Evaluation of a Pumpless Lung Assist Device in Hypoxia-Induced Pulmonary Hypertension in Juvenile Piglets


Persistent pulmonary hypertension is an important cause of mortality and morbidity in term infants. The lung assist device (LAD) is a novel, pumpless, low-resistance extracorporeal oxygenator to supplement mechanical ventilation. The LAD may be associated with fewer complications compared to conventional extracorporeal membrane oxygenation. The objective is to test the feasibility and efficacy of the LAD in juvenile piglets with hypoxia-induced pulmonary hypertension. Pulmonary hypertension was acutely induced by hypoxia in six 3 – 4 week old acutely instrumented and intubated piglets. The LAD was attached between a carotid artery and jugular vein. Gas exchange and hemodynamic variables including pulmonary arterial pressure and cardiac output were measured. Successful LAD cannulation was achieved without complications in all animals. Extracorporeal shunt flow through the device averaged 18% of cardiac output. The LAD achieved oxygen delivery of 20% of total oxygen consumption. Pulmonary arterial pressure was reduced by 35% from 28±5 to 18±4 mm Hg (p<0.05) and systemic PaO2 increased by 33% from 27±2 to 36±4 mm Hg (p<0.05). Other hemodynamic variables remained stable. The novel LAD shows feasibility and efficacy in improving gas exchange and reducing pulmonary arterial pressures in a juvenile animal model of hypoxia-induced pulmonary hypertension.

Persistent pulmonary hypertension of the newborn (PPHN) is a serious clinical problem, affecting approximately 0.2% of liveborn infants, and is associated with high morbidity and mortality (1). PPHN is characterized by delayed relaxation of the pulmonary vascular bed. The high pulmonary vascular resistance causes pulmonary hypertension, leading to right-to-left shunting resulting in severe hypoxemia (2). The mainstay of management is supportive care with mechanical ventilation in addition to adjunctive therapies such as high-frequency ventilation, surfactant therapy, inhaled nitric oxide, and extracorporeal membrane oxygenation (ECMO) (36). Due to serious inherent risks, such as coagulopathy and the need for systemic anticoagulation, which predisposes to systemic and intracranial hemorrhage, ECMO is currently reserved for newborn infants with reversible pulmonary disease in whom conventional or high frequency ventilation with inhaled nitric oxide has failed. An essential element in the ECMO circuit is the mechanical pump which causes shear stress injury to blood components (710).

Advances in membrane and oxygenator technology have led to low-resistance membrane gas exchange devices that allow adequate flow even when the circuit is not driven by a pump. These pumpless extracorporeal lung assist devices (pECLA) and pumpless extracorporeal CO2 removal devices have been evaluated in animal studies (1114) and adult patients (1519). The Lung Assist Device (LAD; MC3, Ann Arbor, Michigan) was initially designed as an artificial lung serving as a bridge for lung transplantation, to be placed in series with the main pulmonary artery (20,21). The LAD could support full gas exchange, and survival up to 7 days with complete diversion of pulmonary blood flow through the LAD with maintenance of hemodynamic stability has been shown with this device (22). Modifications of the LAD design have greatly improved cardiac function and initial survival in a healthy ovine model (23). In a prospective, randomized, unblinded study, the artificial lung decreased ventilator-induced lung injury in a severe ARDS model and improved 5-day survival (24). Recently, the thoracic artificial lung was found to be suitable for 7-day attachment (25) and thirty-day respiratory support (26).

In the current study, we used an acutely instrumented 3–4 week old piglet model to evaluate the feasibility and efficacy of the LAD in improving gas exchange and reducing hypoxia-induced pulmonary arterial hypertension. We hypothesized that the LAD would improve gas exchange and reduce hypoxia-induced pulmonary hypertension in a juvenile whole animal model.


The Institutional Animal Care and Use Committee of the University of Alabama at Birmingham approved the research protocol. All animal care and handling were in accordance with the guidelines of the National Institutes of Health (27).

The LAD Design

The LAD was based on MC3’s proprietary radial flow design. The device was composed of a series of parallel microporous hollow fibers (X30 240 Membrana-Celgard, Inc., Charlotte, NC) potted at both ends to separate the gas flow inside the fibers from the blood flowing outside the fibers. The blood entered an inlet centrally located within the fiber bundle, flowed radially over the microporous hollow fibers, and then exited a tangentially placed outlet (Figure 1). The membrane surface area of the LAD used in this study was 0.5 m2. Previous bench studies of the LAD have characterized the relationship between surface area, gas exchange, blood flow rate, and sweep gas flow rate. MC3 was able to adapt the housing geometry to alter the device size and fiber bundle surface area to meet our requirements of smaller priming volume (60 ml) and gas exchange rates for the piglet model.

Figure 1
Schematic of the device design along the long axis showing blood entering centrally and flowing radially over the parallel microporous hollow fibers while the sweep gas runs inside the fibers in a counter-current fashion. (Courtesy of MC3, Ann Arbor, ...


Six 3–4 week old piglets were heavily sedated with pentobarbital, endotracheally intubated, and mechanically ventilated using an Infant Star® ventilator (Infant Star 950, Nellcor Puritan Bennett, Pleasanton, CA, USA). Animals were placed on a table with a heating pad to maintain body temperature at 37–38°C. Pentobarbital (10 mg/kg/h i.v.) and diazepam (1 mg/kg/h i.v) was used for sedation. The aorta, inferior vena cava, and pulmonary arteries were catheterized as previously described (28). Briefly, 5-Fr catheters were inserted into the thoracic aorta and inferior vena cava via the right femoral artery and right femoral vein, respectively, through a right groin incision. Following a left thoracotomy between the second and third ribs, the pericardium was opened. A 10 or 12 mm precalibrated ultrasonic flow transducer (HT110, Transonic systems, Ithaca, NY, U.S.A.) was affixed around the main pulmonary artery. A 5- Fr polyurethane catheter was then placed into the main pulmonary artery using purse-string sutures, and a similar catheter was placed into the left atrial appendage.

The left carotid artery and internal jugular vein were surgically exposed, and cannulated with low-resistance cannulae (Medtronic, Minneapolis, MN, USA) of 5 or 7 Fr and 8 or 10 Fr diameter, respectively. The LAD was primed with 60 ml of heparinized normal saline, debubbled of air, and attached to the vascular cannulae. Animals were anticoagulated with a bolus of 50 units/kg of heparin after cannulation.

Experimental Protocols

Experiments were performed on sedated piglets after stabilization following surgery. After baseline recording of hemodynamic measurements for 15 minutes and ensuring stability of the recordings, hypoxia was induced by combining air with nitrogen to attain a delivered endotracheal FiO2 of 0.10. Final oxygen concentration was measured using an oxygen analyzer placed in series with the ventilator tubing. After additional continuous recording for 15 minutes and ensuring stability of the recordings, 100 % oxygen at 1 L/min was allowed to flow through the LAD for 15 minutes. The oxygen flow through the LAD was then turned off for 15 minutes for a second baseline recording during hypoxia, followed by resumption of oxygen flow through the LAD. The duration of the experiment for each animal was limited to 2 hours, following which the animal was euthanatized with an overdose of pentobarbital.

Blood samples

Femoral arterial as well as pre- and post-LAD blood gases were monitored using a blood gas analyzer (ABL 700 series, Radiometer America Inc., Westlake, OH, USA) at the following time points: at baseline, 15 minutes after initiation of hypoxia, and 15 minutes after initiation of flow through the LAD.


Throughout the experiment, the heart rate, systemic arterial pressure (SAP), pulmonary arterial pressure (PAP), left atrial pressure (LAP), pulmonary blood flow (=cardiac output; CO) were recorded using pressure transducers (Spectramed P23XL, Oxnard, CA) attached to physiologic recorders (Gould-Brush 2400S, Oxnard, CA) and the ultrasonic flow transducer. Oxygen saturation was continuously monitored by pulse oximetry (N-100 C, Nellcor, Pleasanton, CA, USA) on a forelimb. Extracorporeal flow was continuously measured by another ultrasonic flow probe (12XL, Transonic systems, Ithaca, NY, U.S.A.) placed on the arterial side of the LAD. Pulmonary vascular resistance index (PVRI) was calculated as (PAP-LAP)/CI and systemic vascular resistance index (SVRI) was calculated as (SAP-CVP)/CI [cardiac index (CI) = cardiac output (CO) (ml/min)/weight (kg)]. The oxygen consumption (VO2) was calculated by measuring cardiac output, femoral arterial oxygen tension and saturation, femoral venous oxygen tension and saturation, and hemoglobin concentration, and calculations using the Fick equation [VO2 = CO × (CaO2 − CvO2)]. The capacity of the LAD to provide additional oxygen was calculated as the product of the flow rate through the device times the difference between the post-device oxygen content and pre-device oxygen content [oxygen transfer (ml/min) = LAD flow rate (ml/min) × difference in oxygen content as volume %].

Statistical analysis

Data were expressed as mean ± SEM and analyzed by SigmaStat v. 3.5 (Jandel Scientific, San Rafael, CA). A p value of <0.05 was considered statistically significant.

Hemodynamic variables, blood gases, pulmonary arterial pressure, cardiac output, and flow through the device before and after the experimental conditions were analyzed using ANOVA. If significant differences were found by repeated measures ANOVA, multiple comparisons by Holm-Sidak’s test (for hemodynamic variables) or Tukey’s test (for blood gases) were performed.


The six piglets were healthy 3–4 week old female piglets weighing 4–6 kg. Successful LAD cannulation was achieved in all 6 animals with no complications. Five animals remained alive throughout the experiment, and one died during the hypoxic exposure.

At baseline, all the animals were ventilated at FiO2 of 1.0 with corresponding mean PaO2 of 453 ± 21 mm Hg, while PAP and PVRI were 11 ± 1 mm Hg and 0.041 ± 0.004 mm Hg*ml/min/kg, respectively. Hypoxia significantly reduced PaO2 and PaO2/FiO2 (Table 1) with a parallel increase in PAP and PVRI (Figure 2).

Figure 2
Effect of the lung assist device (LAD) on pulmonary arterial pressure (PAP), pulmonary vascular resistance index (PVRI), mean systemic arterial pressure (SAP), and systemic vascular resistance index (SVRI). Hypoxia: piglet exposed to endotracheal 10% ...
Table 1
Blood gases changes secondary to inducing hypoxia or running sweep gas (100% O2) through the lung assist device (LAD). Hypoxia: piglet exposed to endotracheal 10% oxygen without sweep gas flow through the LAD. LAD on: piglet exposed to endotracheal 10% ...

Insertion of the LAD did not significantly affect the cardiac output. Blood flow through the device was stable throughout the experiment (Figure 3). Extracorporeal shunt flow through the LAD was proportionally related to the catheter size. It increased from 7% of CO when 5 and 7 Fr catheters were used to cannulate the carotid artery and jugular vein respectively to 20% when 8 and 10 Fr catheters were used to cannulate the same respective vessels (Figure 4). The LAD increased mean PaO2 from 37±13 (pre-device) to 501±90 mm Hg (post-device) and decreased PaCO2 from 28±8 (pre-device) to 7±3 (post-device)(p<0.05). Pre-device arterial oxygen saturation averaged 78±7%. Mean oxygen consumption VO2 was 6 ± 1 ml O2/kg/min, while the mean LAD capacity to add oxygen was 1.2 ± 0.5 ml O2/kg/min or 20% of total oxygen consumption.

Figure 3
Extracorporeal blood flow through the lung assist device (LAD) and cardiac output (CO) over time. Filled circles represent flow through the LAD. Open circles represent CO. All values are mean ± SEM.
Figure 4
Percentage of extracorporeal blood flow through the lung assist device (LAD) of cardiac output (CO) over time. All values are mean ± SEM.

The passage of 100% oxygen at 1 lpm through the LAD while endotracheal FiO2 was maintained at 0.10 reproducibly reversed the hypoxia effect on PaO2, PaO2/FiO2, PAP, and PVRI (Figure 5). Mean PAP, as well as PVRI were reduced by 35% from 28±5 to 18±4 mm Hg and from 0.147±0.030 mm Hg*ml/kg/min to 0.088±0.001 mm Hg*ml/kg/min, respectively (p<0.05). PaO2 increased by 33% from 27±2 to 36±4 mm Hg (p<0.05), and mean systemic CO2 tension (PaCO2) decreased from 35±7 to 29±8 mm Hg (p>0.05). Mean SAP decreased by a statistically insignificant 14% from 70±8 to 60±5 mm Hg, and SVRI from 0.45±0.05 mm Hg*ml/kg/min to 0.38±0.03 mm Hg*ml/kg/min (p>0.05), but no vasopressors were required. No significant changes were noted in the pH or calculated bicarbonate (Table 1).

Figure 5
Tracing of pulmonary arterial pressure in a representative piglet upon exposure to hypoxia (10% O2) followed by initiation of oxygen flow through the lung assist device (LAD). Hypoxia increased pulmonary arterial pressure, which was lowered by the LAD. ...


We evaluated the feasibility and efficacy of a pumpless low-resistance membrane oxygenator in reducing pulmonary arterial pressure and improving gas exchange in hypoxia-induced pulmonary hypertension in juvenile piglets. We intended to study the LAD under experimental conditions that resembled the clinical situation of PPHN. The animal model used has been extensively studied by our group and shown to be a reliable model for the study of pulmonary hypertension (2830).

ECMO provides for comprehensive gas exchange but is associated with many inherent complications including consumption of blood components due to the large surface area of the ECMO circuit and shear stress effects of the mechanical pump. In theory, the LAD reduces these risks due to its small surface area that limits the interaction of blood with foreign material while still providing adequate surface area for partial gas exchange, and its low resistance that eliminates the need for a mechanical pump. An increasing recent emphasis on maintenance of large surface areas and short path lengths has enabled the development of low-resistance membrane oxygenators (pressure drop through the device < 10 mm Hg) such as those used in the LAD. Gas exchange through these devices depends on sweep gas flow through the device, surface area of the hollow fiber membrane, and the extent of extracorporeal shunt. Compared to other commercially available gas exchangers, the LAD offers several advantages including significantly smaller priming volume and pressure drop, hence reducing blood damage and allowing its application in situations with less than optimal cardiac output and hemodynamic status. Due to the low resistance of the LAD and very efficient gas exchange through the hollow fiber membrane, the magnitude of the extracorporeal shunt is the most important variable in gas exchange. This is primarily determined by cardiac output and luminal diameter of the catheters used to cannulate the vessels. The amount of oxygen added to the venous blood stream depended on the blood flow rate through the device and the magnitude of desaturation of hemoglobin. The lower the oxygen saturation entering the LAD, the higher the amount of oxygen added to it, as the LAD adds a significant amount of dissolved oxygen thereby significantly increasing the oxygen content of blood leaving the device compared to the blood entering it [VO2 = flow × [(1.34×hemoglobin×SaO2) + (0.0031×PaO2)] − [(1.34× hemoglobin×SvO2) + (0.0031×PvO2)]]. These findings were consistent with the results of Zick et al. (14) who showed that oxygen delivery linearly correlated with the arterial oxygen saturation and the oxygenator blood flow.

Hemodynamically, the animals had stable heart rate and systemic arterial pressure after connecting the LAD and had sufficient and stable blood flow through the device. Under optimal in vitro experimental conditions and extrapolating from higher flow rate results, the LAD was expected to provide 2–2.5 ml O2/kg/min. In our current in vivo study, the oxygen delivery through the LAD was significantly increased by the higher shunt flow rate that was achieved with increasing catheter size. Increasing the catheter size from 5 Fr to 7 Fr internal diameter on the arterial side and from 8 Fr to 10 Fr on the venous side enabled us to improve the device oxygen delivery by six fold, up to a maximum of 40% of total oxygen consumption. Increasing the catheter sizes and hence the flow through the device to around 20% of the cardiac output was well tolerated by the animals as there was no significant change in arterial pH or other systemic hemodynamic variables. Although hemoglobin oxygen saturation is a more important determining factor in calculating blood oxygen content and hence oxygen delivery, we found that the capability of the LAD to improve oxygenation was positively affected by additional factors. First, the LAD provided a significant amount of dissolved oxygen, an addition of up to 14% compared to fully oxygen-saturated hemoglobin, as described in the previous paragraph. Secondly, although similar devices are known for their capacity to remove CO2, PaCO2 decreased only by 17% in our study. This may be due to the fact that PaCO2 at the start of the experiment was low. Still, this decrease in PaCO2 may have positively affected oxygenation by increasing pH, thereby shifting the oxygen dissociation curve to the left and improving hemoglobin oxygen saturation. The efficacy of CO2 removal would have been more pronounced if permissive hypercapnia was allowed. PAP and PVRI were significantly reduced while PaO2 and PaO2/FiO2 were significantly increased in all animals. The effect of the LAD on PAP was immediate as PaO2 consistently increased whenever sweep gas through the LAD was turned on.

Our study had several limitations. Juvenile piglets may have different hemodynamic responses to hypoxia or to extracorporeal shunts compared to newborn humans. However, our laboratory has demonstrated similar responses to hypoxia in acutely instrumented newborn piglets (2830). Additionally, as this experiment was a feasibility study over a limited period of time, coagulation variables that would be of importance over a longer duration of extracorporeal support were not evaluated. In this preparation, the use of the LAD required cannulation of the carotid artery, which has been associated with complications when used in conventional arteriovenous ECMO. Finally, the LAD depends on normal or near-normal cardiac function and hence would be much less effective in low cardiac output situations such as cardiogenic shock.

In summary, we demonstrated the feasibility and efficacy of a pumpless low-resistance hollow fiber membrane oxygenator in the management of hypoxia-induced pulmonary hypertension in juvenile piglets. Although oxygen delivery through the LAD comprised only 20% of total oxygen consumption, the contribution of the LAD to oxygenation was significant with a subsequent reduction in pulmonary vasoconstriction secondary to hypoxia. Our results may have been accentuated by the fact that the induced arterial hypoxemia was severe. However, the PaO2 values in our study are similar to those seen in the clinical situation of PPHN. Full oxygenation through the LAD would require a higher shunt flow rate that may not be feasible with the currently available technology or cardiac output. In order to reduce the risks of carotid artery cannulation, a trial of the LAD through the umbilical vessels is required. Still, this technology has potential as a simple, relatively inexpensive adjunct in the management of severe neonatal respiratory failure, perhaps reducing the need of conventional ECMO, or as a tool to be used during transport. Further studies on efficacy, safety, and biocompatibility are required before human clinical trials can be performed.


Financial Support: Funded in part by PHA 0526041H, K08 HD 046513, R01 HL092906, and a grant from the Children’s Center for Research and Innovation. No financial support from MC3 Corp. (Ann Arbor, Michigan, USA).


cardiac index
cardiac output
extracorporeal membrane oxygenation
lung assist device
persistent pulmonary hypertension of the newborn
pulmonary arterial pressure
pulmonary vascular resistance index
oxygen consumption


No conflict of interest.


1. Walsh-Sukys MC, Tyson JE, Wright LL, Bauer CR, Korones SB, Stevenson DK, Verter J, Stoll BJ, Lemons JA, Papile LA, Shankaran S, Donovan EF, Oh W, Ehrenkranz RA, Fanaroff AA. Persistent pulmonary hypertension of the newborn in the era before nitric oxide: practice variation and outcomes. Pediatrics. 2000;105:14–20. [PubMed]
2. Konduri GG. New approaches for persistent pulmonary hypertension of newborn. Clin Perinatol. 2004;31:591–611. [PubMed]
3. UK Collaborative ECMO Trial Group. UK collaborative randomised trial of neonatal extracorporeal membrane oxygenation. Lancet. 1996;348:75–82. [PubMed]
4. Kinsella JP, Truog WE, Walsh WF, Goldberg RN, Bancalari E, Mayock DE, Redding GJ, de Lemos RA, Sardesai S, McCurnin DC. Randomized, multicenter trial of inhaled nitric oxide and high frequency oscillatory ventilation in severe, persistent pulmonary hypertension of the newborn. J Pediatr. 1997;131:55–62. [PubMed]
5. Inhaled nitric oxide in full-term and nearly full-term infants with hypoxic respiratory failure. The Neonatal Inhaled Nitric Oxide Study Group. N Engl J Med. 1997;336:597–604. [PubMed]
6. Clark RH, Gertsmann DR, Jobe AH, Moffitt St, Slutsky AS, Yoder BA. Lung injury in neonates: causes, strategies for prevention, and long-term consequences. J Pediatr. 2001;139:478–486. [PubMed]
7. Zwischenberger JB, Nguyen TT, Upp JR, Jr, Bush PE, Cox CS, Jr, Delosh T. Complications of neonatal extracorporeal membrane oxygenation: collective experience from the extracorporeal life support organization. J Thorac Cardiovasc Surg. 1994;107:838–849. [PubMed]
8. Plotz FB, van Oeveren W, Bartlett RH, Wildevuur CR. Blood activation during neonatal extracorporeal life support. J Thorac Cardiovasc Surg. 1993;105:823–832. [PubMed]
9. Hirthler MA, Blackwell E, Abbe D, Doe-Chapman R, Le Clair Smith C, Goldthorn J, Canizaro P. Coagulation parameter instability as an early predictor of intracranial hemorrhage during extracorporeal membrane oxygenation. J Pediatr Surg. 1992;27:40–43. [PubMed]
10. Stallion A, Cofer BR, Rafferly JA, Ziegler MM, Ryckman FC. The significant relationship between platelet count and haemorrhagic complications on ECMO. Perfusion. 1994;9:265–269. [PubMed]
11. Iglesias M, Junglebluth P, Petit C, Matute MP, Rovira I, Martinez E, Catalan M, Ramirez J, Macchiarini P. Extracorporeal lung membrane provides lung protection than conventional treatment for severe postpneumonectomy noncardiogenic acute respiratory distress syndrome. J Thorac Cardiovasc Surg. 2008;135:1362–1371. [PubMed]
12. Nielsen ND, Kjaergaard B, Koefoed-Nielsen J, Steensen CO, Larsson A. Apneic oxygenation combined with extracorporeal arteriovenous carbon dioxide removal provides sufficient gas exchange in experimental lung injury. ASAIO J. 2008;54:401–405. [PubMed]
13. Dembinski R, Hochhausen N, Terbeck S, Uhlig S, Dassow C, Schneider M, Schachtrupp A, Henzler D, Rossaint R, Kuhlen R. Pumpless extracorporeal lung assist for protective mechanical ventilation in experimental lung injury. Crit Care Med. 2007;35:2359–2366. [PubMed]
14. Zick G, Frerichs I, Schädler D, Schmitz G, Pulletz S, Cavus E, Wachtler F, Scholz J, Weiler N. Oxygenation effect of interventional lung assist in a lavage model of acute lung injury: a prospective experimental study. Crit Care. 2006;10:R56. [PMC free article] [PubMed]
15. Flörchinger B, Philipp A, Flose A, Hilker M, Kobuch R, Rupprecht L, Keyer A, Pühler T, Hirt S, Wiebe K, Müller T, Langgartner J, Lehle K, Schmid C. Pumpless extracorporeal lung assist: a 10-year institutional experience. Ann Thorac Surg. 2008;86:410–417. [PubMed]
16. Hommel M, Deja M, von Dossow V, Diemel K, Heidenhain C, Spies C, Weber-Carstens S. Bronchial fistulae in ARDS patients: management with an extracorporeal lung assist device. Eur Respir J. 2008;32:1652–1655. [PubMed]
17. Walles T. Clinical experience with the iLA membrane ventilator pumpless extracorporeal lung-assist device. Expert Rev Med Devices. 2007;4:297–305. [PubMed]
18. Zimmermann M, Philipp A, Scmid FX, Dorlac W, Arlt M, Bein T. From Baghdad to Germany: use of a new pumpless extracorporeal lung assist system in two severely injured US soldiers. ASAIO J. 2007;53:e4–e6. [PubMed]
19. Bein T, Scherer MN, Philipp A, Weber F, Woertgen C. Pumpless extracorporeal lung assist (pECLA) in patients with acute respiratory distress syndrome and severe brain injury. J Trauma. 2005;58:1294–1297. [PubMed]
20. Fazzalari FL, Montoya JP, Bonnell MR, Bliss DW, Hirschl RB, Bartlett RH. The development of an implantable artificial lung. ASAIO J. 1994;40:M728–M731. [PubMed]
21. Lynch WR, Montoya JP, Brant DO, Schreiner RJ, Iannettoni MD, Bartlett RH. Hemodynamic effect of a low-resistance artificial lung in series with the native lungs of sheep. Ann Thorac Surg. 2000;69:351–356. [PubMed]
22. Lick SD, Zwischenberger JB, Alpard SK, Witt SA, Deyo DM, Merz SI. Development of an ambulatory artificial lung in an ovine survival model. ASAIO J. 2001;47:486–491. [PubMed]
23. Lick SD, Zwischenberger JB, Wang D, Deyo DM, Alpard SK, Chambers SD. Improved right heart function with a compliant inflow artificial lung in series with the pulmonary circulation. Ann Thorac Surg. 2001;72:899–904. [PubMed]
24. Zwischenberger JB, Wang D, Lick SD, Deyo DJ, Alpard SK, Chambers SD. The paracorporeal artificial lung improves 5-day outcomes from lethal smoke/burn-induced acute respiratory distress syndrome in sheep. Ann Thorac Surg. 2002;74:1011–1016. [PubMed]
25. Sato H, Griffith GW, Hall CM, Toomasian JM, Hirschl RB, Bartlett RH, Cook KE. Seven day artificial lung testing in an in-parallel configuration. Ann Thorac Surg. 2007;84:988–994. [PubMed]
26. Sato H, Hall CM, Lafayette NG, Pohlmann J, Padiyar N, Toomasian JM, Haft JW, Cook KE. Thirty-day, in parallel artificial lung testing in sheep. Ann Thorac Surg. 2007;84:1136–1143. [PubMed]
27. Institute of Laboratory Animal Research Commission on Life Sciences, National Research Council. The Guide for the Care and Use of Laboratory Animals. 7th edition. Washington D.C.: National Academy Press; 1997.
28. Ambalavanan N, Bulger A, Murphy-Ullrich J, Oparil S, Chen YF. Endothelin-A receptor blockade prevents and partially reverses neonatal hypoxic pulmonary vascular remodeling. Pediatr Res. 2005;57:631–636. [PMC free article] [PubMed]
29. Ambalavanan N, Philips JB, 3rd, Bulger A, Oparil S, Chen YF. Endothelin-A receptor blockade in porcine pulmonary hypertension. Pediatr Res. 2002;52:913–921. [PubMed]
30. Ambalavanan N, Bulger A, Ware J, Philips JB., 3rd Hemodynamic effects of levcromakalin in neonatal porcine pulmonary hypertension. Biol Neonate. 2001;80:74–80. [PubMed]