Inaccuracies in selected ion monitoring determination of isotope ratios obviated by profile acquisition: Nucleotide 18O/16O measurements1

Abstract

Precise and accurate measurements of isotopologue distributions (IDs) in biological molecules are needed for the determination of isotope effects, quantitation by isotope dilution and quantifying isotope tracers employed in both metabolic and biophysical studies. While single ion monitoring (SIM) yields significantly greater sensitivity and signal/noise than profile mode acquisitions, we show that small changes in the SIM window width and/or center can alter experimentally determined isotope ratios by up to 5%, resulting in significant inaccuracies. This inaccuracy is attributed to mass granularity, the differential distribution of digital data points across the m/z ranges sampled by SIM. Acquiring data in the profile mode and fitting the data to an equation describing a series of equally spaced and identically shaped peaks eliminates the inaccuracies associated with mass granularity with minimal loss of precision. Additionally a method of using the complete ID profile data that inherently corrects for “spillover” and for the natural abundance ID has been used to determine 18O/16O ratios for 5′,3′-guanosine bis-[18O1]phosphate and TM[18O1]P with precisions of ~0.005. The analysis protocol is also applied to quadrupole-time of flight tandem mass spectrometry using [2-18O]uridine and 3′-UM[18O1]P which enhances signal/noise and minimizes concerns for background contamination.

Keywords: isotope ratio, isotopologue distribution, mass isotopomer distribution, tandem mass spectrometry, heavy atom isotope effect, oxygen-18, nucleotide, mass granularity, whole molecule mass spectrometry

Heavy atom isotope effects generated by substituting 13C, 15N or 18O are important tools for probing the mechanisms and transition states of chemical reactions [1], including enzyme-catalyzed reactions [27]. Because the observed effects for isotopic substitution of heavy atoms are extremely small, typically a few percent, they cannot be determined by direct comparison of experimentally determined rate constants. Instead, heavy atom isotope effects are generally measured by internal competition methods that compare the isotopic composition of the reactant and product [3, 8]. These measurements require the ability to determine isotope ratios with precisions better than 1 percent. Such performance is typically achieved using isotope ratio mass spectrometry (IRMS)1 or dual-label scintillation counting of radioactive isotopes [3, 9]. While these methods have proven useful, both possess inherent limitations that restrict their application. IRMS can only analyze small, non-polar gasses such as CO2, N2, and H2; thus, this technique can only be used when the atom of interest can be quantitatively converted to one of these gasses. Scintillation counting requires substantial double-labeling schemes involving specific isotopic pairs (usually 3H and 14C, although others can be used [10]), again limiting the molecules that can be analyzed.

One alternative to these methods is whole molecule mass spectrometry, a technique that was used as early as the 1970s for isotope ratio analysis of formyl-methyl ester and ethanol by gas chromatography quadrupole mass spectrometry (QMS) [11, 12]. Today, the development of gentler ionization methods such as electrospray ionization (ESI) and matrix assisted laser desorption ionization (MALDI) allow larger molecules to be analyzed. Coupled with QMS, ESI makes possible the analysis of isotope ratios for numerous biologically relevant organic molecules with molecular masses in excess of 500 Da [1316]. However, several design features prevent QMS from achieving the precision and accuracy obtained by IRMS. The sources of error in the ESI-QMS isotope ratio measurements include noise (from both the ion source and detector), background, intensity fluctuations, and the inability to capture the entire ion current arising from individual isotopologues2. Some of these effects can be minimized by working with isotope ratios near unity, utilizing background subtraction, and/or injecting analyte by continuous direct infusion.

To enhance sensitivity, with concomitant improvements in signal/noise and precision, isotope ratio QMS studies often use single ion monitoring (SIM) to maximize the amount of time the QMS acquires data with the greatest intensity from the isotopologue peaks of interest. With singly charged ions, the peaks of an isotopologue ion cluster (IIC) are separated by integral increments in m/z. While SIM acquisitions maximize the sensitivity of detection, the studies presented herein demonstrate that this approach is inherently problematic since current QMS instruments collect data at digitally-determined, discrete m/z values which introduces a mass granularity. A result of this mass granularity is that the separation between the discrete m/z values that are analyzed does not permit the peaks arising from individual isotopologues to be sampled at identical positions across the peak. This inherent feature of QMS can compromise the accuracy of isotope ratios measured by SIM [17]. In our studies, shifting the center of the m/z range monitored in SIM analyses by as little as 0.01 could result in variations in the measured isotope ratio of up to 5%. These variations appear to result from differences in the number of the digitized m/z data points averaged by the SIM function for different peaks in the IIC as implemented by the Finnigan Xcalibur software. We demonstrate that this technical difficulty is alleviated by acquiring data in the Profile-Scan mode and fitting the resulting profile data to an equation that accurately describes a series of identically shaped peaks that comprise the IIC. The equation resulting from the fitting process is analytically integrated to determine the relative areas of each peak in the ID, which provides the necessary data to calculate the isotope effect. While the precision obtained by this method is slightly less than that which might be obtained by IRMS, the protocol described here removes a critical source of inaccuracy and yields isotope ratios of sufficient precision for analysis of isotope effects as small as 1.005. Importantly, the ability to analyze isotope ratios in nucleotides will permit the application of heavy atom isotope effect studies to enzymes involved in DNA and RNA metabolism.

Time of flight mass analyzers, now routinely available both as MALDI-TOF and as the mass analyzer for tandem mass spectra in Q-TOF instruments, inherently generate profile scan data so the procedure described is readily applicable to time-of flight data. The peak fitting algorithm developed obviates the necessity of correcting for overlapping peaks in the IIC. An additional benefit of the algorithm presented is that it fits the entire IIC of the labeled compound with its enrichment as the sole variable, automatically accounting for the natural abundance contribution to the ID based on the elemental composition of the analyte.

Materials and Methods

5′-Thymidine mono[nonbridging-18O1]phosphate

Samples of 18O enriched thymidine monophosphate (TMP) were prepared by hydrolysis of thymidine-5′-p-nitrophenyl phosphate (Sigma) in H218O (~50%) water (Isotec) catalyzed by 0.3 M MgCl2, pH 9.1 (50 mM Tris HCl) [18]. Hydrolysis reactions were placed in flame-sealed glass ampoules and incubated at 37°C in a dry incubator for 62 days.

5′,3′-Guanosine bis[5′nonbridging-18O1]phosphate

Samples of 18O enriched pGp were prepared enzymatically by incubating pre-tRNAasp from Bacillus subtilis (pre-tRNA) with E. coli ribonuclease P (RNase P) in H218O (~50%) water. RNase P reactions, 50 mM MES, pH 6, 1 M NaCl and 25 mM MgCl2 were thermostatted at 37°C [19]. RNAs were synthesized by in vitro transcription essentially as described [20]. RNAs were purified by denaturing polyacrylamide gel electrophoresis using gels containing 6% acrylamide and 7 M urea (National Diagnostics). RNA was visualized by UV shadowing, excised from the gel and eluted into 0.3 M sodium acetate, 10 mM Tris HCl (pH 8), 1 mM EDTA, and 0.1% SDS overnight. After hydrolysis catalyzed by RNase P, the tRNA product with a single 18O incorporated into the scissile phosphate was purified by gel electrophoresis prior to digestion overnight with 17 units of ribonuclease T2 in 50 mM NaOAc, pH 4.7, generating the unique [5′-18O1]pGp

[2-18O]Uridine

[2-18O]Uridine was generated by hydrolysis of 250 μmoles of 2,2′-anhydrouridine (Sigma) with 10 mM sodium hydroxide in dimethylformamide for 5 h at 65 C. Uridine was purified from unreacted anhydrouridine and arabinouridine by HPLC.

3′-Uridine-[nonbridging-18O1]phosphate

To enrich one of the non-bridging oxygens of 3′-UMP with 18O, 2′,3′-cyclic UMP was hydrolyzed in base (pH 9.5) in H218O. The reaction was incubated overnight at 65 °C and the resultant 2′-UM[nonbridging-18O]P and 3′-UM[nonbridging -18O]P were resolved by HPLC.

HPLC purification of nucleosides and nucleotides

3′-UM[18O1]P, 5′-TM[18O1]P and [5′-18O1]pGp were purified by HPLC (300 mm × 3.9 mm, 10 μm C18 packing) eluted isocratically using 50 mM diisopropylethylamine (Aldrich), 1% acetic acid in water (pH 4.3) as the mobile phase. For [2-18O]uridine, samples were resolved by a gradient of acetonitrile in 0.2 M ammonium acetate (5% – 20% acetonitrile in 60 min.). Appropriate fractions containing either uridine, UMP, TMP or pGp were collected. Samples were desalted using multiple rounds of lyophilization and resuspension in water. Prior to mass spectrometric analysis, TMP samples were resuspended in methanol:water (50:50) to a concentration of ~50 μM and pGp samples were resuspended in methanol:water:acetic acid (50:50:0.1) to a concentration of ~100 μM.

Electrospray ionization quadrupole mass spectrometry

ESI-QMS measurements were performed using a Finnigan TSQ 7000 instrument controlled by Xcalibur software in negative ion mode. Samples were injected by direct infusion, flow rate 10 μL/min. The electron photomultiplier voltage was 1440 V and the pcthresh signal cutoff was set to 4000. Care was taken to ensure that at the selected pcthresh cutoff, signal noise could be observed even with the source voltage turned off to ensure linearity, even at low signal intensities. The default value of pcthresh was too large and preferentially reduced the intensity of low abundance peaks in the IIC. The capillary temperature for the ion source was 250°C. Unless specified, SIM acquisitions were performed by entering a 0.1 m/z window, which the Xcalibur software defaulted to the minimum 0.16 m/z, and a 0.1 second dwell time. Centers for SIM acquisitions were varied as indicated in the Results and Discussion section. Narrow-range profile SIM scans were acquired to characterize the number of m/z settings used as a function of SIM width and center. Scan acquisitions for pGp covered a range of 439–449 m/z in one second with nominally 25 data points for each increment of 1 m/z. Scan acquisitions for TMP covered 318–328 m/z in one second with nominally 250 points. Data for all acquisitions were averaged over >200 scans. For both types of acquisitions, blank acquisitions, consisting of either methanol:water:acetic acid (50:50:0.1) for pGp samples or methanol:water (50:50) for TMP samples, were subtracted from the sample acquisitions.

Time of flight mass spectrometry

The application of the analysis of isotope ratios by analysis of IDs was extended to tandem-quadrupole time of flight mass spectrometry by analysis of [2-18O]uridine and UM[18O1]P with an Applied Biosystems Q-STAR. All analyses were performed by direct infusion of the nucleoside/nucleotide in 50% acetonitrile 1% formic acid. To determine that the complete IIC is passed through the quadrupole with no fractionation the collision energy was set at 5 V, the mass window center was set in the middle of the IIC and the lo-resolution setting was increased until the relative intensities of the cluster peaks became invariant. The collision energy was then varied to optimize the intensity of the product ion to be analyzed. A series of tandem mass spectra of [2-18O]uridine and UM[18O1]P ~ 50 μM, were acquired with the 18O/16O ratio varied by serial additions of 18O-labeled material. The time of flight spectra were fit to a series of Gaussian peaks as described below.

Non-linear curve fitting

The relative areas of peaks in an identified IIC were determined by non-linear least squares analysis using Origin software. For profile data acquired with different mass spectrometers, the first task is to determine the peak shape. Our data from time-of-flight mass spectrometers, including both MALDI and ESI sources, yielded best fits to Gaussian peak shapes. The individual isotopic peaks in the profile data obtained by ESI-QMS were asymmetric leading peaks typical of quadrupole mass filters [21] but were well fit by an asymmetric double sigmoidal equation where the different leading and lagging exponential decays were accounted for by the terms w2 and w3, respectively, as shown in equation 2.

equation M1
(2)

All of the profile data points were included in the analysis. Equation 3 then describes the entire IIC of n peaks where Ai and xci represent the amplitude and center of the ith peak, respectively.

equation M2
(3)

Constraining w1, w2 and w3 to be invariant results in identically shaped peaks, while for Gaussian peaks only the peak width needs to be invariant. The centers of the individual peaks, xci, may be fixed at a constant interval (typically 1.000 m/z for singly charged ions). These assumptions are based on the premise that the features of the mass separation that lead to peak broadening will behave identically for the molecules containing heavy isotopes3, This simplification of the fitting procedure is validated by the excellent fits obtained. The areas corresponding to the individual sequential peaks of the IIC, are proportional to the amplitudes, Ai. Two advantages of this procedure are: (1) peak overlaps are correctly accounted for and (2) no peak truncation is required.

Simulated mass spectra

To investigate the required signal/noise and the effects of mass granularity, simulated IDs were generated using Excel. IDs were generated according to equation 4; noise was added using the random number function “RAND(#)” whose values were scaled to obtain the desired S/N.

equation M3
(4)

When random error was introduced, each simulation was run 5 times. When analyzing mass granularity, the data points were separated by 0.039 m/z, and distributed so that a data point was positioned at the maximal y value for the second isotopic peak.

Prediction of IDs

As described in the accompanying paper[22], the area of each peak in the ID normalized to the area of the monoisotopic mass peak, M0, can be calculated from the elemental composition of the analyte by equation 5:

equation M4
(5)

Where Mn/M0 is the area of the peak in the ID that has n neutrons in excess of the monoisotopic mass, normalized to the M0, and m1 and m2 are determined by the number of atoms of each element present, e, and the natural abundance of the elements containing 0, 1 and 2 additional neutrons, represented by the mol fractions xe0, xe1 and xe2, respectively. When a sample enriched at a limited number of sites is being analyzed which is the case in isotope effect studies and when isotope coded affinity tags are employed in proteomics [23], the observed ID is a summation of the natural abundance IDs for the unlabeled compound and each enriched peak whose contributions are weighted by their ratio to the unlabelled mol fraction. In the experimental cases we describe, each labeled compound contains a single enriched 18O atom, so the prediction of the ID will be a sum of the unlabeled and labeled IDs, with the labeled ID weighted by the mol fraction ratio, X18 / X16, as shown in equation 6.

equation M5
(6)

To account for the mass offset introduced by the 18O-label, n′ = n−2 (if n′ <0 then Mn′ =0) while m1 and m2 are calculated with one less O atom. The experimentally sought value, X18/X16, is determined by minimizing the rms for the predicted and experimental IDs as shown in equation 7.

equation M6
(7)

An Excel spreadsheet that performs the calculation of m1 and m2 and provides a nonlinear least squares fit of an experimental ID to equation 7 to determine X18/X16 is available online as supplementary material (“18O KIE calc from ID.xls”).

Results and Discussion

ESI-QMS analysis of isotope ratios in nucleotides

A key limitation to mechanistic studies of heavy atom isotope effects has been the requirement for the high precision afforded by IRMS which necessitates the conversion of the atom of interest in the substrates and/or products to CO2, N2 or H2. Advances in ESI-MS instrumentation permit increasing precision in whole molecule measurements of isotopic composition. In addition to numerous biochemical applications, use of ESI-MS to monitor isotope ratios now allows these instruments to be used for the analysis of isotope effects of reactions involving macromolecular substrates [16, 18, 24, 25]. These studies can be extended to numerous reactions in the metabolism of DNA and RNA which involve the hydrolysis of phosphodiester bonds if the IDs of nucleotides can be determined with the requisite precision.

In these reactions hydrolysis results in incorporation of a solvent water molecule into the reactive phosphate. Thus, isotope effects on this reaction can in principle be measured by comparison of the isotopic composition of the water and the product. However, the nucleophilic oxygen is one of four in the product phosphate, and additional oxygens are present on the ribose and nucleobase of the associated nucleoside. Thus, isolation of the nucleophilic oxygen for IRMS analysis would be quite challenging indeed. However, whole molecule mass spectrometry of the resultant nucleotide monophosphates (TMP and pGp) could be used to determine these effects if the 18O/16O ratio can be measured with sufficient precision and accuracy. Pursuing these studies led us to identify a significant problem associated with the accuracy of SIM measurements of isotope ratios.

Effect of mass granularity on observed isotope ratios

SIM acquisitions are routinely used to determine IDs by QMS. However, in the course of analyzing the 18O content of 18O enriched 5′,3′-guanosine bis-phosphate (pGp) by SIM, large effects (up to 5%) on the measured relative intensities of peaks corresponding to the M0 (442 m/z) and M2 (444 m/z) were observed in preliminary control analyses which resulted from uniform small changes (± 0.01 m/z) for the centers of SIM values entered in the SIM table. As shown in Table 1, when the SIM center was offset by only 0.02 there was a > 2% change in the measured M2/M0 ratio.

Table 1
Changes in the relative intensities of M+2 and M+3 for 18O enriched pGp with 0.01 m/z Shifts in SIM Centers

The dependence of the isotope ratio on the chosen SIM values was surprising because the observed peak shapes between the M0 and M2 peaks are essentially identical as shown in Figure 1. Therefore, small changes in the peak center of the same magnitude and direction, were not anticipated to result in significant disparities in the measured isotope ratio. Additionally, the precision of the isotope ratio measurements for two separate pGp samples were very good when maintaining the same SIM table throughout the analysis, with standard deviations of 0.5 % (Table 1S, supplementary material). These two observations strongly suggest that the differences in the measured M2/M0 ratios result not from a lack of precision in the measurements, but from an error in accuracy. This is a particularly pernicious problem, as one day a test of linearity with isotope dilution may work very well, and in subsequent days a shift in calibration of 0.02 m/z could alter the observed ratios by as much as 5%, but the isotope dilution curves determined separately each day would both be linear.

Figure 1
Non-linear curve fit of data from a scan acquisition of 18O enriched TMP

Most commercial quadrupole mass analyzers separate ions into discrete digital m/z values. For the Finnigan TSQ 7000, 65,535 (216-1) digital values exist for an m/z range of about 2,600. Thus, the difference between sequential m/z settings is approximately 0.0397. This mass granularity can influence the measured peak intensity ratios. A simulation of the calculated intensities that would be observed for two sequential isotopic peaks with a theoretical ratio of 1.000 and each data point separated by 0.0397 m/z highlights the problems introduced by mass granularity as shown in Figure 2. The simulation demonstrates that a significant inaccuracy would be introduced if the isotope ratio were determined only by the ratio of the maximum intensity data points (i.e. 0.998/0.982 ≠ 1.000). Note also that the distribution of data points around the lower mass peak differs from that about the higher mass peak is asymmetrical. This difference in data point distribution makes it impossible to define a set peak width for summation that will produce an accurate isotope ratio. A SIM width of 0.16 m/z, corresponding to the minimum allowed by Xcalibur software, would contain either four or five data points. Figure 2 illustrates that this can result in substantially inaccurate relative areas (0.912 vs 0.864). Additionally, this simulation indicates that a 0.02 m/z shift in the SIM center would alter the m/z settings included in the SIM data acquisition (circled points in Figure 2). The change in data points included in the SIM average again alters the calculated M2/M0 ratio. These effects of mass granularity on the accuracy of the isotope ratios are exacerbated by increased resolution, so these simulated effects may be greater than those in our experimental results. Nonetheless, differences in measured isotope ratios resulting from small shifts in the center of the SIM measurements, and the consequent potential for inaccuracies, should be anticipated.

Figure 2
Simulated distribution of the digital data points for two Gaussian peaks separated by 2 on the m/z axis

To more closely assess the two potential errors introduced by mass granularity, SIM data were acquired in profile mode allowing access to all of the data points acquired in contrast to centroid collection which provides only an average of the data points. Data were acquired for TM[18O]P, m/z = 321 and 323). Like the results for pGp, these samples displayed significant changes in the measured M2/M0 ratio with small changes in the SIM center (data not shown). The m/z ranges reported in the SIM profile mode provided clear evidence of the m/z settings averaged to obtain the reported intensity for each peak (Table 2). A plot of intensity against m/z obtained from the SIM-profile mode acquisition with a nominal 0.1 SIM width, displays several distinct regions of relatively stable intensity with abrupt changes as indicated by arrows in Figure 3. We attribute the abrupt changes in intensity to changes in the m/z data points associated with the SIM center, as suggested in Figure 2. Examination of the tabular profile data indicated that the actual SIM widths always defaulted to a minimum of 0.16 m/z. This result was important because it demonstrated that each SIM acquisition (even in profile mode) required data acquisition at 4 or 5 separate m/z settings whose values were averaged, and which points were included in the average could vary when the selected SIM center was translated by as little as 0.01 m/z.

Figure 3
SIM Profile data collected from 18O enriched TMP samples
Table 2
Output m/z ranges for a SIM table input for consecutive input SIM centers and 0.10 m/z SIM windows

Examination of the m/z ranges for each SIM peak revealed an additional concern for the determination of isotope ratios by SIM as the ranges differed for SIM peaks with centers separated by 2.000 m/z by 0.02 (Table 2). The m/z range shifts after every 0.04 increment of the SIM center. This increment coupled with the 0.02 offset, results in significant changes in the M2/M0 ratio occurring with every 0.02 m/z change in the SIM centers, as one of the SIM ranges is changed. For example from Table 2, analyzing the ratio of intensities at m/z = 323 and m/z = 321, measured results with the SIM center at X.00 and X.01 (where X is either 321 or 323) should produce very similar results because the same m/z ranges are analyzed for both peaks. However, once the SIM center is changed to X.02, the m/z range analyzed for the 323 peak has been shifted 0.04 m/z, whereas the m/z range for the 321 peak remains unchanged. This change in included data points will cause a significant change in the measured 323 peak intensity while the measured 321 peak intensity necessarily remains constant, thus resulting in a significant change in the measured 323/321 (i.e., M2/M0) ratio. Continuing to increment the SIM-center, the m/z ranges for both peaks remain constant when the SIM-centers are increased from X.02 to X.03 but when increased to X.04, the m/z range now shifts only for the 321 peak, causing a second significant change in the observed M2/M0 ratio.

Determination of isotope ratios by curve fitting data from scan acquisitions

Next we show that the potential for significant inaccuracies in the measurement of isotope ratios by SIM data acquisitions can be overcome by acquiring data in scan mode and integrating the area of the entire m/z peak. By acquiring data for the entire peak, as opposed to a small portion of the peak, the differential contributions of individual corresponding data points between the isotopic peaks becomes insignificant as shown in the simulation of Figure 2. However, multiple possibilities for integrating the scan data exist with each possessing idiosyncratic strengths and weaknesses.

One common integration strategy is to sum each individual point comprising the peak over a range of 1.00/z, where z is the charge of the analyzed ion. This strategy includes the greatest number of data points for the analysis of the peak intensity. The major analytical difficulty with this approach is accounting for peak overlap, i.e. when adjacent peaks are not baseline separated. Signal overlap can contribute to both inaccuracies and imprecision in measured isotope ratios. With QMS, signal overlap occurs even below 500 m/z as shown in Figure 1. Modern MALDI-TOF and quadrupole-TOF instruments have much higher mass resolutions, but even with these instruments peak overlap can become a problem at higher m/z or with multiply charged ions.

A second strategy which reduces the potential for significant signal overlap is to sum the data points corresponding to only the upper half of each individual peak [26]. This strategy is a compromise that increases the number of data points contributing to the measured peak intensity relative to SIM measurements which reduces the potential of inaccuracy introduced from mass granularity but limits the peak width analyzed to minimize signal overlaps. This protocol has inherent technical difficulties however in determining what constitutes the “upper-half” which may lead to inaccuracies as well.

We advocate fitting the scan data to an equation that describes a series of peaks separated by 1.000/z with identical peak shapes utilizing all of the scan data points acquired across the IIC, an approach adapted from NMR peak integration [27]. This method minimizes the contributions to uncertainty in the measured isotope ratio by mass granularity and peak overlap. An example is shown in Figure 4 where simulated IDs with overlapping Gaussian peaks and added random errors are analyzed. The simulations show that with 60 data points / m/z and a signal/noise ratio of 25 that random noise will introduce an error of < 0.5%. Analysis of these IDs also demonstrates that inaccuracies are introduced due to peak overlap by determining the area of the peaks by simple summation or by utilizing only the upper half of the peaks (data not shown).

Figure 4
Fits of sequential Gaussians can account for spillover with non-baseline resolved peaks

The functional form used to define the peak shape is both instrument and tuning-dependent, while the parameters of the functional form will be depend on both the tuning and mass range of the ID to be analyzed. For QMS, Figure 1 depicts the data and curve fit for the mass spectrum of [18O]TMP to an equation describing the sum of five peaks. The functional form of the peak shape given in equation (2) was determined from the analysis of unlabeled TMP. Various functional forms were tried and the final form chosen provided both the best fit of the profile scan mass spectrum as well as accurate natural abundance values for M1/M0 and M2/M0. The fit to the equation yielded an R2 of 0.9990. The fits for three analyses for each of three different TMP samples ranged from 0.9987 to 0.9991.

Comparison of the precision and accuracy of the SIM and Scan methods

To address the relative accuracy and precision of the isotope ratios determined by SIM and Scan analyses, several methods utilizing both SIM and Scan acquisition modes were used to calculate the isotope ratios for two separate 18O enriched TMP samples using. Three separate SIM methods were utilized. The standard “SIM” method included data from the entire SIM m/z range. The “Common” method included only the data from the m/z ranges common to both peaks. The “SIM Max” method included only data that corresponded to the maximal ion intensity for each peak. These methods were applied to data acquired from SIM tables centered at both 32X.04 and 32X.08, where X can be either 1 or 3.

Four methods for determining peak intensity were applied to data acquired in the profile mode. The “Equation Fit” method fits the scan data to an equation describing a sum of five peaks as described above. The “Peak” method includes only the data point corresponding to the maximal intensity of each peak. The “½Sum” method sums all data points corresponding to the upper half of each peak, and the “FSum” method sums all the data points for each peak within 0.5 m/z from the peak center. The results of calculating the ratio of peak intensities at m/z = 321 and 323 for each of the three [18O]TMP samples are summarized in Table 3.

Table 3
Comparison of M2/M0 determined with different SIM and Profile methods.

Precision is often more critical than accuracy for isotope ratio measurements. For example, in studies examining isotope ratios in several samples of the same molecule, identical m/z channels will be used for the analysis of all samples. Thus, errors in accuracy will remain constant throughout the study and cancel out in the final analysis. In such cases, precision, and not accuracy will limit the utility of the instrument. SIM analysis, because it maximizes analysis time on the m/z values of maximal signal, is generally considered to provide the best precision. However, recognizing that temperature and other factors can result in drift of the mass calibration, these measurements should not be determined on separate days, when the mass axis calibration could vary by 0.02 m/z.

The precision achieved using the Equation Fit method of analyzing data acquired in profile mode compares favorably with the precision of the SIM methods. The standard deviation of the SIM methods for ratios near 1 ranged from 0.002 to 0.009 with a median value of 0.005 and a mode of 0.004. The standard deviations for the Equation Fit analysis of samples 1 and 3 are 0.003 and 0.006, respectively. This level of precision is comparable to the SIM analyses. The standard deviation for sample 2, = 0.014, is substantially larger. However, this large value appears to result from an outlier point in the third trial, as evidenced by its abnormally large ratio for all three scan methods. Overall, precision comparable to SIM analysis appears readily attainable using the Equation Fit method. The worst precision was observed for ratios calculated using the “Peak” method. This observation is not surprising considering the peak intensities come from a single data point acquired in profile mode. Precisions for the two sum methods were slightly better than the Equation Fit method, demonstrating the data acquired in profile mode by the TSQ display outstanding reproducibility overall. However, these sum methods would require a correction for peak overlap to maximize accuracy.

Unfortunately, no absolute standard of comparison exists to assess the accuracy of the various methods for isotope ratios near one. The M2/M0 ratio calculated from the Equation Fit analysis was used as the standard for comparison for the other methods for two reasons. First, it minimizes two known sources of inaccuracy, mass granularity and peak overlap. Second, the method produces ratios for the intensity of m/z = 322 and 321 (M1/M0) in excellent agreement with the values expected due to the natural abundance of 2H, 13C, 15N, and 17O (Table 4). Thus, the method does not appear to introduce a new source of inaccuracy. The SIM methods using the 32X.08 SIM table and the “½Sum” profile method most closely approximated the Equation Fit values. The potential inaccuracy of the SIM methods is demonstrated again by noting the difference in the results using 32X.04 and 32X.08 SIM tables. The close correspondence of the ½Sum method is not surprising as it holds the same advantages of the Equation Fit method, but to a somewhat lesser extent. Although the ½Sum method does not utilize all of the data points from profile acquisitions, it does utilize at least six data points for each peak. This number of data points appears to be sufficient to minimize sources of error from mass granularity. Additionally, because it excludes points at the peak edges, the ½Sum method also minimizes error introduced from signal overlap which is detectable in the FSum analysis. The errors introduced by peak overlap will increase with both higher molecular masses and lower resolution.

Table 4
Natural abundance measurements for TMP and pGp using the Equation Fit profile method.

Isotope ratio measurements by quadrupole instruments are usually most accurate when they are near unity [17]. This generalization results from the equal contribution of errors in background corrections to the intensity of each peak. For ratios far from unity, such contributions affect the less intense peak to a greater extent than the more intense peak, introducing artifacts into the measured ratios. However, measurement of such non-unity ratios is very useful for several applications such as nutritional trace labeling studies [2830]. Additionally, measuring the predicted M1/M0 value represents an important internal control for judging the accuracy of an ESI-QMS quantitation method.

Samples of 18O enriched pGp allow these analyses to be made on a second molecule to test the general applicability of this technique. For pGp, M1/M0 is measured from the peaks at m/z = 443 and 442. The results are presented in Table 4. In all cases, the average measured values are in very close agreement with the predicted values, coming within 0.001 for a relative standard deviation of 0.021. Additionally, the precision of the measurements is satisfactory, with the standard deviation ranging from 0.001 to 0.004. Interestingly, the measured values for the M2/M0 relative intensity displayed the same absolute accuracy and precision as the M1/M0 ratios despite being only one-fourth the magnitude. This result suggests the method can detect small levels of enrichment, even for very low natural abundance levels. The ability to detect such low enrichment may be extremely valuable for measuring the incorporation of isotopically enriched amino acids into proteins [3133].

Overall, the results presented in Table 4 compare favorably with similar work for derivatized amino acids [3436]. These studies used ESI with either ion trap or QMS, as well as fast atom bombardment ionization-QMS. The best results reported from these studies utilized ESI-QMS with separation of derivatized forms of the amino acids arginine and citrulline by HPLC occurring separately from the mass spectrometric analysis [34]. The results presented here indicate a similar method involving offline HPLC purification combined with ESI-QMS analysis also yields extremely accurate and precise natural abundance for nucleotides.

Application to time-of-flight mass analyzers and tandem mass spectrometry

Acquiring QMS data in scan mode obviates the inaccuracies potentially introduced by SIM data acquisition. Time-of-flight mass analyzers inherently acquire data in scan mode where Gaussian peaks shapes are routinely obtained [37]. Time-of-flight mass analyzers have been adopted to ICP isotope ratio measurements [38, 39] where the performance of quadrupole SIM and time-of-flight data acquisition have been compared [40]. Mass bias attributable to mass dependent detection efficiencies can affect the accuracy of the measurements in time-of- flight instruments, however these effects can be minimized [41, 42].

Tandem mass spectrometry has been applied previously to the analysis of IDs [43, 44], however these efforts were directed to characterizing positional isotopomers by fragmenting a single peak from the ID of the parent ion. There are three significant advantages to employing tandem mass spectrometry to the determination of isotope ratios from IDs. First, the potential for contamination is dramatically decreased; second, the background noise is reduced to near zero, enhancing the signal/noise and reducing the errors contributed by background subtraction; and third, the reduction in mass both reduces the contribution of natural abundance isotopes and enhances the resolution of the peaks in the ID as shown in Figure 5.

Figure 5
Tandem mass spectrum of the protonated uracil product ion derived from [2-18O]uridine with the experimental isotopologue ion cluster fit to a series of 5 Gaussian peaks.

The determination of isotope ratios by tandem mass spectrometry faces four prominent objections. First, the fragmentation procedure should not induce a scrambling of the isotopic label. For the 18O nucleotides this is not anticipated to be a problem, although it is a major consideration for identifying 2H present in solvent exchangeable positions [45]. Second, isotope effects on fragmentation can alter the measured isotopic composition in the product ion from that present in the precursor ion [46]. This will be a more significant problem with 2H and when the isotopically labeled site results in a primary isotope effect on the major fragmentation pathway. In the nucleotide studies described, the remote 18O kinetic isotope effects on fragmentation are anticipated to be significantly smaller than the experimental accuracy of the measurements. Third, isotopic fractionation can occur if the entire IIC of precursor ions is not passed through the first mass analyzer and collision chamber. This can be checked by obtaining the time-of-flight mass spectrum of the precursor cluster as a function of isolation width while using a minimal fragmentation energy. When increasing the isolation width results in no further change in the relative intensities of the peaks in the IIC, the accuracy of the isotope ratios can be tested by comparison to the predicted natural abundance spectrum. In the accompanying manuscript the accuracy for peptide fragmentation is documented. Fourth, instrumental mass bias resulting from mass dependent detection efficiencies, can affect the accuracy of the measurements. Because the mass range of the ID

The ID obtained for the ribosyl-3-phosphate[+1] ion by fragmentation of the protonated IIC of 3′-UM[18O1]P is shown in Figure 5. The nonlinear least squares fit of the ID to five sequential Gaussians separated by 1.00 m/z with identical peak widths is shown. No attempts were made to use only the top half as Gaussian peaks can be integrated across the entire peak with minimal loss of precision [47]. The peak areas obtained for each Gaussian were fit to equation 7 with the mol fraction ratio (X 18O/X 16O) the only adjustable parameter in Excel using “18O KIE calc from ID.xls” included as Supplementary Material. The mol fraction ratio was varied systematically by sequential additions of 3′-UM[18O1]P which increases X 18O/X 16O of the analyzed solution. The result of these analyses is shown in Figure 6. The linearity of the data and the standard deviations of the replicates indicate that the precision of the isotope ratios obtained from the product ion should be sufficient to permit heavy atom isotope effects to be determined on nucleotides. Similar results were obtained for the protonated base product ion when [1-18O]uridine was analyzed (data not shown). The procedures described to obtain and analyze IDs by tandem MS for nucleotides (and peptides [22]) will permit heavy atom isotope effect studies to be conducted on these biomacromolecular substrates.

Figure 6
Graph of XO18 /XO16 observed as calculated by a non-linear least squares fit to equation 7 (solid circles) versus the value calculated based on mass balance.

Supplementary Material

Acknowledgments

We appreciate the assistance of Steven Ingalls for assistance with data acquisition on the Finnigan TSQ 7000. Footnotes:

Footnotes

1This work supported by NIH grants GM56740 (M.E.H.), R33 DK07029 (VEA & SP) and a HHMI predoctoral fellowship (A.G.C.).

1Abbreviations used: IRMS, isotope ratio mass spectrometry; ESI, electrospray ionisation; ID, isotopologue distribution; IIC, isotopologue ion cluster; MALDI, matrix assisted laser desorption ionisation; QMS, quadrupole mass spectrometry; SIM, selected ion monitoring

2Isotopologues are molecules that differ by the presence of different isotopes resulting in an isotopologue ion cluster in mass spectra. The areas of all of the isotopologue peaks comprise the isotopologue distribution. The area of the peak associated with each isotopologue is denoted Mn, where n is the number of neutrons in excess of that present in the monoisotopic mass ion, M0.

3It is recognized at higher mass resolutions that additional peak broadening occurs in the ID due to the presence of multiple isotopologues and the differences in mass associated with the substitution of 2H (1.006), 13C (1.003), 15N (0.997) and 18O (2.004). In this case w will increase with n [48].

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final citable form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

1. Fry A. Heavy atom isotope effects in organic reaction mechanism studies. In: Collins CJ, Bowman NS, editors. Isotope effects in chemical reactions. Van Nostrand-Reinhold; Princeton, N.J: 1970. pp. 364–414.
2. Cleland WW. Determining the chemical mechanisms of enzyme-catalyzed reactions by kinetic studies. Adv Enzymol Relat Areas Mol Biol. 1977;45:273–387. [PubMed]
3. O’Leary MH. Determination of heavy-atom isotope effects on enzyme-catalyzed reactions. Meth Enzymol. 1980;64:83–104. [PubMed]
4. Cook PF, editor. Enzyme mechanism from isotope effects. CRC Press; Boca Raton, FL: 1991.
5. Schramm VL. Enzymatic transition states and transition state analog design. Annu Rev Biochem. 1998;67:693–720. [PubMed]
6. Cleland WW. The use of isotope effects to determine enzyme mechanisms. Arch Biochem Biophys. 2005;433:2–12. [PubMed]
7. Kohen A, Limbach HH, editors. Isotope effects in chemistry and biology. CRC; Boca Raton, FL: 2006.
8. Bigeleisen J, Wolfsberg M. Theoretical and experimental aspects of isotope effects in chemical kinetics. Adv Chem Phys. 1958;1:15–76.
9. Schramm VL. Enzymatic transition-state analysis and transition-state analogs. Meth Enzymol. 1999;308:301–355. [PubMed]
10. Matsson O. Isotope effects for exotic nuclei. In: Kohen A, Limbach HH, editors. Isotope effects in chemistry and biology. CRC Press; Boca Raton: 2006. pp. 417–431.
11. Sawyer CB, Kirsch JF. Kinetic isotope effects for reactions of methyl formate-methoxyl-18O. J Am Chem Soc. 1973;95:7375–7381.
12. Sawyer CB, Kirsch JF. Kinetic isotope effects for the chymotrypsin catalyzed hydrolysis of ethoxyl-18O labeled specific ester substrates. J Am Chem Soc. 1975;97:1963–1964. [PubMed]
13. Bahnson BJ, Anderson VE. Crotonase-catalyzed β-elimination is concerted: A double isotope effect study. Biochemistry. 1991;30:5894–5906. [PubMed]
14. Berti PJ, Blanke SR, Schramm VL. Transition state structure for the hydrolysis of NAD+ catalyzed by diphtheria toxin. J Am Chem Soc. 1997;119:12079–12088. [PMC free article] [PubMed]
15. Goshe MB, Chen YH, Anderson VE. Identification of the sites of hydroxyl radical reaction with peptides by hydrogen/deuterium exchange: Prevalence of reactions with the side chains. Biochemistry. 2000;39:1761–1770. [PubMed]
16. Seila AC, Okuda K, Nunez S, Seila AF, Strobel SA. Kinetic isotope effect analysis of the ribosomal peptidyl transferase reaction. Biochemistry. 2005;44:4018–4027. [PubMed]
17. Begley IS, Sharp BL. Characterisation and correction of instrumental bias in inductively coupled plasma quadrupole mass spectrometry for accurate measurement of lead isotope ratios. J Anal At Spectrom. 1997;12:395–402.
18. Cassano AG, Anderson VE, Harris ME. Analysis of solvent nucleophile isotope effects: Evidence for concerted mechanisms and nucleophilic activation by metal coordination in non-enzymatic and ribozyme catalyzed phosphodiester hydrolysis. Biochemistry. 2004;43:10547–10559. [PubMed]
19. Kaye NM, Zahler NH, Christian EL, Harris ME. Conservation of helical structure contributes to functional metal ion interactions in the catalytic domain of ribonuclease P RNA. J Mol Biol. 2002;324:429–442. [PubMed]
20. Christian EL, McPheeters DS, Harris ME. Identification of individual nucleotides in the bacterial ribonuclease P ribozyme adjacent to the pre-tRNA cleavage site by short-range photo-cross-linking. Biochemistry. 1998;37:17618–17628. [PubMed]
21. Blaum K, Geppert C, Muller P, Nortershauser W, Wendt K, Bushaw BA. Peak shape for a quadrupole mass spectrometer: Comparison of computer simulation and experiment. Int J Mass Spectrom. 2000;202:81–89.
22. Wang B, Sun G, Anderson DR, Jia M, Previs S, Anderson VE. Mass isotopomer distributions of peptide fragment ions by tandem mass spectrometry: Quantitation of low levels of deuterium incorporation. Anal Biochem. 2007 submitted. [PMC free article] [PubMed]
23. Gygi SP, Rist B, Gerber SA, Turecek F, Gelb MH, Aebersold R. Quantitative analysis of complex protein mixtures using isotope-coded affinity tags. Nat Biotechnol. 1999;17:994–999. [PubMed]
24. Cassano AG, Anderson VE, Harris ME. Evidence for direct attack by hydroxide in phosphodiester hydrolysis. J Am Chem Soc. 2002;124:10964. [PubMed]
25. Park H, Girdaukas GG, Northrop DB. Effect of pressure on a heavy-atom isotope effect of yeast alcohol dehydrogenase. J Am Chem Soc. 2006;128:1868–1872. [PubMed]
26. Begley IS, Scrimgeour CM. High-precision δ2H and δ18O measurement for water and volatile organic compounds by continuous-flow pyrolysis isotope ratio mass spectrometry. Anal Chem. 1997;69:1530–1535.
27. Kumar A, Sotak CH, Dumoulin CL, Levy GC. Software for deconvolution of overlapping spectral peaks and quantitative analysis by carbon-13 Fourier transform NMR spectroscopy. Computer-Enhanced Spectroscopy. 1983;1:107–114.
28. Reid S, Shackleton C, Wu K, Kaempfer S, Hellerstein MK. Liquid chromatography/mass spectrometry of plasma glucose and secreted glucuronate for metabolic studies in humans. Biomed Environ Mass Spectrom. 1990;19:535–540. [PubMed]
29. Patterson BW, Zhang XJ, Chen Y, Klein S, Wolfe RR. Measurement of very low stable isotope enrichments by gas chromatography/mass spectrometry: Application to measurement of muscle protein synthesis. Metabolism. 1997;46:943–948. [PubMed]
30. Katanik J, McCabe BJ, Brunengraber DZ, Chandramouli V, Nishiyama FJ, Anderson VE, Previs SF. Measuring gluconeogenesis using a low dose of 2H2O: Advantage of isotope fractionation during gas chromatography. Am J Physiol Endocrinol Metab. 2003;284:E1043–1048. [PubMed]
31. Papageorgopoulos C, Caldwell K, Shackleton C, Schweingrubber H, Hellerstein MK. Measuring protein synthesis by mass isotopomer distribution analysis (MIDA) Anal Biochem. 1999;267:1–16. [PubMed]
32. Papageorgopoulos C, Caldwell K, Schweingrubber H, Neese RA, Shackleton CH, Hellerstein M. Measuring synthesis rates of muscle creatine kinase and myosin with stable isotopes and mass spectrometry. Anal Biochem. 2002;309:1–10. [PubMed]
33. Busch R, Kim YK, Neese RA, Schade-Serin V, Collins M, Awada M, Gardner JL, Beysen C, Marino ME, Misell LM, Hellerstein MK. Measurement of protein turnover rates by heavy water labeling of nonessential amino acids. Biochim Biophys Acta. 2006;1760:730–744. [PubMed]
34. Lagerwerf FM, Wever RM, van Rijn HJ, Versluis C, Heerma W, Haverkamp J, Koomans HA, Rabelink TJ, Boer P. Assessment of nitric oxide production by measurement of [15N]citrulline enrichment in human plasma using high-performance liquid chromatography-mass spectrometry. Anal Biochem. 1998;257:45–52. [PubMed]
35. van Eijk HM, Rooyakkers DR, Soeters PB, Deutz NE. Determination of amino acid isotope enrichment using liquid chromatography-mass spectrometry. Anal Biochem. 1999;271:8–17. [PubMed]
36. Hess S, van Beek J, Pannell LK. Acid hydrolysis of silk fibroins and determination of the enrichment of isotopically labeled amino acids using precolumn derivatization and high-performance liquid chromatography-electrospray ionization-mass spectrometry. Anal Biochem. 2002;311:19–26. [PubMed]
37. Guilhaus M. Principles and instrumentation in time-of-flight mass spectrometry. Physical and instrumental concepts. J Mass Spectrom. 1995;30:1519–1532.
38. Tian X, Emteborg H, Adams FC. Analytical performance of axial inductively coupled plasma time of flight mass spectrometry (ICP-TOFMS) J Anal At Spectrom. 1999;14:1807–1814.
39. Koumenis IL, Vestal ML, Yergey AL, Abrams S, Deming SN, Hutchens TW. Quantitation of metal isotope ratios by laser desorption time-of-flight mass spectrometry. Anal Chem. 1995;67:4557–4564. [PubMed]
40. Vazquez Pelaez M, Costa-Fernandez JM, Sanz-Medel A. Critical comparison between quadrupole and time-of-flight inductively coupled plasma mass spectrometers for isotope ratio measurements in elemental speciation. J Anal At Spectrom. 2002;17:950–957.
41. Axelsson J, Scrivener E, Haddleton DM, Derrick PJ. Mass discrimination effects in an ion detector and other causes for shifts in polymer mass distributions measured by matrix-assisted laser desorption/ionization time-of-flight mass spectrometry. Macromolecules. 1996;29:8875–8882.
42. Benkhedda K, Goenaga Infante H, Adams FC. Determination of total lead and lead isotope ratios in natural waters by inductively coupled plasma time-of-flight mass spectrometry after flow injection on-line preconcentration. Analyt Chim Acta. 2004;506:137–144.
43. Jeffrey FM, Roach JS, Storey CJ, Sherry AD, Malloy CR. 13C isotopomer analysis of glutamate by tandem mass spectrometry. Anal Biochem. 2002;300:192–205. [PubMed]
44. Rantanen A, Rousu J, Kokkonen JT, Tarkiainen V, Ketola RA. Computing positional isotopomer distributions from tandem mass spectrometric data. Metab Eng. 2002;4:285–294. [PubMed]
45. Hoerner JK, Xiao H, Dobo A, Kaltashov IA. Is there hydrogen scrambling in the gas phase? Energetic and structural determinants of proton mobility within protein ions. J Am Chem Soc. 2004;126:7709–7717. [PubMed]
46. Green JR, Cooks RG. Inverse heavy-atom kinetic isotope effects in chloroalkanes. J Phys Chem A. 2004;108:10039–10043.
47. Piepmeier EH. Improving the precision of the integral of a transient absorbance peak. Anal Chem. 1976;48:1296–1300.
48. Yergey J. A general approach to calculating isotopic distributions for mass spectrometry. Int J Mass Spectrom Ion Phys. 1983;52:337–349.