PMCCPMCCPMCC

Search tips
Search criteria 

Advanced

 
Logo of brjopthalBritish Journal of OphthalmologyVisit this articleSubmit a manuscriptReceive email alertsContact usBMJ
 
Br J Ophthalmol. 2007 July; 91(7): 979–980.
PMCID: PMC1955641

Agreement between two non‐contact specular microscopes: Topcon SP2000P versus Rhine‐Tec

Endothelial cell density (ECD) assessment with the noncontact Topcon SP2000P specular microscope is known to be reliable when the automated mode is followed by manual corrections (touched‐up mode). We compared its agreement with Rhine‐Tec, a new noncontact specular microscope, in 270 eyes of 160 patients, by comparing the ECD measured in the automated and touched‐up modes with that of Topcon touched‐up. Good agreement existed between either touched‐up modes with a mean difference of only 2 cells/mm2 95% CI (−27; 23) whereas agreement with the Rhine‐Tec automated mode was poor with an overestimation by a mean of 226 cells/mm2 95% CI (172; 281).

Background

The Topcon SP2000P non‐contact specular microscope (Topcon, Tokyo, Japan) is widely used to measure corneal endothelial cell density (ECD). It uses a cell contour recognition algorithm based on contrast differences, and ECD derived from an automated delineation of cell boundaries with manual correction of inaccurately drawn cells (“touched‐up” mode) has been validated.1,2 A new commercially available non‐contact specular microscope Rhine‐Tec (Rhine‐Tec, Krefeld, Germany) determines ECD by a cell‐centre method consisting of identification of centres of contiguous cells either in an automated or a touched‐up mode using an operator‐defined cell centre. We aimed to determine its agreement with the Topcon touched‐up mode.

Methods

The central endothelium of 270 eyes of 160 subjects (healthy subjects and those operated on for cataract or penetrating keratoplasty to explore a wide range of ECDs) was imaged, first with the Topcon microscope and immediately afterwards with the Rhine‐Tec specular microscope. On each sharpest image, the ECD and percentage of hexagonal cells were determined on the largest possible area by a trained ophthalmologist blinded to the clinical antecedents and choosing the device (Topcon or Rhine‐Tec) randomly. Both automated and touched‐up modes were used for Rhine‐Tec, and the Topcon touched‐up served as the reference.1,2 Agreement was determined using the Bland–Altman method.3 Means were compared using paired t tests (or non‐parametric Wilcoxon's test when required) and results considered significant when p<0.05.

Results

Mean (standard deviation (SD)) number of cells counted was 129 (46) (range 24–219, median 144) for Topcon and 137 (44) (range 47–279, median 136) for Rhine‐Tec (p<0.001). Despite a good correlation between the Topcon touched‐up and Rhine‐Tec automated modes (y = 0.37x+1497, r = 0.8, p<0.001), the agreement was poor, with Rhine‐Tec overestimating ECD by a mean of 226 cells/mm2 95% CI (172 to 281) (fig 11).). A continuous bias in ECD determination characterised by overestimation of low ECDs and underestimation of high ones (p<0.001) was observed with the Rhine‐Tec automated mode, with maximal effects of this drift visible for low (<1800) and high (>2500 cells/mm2) ECDs. Correlation between the either touched‐up modes was good (y = 0.82x+377, r = 0.93 p<0.001) and agreement was far better with a mean difference of 2 cells/mm2 95% CI (−27 to 23) and narrower bounds of agreements (fig 22).). Although the tendency to overestimate low and underestimate high ECDs persisted, it was significantly lower. Mean (SD) percentage hexagonality (calculated in touched‐up modes) was 58 (12) (range 14–100, median 59) and 41 (8) range 14–69, median 41) for Topcon and Rhine‐Tec, respectively (p<0.001).

figure bj99374.f1
Figure 1 Difference against mean graph showing the agreement between the endothelial cell densities (ECDs) obtained with the Topcon touched‐up mode compared with the Rhine‐Tec automated mode. ECD measured in Topcon touched‐up ...
figure bj99374.f2
Figure 2 Difference against mean graph showing the agreement between the endothelial cell densities (ECDs) obtained with the Topcon and Rhine‐Tec touched‐up modes. The solid horizontal line indicated the mean difference between ...

Comments

Both non‐contact specular microscopes assess ECD on small cell samples. As observed with Topcon,1,2,4 the Rhine‐Tec automated mode is also not reliable. However, after manual corrections, Rhine‐Tec shows an acceptable agreement with Topcon. Regarding the two bounds of agreement on this sample size, a consistent individual variation remains. The reliability of hexagonality assessment using Topcon was already dealt with.1 The difference observed with Rhine‐Tec remains unexplained and is probably linked to a different image analysis algorithm (proprietary data not available). Information regarding the coefficient of variation of the cell area would be helpful to further evaluate this device.

Footnotes

Competing interest: None.

Presented in part at the 18th annual meeting of the European Eye Bank Association held on 20–21 January 2006 in Venice, Italy, and 112th annual meeting of the French Society of Ophthalmology held on 6–10 May 2006 in Paris, France.

References

1. Cheung S W, Cho P. Endothelial cells analysis with the Topcon specular microscope SP‐2000P and IMAGEnet system. Curr Eye Res 2000. 21788–798.798 [PubMed]
2. van Schaick W, van Dooren B T, Mulder P G. et al Validity of endothelial cell analysis methods and recommendations for calibration in Topcon SP‐2000P specular microscopy. Cornea 2005. 24538–544.544 [PubMed]
3. Bland J M, Altman D G. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet 1986. 1307–310.310 [PubMed]
4. Vecchi M, Braccio L, Orsoni J G. The Topcon SP 1000 and IMAGEnet systems. A comparison of four methods for evaluating corneal endothelial cell density. Cornea 1996. 15271–277.277 [PubMed]

Articles from The British Journal of Ophthalmology are provided here courtesy of BMJ Publishing Group