Magnetoencephalography measures magnetic fields near the scalp and infers a distribution of dipole current sources within the cortex. The rationale is that postsynaptic currents occurring synchronously in a population of tens of thousands of cortical pyramidal cells can be modeled as a single current dipole. Because the cells are aligned almost perfectly parallel to each other, the dipolar fields of all the neurons add constructively. Unfortunately it is difficult to learn the microscopic distribution or even amplitude of the currents based on the detected MEG field pattern since MEG measures fields at the scalp (far from the dipole) where the structure of the dipole is not easily discernable. For example, it can be difficult to distinguish between a single larger dipole or a group of parallel small dipoles (
Hämäläinen et al., 1993). Also, there could be cancellation among local dipoles in the folded cortex (
Hämäläinen et al., 1993). Nevertheless, an estimate of the equivalent dipole current strengths can be made within the assumption of the MEG reconstruction and typical numbers range from 5 to 120 nAm with many robust stimulus paradigms giving dipole sources in the tens of nAm range (
Hari, 1991). The temporal characteristics of the MEG-detected dipoles show a range of responses, most typically temporally asymmetric bipolar and multi-phasic responses with significant spectral power at frequencies below 100 Hz (
Niedermeyer and Lopes da Silva, 1999). In addition to the transient evoked response, oscillatory fields can be evoked, for example using auditory click- trains (
Lin et al., 2004). Thus in this case, a narrow-band peak at the stimulus repetition frequency is clearly visible in the MEG-detected fields. Similar narrow-band fields have been detected by MEG during visual paradigms with alternating checkerboard stimuli (
Tallon-Baudry et al., 2001).
The direct detection of neuronal activity using MR imaging techniques has been pursued for over a decade (
Bandettini et al., 2005;
Bodurka and Bandettini, 2002;
Bodurka et al., 1999;
Konn et al., 2003;
Konn et al., 2004;
Parkes et al., 2007;
Petridou et al., 2006;
Singh, 1994;
Xiong et al., 2003) and the progress has recently been reviewed (
Bandettini et al., 2005). Most of the proposed techniques are based on the detection of phase shifts or dephasing caused by the extremely small changes in the local B
0 field induced by local neuronal currents. The phase shift or dephasing of the detected MR magnetization accumulates during the TE period of either a gradient echo or spin echo sequence. The phase shift experienced by a given spin within the voxel, γ
![[multiply sign in circle]](/corehtml/pmc/pmcents/otimes.gif)
B
neuralTE, can be either positive or negative depending on the sign of the
z-component of the field produced by the local current. The total dephasing measured in the image is proportional to the integral of this phase shift over the voxel volume and its temporal integral over the sensitive period of the sequence. The phantom findings are generally encouraging, with detectable local magnetic fields as small as 0.2 nT (
Bodurka and Bandettini, 2002) and current dipoles as small as 6.3 nAm (
Konn et al., 2003), but the
in vivo literature contains some positive findings (
Chow et al., 2006;
Xiong et al., 2003), but mostly negative findings (
Chu et al., 2004), including a failure to reproduce the the paradigm of Xiong et al (
Parkes et al., 2007).
Understanding the microscopic spatial distribution and time course of the neuronal currents is critical to assessing the potential ability of these MR-based methods to detect neuronal activation. The phase sensitive portion of the MR sequence must be properly timed relative to the transient or oscillating neuronal current to insure that cancellation of the induced phase during a biphasic or multi-phasic current is minimized. For example, the phase sensitive portion of the sequence (the TE period of a gradient recalled echo sequence) might be limited to a single lobe of a biphasic response. Or, for a spin echo sequence, the timing of the 180° pulse can be chosen to occur at the transition between the two phases of the response allowing the biphasic signal to be detected without cancellation (
Singh, 1994). Finally, the conventional methods using T2 and T2* are only sensitive to the component of the local field along the applied external B
0 field.
The spatial distribution of the current within the voxel can also lead to cancellation of the phase shift induced in the MR signal. Even a DC current in a small, straight wire running through the center of the voxel will not induce a net phase shift in the MR signal in that voxel; the current must be offset from the voxel center to produce a net phase accrual (
Bandettini et al., 2005). This arises since the phase is sensitive to the sign of the magnetic field and thus the integral of the phase shift over the voxel is zero for this symmetric case. In other cases, the magnetic field is expected to cancel for points within the voxel. For example, for a uniform distribution of discrete parallel wires there are some areas where the magnetic field itself cancels, but the integral of the phase shift over the region cancels everywhere but at the edge of the distribution (
Park and Lee, 2007). In this case the signal will experience magnitude dephasing but not a net phase shift. In a recent simulation study,
Park et.al. (2007) numerically calculated the effects of neuronal magnetic fields on voxel magnitude and phase, simulating both dendritic as well as axonal currents. They estimate average neuronal fields of 0.3 nT for simultaneous dendritic currents.
Recently, the use of ultra-low-field MRI has been proposed to detect neuronal currents (
Kraus, 2006) (
Kraus et al., 2008). An extremely low B
0 measurement field allows resonant interactions between the neuronal currents and spin magnetization if γB
0 is brought into the frequency range of the neuronal currents. Although their first measurements were at significantly higher fields, Kraus and colleagues proposed static measurement fields weaker than 0.025 Gauss using a SQUID magnetometer to detect the MR signals at a Larmor frequency of less than 100 Hz (
Matlachov et al., 2004;
Volegov et al., 2004). In this scenario, the transient neuronal fields can possibly act as resonant excitation pulse providing the initial excitation of the proton magnetization. Thus, the neuronal current supplies the local B
1 field for excitation and no transverse magnetization should be detected in locations without neuronal currents. Contrary to the high-field MRI-based phase shift approaches mentioned above, multi- phasic or oscillatory currents are beneficial for increasing the tip angle and thus the detected response— provided they have significant spectral power at the Larmor frequency. However, like the high-field phase shift methods, the technique is still susceptible to cancellation effects from the spatial symmetry of the currents within a voxel. Also, the method is sensitive only to the component of the neuronal field perpendicular to the
z axis and the effect is still dependent on the sign of the neuronal current. Thus, a small, straight wire running through the center of the voxel along the static magnetizing field direction produces an excitation giving transverse magnetization along opposite directions in the
x-y plane on opposite sides of the wire. Thus, no net signal from the voxel is expected for this situation. The low-field method is also adversely affected by the small Boltzmann population and other technical confounds of the low-frequency MRI apparatus.
In this work, we introduce a new approach to effectively lower the Larmor frequency below 100 Hz in high-field MR systems based on the spin-locking mechanism (
Abragam, 1961). During a spin-lock, the magnetization is locked in the rotating frame by an RF field, the so-called spin-lock field B
1lock. The effective resonance frequency in the rotating frame during the spin-lock condition is very close to ω
1lock = γB
1lock, deviating only by a potential Bloch-Siegert shift (
Bloch and Siegert, 1940). This latter small shift in resonance frequency results from the counter- rotating field component of a linearly polarized time varying field. The Larmor frequency in the rotating frame can be chosen by setting the RF power of the spin-lock pulse. In this work, we acquire phantom data with γB
1lock ranging from 20 Hz to 200 Hz. T
1ρ relaxation occurs in the spin-lock state from external magnetic field fluctuations with a
z component at the resonant frequency γB
1lock and harmonics.
Redfield described the saturation of the spin-locked magnetization with an external coil and audio frequency source as “rotary saturation”. (
Redfield, 1955) Redfield analyzed rotary saturation in solids using B
1lock fields large compared to the local fields. They derived a steady state magnetization change in the detected spin locked magnetization M
ρ compared to its initial value M
0 for the solid state which is given by:
Where, B
rotarysat is the applied rotary saturation field (a function hopefully carried out by the oscillating neuronal fields), T
1ρ is the spin-lattice relaxation in the spin-locked state and f(ω
1lock) is the Lorentzian shape function which has a value of 1s on resonance, when γB
1lock = γB
rotarysat (
Abragam, 1961).
Since
Eq. 1 was derived for solids and describes a steady state effect, it cannot be used to accurately model the SIRS mechanism. In order to model the effect in liquids and for transitory applications of a rotary saturation field, we model the Rotary saturation effect as a coherent rotation away from the B
1lock direction by the rotary saturation field (neuronal field). The equations of motion for the magnetization in the spin-lock state are formulated in the doubly rotating frame as:
where H
eff is the effective field formed from the applied B
1lock and the off-resonance pseudo-field
![[multiply sign in circle]](/corehtml/pmc/pmcents/otimes.gif)
ω/γ, Mρ denotes the equilibrium magnetization in the rotating frame, which is very small since B
1lock is small. Here H
eff is the relevant axis toward which the magnetization is at equilibrium, is not the static magnetic field direction. If both the spin-lock RF and the Rotary Saturation field B
rotarysat is on resonance, then H
eff is B
1lock. B
rotarysat is the audio-frequency field applied orthogonal to the spin-lock axis (here B
1lock is taken to be along the x-axis in the singly rotating frame). To preserve the analogy to the singly rotating from, the axes of the doubly rotating frame are renamed so that the z-axis is the equilibrium direction. Thus in the doubly rotating frame, the z axis is the direction B
1lock. Then B
rotarysat must be applied orthogonal to B
1lock to produce a rotation. The magnetization relaxes along H
eff with the time constant T1ρ, and in the other two directions with the time constant T
2*. We propose to utilize spin-locking to sensitize the spins to neuronal magnetic fields oscillating at the Larmor frequency in the rotating frame, where γB
1lock ranges between 20 and 500Hz, as set by the applied B
1lock. Oscillating neuronal currents with spectral power at γB
1lock are then capable of producing rotary saturation of the spin-locked magnetization. Thus, the neuronal currents produce a resonant saturation effect on the MR signal in the rotating frame during the spin-lock state, and the spin-lock state can be tuned to be sensitive to a frequency of interest by adjusting γB
1lock. In practice, we anticipate choosing γB
1lock to the expected frequency of the neuronal oscillations to sensitize the sequence to these frequencies. Thus the neuronal currents induced by the stimulus create the rotary saturation effect. In analogy with the classic rotary saturation experiment of
Redfield (1955), we refer to this potential method as Stimulus-Induced Rotary Saturation (SIRS).
To validate the potential of SIRS to detect local current dipoles associated with neuronal activation, we performed two different experiments. The first is the measurement of rotary saturation spectra to characterize the rotary saturation effect from a current dipole. Secondly, we performed a block-design experiment in order to establish detection limits of the method compared to conventional phase-sensitive methods.