Heart. 2007 November; 93(11): 1426–1432.
Published online 2007 February 3. doi:  10.1136/hrt.2006.087445
PMCID: PMC2016892

The interaction of interventricular pacing intervals and left ventricular lead position during temporary biventricular pacing evaluated by tissue Doppler imaging

Abstract

Objective

To determine the effects of interventricular pacing interval and left ventricular (LV) pacing site on ventricular dyssynchrony and function at baseline and during biventricular pacing, using tissue Doppler imaging.

Methods

Using an angioplasty wire to pace the left ventricle, 20 patients with heart failure and left bundle branch block underwent temporary biventricular pacing from lateral (n = 20) and inferior (n = 10) LV sites at five interventricular pacing intervals: +80, +40, synchronous, −40, and −80 ms.

Results

LV ejection fraction (EF) increased (mean (SD) from 18 (8)% to 26 (10)% (p = 0.016) and global mechanical dyssynchrony decreased from 187 (91) ms to 97 (63) ms (p = 0.0004) with synchronous biventricular pacing compared to unpaced baseline. Sequential pacing with LV preactivation produced incremental improvements in EF and global mechanical dyssynchrony (p<0.0001 and p = 0.0026, respectively), primarily as a result of reductions in inter‐LV–RV dyssynchrony (p = 0.0001) rather than intra‐LV dyssynchrony (NS). Results of biventricular pacing from an inferior or lateral LV site were comparable (for example, synchronous biventricular pacing, global mechanical dyssynchrony: lateral LV site, 97 (63) ms; inferior LV site, 104 (41) ms (NS); EF: lateral LV site, 26 (10)%; inferior LV site, 27 (10)% (NS)). ECG morphology was identical during biventricular pacing through an angioplasty wire and a permanent lead.

Conclusions

Sequential biventricular pacing with LV preactivation most often optimises LV synchrony and EF. An inferior LV site offers a good alternative to a lateral site. Pacing through an angioplasty wire may be useful in assessing the acute effects of pacing.

Keywords: biventricular pacing, lead position, interventricular pacing interval, tissue Doppler imaging

Biventricular pacing is an effective adjuvant treatment for selected patients with chronic refractory heart failure and left bundle branch block.1,2 Around 20–30% of patients implanted on conventional indications fail to improve clinically.3 The selection of the relative timing of left ventricular (LV) and right ventricular (RV) stimulation, and left ventricular lead position may be important in optimising cardiac performance.4,5

Using tissue Doppler derived measures of ventricular dyssynchrony and an angioplasty wire pacing technique, we examined the combined influence of temporary left ventricular pacing from different coronary sinus tributaries over a range of interventricular pacing intervals, in the same patients.

Methods

All patients had severely impaired left ventricular function (ejection fraction [less-than-or-eq, slant]30%), heart failure (New York Heart Association functional class III–IV), and sinus rhythm with left bundle branch block (QRS [gt-or-equal, slanted]120 ms). Each patient gave informed consent for this study, which was approved by the local ethics committee.

Temporary pacing protocol

Quadripolar catheters were positioned at the high right atrium and right ventricular apex. A 6 F AL1 guide catheter was used to cannulate the coronary sinus from a femoral approach. Coronary sinus angiography was acquired in three orthogonal views. A 0.014 inch angioplasty wire (Choice, Scimed) insulated with an uninflated angioplasty balloon (Maverick Scimed) advanced to expose <1 cm terminal tip, was used to pace the lateral free wall (n = 20). The quadripolar catheters and angioplasty wire were connected to a biventricular pacemaker (Medtronic InSync III 8042). Each electrode was tested in VVI or AAI mode. Bipolar configurations were used for atrial and right ventricular pacing. The right ventricular ring was the indifferent electrode for the unipolar left ventricular lead for left ventricular sensing and pacing. Biventricular VDD pacing with an output twice the capture threshold, or below the anodal capture threshold, was programmed to ensure continuous pacing. Biventricular pacing at five different interventricular pacing intervals (−80, −40, synchronous, +40 and +80 ms referenced to right ventricular pacing) was carried out with continuous recording of surface and intracardiac electrograms.

Mean unpaced PR interval was 188 (30) ms. Atrioventricular delays of 80–120 ms were programmed during synchronous biventricular pacing,5,6,7 but were not individually optimised. The pacemaker used in this study delivers the first ventricular stimulus at the programmed atrioventricular interval and delivers the second stimulus at the programmed interventricular interval. During left ventricular preactivation, the second right ventricular stimulus is delayed with a constant left sided atrioventricular delay. During right ventricular preactivation, the second left ventricular stimulus is delayed and the atrioventricular delay shortened to maintain left sided atrioventricular synchrony. With each pacing configuration, a progressive change in QRS morphology confirmed ventricular capture, while comparison with QRS morphology during VVI and VDD pacing excluded the possibility of fusion. After five minutes of pacing for each configuration, echocardiographic data were acquired.

A second inferior left ventricular site was paced through an angioplasty wire in the middle cardiac vein (n = 10) and the protocol was repeated. The anatomical site of pacing (lateral or inferior) was defined by the final wire tip position determined by orthogonal fluoroscopic projections. The calibrated surface measurement of distance between left ventricular tip positions confirmed that left ventricular pacing was from disparate sites.

In four patients, the anterior left ventricular wall was paced through the great cardiac vein and the protocol was repeated.

Angioplasty wire

Temporary pacing using an angioplasty wire has been described during coronary intervention, and recently during biventricular pacing.8,9,10 The angioplasty wire is insulated along the shaft with impedance >20 megaohms to the distal 3 cm tip. The balloon was advanced to expose <1 cm of the distal tip, providing a unipolar electrode. Pacing from the wire was validated during five pacemaker implants. Comparable QRS morphology was observed during left ventricular pacing through an angioplasty wire and through an over‐the‐wire lead (4193 Medtronic) advanced under fluoroscopy to the same position. Diaphragmatic pacing was not observed in any of these five implants.

Echocardiographic data acquisition

All echocardiographic data were acquired using an ATL 5000 ultrasound system. Patients had a baseline conventional echocardiographic examination. The ejection fraction was calculated using Simpson's equation and the apical four chamber view.11

Tissue Doppler imaging enables quantitative assessment of regional systolic function and ventricular dyssynchrony.12,13,14 Pulsed wave tissue Doppler traces were recorded with a surface electrocardiogram during end expiration, using apical four, two, and three chamber views with a 5 mm sample volume on the basal septum, lateral, anterior, inferior, and posterior left ventricular walls, and from the right ventricular free wall at the level of the tricuspid valve annulus. Tissue Doppler imaging was undertaken at baseline, during each different pacing configuration, and at each left ventricular pacing site. Data were analysed offline using HDI lab software (ATL Corporation).

Tissue Doppler imaging measurements

From each tissue Doppler trace, peak systolic velocity and electromechanical delay were measured, with mean values calculated from 3 cardiac cycles. A single, blinded operator analysed all data.

Electromechanical delay was defined as the time from QRS onset to the beginning of systolic contraction (S wave). The longest electromechanical delay within the left ventricle defined the latest site of left ventricular contraction. Maximum dispersion in electromechanical delay between different left ventricular segments defined individual intra‐LV dyssynchrony. Inter‐ LV–RV dyssynchrony was calculated as maximal dispersion in left and right ventricular electromechanical delay. The sum of inter‐LV–RV and intra‐LV dyssynchrony provides a measure of global mechanical dyssynchrony and is a good predictor of left ventricular functional recovery and reverse remodelling after biventricular pacing.13 Systolic velocity was measured as the peak of the S wave (cm/s).

Reproducibility

The reproducibility of systolic velocity, electromechanical delay, and ejection fraction were assessed in 15 random examinations as the mean difference between two independent measurements carried out on different occasions by one observer (intraobserver variability) and between two independent blinded observers (interobserver variability).

Statistical analysis

Continuous data are expressed as the mean (SD) unless specified otherwise. To compare effects of interventricular pacing intervals for each pacing site, we used repeated measures analysis of variance (ANOVA) and the paired t test. Correlation coefficients were calculated by the Pearson product‐moment method. Statistical analyses were made with SPSS version 10.0 and Statview version 5.0. A probability (p) value of <0.05 was considered significant.

Results

Patient characteristics

The baseline characteristics of twenty patients are summarised in table 11.. Ninety five per cent of the patients were on either angiotensin converting enzyme (ACE) inhibitors or angiotensin receptor blockers, 50% were on β blockers, and 65% were on spironolactone. All the patients required oral diuretics.

Table thumbnail
Table 1 Patient baseline demographic variables

All patients were paced from a lateral left ventricular site, and 10 of 20 (50%) were paced from a second inferior left ventricular site. In a left anterior oblique projection, the mean separation between lateral and inferior pacing sites was 39 (8) mm. Four patients were paced from an anterior left ventricular site; owing to the small numbers, only limited data are presented. In all patients, right ventricular pacing was from the right ventricular apex.

Reproducibility

There was good reproducibility of measurement for electromechanical delay, systolic velocity, and ejection fraction, with intraobserver coefficient of variance of 3.2%, 6.0%, and 9.0%, respectively, and interobserver coefficient of variance of 7.7%, 8.3%, and 13%.

Effect of pacing configuration on left ventricular ejection fraction

Left ventricular ejection fraction increased between unpaced and synchronous biventricular pacing with a lateral left ventricular pacing site (from 18 (8)% to 26 (10)%, p = 0.016). From an inferior left ventricular pacing site, the ejection fraction increased to 27 (10)% (NS). There was no significant difference in ejection fraction between lateral and inferior left ventricular pacing sites. During pacing there was a gradual increase in ejection fraction with increasing left ventricular preactivation (p<0.0001 by ANOVA), the increases in ejection fraction during left ventricular preactivation being significantly greater than synchronous biventricular pacing (fig 11).). Left ventricular end diastolic diameter was unchanged from baseline during each of the pacing configurations.

figure ht87445.f1
Figure 1 Mean ejection fraction (EF) at baseline and during biventricular pacing from lateral and inferior left ventricular (LV) pacing sites, at each of five interventricular pacing intervals. Error bars = SEM; p values are shown ...

Effect of pacing configuration on regional electromechanical delay and mechanical dyssynchrony

During the baseline unpaced state, mean right ventricular contraction occurred notably ahead of left ventricular contraction. Within the left ventricle, mean septal contraction occurred earliest, followed by anterior and inferior left ventricular wall contraction, with posterior and lateral left ventricular walls significantly delayed, accounting for inter‐LV–RV, intra‐LV, and global mechanical dyssynchrony.

During biventricular pacing with right ventricular preactivation from both lateral and inferior left ventricular pacing sites, mean right ventricular contraction was modestly delayed by approximately 20 ms, while mean lateral left ventricular wall contraction was advanced. Onset of mean septal contraction was unchanged during pacing from a lateral left ventricular pacing site but was advanced during pacing from the inferior left ventricular wall.

During synchronous biventricular pacing from a lateral or inferior left ventricular site, mean right ventricular contraction was delayed by approximately 50 ms. With an inferior left ventricular pacing site, mean septal contraction was delayed while electromechanical delays in other regions were virtually unchanged. With a lateral left ventricular pacing site, mean septal contraction was substantially delayed, lateral left ventricular wall contraction was minimally delayed, while mean electromechanical delay for the remaining left ventricular regions increased by approximately 15 ms.

With left ventricular preactivation from an inferior left ventricular pacing site, mean right ventricular contraction was significantly delayed, with a small advance in mean lateral left ventricular wall contraction and little change in the remaining left ventricular regions. With a lateral left ventricular pacing site there was a marked delay in mean right ventricular contraction, beyond that seen during synchronous pacing. There was a minor delay in mean lateral wall electromechanical delay, with significant delays in the remaining left ventricular regions.

These changes in mean regional electromechanical delays explain the reductions in intra‐LV, inter‐LV–RV, and global mechanical dyssynchrony observed with different pacing configurations when compared with the baseline unpaced state ((figsfigs 2 and 33;; table 22).

figure ht87445.f2
Figure 2 Mean regional right and left ventricular (RV and LV) electromechanical delays (ms) with no pacing and during biventricular pacing from a lateral and inferior LV pacing site during maximal RV preactivation (A), synchronous biventricular ...
figure ht87445.f3
Figure 3 Mean global mechanical dyssynchrony (sum of inter‐LV–RV and intra‐LV dyssynchrony) at baseline and during biventricular pacing from lateral and inferior left ventricular pacing sites, at each of five interventricular ...
Table thumbnail
Table 2 Effects on left ventricular pacing site and interventricular pacing interval on intra‐LV and inter‐LV–RV dyssynchrony

Intra‐LV and inter‐LV–RV dyssynchrony were significantly reduced with synchronous biventricular pacing from both left ventricular pacing sites. Altering the interventricular pacing interval made no significant difference to intra‐LV dyssynchrony, but inter‐LV–RV dyssynchrony decreased with increasing left ventricular preactivation (p = 0.0001 by ANOVA). There was no significant difference between lateral and inferior left ventricular pacing sites for any interventricular pacing interval. (table 22)

Global mechanical dyssynchrony (sum of inter‐LV–RV and intra‐LV dyssynchrony) significantly decreased with synchronous biventricular pacing, independent of left ventricular pacing site (187 (91) ms, unpaced, to 97 (63) ms, p = 0.0004, lateral left ventricular site; and 104 (41) ms, p = 0.018, inferior left ventricular site). There was a gradual decrease in global mechanical dyssynchrony with increasing left ventricular preactivation, with no significant difference between lateral and inferior left ventricular pacing sites (fig 33).

Optimal resynchronisation

From a lateral left ventricular pacing site, biventricular pacing with left ventricular preactivation provided optimal mechanical resynchronisation or lowest global mechanical dyssynchrony in 15 of the 20 patients (75%) and in eight of 11 patients (73%) with dilated cardiomyopathy. Optimal resynchronisation occurred in a minority of patients (3/20 or 15%) during synchronous biventricular pacing, and in two of 20 patients (10%) with right ventricular preactivation, both of whom had ischaemic heart disease. From an inferior left ventricular pacing site the interventricular pacing interval producing optimal resynchronisation was more variable: three of 10 (30%) with right ventricular preactivation; two of 10 (20%) with synchronous pacing; and five of 10 (50%) with left ventricular preactivation. In all patients with ischaemic heart disease paced from a lateral left ventricular site, optimal resynchronisation occurred during sequential biventricular pacing (two of nine (22%) with right ventricular preactivation, and seven of nine (78%) with left ventricular preactivation).

During synchronous biventricular pacing from an anterior left ventricular site (n = 4) there was no significant difference in ejection fraction, intra‐LV, inter‐ LV–RV, or global mechanical dyssynchrony compared with the unpaced state. These data were not included in further analyses.

The effect of pacing configuration on systolic velocity

Mean septal systolic velocity increased with synchronous biventricular pacing (3.5 (1.0) cm/s unpaced to 4.6 (1.7) cm/s (p = 0.002), lateral left ventricular site; and 4.8 (0.9) cm/s (p = 0.08), inferior left ventricular site). There was a significant increase in septal systolic velocity with increasing left ventricular preactivation (p = 0.0002) without any difference between lateral and inferior left ventricular pacing sites

ECG and mechanical dyssynchrony

QRS duration correlated positively with inter‐LV–RV dyssynchrony (r = 0.5, p = 0.028), but not with intra–LV dyssynchrony (r = 0.27, p = 0.26) when unpaced. There were no correlations between ECG and mechanical dyssynchrony during biventricular pacing.

Complications

There were no complications during the study.

Follow up data

Following this temporary pacing study, 11 of 20 patients (55%) were subsequently enrolled in the CARE‐HF multicentre trial,15 with six patients randomised to biventricular pacing and five to medical treatment. Of the remaining nine patients, one died before permanent pacing from a non‐cardiac cause, one developed a device related infection necessitating system extraction at two months, and in one patient coronary sinus lead placement at permanent implantation was not possible owing to the small calibre of suitable venous tributaries, which had been identified as a likely problem during the temporary pacing study. Chronic follow up data at six months post‐biventricular pacemaker insertion are presented for 12 of 20 patients who underwent this temporary pacing study.

Individual data acquired during the temporary pacing study were used to guide left ventricular lead placement at implant, and in the programming of interventricular pacing intervals. Eleven of the 12 patients (92%) had implantation at a lateral left ventricular pacing site. Before discharge, the interventricular pacing interval was programmed according to the lowest measure of global mechanical dyssynchrony for the implanted left ventricular site during the acute study. At six months post‐implant, 11 of the 12 patients (92%) were symptomatically improved, with a significant reduction in mean baseline NYHA functional grade from (mean (SD)) 3.5 (0.5) to 2.4 (0.5) during biventricular pacing (p<0.001). Chronic follow up data appeared to correspond well with acute data presented in this study. The left ventricular ejection fraction was significantly increased from baseline (18 (6)% to 24 (8)%, p<0.001), while intra‐LV, inter‐LV–RV, and global mechanical dyssynchrony were significantly reduced from baseline (79 (60) to 42 (24) ms, p = 0.04; 123 (52) to 62 (34) ms, p<0.001; and 202 (107) to 104 (59) ms, p = 0.03, respectively). In the patient with no symptomatic improvement at follow up who was enrolled in the CARE‐HF study (patient 7 in table 11),), there was minimal evidence of baseline dyssynchrony with no significant changes in these variables, either during the temporary pacing study or at follow up.

Discussion

Several studies have now shown sequential biventricular pacing may be superior to simultaneous right ventricular and left ventricular stimulation.5,16,17,18 There has been conflicting evidence over the importance of left ventricular lead position in studies which have largely compared lateral versus anterior left ventricular sites.4,19,20,21 However, the combined effects of different interventricular pacing intervals and left ventricular stimulation sites have yet to be fully defined. This study provides a direct comparison within the same patients of different interventricular pacing intervals, suggesting that sequential biventricular pacing, most often using left ventricular preactivation, may further optimise the effects on mechanical dyssynchrony and systolic performance. In addition, it directly compares the effect on mechanical function of biventricular pacing from lateral and inferior left ventricular pacing sites, without evidence that one left ventricular site is more favourable than the other.

Effect of interventricular pacing intervals

Sogaard et al5 found that sequential biventricular pacing was superior to synchronous biventricular pacing in enhancing left ventricular systolic and diastolic performance: nine of 20 patients had delayed contraction of the posterolateral wall and improved with left ventricular lead preactivation, and 11 of 20 had delayed contraction of the septum or inferior wall and showed optimal benefit with right ventricular lead preactivation. The benefits were shown within a small range of interventricular pacing intervals from +20 to −20 ms, and further increases in preactivation were deleterious to left ventricular function. In contrast, our mean data suggest that sequential biventricular pacing with right ventricular lead preactivation was typically better than unpaced, but less beneficial than synchronous biventricular pacing, while the greatest mechanical benefit was seen with left ventricular lead preactivation. One explanation for these discrepancies may be differences in baseline patient characteristics. In the study by Sogaard, despite having left bundle branch block, fewer than 50% of patients showed delayed lateral or posterior left ventricular wall contraction (and those that did, benefited from left ventricular lead preactivation), whereas 75% of our patients had this pattern at baseline. From either left ventricular pacing site the lowest mean measures of dyssynchrony and hence the most optimally coordinated contractions were seen with left ventricular preactivation.

During atrio‐biventricular pacing there is fusion between intrinsic excitation of the septum along the native right bundle branch and excitation of the distal septum from right ventricular pacing, combined with the premature pacing of the left ventricle.22 The pattern of electrical activation is dependent on propagation of the three wavefronts, regions of slow conduction, with spread of excitation within the left ventricle through both specialised and non‐specialised conducting tissue. The resulting mechanical contraction is further influenced by heart failure aetiology, myocardial architecture, the presence of scar tissue,23,24 the position of the pacing electrodes, and the atrioventricular and interventricular pacing intervals.

Biventricular pacing with left ventricular preactivation was seen to maximally enhance mean left ventricular ejection fraction and septal systolic velocity, without significant effect on other left ventricular and right ventricular systolic velocities. Isolated left bundle branch block is known to reduce septal regional ejection fraction while sparing that of the apex and lateral walls,25 and biventricular pacing may simply correct contractility in the regions in which it is impaired. During biventricular pacing there is slow progression of anterolateral wall shortening, with more prominent septal wall shortening, which may arise as a result of preloading and prestretching of septal regions remote from the site of preactivation.26 Adjustment of the interventricular pacing interval may help to optimise septal myocardial stretch and hence contractility.

Influence of left ventricular pacing site

Consistent with previous studies,4 we found that the lateral and posterior left ventricular walls were the most frequent sites of delayed regional contraction (100% of patients with idiopathic dilated cardiomyopathy and 50% of patients with ischaemic heart disease). Published data suggest that the greatest improvement in left ventricular performance is seen when the site of left ventricular pacing is concordant with the site of latest left ventricular contraction.4 Consequently, left ventricular pacing from the lateral free wall has been shown to be superior to an anterior site, where stimulation has been thought to provide the least haemodynamic benefit in the majority of patients.19 In this study, the limited pacing data for four patients from an anterior left ventricular site also failed to show improvement in measures of ventricular dyssynchrony or cardiac function.

Although the inferior left ventricular wall has previously been considered a suboptimal site for left ventricular lead placement, we have demonstrated improvements in both mechanical dyssynchrony and systolic performance.

Temporary left ventricular pacing through an angioplasty wire

Most left ventricular lead delivery systems use over‐the‐wire angioplasty technology. This acute study using an angioplasty guidewire pacing technique provided information about coronary venous anatomy, local capture thresholds, and phrenic nerve stimulation. Previous studies using this technique have shown that individualisation of pacing site configuration can maximise haemodynamic responses and reduce non‐responders.10 This study has further increased our understanding of the mechanisms by which variations in left ventricular pacing site combined with sequential pacing influence mechanical resynchronisation and could be used to guide lead placement and pacemaker programming. The use of non‐invasive tools to reliably predict response to biventricular pacing is still evolving. Performing such a pre‐implant study may enhance the prediction of clinical response. However, reverse remodelling occurs as a chronic process and an apparent lack of acute response assessed by non‐invasive data should be interpreted judiciously.

Study limitations

This study design requires device implantation as a separate procedure to avoid compromising echocardiographic data collection and an increased risk of infection.

In our study, as in previous studies,19,27,28 short atrioventricular delays were programmed. These aimed to minimise fusion, although some degree of fusion cannot entirely be excluded during extreme sequential pacing, while lack of individual optimisation may have had a negative influence on haemodynamic improvement.

The technique of tissue Doppler imaging has some inherent limitations. Traces acquired from an apical view only assess longitudinal axis motion and cannot distinguish between active and passive wall motion. This study used only echocardiographic indices for comparing different pacing configurations. Bordachar et al7 recently showed that haemodynamic and electromechanical responses may vary with pacing configuration; and without haemodynamic data, it is uncertain whether reductions in dyssynchrony would yield an equivalent haemodynamic response.

Reductions in dyssynchrony indices have been observed in patients with reverse remodelling after biventricular pacing.13,29 In our study, with limited chronic data, we speculate that acute reductions in ventricular dyssynchrony will persist chronically and may be used as a guide to predict long term clinical response.

Conclusions

In this acute temporary pacing study of patients with severe heart failure and left bundle branch block, biventricular pacing reduced intra‐LV, inter‐LV–RV, and global mechanical dyssynchrony, and increased septal systolic velocity and left ventricular ejection fraction.

The optimal interventricular pacing interval varies between patients and is ideally assessed on an individual basis. Sequential biventricular pacing with left ventricular preactivation most often achieves maximal mechanical synchronicity and systolic function, particularly when the left ventricular lead is placed laterally.

Pacing from an inferior left ventricular site provides improvements in mechanical dyssynchrony and may offer an alternative when lateral left ventricular lead implantation is not possible.

A new temporary angioplasty wire pacing technique is described, which may have clinical value in assessing the response to biventricular pacing before permanent pacing.

Footnotes

Funding: This work was supported by a research grant from Medtronic Inc.

Competing interests: REL received a research grant from Medtronic Inc and has received sponsorship to attend scientific meetings from Medtronic Inc and Guidant Corporation. AWCC is a consultant on a research grant from Guidant Corporation; he has received sponsorship to attend scientific meetings from Medtronic Inc, Guidant Corporation, and St Jude Medical. JM is a consultant on two research grants from Medtronic Inc and has received sponsorship to attend scientific meetings from Medtronic Inc. DPF has acted as a consultant for Medtronic Inc. NSP has received sponsorship to attend scientific meetings from Medtronic Inc and Guidant Corporation. RJS has received sponsorship to attend medical meetings from St Jude Medical and Guidant Corporation. DWD is a consultant on two research grants from Medtronic Inc; he has received sponsorship to attend scientific meetings from Medtronic Inc and St Jude Medical; he is on the advisory board for St Jude Medical Atrial Fibrillation Division. His partner is an employee of Medtronic Inc.

References

1. Abraham W T, Fisher W G, Smith A L. et al Cardiac resynchronization in chronic heart failure. N Engl J Med 2002. 3461845–1853.1853 [PubMed]
2. Cazeau S, Leclercq C, Lavergne T. et al Effects of multisite biventricular pacing in patients with heart failure and intraventricular conduction delay. N Engl J Med 2001. 344873–880.880 [PubMed]
3. Reuter S, Garrigue S, Bordachar P. et al Intermediate‐term results of biventricular pacing in heart failure: correlation between clinical and hemodynamic data. Pacing Clin Electrophysiol 2000. 231713–1717.1717 [PubMed]
4. Ansalone G, Giannantoni P, Ricci R. et al Doppler myocardial imaging to evaluate the effectiveness of pacing sites in patients receiving biventricular pacing. J Am Coll Cardiol 2002. 39489–499.499 [PubMed]
5. Sogaard P, Egeblad H, Pedersen A K. et al Sequential versus simultaneous biventricular resynchronization for severe heart failure: evaluation by tissue Doppler imaging. Circulation 2002. 1062078–2084.2084 [PubMed]
6. Auricchio A, Stellbrink C, Block M. et al Effect of pacing chamber and atrioventricular delay on acute systolic function of paced patients with congestive heart failure. The Pacing Therapies for Congestive Heart Failure Study Group. The Guidant Congestive Heart Failure Research Group. Circulation 1999. 992993–3001.3001 [PubMed]
7. Bordachar P, Lafitte S, Reuter S. et al Biventricular pacing and left ventricular pacing in heart failure: similar hemodynamic improvement despite marked electromechanical differences. J Cardiovasc Electrophysiol 2004. 151342–1347.1347 [PubMed]
8. Meier B, Rutishauser W. Coronary pacing during percutaneous transluminal coronary angioplasty. Circulation 1985. 71557–561.561 [PubMed]
9. Mixon T A, Cross D S, Lawrence M E. et al Temporary coronary guidewire pacing during percutaneous coronary intervention. Catheter Cardiovasc Interv 2004. 61494–500.500 [PubMed]
10. van Campen L C, Visser F C, de Cock C C. et al Comparison of the hemodynamics of different pacing sites in patients undergoing resynchronization therapy: need for individualization and optimal lead localization. Heart 2006 Jun 27 (epub ahead of print)
11. Schiller N B. Two‐dimensional echocardiographic determination of left ventricular volume, systolic function, and mass. Summary and discussion of the 1989 recommendations of the American Society of Echocardiography. Circulation. 1991;84(3 suppl)I280–I287.I287
12. Yu C M, Chau E, Sanderson J E. et al Tissue Doppler echocardiographic evidence of reverse remodeling and improved synchronicity by simultaneously delaying regional contraction after biventricular pacing therapy in heart failure. Circulation 2002. 105438–445.445 [PubMed]
13. Penicka M, Bartunek J, De Bruyne B. et al Improvement of left ventricular function after cardiac resynchronization therapy is predicted by tissue Doppler imaging echocardiography. Circulation 2004. 109978–983.983 [PubMed]
14. Lane R E, Chow A W C, Chin D. et al Selection and optimisation of biventricular pacing: The role of echocardiography. Heart. 2004;90(suppl 6)vi10–vi16.vi16
15. Cleland J G, Daubert J C, Erdmann E. et al The effect of cardiac resynchronization on morbidity and mortality in heart failure. N Engl J Med 2005. 3521539–1549.1549 [PubMed]
16. Perego G B, Chianca R, Facchini M. et al Simultaneous vs. sequential biventricular pacing in dilated cardiomyopathy: an acute hemodynamic study. Eur J Heart Fail 2003. 5305–313.313 [PubMed]
17. van Gelder B M, Bracke F A, Meijer A. et al Effect of optimizing the VV interval on left ventricular contractility in cardiac resynchronization therapy. Am J Cardiol 2004. 931500–1503.1503 [PubMed]
18. Bordachar P, Lafitte S, Reuter S. et al Echocardiographic parameters of ventricular dyssynchrony validation in patients with heart failure using sequential biventricular pacing. J Am Coll Cardiol 2004. 442157–2165.2165 [PubMed]
19. Butter C, Auricchio A, Stellbrink C. et al Effect of resynchronization therapy stimulation site on the systolic function of heart failure patients. Circulation 2001. 1043026–3029.3029 [PubMed]
20. Dekker A L, Phelps B, Dijkman B. et al Epicardial left ventricular lead placement for cardiac resynchronization therapy: optimal pace site selection with pressure–volume loops. J Thorac Cardiovasc Surg 2004. 1271641–1647.1647 [PubMed]
21. Gasparini M, Mantica M, Galimberti P. et al Is the left ventricular lateral wall the best lead implantation site for cardiac resynchronization therapy? Pacing Clin Electrophysiol 2003. 26162–168.168 [PubMed]
22. Vogt J, Krahnefeld O, Lamp B. et al Electrocardiographic remodeling in patients paced for heart failure. Am J Cardiol. 2000;86(9 suppl 1)K152–K156.K156
23. Gomez A M, Valdivia H H, Cheng H. et al Defective excitation‐contraction coupling in experimental cardiac hypertrophy and heart failure. Science 1997. 276800–806.806 [PubMed]
24. Li Y Y, Feng Y, McTiernan C F. et al Downregulation of matrix metalloproteinases and reduction in collagen damage in the failing human heart after support with left ventricular assist devices. Circulation 2001. 1041147–1152.1152 [PubMed]
25. Grines C L, Bashore T M, Boudoulas H. et al Functional abnormalities in isolated left bundle branch block. The effect of interventricular asynchrony. Circulation 1989. 79845–853.853 [PubMed]
26. Leclercq C, Faris O, Tunin R. et al Systolic improvement and mechanical resynchronization does not require electrical synchrony in the dilated failing heart with left bundle‐branch block. Circulation 2002. 1061760–1763.1763 [PubMed]
27. Kass D A, Chen C H, Curry C. et al Improved left ventricular mechanics from acute VDD pacing in patients with dilated cardiomyopathy and ventricular conduction delay. Circulation 1999. 991567–1573.1573 [PubMed]
28. Auricchio A, Stellbrink C, Sack S. et al Long‐term clinical effect of hemodynamically optimized cardiac resynchronization therapy in patients with heart failure and ventricular conduction delay. J Am Coll Cardiol 2002. 392026–2033.2033 [PubMed]
29. Yu C M, Fung W H, Lin H. et al Predictors of left ventricular reverse remodeling after cardiac resynchronization therapy for heart failure secondary to idiopathic dilated or ischemic cardiomyopathy. Am J Cardiol 2002. 91684–688.688 [PubMed]

Articles from Heart are provided here courtesy of BMJ Group