Fluorescence microscopy below the diffraction limit

Abstract

Fluorescence imaging with conventional microscopy has experienced numerous advances in almost every limiting factor from sensitivity to speed. But improved resolution beyond the fundamental limitation of light diffraction has been elusive until recent years. Now, techniques are available that surpass this barrier and improve resolution up to 10 times over that of conventional microscopy. This chapter provides an overview of these new “super-resolution” imaging methods.

Keywords: STED, SPIM, DLSM, SIM, PALM, STORM, photoactivation, photoswitching

1. Introduction

A basic property of light, diffraction, is a major obstacle for optical microscopy in resolving objects located less than ~one-half wavelength of light from each other. When light is transmitted through an aperture, diffraction disperses it in a radial pattern and produces a wavefront consisting of a series of high and low amplitudes instead of simply maintaining the shape and size of the aperture (Fig. 1A). [1] During transmission through an objective lens, diffraction limits the minimum size of the focal point and results in a three-dimensional point spread function (PSF) of the light. If viewed from the side, the PSF appears as an ellipse with the long axis in the same axis as the transmission. (Fig. 1B) The ellipse has “wings” above and below which also contribute to the diffraction limitation. When the PSF is projected onto a two-dimensional surface, it shows the familiar circular Airy diffraction pattern (Fig. 1C and 1D), which has a center region of high amplitude surrounded by rings of decreasing and increasing amplitude. By not focusing to a point, features located within the central spot remain unresolved and this defines the limitation to resolution in optical imaging.

Figure 1
Diffraction limits the resolving power of light microscopy

The sizes of the PSF and the Airy diffraction pattern are dependent on several parameters, but the most important are the Numerical Aperture (NA), which is the maximum angle of light collection achieved by the objective, and the wavelength of the light (λ) used for imaging. The relationship between these two factors and resolution is described in the equation, d = 1.22λ/2NA, where d is the diameter of the focused spot. Several criteria can be used to define optical resolution, the most common is the Rayleigh limit, which states that two point source objects must be separated by a distance greater than or equal to the distance from the center of the Airy disk to the first minimum of the pattern. Once two structures or molecules are too close together, their individual PSFs cannot be distinguished and they are hence not resolved. (Fig. 1E) The following discusses several techniques that have been able to circumvent this barrier and how they've been implemented in cell and developmental biology. Table 1 also lists several other methods that are not discussed, but have played critical roles in the field of sub-diffraction limited imaging.

Table 1
High-resolution imaging techniques for cell and developmental biology*

2. Image Interference Microscopy (I2M), Incoherent Interference Illumination Microscopy (I3M), and Incoherent Interference Illumination Image Interference Microscopy (I5M)

The InM techniques are interference based techniques using juxtaposed objective lenses focused on the same plane [2]. I2M collects the emission through both objectives, recombines the signals into the same light path, and superimposes the signals on a CCD camera. Since the two light paths are equal in length, interference of the signals produces an interference pattern. A series of images collected at different focal planes (~36 nm steps) can be processed to extract the spatial information in the axial direction. Similar to Standing Wave Fluorescence Microscopy (SWFM), I3M uses illumination through both objectives to produce excitation patterns of nodes and anti-nodes within the focal plane where the beams are constructively interfering [3]. Only the axial resolution is improved since the specimen is evenly illuminated over the entire field of view but selectively excited in axial subsections. I5M is simply a combination of the two and can achieve axial resolutions 3.5-fold better than confocal and 7-fold better than widefield microscopy. [2] All of the images are collected from a large field of view, so the data acquisition per frame is much more rapid than a typical point scanning technique. However, to maintain the sampling frequency dictated by the Nyquist criteria of ~two measurements per resolution limitation, z-sections must be taken at ~35-45 nm intervals. Thus, collection of a set of images throughout an entire specimen can still require several minutes. [2]

3. 4Pi microscopy

4Pi microscopy [4] also uses juxtaposed objectives, but unlike I5M, convergence of the excitation light occurs at a common focal point rather than the focal plane. This results in constructive interference which reduces the possible axial resolution down to ~100 nm and generates point-spread functions ~1.5-fold sharper in the lateral directions and 7-fold sharper in the axial direction compared to confocal imaging [4]. Scanning pixel by pixel makes the image acquisition slow (several minutes), but multi-focal multi-photon excitation and detection using CCD cameras can increase the image speed to that of confocal imaging [5, 6].

In addition to imaging actin fibers at ~100 nm resolution, 4Pi microscopy has been used to image mitochondria in yeast and measure alterations in the mitochondria size and shape of the yeast as they were grown on various carbon sources [5]. Work with the Golgi apparatus monitored the organelle shape, enzyme localization [6], and transport of secretory cargo through the organelle [7]. In addition, coupling 4Pi with correlative electron microscopy produced a potentially revolutionary approach to the study of an organelle with very high interest [7].

4. Microscopy techniques using multiple axes

Confocal theta microscopy (CTM) [8] also relies on the use of two objectives focused on the same point in a specimen, but their axes are arranged perpendicular to each other. Excitation occurs through one objective and emission is collected through the second. This approach takes advantage of the more narrow point spread functions in the lateral direction than in the axial direction. A related technique, Selective Plane Illumination Microscopy (SPIM) [9], uses excitation light focused to a sheet with a cylindrical lens instead of a diffraction-limited spot. The illumination sheet excites an entire plane of a specimen at a time and with the emission objective focused on the same plane, no out-of-focus light is collected because none is created.

A recent improvement in SPIM, digital scanned light sheet fluorescence microscopy (DSLM) [10], produces the illumination sheet with a rapidly scanned beam rather than a cylindrical lens. It has the advantages of homogeneous illumination throughout the light sheet, reduced aberrations, and increased illumination efficiency. SPIM and DSLM are not regarded as super-resolution techniques since the highest NA objective lenses are not used due to their shortened working distances, but the axial resolution is improved over the same objective lens used in a single axis configuration. They are included here because they represent astounding progress in instrumentation design for imaging early developmental processes.

The most prominent examples of SPIM and DSLM imaging were performed on large, living specimen (> 1mm), which present challenges to all fluorescence imaging techniques. A living Medaka (Oryzias latipes) embryo which expressed GFP in somatic and smooth muscles was imaged as proof of principle [9] at ~6 μm axial resolution and ~500 μm depth. These studies entailed three days of embryo imaging including reconstruction of the beating heart. DSLM extended these studies to imaging the first 24 hours of zebrafish development using a nuclear marker, H2B-GFP [10]. Stacks of 400 images were collected at time intervals of ~1 minute in some experiments with resolutions of 300 nm lateral and 1 μm axial. For these experiments, Keller et. al. developed “digital embryos” with ~55 million nuclear positions recorded over a 24 hour period that provide quantifiable information on the spatio-temporal dynamics of cell division and movement in the early embryo. The power of DSLM for developmental biology is evident from the tracking of ~92% of the cells in these embryos, which allows global cell population analysis in addition to study of local areas of cell behavior and differentiation.

5. Stimulated Emission Depletion (STED) microscopy

Stimulated Emission Depletion (STED) microscopy relies on the depletion of the fluorophore's excited state to reduce the detection PSF [11]. (Fig. 2A-C) The sample is raster scanned with two focused laser beams. One of the laser beams (Fig. 2A) excites the fluorophore near its excitation maximum, which is referred to as the excitation laser here. Another laser beam (Fig. 2B) is chosen with a wavelength that falls within the fluorophore emission spectrum, which is discussed here as the STED laser. The illumination sequence is an excitation laser pulse to excite a fluorophore followed immediately by a STED laser pulse of sufficient intensity to saturate and efficiently deplete a subpopulation of molecules in the excited state. The STED pulse is long enough (picosecond scale) to allow relaxation of higher order vibrational energy states to the first excited state and is fast enough to sufficiently deplete the excited state before the molecules in the STED irradiation zone can fluoresce. The power of the excitation beam is approximately that of a typical confocal microscope imaging experiment (103 - 104 W/cm2), but the STED beam power intensity must be high to maintain saturated depletion of the excited state within the STED “doughnut” (106 - 109 W/cm2).

Figure 2
Sub-diffraction limited fluorescence imaging techniques

To deplete only a subpopulation of molecules in the PSF, the STED beam must be focused with a zero node, a region of destructively interfering STED light waves in the center of its PSF that does not depopulated the excited state. Lateral resolution (Fig. 2C) but not axial resolution is improved by creating a lateral zero node in the middle of the STED PSF (Fig. 2B) [12]. To increase resolution in the axial direction, molecules in the excited state are depleted at the top and bottom of a PSF [13]. A combination of the axial and lateral zero nodes reduces the effective PSF of the STED microscope to a sphere of ~40 nm [14].

Most STED publications have used fixed and immunostained specimen [15], although fluorescent protein based imaging has also been shown [16]. Prominent STED examples include imaging of neurons, specifically synaptic vesicle [17] and plasma membrane localized proteins [18, 19]. Generally, imaging was performed at ~60 nm resolution and was simply aimed at obtaining more information about the protein localization within the synapse [20, 21] or within snare clusters in the plasma membrane.

The vastly improved resolution of STED comes with the slow image acquisition caveat associated with raster scanning techniques. This arises from scanning over a field of interest to build the image pixel by pixel coupled with the necessity to maintain a dwell time high enough for discrimination of signal from background at each pixel (~0.3 msec in many experiments) and the small pixel size requirements dictated by the Nyquist criterion. Thus, to image a single plane of 1×1 μm region of a cell of interest using 15×15 nm pixels, an acquisition time of ~1.4 seconds is required [17]. Extend this to a larger 10×10 μm region of a cell and the total time will be many minutes.

The acquisition was improved to video rate (28 frames/sec) to observe the movement of synaptic vesicles with millisecond time resolution and ~60 nm x-y spatial resolution in various parts of an axon. The video rate imaging was accomplished by reducing the field of view to 2.8 × 1.8 μm, decreasing the pixel dwell time to ~3.8 μsec, and increasing the excitation and STED illumination by ~100-1000 fold [22]. Such high irradiation power intensities may lead to increased photobleaching and perhaps to less cell viability. Although cell viability problems were not observed when using 400 MW/cm2 to produce live cell movies of dendritic spines at ~20-40 second time intervals and 25×25 μm frames [23].

6. Structured illumination microscopy (SIM)

Fluorescence emission patterns depend on the spatial distribution of the fluorophores as well as the pattern of the excitation light. By controlling the excitation pattern, high frequency spatial information of the signal generating molecules can manifest as Moiré patterns in the signals [24]. Examples of Moiré patterns are shown in figure 2 in which sets of parallel lines representing specimen fluorophore distributions are multiplied with another set of parallel lines representing the illumination. By rotating the “illumination” patterns to various angles relative to the “sample” patterns, different Moiré fringes are produced in their products (Fig. 2D-I). Although this example is simple compared with SIM, the collection of images operates similarly. The excitation light is passed through a diffraction grating, which produces several diffraction orders of which two are selected to interfere in the focal plane of the objective and produce a sinusoidal pattern of illumination. Since this pattern is known, the images collected at the various phases and orientations can be processed based on this pattern to extract the spatial information underlying the Moiré fringes in the fluorescence distributions.

The original version of SIM had an improved lateral resolution of ~100 nm and was the forerunner for further developments decreasing the axial resolution limit. One of these, 3D-SIM, uses three beams from the diffraction grating to form a three-dimensional interference pattern in the focal plane [25]. By collecting SIM images from multiple focal planes, image processing of 3D-SIM data maintains the ~100 nm lateral information of SIM and extends the axial resolution to ~300 nm, twice better than confocal imaging. A nice example of 3D-SIM is the simultaneous imaging of a cell nucleus stained for DNA, nuclear pore complex proteins, and nuclear lamina [26]. The nuclear pores have a lateral diameter of ~100 nm (plane of the nuclear envelope) and an axial size of ~150-200 nm (perpendicular to the nuclear envelope). Within this volume, which is much smaller than the dimensions of a conventional microscope focal volume, lie the pore complex proteins and lamin B. 3D-SIM can demonstrate that the pore complex proteins located on the cytoplasmic face, such as Nup62, Nup214, and Nup358, are in a different plane from lamin B located on the nucleoplasmic face of the envelope ~50-100 nm away. Nuclear envelopes stained for lamin B and Nup153, which is also located on the nucleoplasmic face, showed the molecules located in distinct lateral domains but within the same plane. [26]

The axial resolution of 3D-SIM is further improved using a combination of SIM and I5M [2]. Like I5M, I5S [27] uses two opposing objectives to maximize collection of fluorescence emission, recombines the light paths for interferometry, and uses interference of the excitation light to increase the axial resolution. The major addition to the instrument is that the three beam structured illumination approach is incorporated in I5S to generate ~100 nm resolution in all axes [27].

Since these techniques rely mainly on widefield imaging and detection with a camera, acquisition of the images used for processing is relatively fast compared with point scanning techniques. However, because the illumination is moved across the specimen to collect at several pattern phases, at several orientations of the pattern, and at several focal planes for 3D information, many images are needed to produce a single SIM image. Thus, the full collection may require seconds to minutes [25] and motion associated with normal cell processes could disrupt the fluorophore distribution, skew the image processing results, and negate any gains in resolution.

7. High density molecular localization techniques

This category of super-resolution techniques contains numerous variations (Table 2) and their differences are often subtle or indistinguishable. All are single molecule imaging techniques that rely on precise molecular localization by determining the center of fluorescence emission through a statistical fit of the photon distribution with the ideal point spread function. If the fluorescence from single molecules is distributed in a Gaussian profile and the background noise is small compared to the molecular signal, the error in the fitted position can be estimated as σxys/N.[28] Here σxy is the standard deviation of a Gaussian approximating the true PSF (~250 nm for light of λ = 550 nm wavelength) and N is the total number of detected photons.

Table 2
Molecular localization techniques*

Precise determination of localization is generally limited to molecules that are separated by the distance required of conventional optics (~250 nm). Otherwise, the hundreds of molecules within the same spot of make difficult the discrimination of one from another in most biological specimen. To circumvent this problem in high molecular density specimen, molecules are required to be initially dark at the activated fluorescence wavelength or to have the capability of switching off. [29] Depending on the fluorophore and the technique employed, the molecules become fluorescent again either by activation of a small pool or by stochastic return from a metastable dark state. The key and common characteristic for all high-density molecular localization techniques is that only a few molecules are “turned on” and fluoresce during a single image. (Fig. 2J-L) By fitting the fluorescent signal from each molecule to a 2D Gaussian distribution, the coordinates xm,ym for the location of the molecule and its uncertainty (σx,y)m are determined. Each molecule is then rendered in a new image as a Gaussian distribution of standard deviation (σx,y)m centered at xm, ym. (Fig. 2M-O)

The probes used in high-density molecular localization have special characteristics that allow maintenance of the necessary low density of detectable molecules, but they differ markedly in their mechanism of switching on and off. PALM [29] and F-PALM [30] have relied on photoactivatable and fluorescent proteins, which are initially dark or can be switched off at the normal imaging wavelengths, but after illumination at ~405 nm, turn on and fluoresce. STORM uses a photoswitching phenomenon observed with several cyanine dyes that are located in close proximity (<2 nm) [31] to each other to turn molecules on and off repeatedly [32]. PALMIRA [33] uses photoswitchable fluorescent proteins, but it and other techniques, dSTORM [34], RPM [35], GSDIM [36], DH-PSF [37], and TL-PALM [38] (Table 2), have used photoswitching into dark states of several conventional fluorescent dyes, such as Cy5, Alexa 647, Alexa488, Texas Red, FITC, Rhodamine110, Oregon Green, Atto532, and Atto565. In addition, the conventional fluorescent proteins, EGFP, EYFP [38], Citrine, and PhiYFP, have also been found to function in a similar manner [36] and several red and orange proteins have been found to undergo photoconversion [39]. Normally these dyes require an initial irradiation step to decrease fluorescence, which can be to simply photobleach enough molecules to get to single molecule density, shift the population into meta-stable dark states or both. While this does produce the necessary low background for single molecule imaging, it increases the light dosage on the specimen and photobleaching reduces the total number of molecules available for imaging.

These techniques are very precise in localizing molecules in the lateral dimensions, but the axial localization has required further approaches. Examples include 3D-STORM, which uses a cylindrical lens to skew the PSF of the single molecule fluorescence in the lateral directions depending on its position within the focal plane [40]. Biplane (BP) F-PALM uses a biplane imaging arrangement in which the image of the same molecule is projected at two different focal planes [41]. Double-helix point spread function (DH-PSF) engineers the PSF of the microscope into two lobes, which are rotated with respect to the image depending on their axial position within the focal plane [37]. And last, interferometric PALM (iPALM) collects light from juxtaposed objectives, recombines the emission from each molecule, and uses interference of the photons based on their location in the focal plane to determine their axial position [42]. Table 2 indicates the capabilities of 3D localization techniques and remarkably, the axial precisions for some approaches equal or even surpass the precisions reported for the lateral values [37, 42].

The temporal constraints on high density molecular localization image acquisition are problematic, but several examples of live cell imaging are published [38, 43-46]. All of these techniques must maintain the molecular density in each frame low enough to isolate single molecules but the density of the localized molecules must be high enough to interpret each super-resolution movie image. In two examples monitoring the movement and formation of macromolecular structures, such as adhesion complexes during cell migration [44], or observing organelle dynamics, such as the morphological changes in endoplasmic reticulum [46], time series containing 25 - 40 sec collection intervals were required while localizing up to 1000 molecules μm-2 frame-1 [44] to produce interpretable images. For live cell studies of single molecule tracking, the compromise is less severe since the experimental goals are to monitor movements of single molecules and plot their trajectories, velocities, and diffusion coefficients on an image of the cell. Such studies include monitoring the distribution and dynamics of viral proteins located at the plasma membrane of mammalian cells, such as Gag, a component of HIV [45], VSVG, a glycoprotein from the vesicular stomatitus virus [45], and hemagglutinin from influenza virus [43], in addition to the recording the dynamics of a structural protein, MreB, inside a living Caulobacter crescentus [38].

High density molecular localization techniques are still developing as an everyday tool for cell biology, so most of the published examples are concerned primarily with proof of principle or introduction of new fluorophores. Favored molecules of interest for proof of principle seem to be actin and microtubules, which make excellent resolution tests since they are distinct well-characterized cellular structures and typically have numerous <25 nm filaments located at various distances from each other. However, the literature contains several other examples that may be more useful for cell biologists in determining the usefulness of molecular localization experiments in their studies. For example, two-color imaging has observed the distribution of the transferrin receptor [47] or microtubules [48] in comparison with clathrin-coated pits as well as the relative distributions of several focal adhesion proteins with each other and with cytoskeleton proteins [49]. 3D molecular localization has provided views of clathrin-coated pits [40], microtubule networks [40, 42], viral glycoproteins [42], focal adhesion complex proteins [42], and mitochondria networks [50, 51]. Finally, an approach similar to correlative light and electron microscopy [52] that may prove most useful for cell biology is overlaying a PALM or other localization technique with an electron micrograph of the same specimen [29].

8. Concluding remarks

Given the broad range of super-resolution techniques that are now available, one common question is “What technique should I use to study my favorite process?” The preceding sections have indicated the characteristics of numerous approaches, but simply answering a few other questions can help narrow the decision. What spatial resolution is necessary? What temporal resolution is necessary? What light dosage can the specimen endure? Are specific fluorophores required for my specimen or experiment? Answers to these will likely steer a cell biologist toward one or the other of the techniques in Table 1.

The question most often raised is “What new information about the cell have we learned from this or that technique?” Often the answer is “not much….yet”. This is not due to the limitations of the technique nor to the limitations of the investigators. These techniques are moving optical imaging of biological specimen into the realm traditionally held by electron microscopy, so some caution is warranted when interpreting the structures and molecular distributions that are now being observed. Thus, the major emphasis has been to assure that the techniques perform as advertised. Perhaps more limiting has been that most of these instruments and their analysis software have been developed by a limited number of brilliant scientists. For many of these instruments, the component parts are commercially available or can be fabricated, which essentially becomes a question of expense. In the case of PALM and other molecular localization techniques, the instrumentation is often already available in the form of TIRF systems, which require simple modifications of extra laser lines and cameras sensitive enough for single molecule imaging. On the other hand, the software for analysis can be a bigger limitation as these are not generally available and when available usually require some knowledge of the programming language for proper use. Thus, cell biologists with little interest or expertise in building optical instrumentation have limited opportunity to perform super-resolution imaging.

However, over the next few years, commercial versions of several instruments discussed here will be brought to market. For instance, Leica Microsystems has developed a commercial version of a STED instrument. The system is current designed to use the dyes ATTO 647N or ATTO 655N using a 635 nm diode laser for excitation and a tunable laser for the depletion. Thus, this version is more limited than latest generation developed by Hell et. al., but still offers resolutions of ~90 nm lateral and ~550 nm axial resolutions. Carl Zeiss will soon offer a system based on structured illumination called HR-SIM, which is designed for a wide range of conventional fluorophores and comes with the software for reconstructing the final images. Another Carl Zeiss system is a molecular localization technique called PAL-M, which also includes a wide range of excitation wavelengths for most of the molecules mentioned earlier. Most importantly, all of these instruments come with the training necessary for their proper use. Thus, it is anticipated that within the next few years as these instruments make their way into laboratories, cell biologists will routinely view their fluorescent specimen with 5-10-fold better resolution than is currently available.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

1. Ditchburn RW. Light. 1991:42–70.
2. Gustafsson MG, Agard DA, Sedat JW. I5M: 3D widefield light microscopy with better than 100 nm axial resolution. J Microsc. 1999;195:10–6. [PubMed]
3. Bailey B, et al. Enhancement of axial resolution in fluorescence microscopy by standing-wave excitation. Nature. 1993;366:44–8. [PubMed]
4. Hell SW, Schrader M, van der Voort HT. Far-field fluorescence microscopy with three-dimensional resolution in the 100-nm range. J Microsc. 1997;187:1–7. [PubMed]
5. Egner A, Jakobs S, Hell SW. Fast 100-nm resolution three-dimensional microscope reveals structural plasticity of mitochondria in live yeast. Proc Natl Acad Sci U S A. 2002;99:3370–5. [PubMed]
6. Egner A, et al. 4Pi-microscopy of the Golgi apparatus in live mammalian cells. J Struct Biol. 2004;147:70–6. [PubMed]
7. Perinetti G, et al. Correlation of 4Pi and electron microscopy to study transport through single Golgi stacks in living cells with super resolution. Traffic. 2009;10:379–91. [PubMed]
8. Stelzer EH, Lindek S. Fundamental reduction of the observation volume in far-field light microscopy by detection orthogonal to the illumination axis: confocal theta microscopy. Optics Communications. 1994;111:536–547.
9. Huisken J, et al. Optical sectioning deep inside live embryos by selective plane illumination microscopy. Science. 2004;305:1007–9. [PubMed]
10. Keller PJ, Stelzer EH. Quantitative in vivo imaging of entire embryos with Digital Scanned Laser Light Sheet Fluorescence Microscopy. Curr Opin Neurobiol. 2009 [PubMed]
11. Hell SW, Wichmann J. Breaking the diffraction resolution limit by stimulated emission: stimulated emission depletion fluorescence microscopy. Optics Letters. 1994;19:780–782. [PubMed]
12. Westphal V, Hell SW. Nanoscale resolution in the focal plane of an optical microscope. Phys Rev Lett. 2005;94:143903. [PubMed]
13. Dyba M, Hell SW. Focal spots of size lambda/23 open up far-field fluorescence microscopy at 33 nm axial resolution. Phys Rev Lett. 2002;88:163901. [PubMed]
14. Schmidt R, et al. Spherical nanosized focal spot unravels the interior of cells. Nat Methods. 2008;5:539–44. [PubMed]
15. Dyba M, Jakobs S, Hell SW. Immunofluorescence stimulated emission depletion microscopy. Nat Biotechnol. 2003;21:1303–4. [PubMed]
16. Willig KI, et al. Nanoscale resolution in GFP-based microscopy. Nat Methods. 2006;3:721–3. [PubMed]
17. Willig KI, et al. STED microscopy reveals that synaptotagmin remains clustered after synaptic vesicle exocytosis. Nature. 2006;440:935–9. [PubMed]
18. Sieber JJ, et al. The SNARE motif is essential for the formation of syntaxin clusters in the plasma membrane. Biophys J. 2006;90:2843–51. [PubMed]
19. Fitzner D, et al. Myelin basic protein-dependent plasma membrane reorganization in the formation of myelin. Embo J. 2006;25:5037–48. [PubMed]
20. Meyer AC, et al. Tuning of synapse number, structure and function in the cochlea. Nat Neurosci. 2009;12:444–53. [PubMed]
21. Kittel RJ, et al. Bruchpilot promotes active zone assembly, Ca2+ channel clustering, and vesicle release. Science. 2006;312:1051–4. [PubMed]
22. Westphal V, et al. Video-rate far-field optical nanoscopy dissects synaptic vesicle movement. Science. 2008;320:246–9. [PubMed]
23. Nagerl UV, et al. Live-cell imaging of dendritic spines by STED microscopy. Proc Natl Acad Sci U S A. 2008;105:18982–7. [PubMed]
24. Gustafsson MG. Surpassing the lateral resolution limit by a factor of two using structured illumination microscopy. J Microsc. 2000;198:82–7. [PubMed]
25. Gustafsson MG, et al. Three-dimensional resolution doubling in wide-field fluorescence microscopy by structured illumination. Biophys J. 2008;94:4957–70. [PubMed]
26. Schermelleh L, et al. Subdiffraction multicolor imaging of the nuclear periphery with 3D structured illumination microscopy. Science. 2008;320:1332–6. [PMC free article] [PubMed]
27. Shao L, et al. I5S: wide-field light microscopy with 100-nm-scale resolution in three dimensions. Biophys J. 2008;94:4971–83. [PubMed]
28. Thompson RE, Larson DR, Webb WW. Precise nanometer localization analysis for individual fluorescent probes. Biophys J. 2002;82:2775–83. [PubMed]
29. Betzig E, et al. Imaging intracellular fluorescent proteins at nanometer resolution. Science. 2006;313:1642–5. [PubMed]
30. Hess ST, Girirajan TP, Mason MD. Ultra-high resolution imaging by fluorescence photoactivation localization microscopy. Biophys J. 2006;91:4258–72. [PubMed]
31. Bates M, Blosser TR, Zhuang X. Short-range spectroscopic ruler based on a single-molecule optical switch. Phys Rev Lett. 2005;94:108101. [PMC free article] [PubMed]
32. Rust MJ, Bates M, Zhuang X. Sub-diffraction-limit imaging by stochastic optical reconstruction microscopy (STORM) Nat Methods. 2006;3:793–5. [PMC free article] [PubMed]
33. Geisler C, et al. Resolution of λ/10 in fluorescence microscopy using fast single molecule photo-switching. Applied Physics A. 2007;88:223–226.
34. Heilemann M, et al. Subdiffraction-resolution fluorescence imaging with conventional fluorescent probes. Angew Chem Int Ed Engl. 2008;47:6172–6. [PubMed]
35. Baddeley D, et al. Light-induced dark states of organic fluochromes enable 30 nm resolution imaging in standard media. Biophys J. 2009;96:L22–4. [PubMed]
36. Folling J, et al. Fluorescence nanoscopy by ground-state depletion and single-molecule return. Nat Methods. 2008;5:943–5. [PubMed]
37. Pavani SR, et al. Three-dimensional, single-molecule fluorescence imaging beyond the diffraction limit by using a double-helix point spread function. Proc Natl Acad Sci U S A. 2009;106:2995–9. [PubMed]
38. Biteen JS, et al. Super-resolution imaging in live Caulobacter crescentus cells using photoswitchable EYFP. Nat Methods. 2008;5:947–9. [PMC free article] [PubMed]
39. Kremers GJ, et al. Photoconversion in orange and red fluorescent proteins. Nat Methods. 2009;6:355–8. [PMC free article] [PubMed]
40. Huang B, et al. Three-dimensional super-resolution imaging by stochastic optical reconstruction microscopy. Science. 2008;319:810–3. [PMC free article] [PubMed]
41. Juette MF, et al. Three-dimensional sub-100 nm resolution fluorescence microscopy of thick samples. Nat Methods. 2008;5:527–9. [PubMed]
42. Shtengel G, et al. Interferometric fluorescent super-resolution microscopy resolves 3D cellular ultrastructure. Proc Natl Acad Sci U S A. 2009;106:3125–30. [PubMed]
43. Hess ST, et al. Dynamic clustered distribution of hemagglutinin resolved at 40 nm in living cell membranes discriminates between raft theories. Proc Natl Acad Sci U S A. 2007;104:17370–5. [PubMed]
44. Shroff H, et al. Live-cell photoactivated localization microscopy of nanoscale adhesion dynamics. Nat Methods. 2008;5:417–23. [PubMed]
45. Manley S, et al. High-density mapping of single-molecule trajectories with photoactivated localization microscopy. Nat Methods. 2008;5:155–7. [PubMed]
46. Andresen M, et al. Photoswitchable fluorescent proteins enable monochromatic multilabel imaging and dual color fluorescence nanoscopy. Nat Biotechnol. 2008;26:1035–40. [PubMed]
47. Subach FV, et al. Photoactivatable mCherry for high-resolution two-color fluorescence microscopy. Nat Methods. 2009;6:153–9. [PMC free article] [PubMed]
48. Bates M, et al. Multicolor super-resolution imaging with photo-switchable fluorescent probes. Science. 2007;317:1749–53. [PMC free article] [PubMed]
49. Shroff H, et al. Dual-color superresolution imaging of genetically expressed probes within individual adhesion complexes. Proc Natl Acad Sci U S A. 2007;104:20308–13. [PubMed]
50. Vaziri A, et al. Multilayer three-dimensional super resolution imaging of thick biological samples. Proc Natl Acad Sci U S A. 2008;105:20221–6. [PubMed]
51. Huang B, et al. Whole-cell 3D STORM reveals interactions between cellular structures with nanometer-scale resolution. Nat Methods. 2008;5:1047–52. [PMC free article] [PubMed]
52. Polishchuk RS, et al. Correlative light-electron microscopy reveals the tubular-saccular ultrastructure of carriers operating between Golgi apparatus and plasma membrane. The Journal of Cell Biology. 2000;148:45–58. [PMC free article] [PubMed]
53. Betzig E, Chichester RJ. Single molecules observed by near-field scanning optical microscopy. Science. 1993;262:1422–1425. [PubMed]
54. Betzig E, et al. Near-field diffraction by a slit: implications for superresolution microscopy. Appl Opt. 1986;25:1890. [PubMed]
55. Betzig E, et al. Near Field Scanning Optical Microscopy (NSOM): Development and Biophysical Applications. Biophys J. 1986;49:269–279. [PubMed]
56. Betzig E, Trautman JK. Near-Field Optics: Microscopy, Spectroscopy, and Surface Modification Beyond the Diffraction Limit. Science. 1992;257:189–195. [PubMed]
57. Betzig E, et al. Breaking the Diffraction Barrier: Optical Microscopy on a Nanometric Scale. Science. 1991;251:1468–1470. [PubMed]
58. Frohn JT, Knapp HF, Stemmer A. True optical resolution beyond the Rayleigh limit achieved by standing wave illumination. Proc Natl Acad Sci U S A. 2000;97:7232–6. [PubMed]
59. Beck M, Aschwanden M, Stemmer A. Sub-100-nanometre resolution in total internal reflection fluorescence microscopy. J Microsc. 2008;232:99–105. [PubMed]
60. Chung E, et al. Two-dimensional standing wave total internal reflection fluorescence microscopy: superresolution imaging of single molecular and biological specimens. Biophys J. 2007;93:1747–57. [PubMed]
61. Bewersdorf J, Schmidt R, Hell SW. Comparison of I5M and 4Pi-microscopy. J Microsc. 2006;222:105–17. [PubMed]
62. Swoger J, Huisken J, Stelzer EH. Multiple imaging axis microscopy improves resolution for thick-sample applications. Opt Lett. 2003;28:1654–6. [PubMed]
63. Hofmann M, et al. Breaking the diffraction barrier in fluorescence microscopy at low light intensities by using reversibly photoswitchable proteins. Proc Natl Acad Sci U S A. 2005;102:17565–9. [PubMed]
64. Heintzmann R, Jovin TM, Cremer C. Saturated patterned excitation microscopy--a concept for optical resolution improvement. J Opt Soc Am A Opt Image Sci Vis. 2002;19:1599–609. [PubMed]
65. Gustafsson MG. Nonlinear structured-illumination microscopy: wide-field fluorescence imaging with theoretically unlimited resolution. Proc Natl Acad Sci U S A. 2005;102:13081–6. [PubMed]
66. van Oijen AM, et al. Far-field fluorescence microscopy beyond the diffraction limit. Journal of the Optical Society of America. 1999;16:909–915.
67. Heilemann M, et al. High-resolution colocalization of single dye molecules by fluorescence lifetime imaging microscopy. Anal Chem. 2002;74:3511–7. [PubMed]
68. Qu X, et al. Nanometer-localized multiple single-molecule fluorescence microscopy. Proc Natl Acad Sci U S A. 2004;101:11298–303. [PubMed]
69. Lidke KA, et al. Superresolution by localization of quantum dots using blinking statistics. Opt Express. 2005;13:7052. [PubMed]