Overtone Mobility Spectrometry (Part 1): Experimental Observations

Abstract

A new method that allows a linear drift tube to be operated as a continuous ion mobility filter is described. Unlike conventional ion mobility instruments that utilize an electrostatic gate to introduce a packet of ions into a drift region, the present approach uses multiple segmented drift regions with modulated drift fields to produce conditions that allow only ions with appropriate mobilities to pass through the instrument. In this way, the instrument acts as a mobility-filter for continuous ion sources. By changing the frequency of the applied drift fields it is possible to tune this instrument to transmit ions having different mobilities. A scan over a wide range of drift field frequencies for a single ion species shows a peak corresponding to the expected resonance time of the ions in one drift region segment and a series of peaks at higher frequencies that are overtones of the resonant frequency. The measured resolving power increases for higher overtones making it possible to resolve structures that were unresolved in the region of the fundamental frequency. We demonstrate the approach by examining oligosaccharide isomers, raffinose and melezitose as well as a mixture of peptides obtained from enzymatic digestion of myoglobin.

Introduction

Separation of a mixture of ions by ion mobility spectrometry (IMS) in linear drift tubes is initiated by gating short packets of ions into the front of a drift region containing a static buffer gas. Different species then migrate through the drift region under the influence of an applied field according to their mobilities in the buffer gas such that ions with high mobilities reach the detector before those with low mobilities.1-6 This approach has attracted considerable attention as a means of analyzing complex mixtures.7-16 Additionally, measured mobilities can be converted to experimental cross sections and comparison of these values with cross sections for model geometries that are calculated can be used to characterize ion structures.17-28

In this paper we introduce a new approach for isolating ions having specific mobilities (or collision cross sections). Ions from a continuous source enter a drift tube with segmented drift regions. The drift fields are modulated at a frequency that allows only those ions having mobilities that are resonant with the experimental conditions to be transmitted through all drift regions. In this way, this device filters away all ions except those with mobilities over the selected narrow range. An unanticipated feature of this approach is the observation that ions can be passed at overtone frequencies; and, the resolving power in higher overtone regions is greater than that observed in the fundamental frequency range. Because of the ability to select ions in different frequency regions, including those that are associated with higher overtones, we refer to the approach as Overtone Mobility Spectrometry (OMS).

In order to demonstrate this approach we have examined two types of systems. First we show that different ions from a simple ESI source are favored at different frequencies. This is done for two trisaccharides (raffinose and melezitose, which are structural isomers). We find that it is relatively straightforward to favor one or the other from a mixture of the two. We then examine the utility of OMS as a means of simplifying a more complicated mixture of tryptic peptides (from a digest of myoglobin).

The separation and selection of ions according to their mobilties is currently an area of significant research activity and a number of other techniques are being developed. These include: differential mobility analysis, in which ions migrate through a flowing gas under the influence of a field;29-31 and, field asymmetric (FA) IMS which separates ions based on differences between high- and low-field mobilities.32-38 Such approaches are finding applications as a means of characterizing a range of analytes, from chemical warfare agents to determining particle sizes, and also have utility as a means of simplifying the analysis of complex mixtures through the removal of chemical noise (or other components of the mixture that differ in mobility).1,10,31,39-45 The OMS approach described here should also be useful for these types of applications. One advantage of the approach is that it is rooted in a fundamental property of the ion (its mobility) making it possible to tune the filter for selection of desired species.

Experimental

General

Detailed descriptions of IMS1,2,4,6,39,46,47 and IMS-MS2,3,48-54 instrumentation as well as theoretical considerations are provided elsewhere.5,6,22-24,55,56 Additionally, instrumentation components including the ion source, ion focusing elements, vacuum system, mass spectrometer, and detection scheme as well as recent instrumentation modifications leading to the current instrument are discussed elsewhere.57-60 Here, the description of the experimental apparatus focuses on the multi-segmented drift tube. Figure 1A shows a schematic diagram of the drift tube mass spectrometer used in these studies. Ions are produced using a home-built electrospray ionization (ESI) source and are continuously focused through an ion funnel (F1) interface into a linear drift region. The drift region is 181 cm in length and can be configured to operate using a uniform field (as described previously)59,60 or alternating fields applied to segmented drift regions (as described in more detail below). Ions that exit the drift tube are focused into the source region of a reflectron time-of-flight (TOF) mass spectrometer for mass analysis.

Figure 1
Schematic diagram of the drift tube mass spectrometer used in these studies (A). The instrument incorporates an ESI ion source, three ion funnels (labeled F1 to F3), a segmented drift region and a time-of-flight mass spectrometer. Upon application of ...

Sample preparation

Raffinose (98% purity), melezitose (99% purity) and horse myoglobin (≥90% purity) were purchased from Sigma (Sigma Chemical, St.Louis) and used without further purification. Raffinose and melezitose (0.25 mg mL-1) were prepared in water:acetonitrile (50:50% by volume) and 2 mM NaCl. Tryptic peptide mixtures were obtained as follows. Myoglobin was initially dissolved in 6 M urea buffer and the sample was diluted to 2 M urea using 0.2 M Tris buffer (pH = 8.0, 10 mM CaCl2) before trypsin digestion. Tryptic peptides were generated by adding TPCK-treated trypsin (2% enzyme by mass to that of the protein) and incubating the solution at 37 °C for 24 hours. Finally, the mixture of peptides was desalted using Oasis hydrophilic-lipophilic balanced (HLB) cartridges (Waters Inc., Milford, MA) and dried on a centrifugal concentrator. Dried peptides were resuspended in water:acetonitrile:acetic acid solution (49:49:2% by volume).

Segmented drift tube and ion funnel assembly

Figure 1B shows a schematic diagram of the segmented drift tube. The mid section of the drift tube has been divided into 22 segments. Of these, 21 segments are identical. Each segment is comprised of five concentric stainless steel rings (14.0 cm o.d. and 7.0 cm i.d.). The first two contain grids (90% transmittance, Ni mesh grid, Buckbee-Mears, St. Paul, MN) and are separated by a 0.18 cm thick Teflon isolator (14.0 cm o.d. and 8.3 cm i.d.). The remaining three lenses are separated by 1.27 cm thick Delrin spacers (14.0 cm o.d. and 8.3 cm i.d.). The rings and spacers are stacked together, sealed with O-rings, and compressed using eight nylon threaded rods; multiple segments are joined to create the mobility filter region. The length of each segment (from the front grid of one segment to the front grid of an adjacent segment) is 5.84 cm and so the total length of the 22 section separation region is 128.5 cm.

One segment near the center of the drift tube is replaced by an ion funnel assembly that is similar to published designs.53,60 This ion funnel is made of 22 0.075 cm thick stainless steel electrodes (12.7 cm o.d.) with aperture diameters that decrease linearly from 7.00 to 0.56 cm; the funnel lenses are separated by 0.18 cm thick Teflon spacers. The funnel is designed such that its total length (5.74 cm) is similar to the individual drift tube segments. The funnel is operated using RF fields of 50-70 Vp-p at 450 kHz and 10 V cm-1 DC field. With the exception of the RF component used to improve ion transmission through the drift tube device, the funnel is utilized here simply as an additional drift segment. Evidence that the ion transit times through the funnel region are similar to the other drift regions is obtained from two observations. First, the placement of an ion funnel in the OMS device at various positions does not significantly change peak widths, peak positions, or peak intensities even when employing two ion funnels. The second observation is that accurate collision cross sections can be obtained from OMS measurements that employ in-line ion funnels. The assembly of lenses and spacers forms a cavity that can be filled with a buffer gas. Helium was used as the buffer gas and typically the pressure was 2.25 Torr, as determined by a capacitance manometer. The buffer gas temperature was ~298 K.

Modulation of drift fields in the segmented drift tube

A uniform, 10 V cm-1 DC field is created across each segment using a combination of applied voltages and appropriate resistors in series (57 kΩ, 1% tolerance). In the present system, fields are applied to pairs of segments (i.e. voltage is applied at the front of each segment pair) to create a sawtooth (Figures 1 and and2)2) voltage gradient down the axis of the drift tube. A second sawtooth is offset from the first by one segment and is turned on while the first sawtooth field is turned off. These fields are alternated at a frequency that is referred to as the drift field application frequency.

Figure 2
Mass spectra obtained upon ESI of raffinose. The bottom spectrum (2a) shows the total mass spectrum obtained when the instrument is operated as a conventional drift tube. The top three mass spectra are obtained by operating the OMS instrument at different ...

As mentioned before, the separation of ions in this instrument occurs based on the mobilities of the ions analyzed. However, in contrast to a conventional drift tube experiment, the separation is recorded in the frequency domain. The frequency here corresponds to the speed at which the two drift field settings or phases (the phase of OMS system refers to the number of distinct field settings) are modulated. For example, for a two-phase system, at 2000 Hz field application frequency, one phase is turned on for 500 μs while the other is turned off for the same amount of time and the two phases are alternatively turned on and off to move the ions across the drift tube. Since the ion transmission is a function of frequency, a plot of frequency versus the signal intensity for a given ion constitutes its OMS spectrum.

For a given ion, the positions of the peaks appearing at higher frequencies are related to multiples of the fundamental frequency. The appearance of these overtone peaks demonstrates that ions can be stable inside the drift tube not only at the fundamental frequency but also at other frequencies that are multiples of the fundamental frequency. For a two-phase system, overtone peaks appear at 3f, 5f, 7f etc. of the fundamental frequency. A thorough explanation for the appearance of the overtone peaks, their positions and the dependence on the number of phases used for the experiment will be given in an accompanying manuscript describing theoretical considerations of the resolving power of OMS.

Results and discussion

Example mass spectra obtained upon ESI of raffinose into segmented drift tube

Figure 2 shows mass spectra that have been recorded upon direct electrospray of a solution containing the three-residue oligosaccharide raffinose. By applying a constant electric field across all segments of the OMS, a mass spectrum (Figure 2a) is obtained containing a number of prominent peaks. The spectrum is dominated by a peak mass-to-charge (m/z) 527. Other smaller peaks at 532.5 and 1031 are also observed. By examining the isotopic patterns for these ions, it is possible to determine their charge state. Isotopes observed for the feature at m/z 527 are spaced by 1 m/z indicating this species is singly charged. Based on the charge state and the m/z value, we assign this ion as the [M+Na]+ monomer ion of raffinose. A spacing of 0.5 Th within the isotopic envelope for the feature at m/z 532.5 indicates that it is a doubly-charged ion. Currently this ion has not been assigned. The peak observed at m/z 1031 can nominally be assigned to the [2M+Na]+ dimer. However, the congested isotope pattern suggests that higher order multimers (e.g., [4M+2Na]2+ and [6M+3Na]3+) are also present. The discussion presented below addresses the assignment of these species.

Upon application of modulated field settings to the OMS drift tube as described above, we observe very different ion distributions. For example, application of the drift field settings at a rate of 2050 Hz yields a mass spectrum in which the 532.5 peak is eliminated. These conditions seem to favor a small population of ions in the ~790-805 m/z range. At ~1400 or 2550 Hz the 1031 or 532.5 peaks are favored. Such a strong frequency dependence of the observed ion distribution was initially very surprising. The result suggests the possibility of obtaining distributions as a function of frequency for each ion by monitoring the intensity of each mass spectral feature as a function of applied frequency.

Examples of the frequency dependence of several different ions

Figure 3 shows several examples of OMS spectra obtained for specific ions upon scanning the applied drift-field frequency. In this case the frequency has been varied from ~0 to ~16000 Hz. The plot shows the ion intensities that are recorded for the 527, 532.5, and 1031 ions mentioned above. Over this range of frequencies, the OMS spectrum for [M+Na]+ (m/z 527) is dominated by a large peak at 2050 Hz. A much smaller peak is observed at 2750 Hz. Smaller peaks that decrease in intensity with increasing frequency are observed at 5977, 9952, and 14194 Hz.

Figure 3
Measured ion intensity (arbitrary units) as a function of field application frequency over a range of ~0 to ~16,000 Hz obtained upon ESI of a 0.25 mg/ml solution of raffinose. The top spectrum corresponds to the sodiated monomer of raffinose. The peak ...

The doubly-charged ion having m/z 532.5 (unassigned ion from the raffinose sample) shows a more complicated pattern of peaks (see Figure 3). The most prominent peak is centered at 2571 Hz and is slightly broader than the peaks observed for the [M+Na]+ ion. Additionally, this peak shows a small reproducible feature at higher frequencies (the shoulder at ~2846 Hz). As the frequency is increased, two additional sets of smaller peaks (at 7383, 7656, and 8555 Hz, as well as 12247, 12734, and 13894 Hz, respectively) are observed. Again, in general, peak intensities decrease with increasing frequency.

The OMS distribution for the m/z 1031 ion shows three sharp peaks at 1398, 1760, and 2068 Hz (Figure 4). Because the isotopic envelope suggested the presence of multiple isobaric species, we have examined the isotopic envelopes of the frequency-resolved ions. This shows that isotopic peaks are spaced by 1.0, 0.5, and 0.33 Th for the features observed at 1398, 1760, and 2068 Hz, respectively. With this charge state information, we assign these species to the singly-charged dimer [2M+Na]+, the doubly-charged tetramer [4M+2Na]2+, and the triply-charged hexamer [6M+3Na]3+ ions, respectively. Additional peaks in the OMS distribution are observed at higher frequencies, including: 4135, 5182, and 6062 Hz. The intensities of these higher-frequency peaks are smaller than the lower-frequency peaks. The isotopic spacing of these peaks allow them to be assigned as [2M+Na]+ (4135 Hz), [4M+2Na]2+ (5182 Hz), and [6M+3Na]3+ (6062 Hz). This spectrum for these ions is very interesting and can be used to develop a working understanding of this instrumentation. Specifically, these data show that multiply-charged multimers of the same nominal m/z value are transmitted at different frequencies. Moreover, the peaks appear as two groups. Each of the peaks in the higher frequency groups appears at a frequency that is about a factor three higher than the lower frequency transmission value. That is, 4135/1398 ≈ 2.96 and 6062/2068 ≈ 2.93. This suggests that each ion is being transmitted at a fundamental frequency as well as at an overtone frequency. The observation of overtone peaks resulted in naming this approach as Overtone Mobility Spectrometry (OMS). Finally, we find it very interesting that although the positions of peaks change by multiples of the fundamental frequency, the peak widths do not. This leads to the observation that the resolving power of OMS increases in higher overtone regions of the spectrum.

Figure 4
Plot of calculated field application frequency with respect to experimental field application frequency for raffinose [M+Na]+ (solid diamond), [2M+Na]+ (open circle), [4M+2Na]2+ (open triangle) and [6M+3Na]3+ (open square). The calculated frequency is ...

Understanding the origin of higher frequency peaks

It is important to understand the origin at the transmission frequencies. For example, using IMS measurements for the [M+Na]+ monomer of raffinose, a drift time (tD) of 15.2 ms is obtained. This is the time required for the ion to traverse the entire 181-cm linear IMS drift tube. Using this information, it is possible to calculate the ion displacement during the time required for application of a single field setting at the frequency giving rise to the most dominant peak (Figure 4) in the OMS spectrum. Here, a frequency of 2050 Hz corresponds to a field application time of ~488 μs. Using the measured drift velocity of the ion (1.19×104 cm s-1), the distance traveled during a single field setting would be 5.81 cm. This is almost identical to the value of one drift segment (in this case 5.84 cm).

Using the measured tD values for species observed in the raffinose sample, it is possible to calculate frequencies for which the peaks are observed in the OMS distribution. Consider the raffinose multimer ions in the OMS spectrum where three distinct features are observed at lower frequency settings (<2500 Hz). The IMS drift distribution profile for this ion also shows three base-line resolved peaks having intensities that increase with increasing drift time. In contrast the peaks in the OMS spectrum exhibit a decreasing intensity pattern for these peaks. Using the method described above, calculated OMS frequencies of 2070, 1760, and 1401 Hz are obtained for peaks with tD values of 15.0, 17.6, and 22.1 ms, respectively. These values are close to the experimental values of 2068, 1760, and 1398 Hz.

Figure 4 shows a plot of calculated frequencies versus experimentally determined frequencies for a number of the ions observed from the raffinose sample. We provide two examples of this plot. The first (Figure 4a) assumes no overtones. In this case only the fundamental experimental transmission frequencies correlate directly with values calculated from IMS measurements. If the higher-frequency (>3000 Hz) features in the OMS distributions are multiplied by the appropriate integer value (3, 5, or 7), we obtain Figure 4b. The correlation coefficient for the plot shown in Figure 4 is 0.9996 and the slope is 1.0063 indicating that in all cases, multiples of the calculated frequency for the same ion give rise to the observed peaks at higher frequencies in the OMS spectra. Hereafter, we refer to the frequency corresponding to an ion traversing one drift tube segment as the fundamental frequency (ff) and the higher frequency settings as overtones. For these data, peaks associated with overtone frequencies of 3f, 5f, and 7f are observed (Figure 3).

Observation of enhanced resolving power for higher overtone regions

The resolving power of OMS is defined by (f/Δf), where f corresponds to the frequency at which maximum ion intensity is transmitted, and Δf is taken as the width of the OMS peak at half maximum. In general the OMS resolving power is observed to increase with increasing overtone number. Consider the peaks for the [2M+Na]+, [4M+2Na]2+, and [6M+3Na]3+ ions having m/z 1031 shown in Figure 3. The OMS resolving power values obtained for the three ff peaks are 15.3, 15.1 and 15.4 for the dimer, tetramer, and hexamer, respectively. The resolving power values obtained for the 3f overtone peaks for the same ions (in the same order) are 40.3, 35.2 and 41.6. This type of increase in resolving power with overtone number is typical. A separate example is the doubly-charged ion (m/z 532.5). The ff setting exhibits a single, relatively broad peak and yields a measured resolving power of 10.8. The resolving power for the 3f overtone peaks give values of 27.6, 28.6 and 39.6, for peaks centered at 7383, 7656, and 8555 Hz, respectively. The resolving power for the peaks at 12247 and 12734 Hz is 44.0 and 62.8, respectively. There is a small signal at 13874 Hz, the expected position for the 5f overtone of the third peak which suggests even higher resolving power. These results demonstrate the utility of operating the instrument at higher overtone frequencies to isolate ions of specific mobilities.

An example separation of the structural isomers of raffinose and melezitose

As another example, we illustrate the utility of enhanced resolving power at higher overtones by examining the separation of structural isomers. Figure 5 shows OMS distribution plots for raffinose (top), melezitose (middle) and a mixture at 3:1 raffinose:melezitose mixture (bottom). From high-resolution IMS measurements, we know that raffinose has a slightly lower mobility than melezitose, allowing a mixture of these two ions to be separated in a conventional drift tube with sufficient resolving power.61 The frequency spectrum obtained for the trisaccharide mixture also shows that. Although, the two isomers are not resolved in the fundamental region of the spectrum, they are partially and fully resolved at the 3f and 5f overtone frequency regions, respectively. Thus, even at this very early stage of development, this becomes an interesting approach. In this case, a very simple means of selecting isomers would be useful for MS/MS types of analyses.

Figure 5
A comparion of measured ion intensities as a function of field application frequency over a rang of ~0 to ~14000 Hz for raffinose, melezitose and a mixture of raffinose and melezitose at 3:1 ratio. The data is collected using a two-phase wave driving ...

Separation of peptides from a tryptic digest of myoglobin

As a final example, we consider a more complex mixture of tryptic peptides generated upon enzymatic digestion with trypsin. Several results are shown in Figure 6. When the drift tube is configured as a traditional IMS instrument, we observe the complicated pattern of peaks associated with forming the [M+H]+, [M+2H]2+, and [M+3H]3+ peptides upon electrospray. For reference, we have indicated the positions and assignments of several large peaks, [ALELFR+2H]2+ at m/z = 374.7, [HGTVVLTALGGILK+3H]3+ at m/z = 460.2, and [YLEFISDAIIHVLHSK+3H]3+ at m/z = 628.9. When the multiple drift region configuration is employed, it is straightforward to find conditions where only a narrow band of mobility-selected ions is transmitted. For example, mass spectra recorded at 6000, 7000, or 8000 Hz are largely dominated by single peaks corresponding to [YLEFISDAIIHVLHSK+3H]3+, [HGTVVLTALGGILK+3H]3+, or [ALELFR+2H]2+, respectively. These data illustrate the selectivity of the approach. This could have utility in applications that seek to resolve specific conformations or isomers for subsequent analysis. For example, an OMS selection followed by collision-induced dissociation (CID) and MS analysis.

Figure 6
Mass spectra obtained upon ESI of a mixture of tryptic peptides of myoglobin at application frequencies of 6000, 7000 and 8000 Hz as well as the total mass spectrum (bottom). At different field application frequencies different ions are stable inside ...

To further illustrate the application of this technique for complex sample analysis, we have analyzed a tryptic digest of three proteins; myoglobin (horse), hemoglobin (human) and cytochrome c (human) and spiked with substance p peptide. A 2D frequency (m/z) distribution for the above tryptic digest is shown in the Supplementary Information section (Supplementary Figure 1). Here, the peak capacity for the fundamental peak range is calculated to be ~7 while for the 5th overtone level the peak capacity is ~27 and for the 9th overtone level the peak capacity is ~36. Therefore, it is important to note the advantage of this technique for complex sample analysis especially when using higher overtone range.

Ion signal

One final issue that we have not addressed so far is associated with ion signal. Ultimately the resolution that is obtained with this technique comes at the cost of a decreased signal. Some loss is incurred because this method acts as a filter. Thus, at a given transmission frequency setting, many ions will be lost. This is analogous to losses associated with selection of ions in a quadrupole mass filter. Another source of ion loss in the present work is due to the use of grids between each drift region. Although we have used 90% transmittance grids for this proof of concept, we anticipate that higher transmission grids (to a limit of gridless) designs will further improve this system. We are currently working on these designs.

Summary

A new ion-mobility separation approach that utilizes a segmented drift tube has been presented. In this approach, linear voltage gradients are applied to pairs of drift region segments in a fashion that creates a sawtooth pattern down the axis of the drift tube. The position of the sawtooth field is shifted between two sets of drift tubes such that only ions having mobilties that are in resonance with the switching frequency are passed through the instrument. In this way the device acts as a mobility-filter. The approach has been demonstrated using several ions that are formed upon ESI of the oligosaccharide isomers raffinose and melezitose. A surprising finding is that it is possible to transmit ions at overtone frequencies and the peaks that are observed at higher overtones are sharp. This result is striking and raises the possibility of developing relatively simple, high resolving power mobility filters. The ability to select ions from continous sources based on their mobilities is likely to find applications when combined with other scanning mass spectrometric devices, such as quadrupoles and trapping instruments. Further work involving the development of these instruments is ongoing.

Supplementary Material

01

Acknowledgments

The development of new instrumentation is supported in part by a grants from the National Institutes of Health (AG-024547-01 and P41-RR018942), and the METACyte initiative funded by a grant from the Lilly Endowment. The authors are grateful for numerous stimulating discussions with their colleagues, Liang-shi Li, Caroline C. Jarrold, and Gary M. Hieftje about overtones and instrumentation in general. We also thank John Poehlman and Andrew Alexander for technical support.

Footnotes

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

1. For a review of IMS techniques see for example (and references therein): St. Louis RH, Hill HH., Jr Ion mobility spectrometry in analytical chemistry. Crit Rev Anal Chem. 1990;21:321–355.
2. For a review of IMS techniques see for example (and references therein): Clemmer DE, Jarrold MF. Ion mobility measurements and their applications to clusters and biomolecules. J Mass Spectrom. 1997;32:577–592.
3. For a review of IMS techniques see for example (and references therein): Hoaglund-Hyzer CS, Counterman AE, Clemmer DE. Anhydrous protein ions. Chem Rev. 1999;99:3037–3079. [PubMed]
4. For a review of IMS techniques see for example (and references therein): Collins DC, Lee ML. Developments in ion mobility spectrometry-mass spectrometry. Anal Bioanal Chem. 2002;372:66–73. [PubMed]
5. Revercomb HE, Mason EA. Theory of plasma chromatography/gaseous electrophoresis. Anal Chem. 1975;47:970–983.
6. Mason EA, McDaniel EW. Transport Properties of Ions in Gases. Wiley; New York: 1988.
7. Liu X, Valentine SJ, Plasencia MD, Trimpin S, Naylor S, Clemmer DE. Mapping the human plasme proteome by SCX-LC-IMS-MS. J Am Soc Mass Spectrom. 2007;18:1249–1264. [PMC free article] [PubMed]
8. Valentine SJ, Plasencia MD, Liu X, Krishnan M, Naylor S, Udseth HR, Smith RD, Clemmer DE. Toward plasma proteome profiling with ion mobility-mass spectrometry. J Proteome Res. 2006;5:2977–2984. [PubMed]
9. Valentine SJ, Counterman AE, Hoaglund-Hyzer CS, Clemmer DE. Gas-phase separations of protease digests. J Am Soc Mass Spectrom. 1998;9:1213–1216. [PubMed]
10. Valentine SJ, Kulchania M, Barnes CAS, Clemmer DE. Multidimensional separations of complex peptide mixtures: a combined high-performance liquid chromatography/ion mobility/time-of-flight mass spectrometry approach. Int J Mass Spectrom. 2001;212:97–109.
11. Matz LM, Dion HM, Hill HH. Evaluation of capillary liquid chromatography-electrospray ionization ion mobility spectrometry with mass spectrometry detection. J Chromatogr A. 2002;946(12):59–68. [PubMed]
12. Matz LM, Asbury GR, Hill HH. Two-dimensional separations with electrospray ionization ambient pressure high-resolution ion mobility spectrometry/quadrupole mass spectrometry. Rapid Commun Mass Spectrom. 2002;16(7):670–675. [PubMed]
13. Ruotolo BT, Gillig KJ, Stone EG, Russell DH. Peak capacity of ion mobility mass spectrometry: separation of peptides in helium buffer gas. J Chromatogr B. 2002;782(12):385–392. [PubMed]
14. McLean JA, Ruotolo BT, Gillig KJ, Russell DH. Ion mobility-mass spectrometry: a new paradigm for proteomics. Int J Mass Spectrom. 2005;240(3):301–315.
15. McLean JA, Ridenour WB, Caprioli RM. Profiling and imaging of tissues by imaging ion mobility-mass spectrometry. J Mass Spectrom. 2007;42(8):1099–1105. [PubMed]
16. Fenn LS, McLean JA. Biomolecular structural separations by ion mobility-mass spectrometry. Anal Bioanal Chem. 2008;391:905–909. [PubMed]
17. Von Helden G, Hsu M–T, Kemper PR, Bowers MT. Structures of carbon cluster ions from 3 to 60 atoms: linears to rings to fullerenes. J Chem Phys. 1991;95(5):3835–3837.
18. Jarrold MF, Constant VA. Silicon cluster ions: evidence for a structural transition. Phys Rev Lett. 1991;67(21):2994–2997. [PubMed]
19. Von Helden G, Hsu M–T, Gotts N, Bowers MT. Carbon cluster cations with up to 84 atoms: structures, formation mechanism, and reactivity. J Phys Chem. 1993;97:8182–8192.
20. Shelimov KB, Hunter JM, Jarrold MF. Small carbon rings: dissociation, isomerization, and a simple model based on strain. Int J Mass Spectrom Ion Phys. 1994;138:17–31.
21. Clemmer DE, Hudgins RR, Jarrold MF. Naked protein conformations: Cytochrome c in the gas phase. J Am Chem Soc. 1995;117:10141–10142.
22. Mesleh MF, Hunter JM, Shvartsburg AA, Schatz GC, Jarrold MF. Structural Information from Ion Mobility Measurements: Effects of the Long-Range Potential. J Phys Chem. 1996;100:16082–16086.
23. Shvartsburg AA, Jarrold MF. An exact hard-spheres scattering model for the mobilities of polyatomic ions. Chem Phys Lett. 1996;261:86–91.
24. Wyttenbach T, Von Helden G, Batka JJ, Jr, Carlat D, Bowers MT. Effect of the long-range potential on ion mobility measurements. J Am Soc Mass Spectrom. 1997;8:275–282.
25. Valentine SJ, Anderson JG, Ellington AD, Clemmer DE. Disulfide-intact and -reduced lysozyme in the gas phase: Conformations and pathways of folding and unfolding. J Phys Chem B. 1997;101:3891–3900.
26. Hudgins RR, Ratner MA, Jarrold MF. Design of helices that are stable in vacuo. J Am Chem Soc. 1998;120:12974–12975.
27. Counterman AE, Clemmer DE. Large anhydrous polyalanine ions: evidence for extended helices and onset of a more compact state. J Am Chem Soc. 2001;123:1490–1498. [PubMed]
28. Myung S, Badman E, Lee YJ, Clemmer DE. Structural transitions of electrosprayed ubiquitin ions stored in an ion trap over .apprx.10 ms to 30 s. J Phys Chem A. 2002;106:9976–9982.
29. Rosell-Llompart J, Loscertales IG, Bingham D, de la Mora JF. Sizing nanoparticles and ions with a short differential mobility analyzer. J Aerosol Sci. 1996;27:695–719.
30. Labowsky M, de la Mora JF. Novel ion mobility analyzers and filters. J Aerosol Sci. 2006;37:340–362.
31. McMurry PH. A review of atmospheric aerosol measurements. Atmospheric Environment. 2000;34:1959–1999.
32. Purves RW, Guevremont R, Day S, Pipich CW, Matyjaszcayk MS. Mass spectrometric characterization of a high-field asymmetric waveform ion mobility spectrometer. Rev Sci Instrum. 1998;69:4094–4105.
33. Purves RW, Guevremont R. Electrospray ionization high-field asymmetric waveform ion mobility spectrometry-mass spectrometry. Anal Chem. 1999;71:2346–2357. [PubMed]
34. Eiceman GA, Tadjikov B, Krylov E, Nazarov EG, Miller RA, Westbrook J, Funk P. Miniature radio-frequency mobility analyzer as a gas chromatographic detector for oxygen-containing volatile organic compounds, pheromones and other insect attractants. J Chromatogr A. 2001;917:205–217. [PubMed]
35. Shvartsburg AA, Tang K, Smith RD. Modeling the resolution and sensitivity of FAIMS analyses. J Am Soc Mass Spectrom. 2004;15:1487–1498. [PubMed]
36. Venne K, Bonneil E, Eng K, Thibault P. Improvement in peptide detection for proteomics analyses using nanoLC-MS and high-field asymmetry waveform ion mobility mass spectrometry. Anal Chem. 2005;77:2176–2186. [PubMed]
37. Shvartsburg AA, Li F, Tang K, Smith RD. High-resolution field asymmetric waveform ion mobility spectrometry using new planar geometry analyzers. Anal Chem. 2006;78:3706–3714. [PMC free article] [PubMed]
38. Shvartsburg AA, Mashkevich SV, Smith RD. Feasibility of higher-order differential ion mobility separations using new asymmetric waveforms. J Phys Chem A. 2006;110:2663–2673. [PMC free article] [PubMed]
39. Henderson SC, Valentine SJ, Counterman AE, Clemmer DE. ESI/ion trap/ion mobility/time-of-flight mass spectrometry for rapid and sensitive analysis of biomolecular mixtures. Anal Chem. 1999;71:291–301. [PubMed]
40. Srebalus CA, Li J, Marshall WS, Clemmer DE. Gas-phase separations of electrosprayed peptide libraries. Anal Chem. 1999;71:3918–3927. [PubMed]
41. Asbury GR, Klasmeier J, Hill HH., Jr Analysis of explosives using electrospray ionization/ion mobility spectrometry (ESI/IMS) Talanta. 2000;50:1291–1298. [PubMed]
42. Asbury GR, Wu C, Siems WF, Hill HH. Separation and identification of some chemical warfare degradation products using electro-spray high resolution ion mobility spectrometry with mass selected detection. Anal Chim Acta. 2000;404:273–283.
43. Steiner WE, Clowers BH, Haigh PE, Hill HH. Secondary ionization of chemical warfare agent simulants: atmospheric pressure ion mobility time-of-flight mass spectrometry. Anal Chem. 2003;75:6068–6076. [PubMed]
44. Counterman AE, Hilderbrand AE, Srebalus Barnes CA, Clemmer DE. Formation of peptide aggregates during ESI: size, charge, composition, and contributions to noise. J Am Soc Mass Spectrom. 2001;12:1020–1035.
45. Steiner WE, Harden CS, Hong F, Klopsch SJ, Hill HH, McHugh VM. Detection of aqueous phase chemical warfare agent degradation products by negative mode ion mobility time-of-flight mass spectrometry [IM(tof)MS] J Am Soc Mass Spectrom. 2006;17:241–245. [PubMed]
46. St. Louis RH, Siems WF, Hill HH., Jr Ion mobility detection after capillary gas chromatography. Lc-Gc. 1988;6:810–814.
47. Chen YH, Hill HH, Jr, Wittmer DP. Analytical merit of electrospray ion mobility spectrometry as a chromatographic detector. J Microcolumn Separations. 1994;6:515–524.
48. Chen YH, Siems WF, Hill HH., Jr Fourier transform electrospray ion mobility spectrometry. Anal Chim Acta. 1996;334(12):75–84.
49. Hoaglund CS, Valentine SJ, Sporleder CR, Reilly JP, Clemmer DE. Three-dimensional ion mobility/TOFMS analysis of electrosprayed biomolecules. Anal Chem. 1998;70:2236–2242. [PubMed]
50. Bluhm BK, Gillig KJ, Russell DH. Development of a Fourier-transform ion cyclotron resonance mass spectrometer-ion mobility spectrometer. Rev Sci Instrum. 2000;71(11):4078–4086.
51. Gillig KJ, Ruotolo B, Stone EG, Russell DH, Fuhrer K, Gonin M, Schultz AJ. Coupling high-pressure MALDI with ion mobility/orthogonal time-of-flight mass spectrometry. Anal Chem. 2000;72:3965–3971. [PubMed]
52. Hoaglund-Hyzer CS, Clemmer DE. Ion trap/ion mobility/quadrupole/time-of-flight mass spectrometry for peptide mixture analysis. Anal Chem. 2001;73:177–184. [PubMed]
53. Tang K, Shvartsburg AA, Lee H-N, Prior DC, Buschbach MA, Li F, Tolmachev AV, Anderson GA, Smith RD. High-sensitivity ion mobility spectrometry/mass spectrometry using electrodynamic ion funnel interfaces. Anal Chem. 2005;77:3330–3339. [PMC free article] [PubMed]
54. Clowers BH, Siems WF, Hill HH, Massick SM. Hadamard transform ion mobility spectrometry. Anal Chem. 2006;78(1):44–51. [PubMed]
55. Mack E. Average cross-sectional areas of molecules by gaseous diffusion methods. J Am Chem Soc. 1925;47:2468–2482.
56. Shvartsburg AA, Hudgins RR, Dugourd P, Jarrold MF. Structural information from ion mobility measurements: applications to semiconductor clusters. Chem Soc Rev. 2001;30:26–35.
57. Hoaglund-Hyzer CS, Lee YJ, Counterman AE, Clemmer DE. Coupling ion mobility separations, collisional activation techniques, and multiple stages of MS for analysis of complex peptide mixtures. Anal Chem. 2002;74:992–1006. [PubMed]
58. Valentine SJ, Koeniger SL, Clemmer DE. A split-field drift tube for separation and efficient fragmentation of biomolecular ions. Anal Chem. 2003;75:6202–6208. [PubMed]
59. Koeniger SL, Valentine SJ, Myung S, Plasencia M, Lee YJ, Clemmer DE. Development of field modulation in a split-field drift tube for high-throughput multidimensional separations. J Prot Res. 2005;4:25–35. [PubMed]
60. Koeniger SL, Merenbloom SI, Valentine SJ, Jarrold MF, Udseth HR, Smith RD, Clemmer DE. An IMS-IMS analogue of MS-MS. Anal Chem. 2006;78:4161–4174. [PubMed]
61. Clowers BH, Dwivedi P, Steiner WE, Hill HH, Bendiak B. Separation of sodiated isobaric disaccharides and trisaccharides using electrospray ionization-atmospheric pressure ion mobility-time of flight mass spectrometry. J Am Soc Mass Spectrom. 2005;16:660–669. [PubMed]