitakawa Y. An ultrasound biomicroscopic study of filtering blebs after mitomycin C trabeculectomy. Ophthalmology. 1995;102:1770–1776. [PubMed]
59. Zhang Y, Wu Q, Zhang M, Song BW, Du XH, Lu B. Evaluating subconjunctival bleb function after trabeculectomy using slit-lamp optical coherence tomography and ultrasound biomicroscopy. Chin Med J. 2008;121:1274–1279. [PubMed]
60 •. Kawana K, Kiuchi T, Yasuno Y, Oshika T. Evaluation of trabeculectomy blebs using 3D cornea and anterior segment optical coherence tomography. Ophthalmology. 2009;116:848–855. Impressive application of 3D OCT for visualization of bleb internal structure. [PubMed]
61. Aptel E, Dumas S, Denis P. Ultrasound biomicroscopy and optical coherence tomography imaging of filtering blebs after deep sclerotomy with new collagen implant. Eur J Ophthalmol. 2009;19:223–230. [PubMed]
62. Khairy HA, Atta HR, Green FD, et al. Ultrasound biomicroscopy in deep sclerotomy. Eye. 2005;19:555–560. [PubMed]
63. Lewis RA, von Wolff K, Tetz M, et al. Canaloplasty: circumferential viscodilation and tensioning of Schlemm’s canal using a flexible microcatheter for the treatment of open-angle glaucoma in adults: interim clinical study analysis. J Cataract Refract Surg. 2007;33:1217–1226. [PubMed]
64. Shingleton B, Tetz M, Korber N. Circumferential viscodilation and tensioning of Schlemm canal (canaloplasty) with temporal clear corneal phacoemulsification cataract surgery for open-angle glaucoma and visually significant cataract. One-year results. J Cataract Refract Surg. 2008;34:433–440. [PubMed]
65. Singh M, Aung T, Aquino MC, Chew PT. Utility of bleb imaging with anterior segment optical coherence tomography in clinical decision-making after trabeculectomy. J Glaucoma. 2009;18:492–495. [PubMed]
66. Bochmann F, Johnson Z, Atta HR, Azuara-Blanco A. Increasing the size of a small peripheral iridotomy widens the anterior chamber angle: an ultrasound biomicroscopic study. Klin Monatsbl Augenheilkd. 2008;225:349–352. [PubMed]
67. Ishikawa H, Inazumi K, Liebmann JM, Ritch R. Inadvertent corneal indentation can cause artifactitious widening of the iridocorneal angle on ultrasound biomicroscopy. Ophthalmic Surg Laser. 2000;31:342–345. [PubMed]
68. Matsunaga K, Ito K, Esaki K, et al. Evaluation and comparison of indentation ultrasound biomicroscopy gonioscopy in relative pupillary block, peripheral anterior synechia, and plateau iris configuration. J Glaucoma. 2004;13:516–519. [PubMed]
69. Carrillo MM, Trope GE, Pavlin C, Buys YM. Use of ultrasound biomicroscopy to diagnose Ahmed valve obstruction by iris. Can J Ophthalmol. 2005;40:499–501. [PubMed]
70. Dietlein TS, Engels BF, Jacobi PC, et al. UBM-guided chamber angle surgery for glaucoma management: an experimental study. Eye. 2003;17:340–345. [PubMed]
71. Dada T, Aggarwal A, Vanathi M, et al. Ultrasound biomicroscopy in opaque grafts with post-penetrating keratoplasty glaucoma. Cornea. 2008;27:402–405. [PubMed]
72. Rothman RF, Sidoti PA, Gentile RC, et al. Glaucoma drainage tube kink after pars plana insertion. Am J Ophthalmol. 2001;132:413–414. [PubMed]
73. Tello C, Liebmann JM, Potash SD, et al. Measurement of ultrasound biomicroscopy images: intraobserver and interobserver reliability. Invest Ophthalmol Vis Sci. 1994;35:3549–3552. [PubMed]
74 •. Ishikawa H, Liebmann JM, Ritch R. Quantitative assessment of the anterior segment using ultrasound biomicroscopy. Curr Opin Ophthalmol. 2000;11:133–139. Excellent review of anterior segment biometry methods. [PubMed]
75. Marchini G. Biometric data and pathogenesis of angle closure glaucoma. Acta Orthop Scand. 2002;80(S236):13–14. [PubMed]
76. Marchini G, Pagliarusco A, Toscano A, et al. Ultrasound biomicroscopic and conventional ultrasonographic study of ocular dimensions in primary angle-closure glaucoma. Ophthalmology. 1998;105:2091–2098. [PubMed]
77. Ramani KK, Mani B, Ronnie G, Joseph R, Lingam V. Gender variation in ocular biometry and ultrasound biomicroscopy of primary angle closure suspects and normal eyes. J Glaucoma. 2007;16:122–128. [PubMed]
78. Nonaka A, Iwawaki T, Kikuchi M, Fujihara M, Nishida A, Kurimoto Y. Quantitative evaluation of iris convexity in primary angle closure. Am J Ophthalmol. 2007;143:695–697. [PubMed]
79. Sakata LM, Sakata K, Susanna R, Jr, et al. Long ciliary processes with no ciliary sulcus and appositional angle closure assessed by ultrasound biomicroscopy. J Glaucoma. 2006;15:371–379. [PubMed]
80. Gazzard G, Friedman DS, Devereux JG, Chew P, Seah SK. A prospective ultrasound biomicroscopy evaluation of changes in anterior segment morphology after laser iridotomy in Asian eyes. Ophthalmology. 2003;110:630–638. [PubMed]
81. Dada T, Mohan S, Sihota R, Gupta R, Gupta V, Pandey RM. Comparison of ultrasound biomicroscopic parameters after laser iridotomy in eyes with primary angle closure and primary angle-closure glaucoma. Eye. 2007;21:956–961. [PubMed]
82. Yoo C, Oh JH, Jung HR. Peripheral anterior synechiae and ultrasound biomicroscopic parameters in angle-closure glaucoma suspects. Korean J Ophthalmol. 2007;21:106–110. [PMC free article] [PubMed]
83. He M, Friedman DS, Ge J, et al. Laser peripheral iridotomy in eyes with narrow drainage angles: ultrasound biomicroscopy outcomes. The Liwan Eye Study. Ophthalmology. 2007;114:1513–1519. [PubMed]
84. Kaushik S, Sumar S, Jain R, Bansal R, Pandav SS, Gupta A. Ultrasound biomicroscopic quantification of the change in anterior chamber angle following laser peripheral iridotomy in early chronic primary angle-closure glaucoma. Eye. 2007;21:735–741. [PubMed]
85. Chalita MR, Li Y, Smith S, et al. High-speed optical coherence tomography of laser iridotomy. Am J Ophthalmol. 2005;140:1133–1136. [PMC free article] [PubMed]
86. Lei K, Wang N, Wang L, Wang B. Morphological changes of the anterior segment after laser peripheral iridotomy in primary angle closure. Eye. 2009;23:345–350. [PubMed]
87. Silverman RH, Ketterling JA, Coleman DJ. High-frequency ultrasonic imaging of the anterior segment using an annular array transducer. Ophthalmology. 2007;114:816–822. [PMC free article] [PubMed]
88. Hu CH, Xu XC, Cannata JM, et al. Development of a real-time, high-frequency ultrasound digital beamformer for high-frequency linear array transducers. IEEE Trans Ultrason Ferroelectr Freq Control. 2006;53:317–323. [PubMed]
89. Cucevic V, Brown AS, Foster FS. Thermal assessment of 40-MHz pulsed Doppler ultrasound in human eye. Ultrasound Med Biol. 2005;31:565–573. [PubMed]
90. Yasuno Y, Madjarova VD, Makita S, et al. 3D and high-speed swept-source optical coherence tomography for in vivo investigation of human anterior eye segments. Opt Express. 2005;13:10652–10664. [PubMed]
91. Pircher M, Gotzinger E, Findl O, et al. Human macula investigated in vivo with polarization-sensitive optical coherence tomography. Invest Ophthalmol Vis Sci. 2006;47:5487–5494. [PubMed]
92. Hitzenberger CK, Gotzinger E, Pircher M. Birefringence properties of the human cornea measured with polarization sensitive optical coherence tomography. Bull Soc Belge Ophtalmol. 2006;302:153–168. [PubMed]
93. Yasuno Y, Yamanari M, Kawana K, Oshika T, Miura M. Investigation of post-glaucoma-surgery structures by 3D and polarization sensitive anterior eye segment optical coherence tomography. Opt Express. 2009;16:3980–3996. [PubMed]
94. Huber R, Wojtkowski M, Fujimoto JG. Fourier domain mode locking (FDML): a new laser operating regime and applications for optical coherence tomography. Opt Express. 2006;14:3225–3237. [PubMed]
95. Potsaid B, Gorczynska I, Srinivasan VJ, et al. Ultrahigh speed spectral/Fourier domain OCT ophthalmic imaging at 70,000 to 312,500 axial scans per second. Opt Express. 2008;16:15149–15169. [PMC free article] [PubMed]